K183567 · Vyaire Medical · BZC · Aug 23, 2019 · Anesthesiology
Device Facts
Record ID
K183567
Device Name
Vyntus/SentrySuite Product Line
Applicant
Vyaire Medical
Product Code
BZC · Anesthesiology
Decision Date
Aug 23, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1880
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Vyntus/SentrySuite Product Line is intended to be used for measurements, data collection and analysis of lung function (PFT) and cardio-pulmonary (CPET) parameters, aiding in the diagnosis of related conditions. The results of the test can be viewed online with the help of a computer screen and can be printed after the test. The test results can be saved for future reference or report generation purposes. The products can be utilized with patients age 4 years and older as long as they can cooperate in the performance - no special limit to patient's sex or height exists. Measurements will be performed under the direction of a physician in a hospital environment, physician's office or similar setting (professional healthcare facilities). A qualified physician has to reassess all Vyntus/SentrySuite measurements. An interpretation by Vyntus/SentrySuite is only significant if it is considered in connection with other clinical findings. Additional for Vyntus ECG: The Vyntus ECG is intended for measuring the surface electrocardiogram (ECG) of a patient. The acquired ECG can be recorded and displayed on the screen or printed on paper. 12-lead ECGs are analyzed automatically and suggestions for the interpretation of the resting ECG can be made by the software. ECG interpretation statements made by the Vyntus/ SentrySuite represent partial qualitative information on the patient's cardiovascular condition and no therapy or drugs can be administered based solely on the interpretation statements. The Vyntus ECG can be used for non-interpretive applications in patients age 4 years and older and a weight of 20 kg or higher. The Vyntus ECG is intended to be used for routine ECG collection, recording both under resting and stress conditions. The measurement is performed by trained healthcare professionals under the direction of a physician in healthcare facilities (e.g. the doctor's office or hospital). The Vyntus ECG is not intended for intracranial use. The Vyntus ECG is not intended for use in an EMS environment (Emergency Medical Services Environment). The Vyntus ECC is not intended for use in home healthcare environments. Automatic interpretation of the ECG is not possible for pediatric and adolescent patients below 16 years of age and for patients with pacemakers.
Device Story
System measures lung function (PFT) and cardio-pulmonary (CPET) parameters; collects surface ECG signals. Inputs: gas exchange (O2 uptake, CO2 production), ventilation, ECG waveforms. Hi/Lo FiO2 option allows measurement with varying inspiratory O2 concentrations. Software (SentrySuite 3.0) processes data, performs automated 12-lead ECG interpretation, and generates reports. Used in hospitals/clinics by trained professionals under physician direction. Output viewed on screen/printed. Physician reassesses all measurements; ECG interpretation provides partial qualitative/quantitative info; no therapy/drugs administered based solely on software output. Benefits: aids diagnosis of pulmonary/cardiovascular conditions.
Clinical Evidence
No clinical data performed. Substantial equivalence supported by bench testing, including basic safety (IEC 60601-1), EMC (IEC 60601-1-2), risk management (ISO 14971), usability (EN 62366), software lifecycle (ISO 62304), biocompatibility (ISO 10993-1), and measurement accuracy testing.
Technological Characteristics
CPET/PFT system with Hi/Lo FiO2 option. Includes HR-valve, breathing bag, DVT, and inhaler adapter. Connectivity: PC-based SentrySuite software. Software version 3.0. Non-sterile. Standards: IEC 60601-1, IEC 60601-1-2, ISO 14971, EN 62366, ISO 62304, ISO 10993-1.
Indications for Use
Indicated for patients age 4+ (ECG 20kg+) for lung function (PFT) and cardio-pulmonary (CPET) parameter measurement/analysis and routine ECG collection (resting/stress). Contraindicated for intracranial use, EMS environments, home healthcare, and automatic ECG interpretation for patients <16 years or with pacemakers.
