Life Scope PT BSM-1700 Series, AY Series, Data Acquisition Unit, LIFE SCOPE BSM 6000 SERIES BEDSIDE MONITORING SYSTEM, Nihon Kohden CSM-1901 BEDSIDE MONITORING SYSTEM
K213316 · Nihon Kohden Corporation · KOI · Dec 29, 2021 · Anesthesiology
Device Facts
Record ID
K213316
Device Name
Life Scope PT BSM-1700 Series, AY Series, Data Acquisition Unit, LIFE SCOPE BSM 6000 SERIES BEDSIDE MONITORING SYSTEM, Nihon Kohden CSM-1901 BEDSIDE MONITORING SYSTEM
Applicant
Nihon Kohden Corporation
Product Code
KOI · Anesthesiology
Decision Date
Dec 29, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.2775
Device Class
Class 2
Indications for Use
The Smart Cable NMT Module and Accessories are indicated for monitoring the relaxation of the patient when neuromuscular blockades are administered. The Smart Cable NMT Module and Accessories are comprised of: - AF-201P NMT Module with Smart Cable - Disposable Electrodes - Main cable - Holder (optional) The Smart Cable NMT Module and Accessories are intended to be used as a system that requires Nihon Kohden compatible electrodes and bedside monitoring systems. The Smart Cable NMT Module and Accessories are intended for use by medical personnel in clinical settings and are available by prescription only.
Device Story
Smart Cable NMT Module (AF-201P) uses EMG technology to monitor neuromuscular transmission; stimulates ulnar nerve via disposable electrode array; records muscle action potential. Input units (BSM-1700/AY Series) acquire physiological signals (ECG, SpO2, NIBP, IBP, CO2, Temp, CO, TOF, BIS) and transfer data to core bedside monitoring units (BSM-6000/CSM-1901). Systems display waveforms/numerics; generate audible/visual alarms for out-of-limit measurements. Used by medical personnel in hospitals (OR, ICU, recovery, emergency). Software update enables input units to recognize Smart Cable NMT Module connection. Output assists clinicians in determining muscle relaxant/anesthetic dosing. Benefits include quantitative assessment of patient relaxation levels.
Clinical Evidence
Bench testing only. Software verification and validation (unit, integration, system-level testing) performed per FDA guidance. Electrical safety and EMC testing conducted per ANSI/AAMI ES60601-1 and IEC 60601-1-2. No clinical data relied upon.
Technological Characteristics
System comprises core unit and input unit (BSM-1700/AY Series). Connectivity via cable or WLAN. Software-based signal processing for physiological parameters (ECG, SpO2, NIBP, IBP, CO2, Temp, CO, TOF, BIS). Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), and IEC 60601-2-40 (electromyographs).
Indications for Use
Indicated for monitoring patient relaxation during neuromuscular blockade administration. Used by medical personnel in clinical settings (OR, recovery, ICU). Compatible with Nihon Kohden bedside monitoring systems and electrodes. Prescription only.
Regulatory Classification
Identification
An electrical peripheral nerve stimulator (neuromuscular blockade monitor) is a device used to apply an electrical current to a patient to test the level of pharmacological effect of anesthetic drugs and gases.
Predicate Devices
Life Scope BSM-6000 Series Bedside Monitors (K201949)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the acronym "FDA" followed by "U.S. FOOD & DRUG ADMINISTRATION" in two lines.
December 29, 2021
Nihon Kohden Corporation % Sunita Teekasingh Sr. Regulatory Affairs Consultant - GSA2 Group LLC Nihon Kohden America. Inc 15353 Barranca Parkway Irvine, California 92618
Re: K213316
Trade/Device Name: Life Scope PT BSM-1700 Series, Data Acquisition Unit, LIFE SCOPE BSM 6000 SERIES BEDSIDE MONITORING SYSTEM, Nihon Kohden CSM-1901 BEDSIDE MONITORING SYSTEM Regulation Number: 21 CFR 868.2775 Regulation Name: Electrical Peripheral Nerve Stimulator Regulatory Class: Class II Product Code: KOI, MHX Dated: November 26, 2021 Received: November 29, 2021
Dear Sunita Teekasingh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpm/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory. ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213316
### Device Name
Smart Cable™ NMT Module and Accessories; Life Scope® BSM-6000 Series Bedside Monitoring Systems; Life Scope® CSM-1901 Bedside Monitoring Systems; Life Scope ® PT BSM-1700 Series and AY Series And Accessories.
