The Flowhandy ZAN100 USB is an open, personal-computer-based spirometry system with optional shutter for measuring and analyzing breath flow, volume, and pressure in adult and pediatric subjects for use by pulmonologists, allergists, general practitioners, and occupationalmedicine practitioners in lung-function diagnosis.
Device Story
Flowhandy ZAN100 USB is a PC-based pneumotachometer spirometry system. Input: breath flow/pressure via hand-held pneumotach with variable-orifice core and differential pressure sensor. Optional solenoid-activated shutter interrupts airflow for pressure measurements. Transformation: PC software calculates spirometric values from pressure drop across orifice. Output: spirometric data displayed, stored, and analyzed on PC. Used in clinical settings (pulmonology, allergy, general/occupational practice) by healthcare professionals. Benefits: enables lung-function diagnosis through standardized breathing maneuvers.
Clinical Evidence
No clinical trials were conducted. Evidence consists of bench testing for safety (EN 60601-1, EN 60601-1-2) and effectiveness (meeting American Thoracic Society requirements). Clinical performance is supported by CE marking and historical operational data in Europe demonstrating safe and effective use under actual conditions.
Technological Characteristics
Pneumotachometer using variable-orifice diaphragm; semiconductor pressure transducers (0-1.4 kPa flow, 0-7 kPa pressure). Materials: POM (polyoxymethylene). Power: 5 VDC via USB. Dimensions: 100x90x45 mm. Connectivity: USB interface to PC. Applied part: Type BF. Enclosure: IPX1. Software-based analysis of flow/volume/pressure.
Indications for Use
Indicated for adult and pediatric patients requiring lung-function diagnosis, including measurement and analysis of breath flow, volume, and pressure.
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
{1}------------------------------------------------
510(k) Number K052176
Date
# Flowhandy ZAN100 USB 510(k) Summary
## Submitter
| Company: | ZAN Messgeräte GmbH<br>Schlimpfhofer Strasse 14, D-97723 Oberthulba, Germany<br>Tel +49 97 36 81810, Fax +49 97 36 818120 |
|-------------------|---------------------------------------------------------------------------------------------------------------------------|
| Contact: | Jim Lewis<br>Ferraris Respiratory Inc, 908 Main Street, Louisville CO 80027, USA<br>Tel 303 666 5555, Fax 303 666 5588 |
| Prepared: | August 2005 |
| Device Name | |
| Trade: | Flowhandy ZAN100 USB |
| Common: | Spirometer |
| Classification: | Diagnostic spirometer, 21 CFR 868.1840; Class II |
| Product code: | 73-BZG |
| Predicate Devices | |
| Trade Name: | MasterScreen Pneumo |
| 510(k) Number: | K933839 |
| Manufacturer: | Erich Jaeger Gmbh |
#### Device Description
The Flowhandy ZAN100 USB is a pneumotachometer spirometry system. The system comprises a Windows personal computer (sold separately) for data collection, analysis, storage, and display; proprietary Betterflow ZAN100 USB software program for the PC; and a Flowlandy, hand-held pneumotach with reusable variable-orifice core, differential pneumatic pressure sensor, and USB interface. An optional computer-controlled shutter may be added to the air passageway to interrupt flow to allow pressure measurements using the on-board pressure sensor.
The ZAN100 measures the pressure drop across the known orifice to indicate the rate of breath flow both in and out of the air passageway. Knowing the flow rate of air from a subject allows the calculation of the most recognized spirometric values for the subject. Closing off the passageway with the optional shutter enables the pressure sensor to take important pressure measurements for the subject's breathing as well.
#### Indications for Use
The Flowhandy ZAN100 USB is an open, personal-computer-based spirometry system with optional shutter for measuring and analyzing breath flow, volume, and pressure in adult and pediatric subjects for use by pulmonologists, allergists, general practitioners, and occupationalmedicine practitioners in lung-function diagnosis.
