The Embla BreathSensor provides a qualitative air flow signal by oral and/or nasal temperature sensitive resistive components for recording onto a recording system in support of airflow analysis and sleep studies. The Embla BreathSensor does not provide any diagnostic conclusion about the patient's condition to the user. The Embla BreathSensor user is a qualified medical practitioner in a hospital environment who will exercise their professional judgment in using this information.
Device Story
Embla BreathSensors are disposable, non-sterile thermistor-based sensors used in sleep studies (polysomnography). Sensors consist of carbon ink temperature-sensitive elements on a polyester substrate, with silver ink interfaces. Applied between the upper lip and nostrils, they detect temperature changes between ambient and exhaled air to provide a qualitative airflow signal. An interface cable connects the sensor to standard PSG recording equipment; active cable versions perform signal filtering and convert thermistor resistance changes to thermocouple-like signals. Used by medical practitioners in hospital settings to support airflow analysis; the device provides no diagnostic conclusions. Benefits include non-invasive, qualitative monitoring of patient respiration during sleep.
Clinical Evidence
Bench testing only. Evaluations included verification of dimensions, material specifications, biocompatibility (cytotoxicity, sensitization, irritation), base resistance, temperature sensitivity, qualitative airflow response, and electromagnetic compatibility (IEC 60601-1-2:2014).
Technological Characteristics
Thermistor-based airflow sensor. Materials: carbon ink (sensing element), silver ink (conductive interface), polyester substrate, and adhesive tape (Micro-Pore/Micro-Foam). Dimensions vary by model (Adult, Small Adult/Child). Connectivity: wired via interface cable to PSG amplifier. Non-sterile, single-patient use. Active interface cables provide signal filtering and thermistor-to-thermocouple signal conversion using internal battery power.
Indications for Use
Indicated for use by qualified medical practitioners in hospital environments to provide qualitative airflow signals via oral/nasal temperature-sensitive resistive components for recording during sleep studies. Contraindicated for patients with adhesive tape allergies.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}------------------------------------------------
September 27, 2019
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Embla Systems (Embla) Sanjay Mehta Director, Quality Assurance & Regulatory Affairs 1 Hines Road, Suite 202 Kanata, ON, CA K2K 3C7
Re: K180001
Trade/Device Name: Embla BreathSensors Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing frequency monitor Regulatory Class: Class II Product Code: MNR Dated: August 26, 2019 Received: August 29, 2019
Dear Sanjay Mehta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Michael Ryan Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180001
Device Name Embla BreathSensors
#### Indications for Use (Describe)
The Embla BreathSensor provides a qualitative air flow signal by oral and/or nasal temperature sensitive resistive components for recording onto a recording system in support of airflow analysis and sleep studies.
The Embla BreathSensor does not provide any diagnostic conclusion about the patient's condition to the user.
The Embla BreathSensor user is a qualified medical practitioner in a hospital environment only who will exercise their professional judgment in using this information.
Type of Use (Select one or both, as applicable)
| [X] Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|--------------------------------------------------|-----------------------------------------------|
|--------------------------------------------------|-----------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
| FORM FDA 3881 (7/17) | Page 1 of 1 | PSC Publishing Services (301)443-6740 EF |
|----------------------|-------------|------------------------------------------|
|----------------------|-------------|------------------------------------------|
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Natus Neurology. The word "natus" is written in a teal sans-serif font, with a registered trademark symbol next to the "s". Below "natus" is the word "neurology" in a smaller, italicized, black sans-serif font. The logo is simple and clean, with a focus on the company name.
### 510(K): K180001, EMBLA BREATHSENSORS
# K180001 510(k) Summary
510(k) Summary This summary is provided to support the 510(k) pre-market notification for the Embla BreathSensors.
| Company Name: | Embla Systems (Embla)<br>1 Hines Road, Suite 202<br>Kanata, ON, CA, K2K 3C7 |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Contact: | Mr. Sanjay Mehta<br>Director, Quality Assurance & Regulatory Affairs<br>Phone: (905) 287-5055<br>Email: Sanjay.Mehta@natus.com |
| Date Summary Prepared: | September 27, 2019 |
| Trade Name: | Embla BreathSensors |
| Common Name: | Thermistor for Air Flow |
| Classification Name: | Breathing Frequency Monitor<br>21 CFR 868.2375<br>Product Code: MNR<br>Class II |
| Predicate Device: | K913749 EdenTrace Airflow 3171, Sleep Lab Airflow 3170<br>EdenTec Corporation (Primary predicate)<br>K042253 SleepSense Sleep Disorder Sensors<br>S.L.P. Ltd |
#### Product Description
The Embla BreathSensors (Model 970 Series) are airflow thermistors used as an accessory to a polysomnogram (PSG) or sleep study recording system. The BreathSensor family is comprised of two different variations: Models 971 Adult and Model 974 Small Adult/Child.