### Indications for Use (Describe)
The Smart Cable NMT Module and Accessories are indicated for monitoring the relaxation of the patient when neuromuscular blockades are administered.
The Smart Cable NMT Module and Accessories are comprised of:
- · AF-201P NMT Module with Smart Cable
- · Disposable Electrodes
- · Main cable
- · Holder (optional)
The Smart Cable NMT Module and Accessories are intended to be used as a system that requires Nihon Kohden compatible electrodes and bedside monitoring systems. The Smart Cable NMT Module and Accessories are intended for use by medical personnel in clinical settings and are available by prescription only. ==============================================================================================================================================================================
Life Scope® BSM-6000 Series Bedside Monitoring Systems
The Life Scope® BSM-6000 Series Bedside Monitoring Systems are intended to monitor, display, and record physiological data to provide cardiac and vital signs monitoring within a medical facility. The device is intended to produce a visual record of the electrical signals produced by the heart and monitor the electrocardiogram to generate visible and/or audible alarms when an arrhythmia exists.
The device is also intended to monitor heart rate, blood oxygen saturation (SpO2), non- invasive blood pressure (NIBP), invasive blood pressure (IBP), body temperature, BIS, cardiac output (CO), oxygen concentration (FiO2), carbon dioxide concentration (CO2), EtCO2, respiratory rate and inspired anesthetic agents and anesthetic gases including N2O, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane.
The device may generate an audible and/or visual alarm when a measured rate falls outside preset limits. The device may also be used to condition and transmit physiological signals via radio frequency. The device can interface to external equipment to display numerical and waveform data and alarms from the external devices. Supported external devices include CO2 monitors, BIS monitors, Anesthetic agents/gases detection system, Anesthesia machine, Ventilators, CCO monitors. TOF monitors, CCO/SvO2 Monitors, EEG monitoring device, tcPO2/tcPCO2 monitors, and external devices which output analog voltage signal.
The device will be available for use by medical personnel on patients within a medical facility on all patient populations. The system requires a BSM-6000 core unit with a compatible input unit: AY Series or Life Scope® PT BSM-1700 Series.
Life Scope® CSM-1901 Bedside Monitoring Systems
The Life Scope® CSM-1901 Bedside Monitoring Systems are intended to monitor, display and record physiological data to provide cardiac and vital signs monitoring within a medical facility. The device is intended to produce a visual record of the electrical signal produced by the heart and monitor the electrocardiogram to generate visible alarms when an arthythmia exists. The device is also intended to monitor heart rate, blood oxygen saturation (SpO2),
{3}------------------------------------------------
non-invasive blood pressure (NIBP), invasive blood pressure (IBP), body temperature, BIS, cardiac output (CO), oxygen concentration (FiO2), carbon dioxide concentration (CO2), EtCO2, respiratory rate, inspired and expired anesthetic agents and anesthetic gases including N20, halothane, isoflurane, enflurane and desflurane. The device also displays patient data from external devices such as ventilators. TOF monitors, and EEG measuring unit. The device may generate and audible and/or visual alarm when a measured rate falls outside preset limits. The device will be available for use by trained medical facility on all patient populations, including adult, neonate, infant, child, and adolescent subgroups.
The system requires a CSM-1901 core unit with a compatible input unit: AY Series or Life Scope® PT BSM-1700 Series.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
### Life Scope® PT BSM-1700 Series and Accessories
The Life Scope® PT BSM-1700 Series and Accessories are intended acquire and transfer electrical impulses from the patient to the main unit of the device. The BSM-1700 Series input unit monitors physiological data and may generate an audible and/or visual alarm when a measured rate falls outside preset limits when discomected from the core unit of the device. The input unit can be removed from one core unit and connected to another devices core unit. The Life Scope® PT BSM-1700 Series can be used in transport mode where data is transferred from one device to another device by using with or without WLAN technology. The input unit can acquire the following parameter signals: Electrocardiogram (ECG), Impedance respiration (Imp Resp), Non-invasive blood pressure (NIBP), Arterial oxygen saturation (SpO2), Carbon dioxide concentration (CO2), Invasive blood pressure (Temp), Cardiac Output (CO), TOF and Bispectral Index (BIS).