Regulatory Classification
Identification
A pulmonary-function data calculator is a device used to calculate pulmonary-function values based on actual physical data obtained during pulmonary-function testing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 23, 2019
Vyaire Medical Elmar Niedermeyer Cnstl. Regulatory Affairs Leibnizstrasse 7 Hoechberg, 97204 De
Re: K183567
Trade/Device Name: Vyntus/SentrySuite Product Line Regulation Number: 21 CFR 868.1880 Regulation Name: Pulmonary-Function Data Calculator Regulatory Class: Class II Product Code: BZC, DPS Dated: July 23, 2019 Received: July 25, 2019
## Dear Elmar Niedermeyer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-
combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for James J. Lee, PhD Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K183567
Device Name
Vyntus/SentrySuite Product Line
### Indications for Use (Describe)
The Vyntus/SentrySuite Product Line is intended to be used for measurements, data collection and analysis of lung function (PFT) and cardio-pulmonary (CPET) parameters, aiding in the diagnosis of related conditions. The results of the test can be viewed online with the help of a computer screen and can be printed after the test results can be saved for future reference or report generation purposes.
The products can be utilized with patients age 4 years and older as long as they can cooperate in the performance - no special limit to patient's sex or height exists. Measurements will be performed under the direction of a physician in a hospital environment, physician's office or similar setting (professional healthcare facilities).
A qualified physician has to reassess all Vyntus/SentrySuite measurements. An interpretation by Vyntus/SentrySuite is only significant if it is considered in connection with other clinical findings.
Additional for Vyntus ECG:
The Vyntus ECG is intended for measuring the surface electrocardiogram (ECG) of a patient. The acquired ECG can be recorded and displayed on the screen or printed on paper. 12-lead ECGs are analyzed automatically and suggestions for the interpretation of the resting ECG can be made by the software. ECG interpretation statements made by the Vyntus/ SentrySuite represent partial qualitative information on the patient's cardiovascular condition and no therapy or drugs can be administered based solely on the interpretation statements.
The Vyntus ECG can be used for non-interpretive applications in patients age 4 years and older and a weight of 20 kg or higher. The Vyntus ECG is intended to be used for routine ECG collection, recording both under resting and stress conditions. The measurement is performed by trained healthcare professionals under the direction of a physician in healthcare facilities (e.g. the doctor's office or hospital). The Vyntus ECG is not intended for intracranial use. The Vyntus ECG is not intended for use in an EMS environment (Emergency Medical Services Environment). The Vyntus ECC is not intended for use in home healthcare environments. Automatic interpretation of the ECG is not possible for pediatric and adolescent patients below 16 years of age and for patients with pacemakers.
Type of Use (Select one or both, as applicable)
| <span></span> | <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|------------------------------------------------------------|
| <span></span> | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Vyaire Medical. The word "Vyaire" is in a dark blue, sans-serif font, with the "TM" symbol in the upper right corner. Below "Vyaire" is the word "MEDICAL" in a smaller, light blue, sans-serif font.
# 510(k) Summary
Colleen Watson
847-8473629
Vyaire Medical Inc.
26125 N. Riverwoods Blvd. Mettawa, IL 60045, USA + 847-786-5998 x110070
Colleen.Okeeffe@vyaire.com
#### I. Type of Submission
Traditional 510(k) Date Prepared: August 21, 2019
| II. | Submitter |
|----------|---------------------|
| Name: | Vyaire Medical GmbH |
| Address: | Leibnizstrasse 7 |
| | D-97204 Hoechberg |
| | Germany |
| Official Correspondent | Elmar Niedermeyer | US Agent |
|------------------------|-----------------------------------------------------------------------------|----------|
| Address: | CareFusion Germany 234 GmbH<br>Leibnizstrasse 7, 97204 Hoechberg<br>Germany | Address |
| Phone: | +49 931 49 72 - 361 | Phone: |
| FAX: | +49 931 49 72 - 62361 | Fax: |
| E-mail | elmar.niedermeyer@vyaire.com | E-mail: |
| III. | Establishment Registration Number |
|---------|-----------------------------------|
| 9615102 | |
#### IV. Common Name or Classification Name
Pulmonary function data calculator (Primary) (CFR 868.1880, Product Code BZC) Electrocardiograph (Secondary) (CFR 870.2340, Product Code DPS)
#### V. Trade Name
Vyntus/SentrySuite Product Line
#### VI. Device Classification
This is a Class II device
#### VII. Classification Panel
73 Anesthesiology Part 868 (Primary) 74 Cardiovascular Part 870 (Secondary)
#### VIII. Reason for Premarket Notification
New medical device, new option Hi/Lo FiO2, Software SentrySuite 3.0 (update)
#### IX. Legally predicate marketed devices
Vyntus / SentrySuite Product Line K150810 Code BZC, DPS Predicate Device Company: Vyaire Medical GmbH
#### X. Reference Devices
K992214 Code BZC, DPS, MWI, MLC Oxycon Pro Reference Device Company: Vyaire Medical GmbH
#### XI. Device Description
Option Hi/Lo FiO2: The High/low FiO2 option of the Vyntus CPX is designed to measure the ventilation and the gas exchange (02 uptake, CO2 production) of a subject with varying inspiratory concentrations of O2.
