The AUDICOR System when used with AUDICOR Sensors on the chest wall and properly attached Holter unit, is intended for use on adults 18 years of age and older in acquiring, analyzing and reporting ECG, heart sound, Sleep Disordered Breathing (SDB) and snoring detection and to provide interpretation of the data in an integrated report for consideration by physicians. The SDB analysis and reporting is intended for use on adult patients only as a screening device for obstructive or mixed apnea to determine the need for evaluation by polysomnography based on the patient's score. Patients screened for SDB should have periods of sleep of at least 4 hours duration during which the ECG is predominantly sinus rhythm in nature. The AUDICOR recording may be obtained at any location specified by a physician including home, hospital or clinic. The device is intended for use only under the direct supervision of a physician.
Device Story
AUDICOR System is a wearable, battery-powered ambulatory monitor capturing ECG, heart sounds, snoring, and activity level via chest-wall sensors and tri-axial accelerometer. Device records up to 48 hours of data; software analyzes signals to detect heart rate, arrhythmias (e.g., atrial fibrillation), heart rate variability, snoring, and sleep-disordered breathing (SDB). Clinicians review automatically detected events and measurements via integrated reports, modifying interpretations as needed. Used in home, hospital, or clinic settings under physician supervision. SDB detection utilizes accelerometer waveform signals to calculate SDB scores, aiding clinical decision-making regarding polysomnography referral. Benefits include non-invasive screening for apnea and cardiac rhythm monitoring.
Clinical Evidence
Prospective clinical study at Taipei Veterans General Hospital. Test set of 77 recordings compared subject device accelerometer-based SDB algorithm against predicate device ECG-based SDB algorithm. Concurrent polysomnography (PSG) served as clinical gold standard for establishing PSG-AHI positive/negative cohorts. Statistical analysis demonstrated substantial equivalence in SDB detection accuracy between subject and predicate devices.
Technological Characteristics
Small, wearable, battery-powered device. Sensors: ECG leads, tri-axial accelerometer (for activity and SDB detection), acoustic sensor (for snoring). Connectivity: Non-volatile memory for data storage and download to analysis module. Standards: EN 60601-1, EN 60601-1-2, ANSI/AAMI EC38, ANSI/AAMI EC57.
Indications for Use
Indicated for adults 18+ as a screening device for obstructive or mixed apnea to determine need for polysomnography. Patients must have at least 4 hours of sleep with predominantly sinus rhythm ECG.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}------------------------------------------------
K131883: 510(k) Notification: Modification to AUDICOR System - CPAM with SDB (Accelerometer Based)
## 510k Summary
As required by 21 CFR, part 807.92
| Submitted By: | Inovise Medical, Inc.<br>8770 SW Nimbus Ave, Suite D<br>Portland, OR 97008-7196<br>Phone 503-431-3821 |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Fax 503-431-3801 |
| Contact: | Earl Anderson |
| | Director, Quality and Regulatory |
| Date Prepared: | June 14, 2013 (Revised April 11, 2014 - Test Summary and Conclusions) |
| Proprietary<br>Name: | Modification to AUDICOR CPAM (Cardiopulmonary Ambulatory Monitor) |
| Common/ Usual<br>Name: | Ambulatory Monitor / Acoustic Cardiograph |
| Classification: | Primary: 868.2375, MNR, Ventilatory Effort Recorder |
| | Secondary: 870.2800, MLO, class II, Electrocardiograph, Ambulatory (with analysis program |
| Performance<br>Standards: | AAMI EC38 and AAMI EC57 |
| Intended Use: | The AUDICOR System when used with AUDICOR Sensors on the chest wall and<br>properly attached Holter unit, is intended for use on adults 18 years of age and<br>older in acquiring, analyzing and reporting ECG, heart sound, Sleep Disordered<br>Breathing (SDB) and snoring detection and to provide interpretation of the data in<br>an integrated report for consideration by physicians. The SDB analysis and<br>reporting is intended for use on adult patients only as a screening device for<br>obstructive or mixed apnea to determine the need for evaluation by<br>polysomnography based on the patient's score. Patients screened for SDB should<br>have periods of sleep of at least 4 hours duration during which the ECG is<br>predominantly sinus rhythm in nature. The AUDICOR recording may be obtained<br>at any location specified by a physician including home, hospital or clinic. The<br>device is intended for use only under the direct supervision of a physician. |
| Device: | The AUDICOR System employs the AUDICOR AM (Home) ambulatory recording |
Description:
device to capture 10-second snapshots or up to 48 continuous hours of physiologic data. ECG, heart sound, sleep-disordered breathing, snoring sounds, and activity level are interpreted from the physiologic signals captured. The AUDICOR System includes software to analyze recorded data, then display and present a summary of patient data over time in a trended format. Notable events are detected and displayed for review by the clinician.
The AUDICOR System analyzes and reports the following parameters:
- Heart rate including bradycardia and tachycardia events .
- Atrial fibrillation .
- ECG beat classification and morphology grouping with user-editing .
- Heart rate variability .
- . Snoring detection
- . Sleep disordered breathing (apnea/hypopnea) events
- . Sleep disordered breathing score
- Activity level .