{2}------------------------------------------------
## Summary of Technological Characteristics
The Flowhandy uses the pressure difference in the air stream across a variable-orifice diaphragm to measure a subject's airflow rate (pneumotachometry) throughout established breathing maneuvers to provide spirometric analysis. The instantaneous flow rate values are acquired on a personal computer, where the data is analyzed, stored, and displayed. The optional shutter is a solenoid-activated valve for interrupting airflow.
### Technical Specifications
.
. .
.
.
♥
.
.
.
.
.
.
.
●
.
●
.
.
- Operating temperature range: 10 to 40 °C Operating humidity: 20 to 90%, non-condensing Storage temperature: 0 to 40 °C Storage humidity: 10 to 90%, non-condensing Material: POM (polyoxymethylene) Enclosure rating: IPX1 Type of applied part: BF (subject floating) Flow sensing Measuring principle: Pressure difference with variable diaphragm (pneumotachometry) Dimensions: 100x90x45 mm Weight: 250 a Power supply: 5 VDC, 0.2 Wmax (powered from USB port) Pressure transducer: Semiconductor; 0 to 1.4 kPa; 0.1% accuracy Volume resolution: <5 ml Flow resolution: <1 ml/s Range: 0 to 15 l/s, bi-directional Max linearity error (corrected): 2.5% Flow resistance: <0.03 kPa/l/s Effective dead space: <50 ml
Shutter valve for mouth/nasal pressure reading Dimensions: ● 180x50x80 mm
- Weight: . 420 q Additional dead space: . <30 ml
- Mouth/Nasal pressure transducer: .
- Semiconductor; 0 to 7 kPa; 0.2% accuracy Power supply: 5 VDC, 0.5 Wmax (powered from USB port)
## Summary of Non-Clinical Performance Data
#### Safety
.
The Flowhandy was examined and bench tested by third-party examiners to demonstrate conformance to three recognized international consensus standards for safety of medical electrical equipment: EN 60601-1 for general medical-device safety of medical electrical safety, and EN 60601-1-2 for electromagnetic compairity,
#### Effectiveness
ln-house and third-party testing demonstrate that Flowhandy measurement performance meets or exceeded published American Thoracic Society requirements and the claimed performance meeting requirements under stated and anticipated operating conditions.
{3}------------------------------------------------
## Summary of Clinical Performance Data
CE marking and years of successful operation in Europe demonstrate that patients and clinicians can safely and effectively use the Flowhandy under actual-use conditions and that the users guide, product physical design, and other human-factor characteristics of the Flowlandy system are appropriate for the product's intended use.
## Equivalence to Predicate Devices
The Flowhandy design uses the same technologies and design principles as its predicate devices. The system is substantially equivalent to Jaeger's MasterScreen Pneumo: using the same measurement principles; similar construction, material, and energy source; and meeting the same performance characteristics and intended uses.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes representing the three branches of government. The eagle is surrounded by the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" in a circular arrangement.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 1 2 2006
Zan Messgerate GmbH C/O Mr. Jim Lewis Ferraris Respiratory, Incorporated 908 Main Street Louisville, Colorado 80027
Re: K052176
Trade/Device Name: Flowhandy ZAN100 USB Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: II Product Code: BZG Dated: December 27, 2005 Received: December 28, 2005
Dear Mr. Lewis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Lewis
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name: Flowhandy ZAN100 USB
Indications for Use:
The Flowhandy ZAN100 USB is an open, personal-computer-based spirometry system with optional shutter for measuring and analyzing breath flow, volume, and pressure in adult and pediatric subjects for use by pulmonologists, allergists, general practitioners, and occurationalmedicine practitioners in lung-function diagnosis.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
の 2006年の1000年には、1000年には、1000年には、1000年には、1000年には、1000年には、1000年には、1000年には、1000年には、1000年には、1000年には、1000年には1000年に100000000000000000000000000000000000000000000000000000000000000
Chr. M.
: General Hospital
Services
K052176
Intended Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.