WALLGRAFT TRACHEOBRONCHIAL ENDOPROTHESIS WITH UNISTEP PLUS DELIVERY SYSTEM
K021901 · Boston Scientific Scimed, Inc. · JCT · Jul 9, 2002 · General, Plastic Surgery
Device Facts
Record ID
K021901
Device Name
WALLGRAFT TRACHEOBRONCHIAL ENDOPROTHESIS WITH UNISTEP PLUS DELIVERY SYSTEM
Applicant
Boston Scientific Scimed, Inc.
Product Code
JCT · General, Plastic Surgery
Decision Date
Jul 9, 2002
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.3720
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The WALLGRAFT® Tracheobronchial Endoprosthesis is indicated for use in the treatment of tracheobronchial strictures produced by malignant neoplasms.
Device Story
Self-expanding tracheobronchial endoprosthesis; metallic superalloy monofilament wire stent with radiopaque core; braided polyester (PET) graft material bonded to exterior. Delivered via coaxial Unistep Plus Delivery System; exterior tube constrains stent; interior tube features holding sleeve/stent cup for partial deployment and re-constrainment. Radiopaque markers facilitate fluoroscopic visualization during placement. Used by physicians in clinical settings to maintain airway patency in patients with malignant strictures. Device provides mechanical support to airway; prevents occlusion by tumor growth.
Clinical Evidence
Bench testing only. Testing included catheter length, crossing profile, deployment/re-constrainment forces, stent graft securement, tensile testing of various assembly points, deployed stent graft outer diameter, accelerated aging, and biocompatibility. All results confirmed the modified delivery system meets specifications.
Technological Characteristics
Metallic superalloy monofilament wire stent with radiopaque core; braided polyester (PET) graft. Coaxial delivery system (9F-12F). Radiopaque marker bands for imaging. Self-expanding. Sterilized device. No software/algorithm.
Indications for Use
Indicated for patients with tracheobronchial strictures caused by malignant neoplasms.
Regulatory Classification
Identification
The tracheal prosthesis is a rigid, flexible, or expandable tubular device made of a silicone, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the trachea or trachealbronchial tree. It may be unbranched or contain one or two branches. The metal tracheal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.
Special Controls
The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”
Predicate Devices
WALLGRAFT® Tracheobronchial Endoprosthesis with Unistep™ Delivery System (K000001)
WALLGRAFT® Tracheobronchial Endoprosthesis with Unistep™ Plus Delivery System (K003100)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features an abstract image of an eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV - 3 2006
Boston Scientific Corporation % Ms. Angela Byland Manager, Regulatory Affairs Cardiovascular Two Scimed Place Maple Grove, Minnesota 55311-1566
Re: K021901
Trade/Device Name: Wallgraft® Tracheobronchial Endoprosthesis with Unistep Plus Delivery System
Regulation Number: 21 CFR 878.3720 Regulation Name: Tracheal prosthesis Regulatory Class: II Product Code: JCT Dated: June 7, 2002 Received: June 10, 2002
Dear Ms. Byland:
This letter corrects our substantially equivalent letter of July 9, 2002.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{1}------------------------------------------------
Page 2 - Ms. Angela Byland
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Herbert Lehman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use Statement
| 510(k) Number (if known) | K021901 |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | WALLGRAFT® Tracheobronchial Endoprosthesis with Unistep™ Plus Delivery System |
| Indications For Use | The WALLGRAFT® Tracheobronchial Endoprosthesis is indicated for use in the treatment of tracheobronchial strictures produced by malignant neoplasms. |
Prescription Use:_____________________________________________________________________________________________________________________________________________________________ × (Per 21 CFR §801 Subpart D) OR
Over-The-Counter Use: (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Huld Lume
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
**510(k) Number** \$\underline{K_{02}(90)}\$
{3}------------------------------------------------
K021901 p. lot 3
# 510(k) Summary
| Submitter's<br>Name and<br>Address | Boston Scientific Corporation<br>One Scimed Place<br>Maple Grove, MN 55311 | | |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------|---------|-------------------|
| Contact Name<br>and Information | Angela Byland<br>Manager, Regulatory Affairs<br>Phone: 763-494-2887<br>Fax: 763-494-2981<br>e-mail: bylanda@bsci.com | | |
| Original Date<br>Prepared | June 7, 2002 | | |
| Date Prepared | July 14, 2006 | | |
| Proprietary<br>Name(s) | WALLGRAFT® Tracheobronchial Endoprosthesis with<br>Unistep™ Plus Delivery System | | |
| Common Name | Tracheal Endoprosthesis | | |
| Product Code | JCT | | |
| Classification of<br>Device | Class II, 21 CFR Part 878.3720 | | |
| Predicate Device | WALLGRAFT®<br>Tracheobronchial<br>Endoprosthesis<br>with Unistep™<br>Delivery System | K000001 | June 5, 2005 |
| | WALLGRAFT®<br>Tracheobronchial<br>Endoprosthesis<br>with Unistep™<br>Delivery System | K003100 | December 20, 2000 |
:
B
{4}------------------------------------------------
K021901 f. 2 of 3
Image /page/4/Picture/1 description: The image shows the logo for Boston Scientific. The words "Boston" and "Scientific" are stacked on top of each other. The font is a serif font and the text is black.