APR 1 1 2014
{1}------------------------------------------------
- Heart sound and combined ECG/heart sound measurements .
- Heart rate distributions of heart sound parameters .
The clinician can review automatically detected events and measurements, and modify them as well as modify the interpretative statements generated.
The AUDICOR System has been tested to the applicable requirements of the Test Summary: following standards, and shown to comply.
- EN 60601-1 Standard for Medical Electrical Equipment: General Requirements . for Safety
- EN 60601-1-2 Standard for Medical Electrical Equipment Part 1-2: General . Requirements for Safety - Collateral Standard: Electromagnetic Compatibility -Requirements and Tests
- ANSI/AAMI EC38:Ambulatory Electrocardiographs .
- ANSI/AAMI EC57: Test and Reporting Performance Results of Cardiac . Rhythm and ST-segment Measurement Algorithms
Prospective clinical testing was performed to evaluate the performance of the technology modification for SDB detection. A study was designed and undertaken that compares AUDICOR SDB detection performance of the predicate device (K120462) to that of the subject device. The prospective study was undertaken at Taipei Veterans General Hospital in Taipei, Taiwan. The study resulted in development of a testset of 77 recordings which were used for the performance comparison. Physician-overread results from concurrent polysomnography recordings on the test subjects was the clinical gold-standard used for establishing PSG-AHI positive and PSG-AHI negative cohorts. The ECG-based SDB algorithm of the predicate device was applied to the concurrent ECG data recorded from the subjects. The accelerometer-based SDB algorithm of the subject device was applied to concurrent accelerometer data recorded from the subjects. Using these algorithm results, statistical analyses were undertaken to compare the SDB accuracy of the predicate device to that of the subject device. Substantial equivalence of the subject device's algorithm performance to that of the predicate device was demonstrated through this statistical comparison. (see Attachment 6, "Clinical Study Report", "K131883 - Response to reviewer's request for additional information, dated 09/12/2013"). This statistical equivalency in performance is central to the claim of substantial equivalency with the predicate device.
The reporting of SDB detection of the subject device is substantially equivalent to Substantial AUDICOR CPAM with SDB detection (K120462). Any differences between the Equivalence: subiect device and the predicate do not raise new questoins of safety and effectiveness.
The subject device and the predicate device are technologically equivalent in that Technological the devices: Characteristics:
- 1) Are small, wearable, battery-powered devices
- Acquire ECG data from patients using limited ECG leads 2)
- Acquire a sound signal for detection of snoring 3)
- 4) Analyze the signals from an accelerometer to derive activity level
- 5) Use non-volatile memory to store patient data for download to the analysis module of the system
The subject device is different from the predicates in the following ways:
{2}------------------------------------------------
- 1) The predicate device uses ECG waveform signals for SDB detection. Through a technology change, the subject device accomplishes SDB detection using waveform signals obtained from an extant tri-axial accelerometer housed within the Hemo ambulatory recording device. This accelerometer was previously described within K120462 where it as used for activity level assessment only.
- The combination of non-clinical tests outlined above and described in the Conclusion: Declaration of Conformity together with clinical performance testing, also outlined above, demonstrates the subject device is as safe and as effective and performs substantially equivalent to the predicate device noted.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
## Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 11. 2014
Inovise Medical. Incorporated Mr. Earl Anderson Director, Quality and Regulatory 8770 SW Nimbus Avenue, Suite D Portland, OR 97008
Re: K131883
Trade/Device Name: AUDICOR® CPAM with SDB Regulation Number: 21 CFR 868.2375 Regulation Name: Ventilatory Effort Recorder Regulatory Class: II Product Code: MNR. MLO Dated: March 5. 2014 Received: March 6, 2014
Dear Mr. Anderson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. Iabeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/4/Picture/7 description: The image contains a logo or seal with some text. The text includes "Tejashri Purohit-Sheth, M.D.," "Clinical Deputy Director," and "DAGRID." There is also the word "FOR" at the bottom right of the image. The image appears to be a signature or approval stamp.
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
## Device Name: AUDICOR® CPAM with SDB
Indications For Use:
The AUDICOR System when used with AUDICOR Sensors on the chest wall and properly attached Hotter unit, is intended for use on adults 18 years of age and older in acquiring, analyzing and reporting ECG, heart sound, sleep-disordered breathing (SDB) and snoring detection and to provide interpretation of the data in an integrated report for consideration by physicians.
The SDB analysis and reporting is intended for use on adult patients only as a screening device for obstructive or mixed apnea to determine the need for evaluation by polysomnography based on the patient's score. Patients screened for SDB should have periods of sleep of at least 4 hours duration during which the ECG is predominantly sinus rhythm in nature.
The AUDICOR recording may be obtained at any location specified by a physician including home, hospital or clinic. The device is intended for use only under the direct supervision of a physician.
Prescription Use XXX (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anya C. Harry Page 1 of 1
Image /page/5/Picture/14 description: The image shows the letters FDA in a pixelated style. The letters are large and take up most of the image. The letters are in a sans-serif font. The image is in black and white.
2014.04.11 12:58:52 -04'00'
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.