The ECCOVISION™ is intended to measure the upper respiratory airway by means of acoustic reflection.
Device Story
Eccovision measures upper respiratory airway patency using acoustic reflection technology. System includes control unit, wave tubes (pharyngometer/rhinometer), electronic platform, mouthpieces, and nose tips. Device connects to customer-provided PC running Eccovision application software. Operation involves passing acoustic signals through a probe positioned in the mouth or nose; two microphones in the wave tube capture reflected signals. Software processes incident and reflected sound signals to generate area-distance curves, minimal cross-sectional area, and volume measurements. Used in clinical settings to provide objective airway data; output displayed on screen or printed. Healthcare providers use these measurements to assess airway dimensions and patency, aiding in clinical decision-making regarding respiratory health.
Clinical Evidence
Bench testing only. Validation included electrical safety testing (IEC 60601-1, IEC 60601-1-2), black/white box software testing, and comparative analysis against the legacy core. Oral and nasal measurement accuracy was validated using statistical methods including scatter plots, regression analysis, Bland-Altman analysis, ANOVA, and descriptive statistics to demonstrate volume measurement acceptance criteria within 10%.
Technological Characteristics
Acoustic reflection measurement system. Components: control unit, wave tubes, electronic platform, mouthpieces (Santoprene 281-55), nose tips (Silicone G2705). Connectivity: connects to customer-provided PC. Software: Windows-based application with GUI. Electrical safety: IEC 60601-1, IEC 60601-1-2. Sterilization: uses filter strips.
Indications for Use
Indicated for measuring the upper respiratory airway via acoustic reflection in patients requiring airway patency assessment.
Regulatory Classification
Identification
A rhinoanemometer is a device used to quantify the amount of nasal congestion by measuring the airflow through, and differential pressure across, a patient's nasal passages.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 9, 2017
Sleep Group Solutions % Stephen W. Inglese Owner, CEO Ouality Solutions and Support, LLC Po Box 8271 Holland, MI 49422
Re: K170071
Trade/Device Name: Sgs Eccovision™ Pharyngometer, Sgs Eccovision™ Rhinometer, Sgs Eccovision™ Rhino/Pharyngometer
Regulation Number: 21 CFR 868.1800 Regulation Name: Rhinoanemometer Regulatory Class: Class II Product Code: BXO Dated: September 25, 2017 Received: October 3, 2017
Dear Stephen W. Inglese:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely yours,
# Srinivas Nandkumar -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use Statement 4.0
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
51 O(k) Number (if known) K170071
Device Name ECCOVISIONTM
Indications for Use (Describe)
The ECCOVISION™ is intended to measure the upper respiratory airway by means of acoustic reflection.
Type of Use (Select one or both, as applicable)
IZI Prescription Use (Part 21 CFR 801 Subpart D)
O Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information, Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
PSC .Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
#### 5.0 510(k) Summary
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the
| Device Common Name: | Rhinoanemometer and Pharyngometer |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Proprietary Name: | ECCOVISION™ |
| Submitter: | Sleep Group Solutions<br>2035 Harding Street, #200/201<br>Hollywood, Florida 33020<br>Phone: 305-830-0327 |
| Contact: | Stephen Inglese<br>Consultant<br>Quality Solutions and Support, LLC<br>Phone: 561-251-0876<br>Email: swi@qss-llc.com |
| Date Prepared: | December 11, 2016 |
| Classification Regulation: | 21 CFR §868.1800 Rhinoanemometer<br>Class II |
| Panel: | Anesthesiology |
| Product Code: | BXQ |
| Predicate Device: | K011329 – SGS Eccovision Rhinometer and<br>K921452 - SGS Eccovision Pharyngometer<br>(Submitter's previously cleared devices) |
| Description of Device | The Eccovision™ device is used to obtain an<br>objective measurement of the upper respiratory<br>airway. The device uses acoustic signal processing<br>technology to provide graphical representation of<br>the airway patency as a function of distance from<br>the airway opening. |
| Indication for Use: | To measure the upper respiratory airway by means |
| | of acoustic reflection |
{4}------------------------------------------------
The system consists of a control unit (which connects to customer owned personal computer), and software application, wave tube (one each for the Pharyngometer and Rhinometer) and electronic platform, mouthpieces and nose tips and filter strips. The device performs a dynamic test that determines the dimension of the oral airway past the glottis while the patient is breathing thorough either the mouthpiece or nose tip. A customer provided computer with the loaded Eccovision™ application software then processes the incident and reflected sound signals provides an area-distance curve representing the lumen together with minimal crosssectional area and volume. A measurement is obtained by passing a signal along a probe positioned in the mouth or nose then recover a signal by use of two (2) microphones in the wave tube. The signal is processed by the software and displayed on a screen or relayed to a printer, detailing the cross-sectional area of the
airway as a function of distance from the teeth.