Vyaire Medical GmbH
{4}------------------------------------------------
#### XII. Indications for Use
The Vyntus/SentrySuite Product Line is intended to be used for measurements, data collection and analysis of lung function (PFT) and cardio-pulmonary (CPET) parameters, aiding in the diagnosis of related conditions. The results of the test can be viewed online with the help of a computer screen and can be printed after the test. The test results can be saved for future reference or report generation purposes.
The products can be utilized with patients age 4 years and older as long as they can cooperate in the performance - no special limit to patient's sex or height exists. Measurements will be performed under the direction of a physician in a hospital environment, physician's office or similar setting (professional healthcare facilities). A qualified physician has to reasess all Vyntus/SentrySuite measurements. An interpretation by Vyntus/SentrySuite is only significant if i is considered in connection with other clinical findings.
## Additional for Vyntus ECG:
The Vyntus ECG is intended for measuring the surface electrocardiogram (ECG) of a patient. The acquired ECG can be recorded and displayed on the screen or printed on paper. 12-lead ECGs are analyzed automatically and suggestions for the interpretation of the resting ECG can be made by the software. ECG interpretation statements made by the Vyntus/SentrySuite represent partial qualitative and quantitative information on the patient's cardiovascular condition and no therapy or drugs can be administered based solely on the interpretation statements. The Vyntus ECG can be used for non-interpretive applications in patients age 4 years and older and a weight of 20 kg or higher. The Vyntus ECG is intended to be used for routine ECG collection, recording both under resting and stress conditions. The measurement is performed by trained healthcare professionals under the direction of a physician in healthcare facilities
(e.g. the doctor's office or hospital). The Vyntus ECG is not intended for intracranial use. The Vyntus ECG is not intended for use in an EMS environment (Emergency Medical Services Environment). The Vyntus ECG is not intended for use in home healthcare environments. Automatic interpretation of the ECG is not possible for pediatric and adolescent patients below 16 years of age and for patients with pacemakers.
#### XIIII. Required Components
- Vyntus CPX / ECG with SentrySuite Software ●
- Instruction for Use
- Option Hi/Lo FiO2
- HR-valve 2730 complete ●
- Breathing bag
- DVT complete for High / Low FiO2 ●
- Hans-Rudolph inhaler adapter ●
- Filling assembly for bag ●
{5}------------------------------------------------
#### XIV. 5.14 Summary Table of Comparison
| Comparison to predicate devices SentrySuite Product Line Predicate K150810 &<br>Reference Device Oxycon Pro K992214 | | | |
|---------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| | Oxycon Pro<br>K992214<br>Reference Device | Vyntus/SentrySuite Product Line<br>K150810<br>Predicate | Vyntus/SentrySuite Product<br>Line<br>(with Hi/Lo FiO2) |
| Indication for<br>Use | Oxycon Pro is a software-driven<br>medical device for exercise<br>measurements, including<br>ECG ST Segment Analysis and/or<br>ECG Stress Analysis. It measures<br>the human response to increasing<br>workloads with emphasis on the gas<br>exchange parameters.<br>Measurements include ventilation,<br>oxygen uptake, carbon dioxide<br>excretion, heart rate and derived<br>parameters. The results of the tests,<br>including the ECG wave forms, can<br>be viewed on the computer screen<br>and can be printed during the test.<br>The test results can be saved on the<br>computer hard disk for further<br>referral or report generation<br>purposes.<br>The Oxycon Pro interfaces to a test<br>subject via mouthpiece or a face<br>mask and ECG electrodes. The<br>Oxycon Pro interfaces to a<br>peripheral ergometer or treadmill.<br>The patient population is 4 years<br>and older.<br>The Oxycon Pro is capable of<br>performing computerized ECG<br>interpretation during resting<br>condition.