The BreathSensors are for use with any standard PSG amplifier or recorder. An interface cable connects the BreathSensor to the polysomnography equipment.
Embla BreathSensors are designed to provide a qualitative measure of respirations (flow) from the mouth and/or nose. The user selects the appropriate size of BreathSensor based on their clinical judgment.
The nasal and/or oral temperature sensor elements are carbon ink placed on a substrate. The carbon ink is the temperature sensitive material. Silver ink interfaces with the carbon ink temperature sensing elements to provide a connection to the interface cable.
The Embla BreathSensors are non-sterile, single patient use, disposable devices.
#### Intended Use of the Device
The Embla BreathSensor provides a qualitative air flow signal by oral and/or nasal temperature resistive resistive resistive components for recording onto a recording system in support of airflow analysis and sleep studies. The Embla BreathSensor does not provide any diagnostic conclusion about the patient's condition to the user. The Embla BreathSensor user is a qualified medical practitioner in a hospital environment who will exercise their professional judgment in using this information.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Natus Neurology. The word "natus" is in a teal sans-serif font, and the word "neurology" is in a black italicized sans-serif font. The logo is simple and modern, and it is likely used to represent the company's brand.
# Summary of Technological Characteristics
The BreathSensors are single patient use, disposable sensors applied to a patient below the nose, (above the mouth), to support a qualitative indication of airflow from the nostrils and/or mouth. The BreathSensors are a form of thermistor, indicating airflow by temperature change. The accessory interface cables connect between the BreathSensor and user equipment for display and recording by the user's equipment.
Table 1 provides a side-by-side comparison of the BreathSensors to the predicate device applied to support this premarket notification.
| Table 1: Substantial Equivalence Technical Characteristics | | | | |
|------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | BreathSensors<br>Under Review | EdenTrace Airflow<br>3171, Sleep Lab<br>Airflow 3170<br>(Primary Predicate<br>K913749) | S.L.P. Ltd<br>SleepSense Sleep<br>Disorder Sensors<br>(Predicate K042253) | Equivalence<br>Comments |
| Intended Use | Disposable temperature<br>sensitive sensor to<br>support recording<br>airflow. | Disposable temperature<br>sensitive sensor to<br>support recording<br>airflow. | Disposable temperature<br>sensitive sensor to<br>support recording<br>airflow. | Same |
| Indications for<br>Use | The Embla<br>BreathSensor provides a<br>qualitative air flow<br>signal by oral and/or<br>nasal temperature<br>sensitive resistive<br>components for<br>recording onto a<br>recording system in<br>support of airflow<br>analysis and sleep<br>studies.<br>The Embla<br>BreathSensor does not<br>provide any diagnostic<br>conclusion about the<br>patient's condition to<br>the user.<br>The Embla<br>BreathSensor user is a<br>qualified medical<br>practitioner who will<br>exercise their<br>professional judgment<br>in using this<br>information.<br>Rx only | The airflow signal is<br>provided by oral and/ or<br>nasal temperature<br>sensitive components.<br>Accessory single use<br>airflow sensors are used<br>to provide resistance<br>changes as affected by<br>temperature changes in<br>airflow representing<br>inhaled and exhaled air.<br>Rx only | SleepSense sensors<br>provide a qualitative<br>measure of a patient's<br>physiological<br>parameters for<br>recording onto an FDA-<br>cleared data acquisition<br>system. Their target<br>population: Children<br>and adult patients who<br>are screened during<br>sleep disorder studies.<br>Their environment of<br>use is usually at a sleep<br>laboratory or sometimes<br>at the patient's home.<br>Rx Only | Both the Embla<br>BreathSensors under<br>review and primary<br>predicate Airflow Sensors<br>provide oral and/or nasal<br>temperature sensitive<br>components.<br>The population of use of<br>the Embla BreathSensors<br>under review and primary<br>predicate EdenTec Airflow<br>Sensors are similar.<br>The variations of the<br>primary predicate Airflow<br>Sensors and Embla<br>BreathSensors under<br>review are identified by a<br>Model variation name<br>(Adult, Small Adult /<br>Child).<br>All three devices apply<br>interface cables as<br>applicable to condition the<br>signal for medical<br>professional user's<br>recording equipment.<br>All Rx Only<br>The Embla BreathSensors<br>under review provide<br>additional detail with<br>regard to the indications. |