The Data Acquisition Unit (DAU) is an optional accessory. The DAU is used to communicate between the compatible parent core unit and the input unit using connection cables. The Reys on the DAU allow operation of the bedside monitor remotely. The DAU is only compatible with the Life Scope® PT BSM-1700 Series and the AY Series Input Units. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
AY Series and Accessories
AY Series and Accessories are intended to acquire and transfer electrical impulses from the patient to the core unit of the device. The input unit can acquire the following parameter signals: Electrocardiogram (ECG), Impedance respiration (Imp Resp), Non-invasive blood pressure (NIBP), Arterial oxygen saturation (SpO2), Carbon dioxide concentration (CO2), Invasive blood pressure (IBP), Temperature (Temp), Cardiac Output (CO), TOF and Bispectral Index (BIS). AA Series smart expansion unit adds additional MULTI sockets to an AY Series input unit and can only be used with compatible monitoring systems.
The Data Acquisition Unit (DAU) is an optional accessory. The DAU is used to communicate between the compatible parent core unit and the input unit using connection cables. The keys on the DAU allow operation of the bedside monitor remotely. The DAU is only compatible with the Life Scope® PT BSM-1700 Series Input Units.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| | <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
| | <span> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
#### SECTION 5 510(K) SUMMARY OR 510(K) STATEMENT
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for NIHON KOHDEN. The logo consists of a blue square with a white abstract design on the left, followed by the text "NIHON KOHDEN" in blue. The text is in a sans-serif font and is slightly blurred.
| 5.1 | General Provisions | Manufacturer Name: | Nihon Kohden Corporation |
|-----|--------------------|----------------------|------------------------------------------------------------------------------------------------------------------------------|
| | | Address: | 1-31-4 Nishiochiai, Shinjuku-Ku<br>Tokyo 161-8560, Japan |
| | | Submitter Name: | Nihon Kohden America |
| | | Address: | 15353 Barranca Parkway<br>Irvine, California, USA<br>92618 |
| | | Primary Contact | Sunita Teekasingh<br>GSA2 Group LLC Consultant |
| | | Phone Number | 612-814-7999 |
| | | Email | Sunita.Teekasingh@nihonkohden.com |
| | | Secondary Contact: | Sandra Gadeyne<br>Sr. Director of Quality and Regulatory Affairs |
| | | Phone Number | 949-268-7708 |
| | | Mobile Number: | 949-356-3401 |
| | | Email | Sandra.Gadeyne@nihonkohden.com |
| | | Date of Preparation | 27-Dec-2021 |
| | | Submission Type | Special 510(K) |
| | | Type of Modification | Software Modification |
| | | Predicate | K201949 – cleared 2 May2021<br>• Life Scope® BSM-6000 Series Bedside<br>Monitors;<br>• Nihon Kohden CSM-1901 Bedside Monitor |
5.2 INTRODUCTION
This submission is to obtain clearance for the software modification to the input units for Life Scope® PT BSM 1700 Series and A Y Series Input Units , see Table 5-1 for the summary. The input units required an updated software package to work with the Smart Cable NMT Module and Accessories (K201949).
The software modifications on the input units does not change the safety, performance of the predicate devices.