### Summary of Technological Characteristics
The Eccovision™ device has been designed and manufactured with the same indications for use as the predicate devices. The technical changes apply to the software. The application language being changed from DOS to Windows. This along with minor hardware changes to the Controller allows for a more efficient device.
The testing performed on the device is described as follows:
- 1. Summary: Test 1
- □ Certificate of Electrical Safety for IEC 60601-1-2
- I ITL Product Testing Certificate of Compliance: K170460.00 -Electromechanical Safety
- 2. Summary: Test 2
□ Certificate of Electrical Safety for IEC 60601-1
{5}------------------------------------------------
- □ C170450.01 General Requirements Basic Safety and Essential Performance
- 3. Summary: Test 3
(Oral and Nasal testing was conducted separately)
#### Test description
- D Each test case specifies the test pre-conditions, steps & post conditions
- □ Test traceability i.e. success or deviations will be recorded in Data sheet & Deviation log sheet in the Appendix section
#### Approach
To show that the new core complies with the specification the following approach will be applied
- □ Black box testing focuses on external interfaces
- □ White box testing focuses on internal interfaces
- 4. Summary: Test 4
#### Purpose
This protocol outlines design validation testing to show that the new (Eccovision) system is equivalent to the cleared predicate system. Testing will be performed by an "Evaluator" who is familiar with the business case
#### Approach
To show equivalency between cleared predicate system & new (Eccovision) system the following approach will be applied
- □ Show equivalency between new core & legacy core using a model (fixed ADP)
- D Show that the new system is equivalent to legacy system on real patients
- 5. Summary Test 5
#### Purpose
This protocol outlines design validation testing for the Eccovision UI. Testing will be performed by an "Evaluator" who is familiar with the business case
### Approach
Run a series of test cases made of most common UI use case with known expected results and capture the actual results
{6}------------------------------------------------
- Testing To Prove Acceptance Criterion: Oral and Nasal 6.
To demonstrate the justification of the acceptance criterion of the volume measurements within 10%, analysis was performed on both the Oral and Nasal aspects of the device.
