← Product Code [BZG](/submissions/AN/subpart-b%E2%80%94diagnostic-devices/BZG) · K061571

# ASTRA 300 (K061571)

_Sdi Diagnostics, Inc. · BZG · Jan 31, 2007 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-b%E2%80%94diagnostic-devices/BZG/K061571

## Device Facts

- **Applicant:** Sdi Diagnostics, Inc.
- **Product Code:** [BZG](/submissions/AN/subpart-b%E2%80%94diagnostic-devices/BZG.md)
- **Decision Date:** Jan 31, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.1840
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology

## Indications for Use

The SDI Astra 300 Spirometer is a freestanding laboratory instrument for performing basic lung function tests and oximetry in people of all ages. It is intended to be used by physicians or professional medical personnel for testing in any setting.

## Device Story

ASTRA 300 is a handheld, portable diagnostic spirometer; measures patient breath flow and volume. Uses turbine transducer with infrared interruption to detect flow; internal algorithms calculate pulmonary values. Optional pulse oximetry module measures blood oxygen saturation (SpO2) and pulse rate. Device displays tabular and graphical data on integrated LCD. Operated by physicians or professional medical personnel in various settings. Output assists clinicians in assessing pulmonary condition and making diagnostic decisions.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Handheld portable spirometer; turbine transducer with infrared interruption sensing; optional pulse oximetry module; LCD display; internal software algorithms for data processing.

## Regulatory Identification

A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.

## Predicate Devices

- SDI Spirolab II Spirometer ([K013812](/device/K013812.md))
- Motion Media Technology CareStation 126S ([K031863](/device/K031863.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K061571

Image /page/0/Picture/1 description: The image shows the logo for SDI Diagnostics. The logo consists of the letters "sdi" in a bold, sans-serif font, stacked above the word "DIAGNOSTICS" in a smaller, sans-serif font. A horizontal line separates the letters "sdi" from the word "DIAGNOSTICS".

JAN 3 1 2007

### 510(k) Summary

## [As Required by 21 CFR 807.92]

| Owner / Submitter of 510(k)     | SDI Diagnostics, Inc.<br>Michael J. Boyle - President<br>10 Hampden Drive<br>Easton, MA 02375                                                                                                                                                                                                                                                                                              |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                 | Tel: (508)238-7033, fax (508)230-8497                                                                                                                                                                                                                                                                                                                                                      |
|                                 | e-mail: mjboyle@sdidiagnostics.com                                                                                                                                                                                                                                                                                                                                                         |
| Establishment Registration No.: | 1221256                                                                                                                                                                                                                                                                                                                                                                                    |
| Contact:                        | Cosimo Cariolo<br>e-mail: ccariolo@sdidiagnostics.com                                                                                                                                                                                                                                                                                                                                      |
| Trade Name:                     | SDI ASTRA 300 Diagnostic Spirometer With or Without Pulse<br>Oximetry Function                                                                                                                                                                                                                                                                                                             |
| Common Name:                    | Spirometer                                                                                                                                                                                                                                                                                                                                                                                 |
| Classification Name:            | Spirometer, Diagnostic                                                                                                                                                                                                                                                                                                                                                                     |
| Regulation Number               | 868.1840                                                                                                                                                                                                                                                                                                                                                                                   |
| Classification Panel:           | Anesthesiology                                                                                                                                                                                                                                                                                                                                                                             |
| Regulatory Class:               | II                                                                                                                                                                                                                                                                                                                                                                                         |
| Product Code:                   | BZG, DQA                                                                                                                                                                                                                                                                                                                                                                                   |
| Predicate Devices               | 510(k) #K013812 SDI Spirolab II Spirometer<br>510(k) #K031863 Motion Media Technology CareStation 126S                                                                                                                                                                                                                                                                                     |
| Device Description:             | The ASTRA 300 Diagnostic Spirometer is a hand-held portable<br>diagnostic spirometer for the measurement of patient breath flow<br>and volume. The device uses a turbine transducer that measures<br>flow via infrared interruption. Algorithms are used to determine<br>values based on this flow measurement. Tabular and graphical<br>data are displayed on the spirometer LCD display. |

and the continued to the contraction of the program and the program and the first of the first of the first of the first of the first of the first

HAMPDEN :

ELEPHONE

,  

{1}------------------------------------------------

Device Description (cont'd):

With the optional Pulse Oximeter module, the specific patient parameters oxygen saturation in the blood (SpO2) and pulse rate, both relating to the patient's pulmonary condition, may be measured and displayed.

The state and the commend of the comments of the first to the first to the first to the first

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" are arranged in a circular pattern around the eagle. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Cosimo Cariolo Director of Marketing SDI Diagnostics, Incorporated 10 Hampden Drive Easton, Massachusetts 02375

Re: K061571

JAN 3 1 2007

Trade/Device Name: Astra 300 Regulation Number: 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: II Product Code: BZG Dated: January 29, 2007 Received: January 30, 2007

Dear Mr. Cariolo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

#### Page 2 - Mr. Cariolo

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number: K061571

Device Name: Astra 300

Indications For Use: The SDI Astra 300 Spirometer is a freestanding laboratory instrument for performing basic lung function tests and oximetry in people of all ages. It is intended to be used by physicians or professional medical personnel for testing in any setting.

Prescription Use
(Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Whet Whal

(Division Sign-Off) Division of Anastin Cities Cisneral Hospital, Infection Control, Dental Devices

510(k) Number: K061571

3

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-b%E2%80%94diagnostic-devices/BZG/K061571](https://fda.innolitics.com/submissions/AN/subpart-b%E2%80%94diagnostic-devices/BZG/K061571)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
