EASYONE PRO RESPIRATORY ANALYSIS SYSTEM

K091428 · Ndd Medizitechnik AG · BTY · Jan 27, 2010 · Anesthesiology

Device Facts

Record IDK091428
Device NameEASYONE PRO RESPIRATORY ANALYSIS SYSTEM
ApplicantNdd Medizitechnik AG
Product CodeBTY · Anesthesiology
Decision DateJan 27, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.1890
Device ClassClass 2
AttributesPediatric

Indications for Use

The ndd EasyOne Pro Respiratory Analysis System is designed for conducting lung function measurements in general or specialist practices or in hospitals. The EasyOne Pro Respiratory Analysis System can also be used outside of the laboratory when performing lung function screenings or measurements in occupational medicine. The EasyOne Pro Respiratory Analysis System is used to conduct slow and forced spirometry on adults and children starting at age 4. Measurement of Diffusing Capacity of the lung based on CO (DLCO) can be performed on adults and children starting at age 6.

Device Story

Portable pulmonary function testing system; performs spirometry and DLCO measurements. Inputs: respiratory airflow/gas samples via single-patient-use spirette/barriette. Operation: main unit with integrated touchscreen; hand-held sensor; valve unit for DLCO gas supply. Used in clinics, hospitals, and occupational medicine settings by healthcare professionals. Device computes lung function parameters compared against predicted normal values based on patient age, gender, and height. Built-in quality control ensures test performance. Output: diagnostic lung function data used by clinicians to assess respiratory health. Benefits: enables portable, accurate screening and diagnostic testing for obstructive/restrictive lung conditions.

Clinical Evidence

Bench testing only. Dynamic waveform testing performed for spirometry; DLCO performance validated using DLCO simulator. Comparative measurements against another DLCO testing device confirmed adherence to ATS accuracy and precision recommendations. Electrical safety verified per IEC 60601-1 and IEC 60601-1-2. Biocompatibility confirmed per ISO 10993.

Technological Characteristics

Portable system; main unit with touchscreen and hand-held sensor. Materials: biocompatible (ISO 10993). Connectivity: standalone or networked. Power: electrical. Software: performs automated lung function calculations based on standard predictive equations. Quality control: built-in automated checks.

Indications for Use

Indicated for lung function measurements, including slow and forced spirometry in adults and children aged 4+, and DLCO measurements in adults and children aged 6+.