<br>The intended use locations are<br>either in a physician office, hospital<br>exercise rehabilitation facilities, or<br>similar areas. It is intended to be<br>used by or on the order of a<br>physician or similar qualified health<br>care professional. This device is<br>intended for use in the hospital<br>environment, physician's office, or<br>similar settings. This device is not<br>intended for home use. | The Vyntus/SentrySuite product line is<br>intended to be used for measurements,<br>data collection and analysis of lung<br>function (PFT) and cardio-pulmonary<br>(CPET) parameters, aiding in the<br>diagnosis of related conditions. The<br>results of the test can be viewed online<br>with the help of a computer screen and<br>can be printed after the test. The test<br>results can be saved for future reference<br>or report generation purposes.<br>The products can be utilized with<br>patients age 4 years and older as long as<br>they can cooperate in the performance -<br>no special limit to patient's sex or height<br>exists. Measurements will be performed<br>under the direction of a physician in a<br>hospital environment, physician's office<br>or similar setting (professional<br>healthcare facilities). A qualified<br>physician has to reassess all<br>Vyntus/SentrySuite measurements. An<br>interpretation by Vyntus/SentrySuite is<br>only significant if it is considered in<br>connection with other clinical findings.<br>Additional for Vyntus ECG:<br>The Vyntus ECG is intended for<br>measuring the surface electro-<br>cardiogram (ECG) of a patient. The<br>acquired ECG can be recorded and<br>displayed on the screen or printed on<br>paper. 12-lead ECGs are analyzed<br>automatically and suggestions for the<br>interpretation of the resting ECG can be<br>made by the software. ECG<br>interpretation statements made by the<br>Vyntus/SentrySuite represent partial<br>qualitative and quantitative information<br>on the patient's cardiovas-cular<br>condition and no therapy or drugs can be<br>administered based solely on the<br>interpretation statements. The Vyntus<br>ECG can be used for non-interpretive<br>applications in patients age 4 years and<br>older and a weight of 20 kg or higher.<br>The Vyntus ECG is intended to be used<br>for routine ECG collection, recording<br>both under resting and stress conditions. | Identical<br>(to Predicate K150810) |
| | | trained healthcare professionals under<br>the direction of a physician in healthcare<br>facilities (e.g. the doctor's office or<br>hospital). The Vyntus ECG is not<br>intended for intracranial use. The Vyntus<br>ECG is not intended for use in an EMS<br>environment (Emergency Medical<br>Services Environment). The Vyntus<br>ECG is not intended for use in home<br>healthcare environments. Automatic<br>interpretation of the ECG is not possible<br>for pediatric and adolescent patients<br>below 16 years of age and for patients<br>with pacemakers. | |
| Intended<br>Patient<br>population | The patient population is 4 years<br>and older. | 4 years and older and a weight of 20 kg<br>or higher | Identical<br>(to Predicate K150810) |
| Software<br>Platform | JLab Software | SentrySuite Software | Identical<br>(to Predicate K150810) |
| Software<br>Version | JLab version 5.7 | SentrySuite version 2.13 | SentrySuite version 3.0 |
| Perfor-mance<br>(measuring<br>programs) | Slow Spirometry Forced Spirometry MVV Breath-by-Breath Indirect Calorimetry Resting ECG Stress ECG | Slow Spirometry Forced Spirometry MVV Breath-by-Breath Indirect Calorimetry Resting ECG Stress ECG | Identical |
| | High/Low FiO2 | | Identical<br>(to Reference K992214) |
| | Intrabreath Respiratory Drive P0.1 Mixing Chamber | | |
| Perfor-mance<br>(technical data<br>and accuracy) | Flow measurement<br>Range: 0 - 15 L/s<br>Accuracy: 70 mL/s or 3%<br>Volume measurement<br>Range: 0 - 10 L<br>Accuracy: 50 mL or 2%<br>Ventilation<br>Range: 0 - 300 L/min<br>Accuracy: 0.05 L/min or 2%<br>V'O2, V'CO2<br>Range: 0 - 7 L/min<br>Accuracy: 0.05 L/min or 3%<br>RER<br>Range: 0.6 - 2.0<br>Accuracy: 4% | Flow measurement<br>Range: 0 - 15 L/s<br>Accuracy: 70 mL/s or 3%<br>Volume measurement<br>Range: 0 - 10 L<br>Accuracy: 50 mL or 2%<br>Ventilation<br>Range: 0 - 300 L/min<br>Accuracy: 0.5 L/min or 2%<br>V'O2, V'CO2<br>Range: 0 - 7 L/min<br>Accuracy: 0.05 L/min or 3%<br>RER<br>Range: 0.6 - 2.0<br>Accuracy: 4% | Identical |