| Contraindication | Contraindicated for use<br>on patients who exhibit | None | None | The allergic reaction to<br>tape adds clarification on |
| Table 1: Substantial Equivalence Technical Characteristics | | | | |
| Feature | BreathSensors<br>Under Review | EdenTrace Airflow<br>3171, Sleep Lab<br>Airflow 3170<br>(Primary Predicate<br>K913749) | S.L.P. Ltd<br>SleepSense Sleep<br>Disorder Sensors<br>(Predicate K042253) | Equivalence<br>Comments |
| | allergic reactions to<br>adhesive tape. | | | use. |
| Environment of<br>Use | Hospitals | Hospitals, clinics and<br>home. | Hospitals, clinics (sleep<br>labs) and home. | The Embla Breath Sensors<br>have a hospital<br>environment of use. |
| Sensor<br>variations | BreathSensor<br>Adult, Small Adult /<br>Child | Airflow Sensor<br>Adult, Small Adult,<br>Child, Infant, Preemie | Disposable Thermal<br>Flow Sensor<br>Adult, Child, Extra<br>Small | Similar<br>The medical professional<br>selects the size of<br>BreathSensor that best fits<br>their patient with regard to<br>placement of nasal and oral<br>BreathSensor elements. |
| Temperature<br>Sensing<br>Elements | Breath Sensor<br>Nasal and Oral: Adult,<br>Small Adult / Child, | Airflow Sensor<br>Nasal and Oral: Adult,<br>Small Adult, Child,<br>Infant<br>Oral only: Preemie | Disposable Flow Sensor<br>Nasal and Oral, Oral<br>only options: Adult,<br>Child, Extra Small | Similar<br>The primary predicate<br>variation names of Small<br>Adult and Child are<br>renamed to one variation. |
| BreathSensor:<br>Single patient<br>use, disposable | Yes | Yes | Yes | Same |
| Qualitative<br>temperature<br>output | Yes | Yes | Yes | Same<br>The signal output from the<br>BreathSensor is a<br>qualitative signal based<br>upon the difference in<br>ambient and exhaled air.<br>The medical professional<br>will adjust signal amplitude<br>on the recording<br>equipment. |
| Duration of use | Duration of a sleep<br>study, anticipated to be<br>up to 12 hours | Duration of a sleep<br>study | Duration of a sleep<br>study | Same |
| Patient applied<br>location | Between upper lip and<br>nostrils | Between upper lip and<br>nostrils | Between upper lip and<br>nostrils | Same |
| Sensor materials | Carbon Ink<br>Silver Ink<br>Polyester<br>Tape | Carbon Ink<br>Silver Ink<br>Polyester (Mylar)<br>Tape | Unknown | Carbon ink is the<br>temperature sensitive<br>material. The conductive<br>silver ink conducts the<br>signal to the interface cable<br>input connector. The<br>carbon and silver ink are<br>placed on a polyester<br>substrate. The polyester<br>substrate and ink are |
| Table 1: Substantial Equivalence Technical Characteristics | | | | |
| Feature | BreathSensors<br>Under Review | EdenTrace Airflow<br>3171, Sleep Lab<br>Airflow 3170<br>(Primary Predicate<br>K913749) | S.L.P. Ltd<br>SleepSense Sleep<br>Disorder Sensors<br>(Predicate K042253) | Equivalence<br>Comments |
| | | | | covered by tape. |
| Patient contact<br>material | Micro-Pore,<br>Micro-Foam Tape | Micro-Pore Tape | Unknown | Equivalent<br>Final finished form<br>BreathSensors have been<br>evaluated for cytotoxicity,<br>irritation and sensitization<br>with passing results. |
| Sterility | Supplied non-sterile | Supplied non-sterile | Supplied non-sterile | Same |
| Accessory<br>Interface cables | Yes<br>Active | Yes<br>Passive and active | Yes<br>Passive and active | Similar<br>Active interface cable<br>variations are available.<br>The interface cables<br>connect to the medial<br>professional's amplifier,<br>recording equipment,<br>supporting transfer of the<br>BreathSensor thermistor<br>signal to their equipment. |