Regulatory Classification

Identification

A predictive pulmonary-function value calculator is a device used to calculate normal pulmonary-function values based on empirical equations.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K091425 ## JAN 2 7 2010 #### 510(k) Summary 5 In accordance with 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence. #### Submitter Information: | Submitter name: | ndd Medizintechnik AG<br>Technoparkstrasse 1<br>8005 Zurich, Switzerland | |----------------------------------|--------------------------------------------------------------------------| | United states<br>contact person: | Mr. Jerry Masiello<br>Operations Manager | | | ndd Medical Technologies, Inc | | Phone: | (877) 904.0090 x11 | | Fax: | (978) 824.3853 | | Mailto: | jmasiello@nddmed.com | | Date prepared: | May 11, 2009 | #### Device Name: | Proprietary name: | EasyOne Pro Respiratory Analysis System | |----------------------|--------------------------------------------------| | Common name: | Pulmonary function testing device | | Class: | Class II according to 868.1890 | | Classification name: | Calculator, Predicted Values, Pulmonary Function | | Product code: | BTY | #### Predicate Device: Substantial Equivalence is claimed with the SensorMedics Vmax 229 Series Pulmonary Function Analysis Instrument, K942211. The ndd Medical Technologies EasyOne Pro Respiratory Analysis System is a successor of the EasyOne Spirometer, K993921. Compared to the EasyOne Spirometer the EasyOne Pro Respiratiory Analysis System provides DLCO (Diffusion Capacity) options in addition to spirometry tests. #### Device Description: The ndd EasyOne Pro Respiratory Analysis System is a portable device for performing lung function measurements such as spirometry and diffusion capacity (DLCO) tests. Measured lung function parameters can be compared to predicted normal values that are computed based on gender, age, and body height of the patient; the applied equations are available in literature and/or published standards. The device consists of a compact main unit and a hand-held sensor. Spirometry tests can be performed by connecting the hand-held sensor to the main unit. In order to perform DLCO tests a valve unit must additionally be connected to the hand-held sensor. The valve unit is connected with the main unit with a gas supply tube. {1}------------------------------------------------ The spirette respiratory tube is a mouthpiece for single patient use, which is inserted in the sensor. In the valve unit there is also an accessory for single patient use inserted, which is called barriette. The barriette prevents the passage of microorganisms into EasyOne Pro Respiratory Analysis System. A touch screen is integrated in the main unit that provides the user interface. The main power switch, different connections as well as different ports are located on the rear panel of the main unit. The EasyOne Pro can be used as a stand-alone system and can be connected to a network. The device has built-in quality control to assure correct test performance and equipment function. #### Intended Use: The ndd EasyOne Pro Respiratory Analysis System is designed for conducting lung function measurements in general or specialist practices or in hospitals. The EasyOne Pro Respiratory Analysis System can also be used outside of the laboratory when performing lung function screenings or measurements in occupational medicine. The EasyOne Pro Respiratory Analysis System is used to conduct slow and forced spirometry on adults and children starting at age 4. Measurement of Diffusing Capacity of the lung based on CO (DLCO) can be performed on adults and children starting at age 6. #### Comparison of Technological Characteristics: The device has the same technological characteristics as the predicate devices. Testing was conducted to demonstrate that lung function measurements are as accurate and precise as the methods used in predicate devices. The new as well as the predicate devices meet the ATS recommendations for lung function measurements. The EasyOne Pro Respiratory Analysis System is therefore substantially equivalent to the predicate devices. #### Summary of Testing: Dynamic wave form testing was performed regarding the diagnostic spirometry tests. The DLCO performance was tested by using a DLCO simulator. In addition comparative measurements with another DLCO testing device demonstrate that the EasyOne Pro Respiratory Analysis System meets the ATS recommendations for accuracy and precision for DLCO testing and the intended diagnostic spirometry tests. The device was tested to demonstrate conformance with IEC 60601-1 and IEC 60601-1-2 requirements for electrical safety. The materials used meet the requirements for biocompatibility in accordance with ISO 10993. Software verification and validation revealed that the EasyOne Pro Respiratory Analysis System meets the specified criteria. #### Conclusion: Based on the above, ndd Medical Technologies concluded that the EasyOne Pro Respiratory Analysis System is substantially equivalent to the legally marketed predicate devices and is safe and effective for its intended use. {2}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image contains the text "Public Health Service". The text is in a simple, sans-serif font and is horizontally aligned. The text is likely part of a document or sign. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 # JAN 2 7 2010 ndd Medizintechnik AG C/O Mr. Jerry Masiello Operations Manager ndd Medical Technologies, Incorporated Two Dundee Park Andover, Massachusetts 01810 Re: K091428 Trade/Device Name: EasyOne Pro Respiratory Analysis System Regulation Number: 21 CFR 868.1890 · Regulation Name: Predictive Pulmonary-Function Value Calculator Regulatory Class: II Product Code: BTY Dated: January 21, 2010 Received: January 27, 2010 Dear Mr. Masiello: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings or feathers. The eagle faces to the left. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle. Page 2 - Mr. Masiello Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/cdrh/comp/ for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, h for Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use Statement 510(k) Number (if known): K091428 EasyOne Pro Respiratory Analysis System Device Name: Indications for Use: The ndd EasyOne Pro Respiratory Analysis System is designed for conducting lung function measurements in general or specialist practices or in hospitals. The EasyOne Pro Respiratory Analysis System can also be used outside of the laboratory when performing lung function screenings or measurements in occupational medicine. The EasyOne Pro Respiratory Analysis System is used to conduct slow and forced spirometry on adults and children starting at age 4. Measurement of Diffusing Capacity of the lung based on CO (DLCO) can be performed on adults and children starting at age 6. Prescription Use _ × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) L. Schultheis (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K091428
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...