| | Hi/Lo Fi O2<br>Premixed inspiratory gases up to<br>100% O2 from manually filled bag | | Identical<br>(to Reference K992214) |
| Sterilization | The device along with its<br>accessories is neither supplied<br>sterile nor intended to be sterilized | The device along with its accessories is<br>neither<br>supplied sterile nor intended to be<br>sterilized | Identical |
# Comparison to predicate devices SentrySuite Product Line Predicate K150810 &
{6}------------------------------------------------
{7}------------------------------------------------
## Summary of technological characteristic compared to the predicate / reference device to the table above:
- The new Vyntus/SentrySuite Product Line with option Hi/Lo FiO2 is identical in indication for use compared to the ● predicate Vyntus/SentrySuite Product Line K150810.
- The patient population for the proposed device is identical to the predicate device Vyntus/SentrySuite Product Line with K150810.
- . The Software platform for the proposed device is identical to the predicate Vyntus/SentrySuite Product Line K150810. Both devices use the software SentrySuite and differ only in software version for the proposed device is the SentrySuite version 3.0.
- . The measuring programs & accuracy of the proposed device are identical to the predicate device. Except the new measurement Hi/Lo FiO2 is identical to the reference device Oxycon Pro K992214.
#### XV. Summary of Device Testing
## Non-clinical tests conducted for determination of substantial equivalence:
| | Characteristic | Standard/Test | Results Summary |
|----|---------------------|-------------------------|---------------------------------------------------------------|
| 1. | Basic Safety | IEC 60601-1 | The proposed device passes the applicable tests and standards |
| 2. | EMC Compatibility | IEC 60601-1-2 | The proposed device passes the applicable tests and standards |
| 3. | Risk Management | ISO 14971 | The proposed device passes the applicable tests and standards |
| 4. | Usability | EN 62366 | The proposed device passes the applicable tests and standards |
| 5. | Software life cycle | ISO 62304 | The proposed device passes the applicable tests and standards |
| 6. | Biocompatibility | ISO 10993-1 | The proposed device passes the applicable tests and standards |
| 7. | Accuracy Testing | Accuracy of measurement | The proposed device passes the applicable tests and standards |
## Summary Discussion of Bench Performance Data
The Vyaire Medical GmbH Vyntus SentrySuite Product Line with the option Hi/Lo FiO2 passed all specified test requirements. The validation and verification testing confirmed this device with the option Hi/Lo FiO2 meets user needs and design inputs for PFT and CPET.
Testing also confirmed physical attributes and device performance meet the standards listed in the performance testing summary above. These standards address basic safety, EMC compatibility, software life cycle and biocompatibility. All testing which have been performed demonstrate substantial equivalence to the predicate devices.
## Clinical tests conducted for determination of substantial equivalence and/or of clinical information:
Clinical Performance Data/Information:
Clinical testing was not performed with this device.
## Conclusion drawn from non-clinical and clinical data:
The Vyaire Medical GmbH CPET system with the new option HilLo FiO2 meets the functional claims and intended in the product labeling. The Vyntus/SentrySuite Product Line is substantially equivalent to the predicate devices described in the submission.
#### XVI. Conclusion
Based on the above. Vyaire Medical GmbH concludes that the Vyntus/SentrySuite Product Line with the new option Hi/Lo FiO2 is substantially equivalent to the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.