| Accessory<br>Interface cable:<br>Active Cable | Yes<br>Battery powered,<br>filtering and conversion<br>of sensor thermistor to<br>thermocouple sensor<br>signal. | Yes<br>Battery Powered,<br>filtering and conversion<br>of sensor thermistor to<br>thermocouple sensor<br>signal. | Yes<br>Battery Powered,<br>filtering and conversion<br>of sensor thermistor to<br>thermocouple sensor<br>signal. | Equivalent<br>The active interface cables<br>have sensor and user<br>recording equipment<br>compatible connectors and<br>are battery powered<br>(internal).<br>The active interface cables<br>are battery-powered by an<br>internal non-user<br>replaceable battery. The<br>active interface cables<br>provide filtering and<br>conversion of the<br>thermistor sensor signal to<br>thermocouple sensor signal<br>for presentation to the<br>user's recording equipment |
| Accessory<br>interface cable:<br>Reusable | Yes | Yes | Yes | Same |
| Connection to<br>user's<br>equipment | Yes<br>Accessory Interface<br>Cable | Yes<br>Accessory Interface<br>Cable | Yes<br>Accessory Interface<br>Cable | Same |
| Interface cable<br>variations | Active | Active and passive | Active and passive | Similar<br>The active interface cables<br>interface the BreathSensor |
| Table 1: Substantial Equivalence Technical Characteristics | | | | |
| Feature | BreathSensors<br>Under Review | EdenTrace Airflow<br>3171, Sleep Lab<br>Airflow 3170<br>(Primary Predicate<br>K913749) | S.L.P. Ltd<br>SleepSense Sleep<br>Disorder Sensors<br>(Predicate K042253) | Equivalence<br>Comments |
| | | | | to recording equipment. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Natus Neurology. The word "natus" is in a teal color and in a larger font. Below that, the word "neurology" is in a smaller, black font. There is a registered trademark symbol to the right of the word "natus".
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Natus Neurology. The word "natus" is in a teal color and in a larger font. Below that, the word "neurology" is in a smaller, black font. There is a registered trademark symbol to the right of the word "natus".
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Natus Neurology. The word "natus" is written in a teal, sans-serif font. Below that, the word "neurology" is written in a black, sans-serif font. The logo is simple and modern.
As noted, the differences do not affect substantial equivalence of safety and effectiveness:
- A contraindication is added for patients who may have allergic reactions to adhesive tape. .
- . The BreathSensors are supported by available accessory interface cable(s).
## Performance Tests to Demonstrate Substantial Equivalency
To establish the technical equivalency of the BreathSensors, evaluations were conducted to confirm compliance with performance requirements. Table 2 summarizes evaluations and results supporting substantial equivalence.
| Table 2: Summary of Evaluations Supporting Substantial Equivalence | | |
|--------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|--------|
| Feature | Evaluation Summary | Result |
| Dimensions | Verification of dimensions. | Pass |
| Labels and Labeling | Verification of label and labeling content for variations. | Pass |
| Material | Verification of specified materials. | Pass |
| Biocompatibility of patient<br>contact materials | Supported with final finished form cytotoxicity, sensitization and irritation<br>evaluations. | Pass |
| Resistance | Verify BreathSensor base resistance. | Pass |
| Temperature Sensitivity | Verify minimum temperature sensitivity of resistance change per ℃. | Pass |
| Performance Testing | Comparison with predicate Disposable Flow Sensor, evaluation of qualitative airflow<br>response with Interface Cable. | Pass |
| Electromagnetic<br>compatibility | BreathSensor and Interface Cable compliance with IEC 60601-1-2:2014,<br>4th edition | Pass |
#### Conclusion
The Embla BreathSensors meet performance requirements. The intended use and technology of the Embla BreathSensors are similiar to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.