### Device Description
The Wallgraft Endoprosthesis with Unistep Plus Delivery System is available in the models indicated in the table below.
A brief description of the stent graft and delivery system components follow.
| Stent Graft<br>Diameter | Stent Graft Length | Delivery System<br>Profile |
|-------------------------|----------------------------|----------------------------|
| 6.0 mm | 20 mm, 30 mm, 50 mm, 70 mm | 9F |
| 7.0 mm | 20 mm, 30 mm, 50 mm, 70 mm | 9F |
| 8.0 mm | 20 mm, 30 mm, 50 mm, 70 mm | 9F |
| 9.0 mm | 20 mm, 30 mm, 50 mm, 70 mm | 10F |
| 10.0 mm | 20 mm, 30 mm, 50 mm, 70 mm | 10F |
| 12.0 mm | 30 mm, 50 mm, 70 mm | 11F |
| 14.0 mm | 50 mm, 70 mm | 12F |
Stent Graft Description
The Wallgraft Endoprosthesis (stent graft) consists of a metallic stent comprised of biomedical superalloy monofilament wire with a radiopaque core, braided in a tubular mesh configuration. A platinum nickel micro-cable is incorporated into the stent to enhance radiopacity of the device. A graft material comprised of braided polyester varn (PET) is adhesively bonded to the outside of the metallic stent. This design configuration results in a stent graft that is flexible, compliant, and self-expanding with the barrier characteristics of a tubular graft.
#### Delivery System Description
The Unistep Plus Delivery System consists of a coaxial tube system. The exterior tube serves to constrain the stent graft over the interior tube until retracted during deployment. The coaxial tubes have the capability of re-constraining the stent graft after partial deployment. A holding sleeve and stent cup, attached to the interior tube, aid in the re-constraining process.
Radiopaque marker bands situated adjacent to the proximal and distal ends of the stent graft facilitate imaging during deployment. A radiopaque marker band located on the exterior tube and a limit marker band on the interior tube function as deployment limit markers. Re-constrainment is possible up to the point where the exterior tube marker band is proximally retracted to the location of the interior tube limit marker.
The interior tube of the coaxial system contains a central lumen which will accommodate a 0.035" guide wire. The delivery system will be available in a single working length of 90 cm.
{5}------------------------------------------------
021901 p. 3 of 3
| Intended Use | The WALLGRAFT® Tracheobronchial Endoprosthesis is<br>indicated for use in the treatment of tracheobronchial<br>strictures produced by malignant neoplasms. |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>Characteristics | The modified 9 F - 11 F Unistep Plus Delivery System of the<br>6 mm to 12 mm Wallgraft Endoprosthesis will be<br>manufactured in a substantially equivalent manner to the<br>currently marketed 12 F delivery system of the 14 mm<br>Wallgraft Tracheobronchial Endoprosthesis with Unistep<br>Plus Delivery System, cleared to market under K003100,<br>December 20, 2000. |
| Performance<br>Testing | Testing was conducted to verify that the modified 9 F - 11 F<br>delivery system met product specifications. The following<br>testing was performed:<br>Total Catheter Length Catheter Crossing Profile Deployment Force Reconstrainment Force Stent Graft Securement Hub to Stainless Steel Tube Tensile Distal Tip Tensile Inner Member Assembly Tensile Valve Body to Exterior Tube Tensile Deployed Stent Graft OD Post Accelerated Aging Testing Biocompatibility All test results verified that the modified 9 F - 11 F delivery<br>system is adequate for its intended use. The modified<br>device is considered substantially equivalent to the currently<br>marketed 9 F - 11 F delivery system (K000001) and the 12<br>F delivery system (K003100). |
| Conclusion | In summary, Boston Scientific Corporation has<br>demonstrated that the WALLGRAFT® Tracheobronchial<br>Endoprosthesis with Unistep™ Plus Delivery System is<br>substantially equivalent based on design, test results, and<br>indications for use to the predicate devices. |
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.