### Analysis Events - Oral
Reports C1 - C7 Demonstrates the oral analysis performed:
- C1 Oral Analysis Report Summary RPT 0180
- C2 Oral Scatter Plots RPT 0100
- C3 Oral Regular Regression Analysis RPT 0110
- C4 Oral Multi Variate Mix Regression Analysis RPT 0120
- C5 Oral Bland Altman Analysis RPT 0140
- C6 Oral Anova RPT 0150
- C7 Oral Descriptive Statistics Analysis RPT 0160
#### Analysis Events - Nasal
Reports C8 - C14 Demonstrates the nasal analysis performed:
- C8 Nasal Analysis Report Summary RPT 0280
- C9 Nasal Scatter Plots RPT 0200
- C10 Nasal Regular Regression Analysis RPT 0210
- C11 Nasal Multi-Variate Mix Regression Analysis RPT 0220
- C12 Nasal Bland Altman Analysis RPT 0240
- C13 Nasal Anova RPT 0250
- C14 Nasal Descriptive Statistics Analysis RPT 0260
### Substantial Equivalence:
Based on the identical indication, similar technological characteristics, and results of performance testing, the Eccovision System is substantially equivalent to the previously cleared in both K011329 and K921452 demonstrated in Table 1:
#### Table 1: Device Comparison Table
| | Proposed Device | Predicate Device | Predicate Device |
|------------------------------|-------------------------------------------------------|----------------------------------------------|----------------------------------------------|
| 510(k) Number | TBD – Eccovision™<br>(Pharyngometer or<br>Rhinometer) | K921452 – SGS<br>Pharyngometer | K011329 – SGS<br>Rhinometer |
| Submitter | Sleep Group Solutions | Sleep Group Solutions -<br>Hood Laboratories | Sleep Group Solutions -<br>Hood Laboratories |
| Classification<br>Regulation | 868.1800 | 868.1800 | 868.1800 |
| Product Code | BXQ | BXQ | BXQ |
{7}------------------------------------------------
| System Function to<br>capture data | Both Mouth piece or Nose<br>tip via the Wave Tube | Mouth piece via the Wave<br>Tube | Nose tip via the Wave<br>Tube |
|--------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| Indication | Intended to measure the<br>upper respiratory airway<br>by acoustic reflection | Intended to measure the<br>upper respiratory airway<br>by acoustic reflection | Intended to measure<br>the upper respiratory<br>airway by acoustic<br>reflection |
| Computer<br>Requirements -<br>Hardware and<br>Software (Operating<br>System, device<br>application software) | A customer owned<br>computer with the<br>Eccovision software<br>application loaded by the<br>customer. | A provided computer<br>with installed device<br>application software | A provided computer<br>with installed device<br>application software |
| Control Unit -<br>Hardware | Provided hardware which<br>connects to the Computer.<br>Modified with the redesign<br>of the PCB (Print Circuit<br>Board) configuraton | Provided hardware<br>which connects to the<br>Computer. | Provided hardware<br>which connects to the<br>Computer |
| Wave Tube - Hardware | • Supporting the<br>Pharyngometer -<br>Connects to the<br>Electronic Platform<br>• Supporting the<br>Rhinomenter - Connects<br>to the Control Unit | Supporting the<br>Pharyngometer -Connects<br>to the Electronic Platform | Supporting the<br>Rhinomenter -<br>Connects to the<br>Control Unit |
| Electronic Platform -<br>Hardware | Supporting the<br>Pharyngometer - Connects<br>to the Control Unit | Supporting the<br>Pharyngometer – Connects<br>to the Control Unit | NA |
| Application Software<br>- Language | Windows with GUI | DOS | DOS |
| User Manual –<br>General Instructions | Instructions to support<br>combined Oral<br>(Pharyngometer) and<br>Nasal (Rhinometer)<br>airway data capture | Instructions to support<br>Oral (Pharyngometer)<br>airway data capture | Instructions to<br>support Nasal<br>(Rhinometer) airway<br>data capture |
| Filters | The Wave Tube filter<br>design and material<br>remains the same as<br>originally indicated | 3M polypropylene filter | NA |
| Mouth Piece Design<br>and Material | The Mouth piece design<br>and material remains the<br>same as originally<br>indicated. | Mouth piece design<br>developed with<br>Santoprene®, 281-55<br>thermoplastic rubber<br>medical grade. | NA |
| Nose Tip Design and<br>Material | The Nose tip design and<br>material remains the same<br>as originally indicated. | NA | Silicone G2705 from<br>Kraton |
Changes are noted in BOLD
{8}------------------------------------------------
## Substantial Equivalence Summary
The results of the testing as demonstrated in Test 1 through Test 5 of electrical safety, application (supporting both hardware and software) and proven acceptance criterion testing reports C1 - C8 for the oral functionality and reports C9 - 14 for the nasal functionality of the Eccovision System that the device is safe and as effective, and performs as or better than the predicate devices cleared in K011329 and K921452.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.