| Smart Cable<br>NMT Module<br>and Accessories<br>K201949 | Parent device Core Unit<br>software version | Input Unit Software version AY Series or<br>BSM-1700 Series |
|---------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| AF-201P<br>Software<br>Version 3.0-<br>no changes to the<br>original software<br>under<br>K201949 | Life Scope®BSM-6000 Series<br>Bedside Monitoring Systems<br>(K201949) Software v08-31<br><br>Life Scope® CSM-1901<br>Bedside Monitoring Systems<br>(K201949) Software v01-32<br><br>Life Scope® BSM-3000 Series Bedside Monitors (K201949) Software v08-31 | AY Series v08-31<br>AY-631P<br>AY-633P<br>AY-651P<br>AY-653P<br>AY-671P<br>AY-673P<br><br>OR<br><br>BSM-1700 Series<br>v02-66<br>BSM-1733<br>BSM-1753<br>BSM-1773 |
{7}------------------------------------------------
In addition to the software changes, Nihon Kohden (NK) is updating the intended use for the Life Scope® BSM-6000 Series Bedside Monitoring Systems, and Life Scope® CSM-1901 Bedside Monitoring Systems:
- A statement was added to the parent devices to clarify these systems require both a core unit and an ● input unit.
- The Life Scope® PT BSM-1700 Input unit and the AY Series Input unit and accessories were added to the parent device intended use.
| Compatible Parent Bedside<br>Monitoring Core Units | Input Model | Configuration | |
|-----------------------------------------------------------------------------------------------------------------|-------------|-----------------|-------------------------------------------------|
| | | Multi Connector | SpO2 Probe |
| Life Scope® BSM 6000<br>Series Bedside Monitoring<br>Systems (K201949) | AY-631P | 1 | Masimo |
| | AY-633P | 3 | |
| | BSM-1733 | 3 | |
| Life Scope® CSM-1901<br>Bedside Monitoring Systems<br>(K201949) | AY-651P | 1 | Nellcor |
| | AY-653P | 3 | |
| | BSM-1753 | 3 | |
| Life Scope® G5 Series<br>Bedside Monitoring Systems<br>(K203435) | AY-671P | 1 | Nihon Kohden |
| | AY-673P | 3 | |
| | BSM-1773 | 3 | |
| Life Scope® G7 Series<br>Bedside Monitoring Systems<br>(K203435) | | | |
| Life Scope® CSM-1901 and<br>BSM-6000 Series Bedside<br>Monitoring Systems<br>(K201949)<br>Data Acquisition Unit | JA 690PA | N/A | Unit only used<br>with CSM-1901<br>and BSM-6000 |
| | JA-694PA | 4 | |
## Table 5-2: Parent Devices: Core Unit and Input unit configurations
- 5.3 SMART CABLE NMT MODULE AND ACCESSORIES
The Smart Cable NMT Module and Accessories were previously cleared under K201949, this system requires a compatible monitoring system for the NMT module to work. The input unit software has been updated to recognize the Smart Cable NMT Module connection. There have been no modifications to the Intended Use, safety or performance for Smart Cable NMT Module and Accessories as a result of the input unit software update.
{8}------------------------------------------------
| Subject Device | Trade Name: | Smart Cable NMT Module and Accessories |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Common Name: | Electrical peripheral nerve stimulator |
| | Classification Name | Stimulator, Nerve, Peripheral, Electric |
| | Classification: | Class II |
| | Product Code: | KOI |
| | Regulation Number: | 21 CFR 868.2775 |
| Predicate Device | Trade Name: | Smart Cable NMT Module and Accessories |
| | Marketing Names: | NMT Smart Cable TOF Pod; Smart Cable NMT Pod, with EMG Support; NMT Pod TOF Pod; TOF Smart Pod; NMT Smart Pod, NMT Module, AF-201P, Smart Cable NeuroMuscular Transmission Pod, NMT Pod and Train of Four Pod, Disposable electrodes, EMG electrodes, NMT electrodes |
| | Classification Name: | Electrical peripheral nerve stimulator |
| | Premarket Notification | Stimulator, Nerve, Peripheral, Electric |
| | Product Code: | KOI |
| | Device Classification: | Class II |
| | Regulation Number: | 21 CFR 868.2775 |
| | Manufacturer: | Nihon Kohden |
| | 510K number: | K201949 |
| Device Description | The Nihon Kohden Smart Cable NMT Module (NMT Module) and Accessories is an optional accessory for the Nihon Kohden bedside monitoring systems. The Smart Cable NMT Module and Accessories TOF (Train of Four) are based on EMG technology. With this system, the user can apply electrical stimulation on the ulnar nerve to detect the muscle's action potential. The reaction to the electrical impulse can be visualized on the connected monitoring system. The Smart Cable NMT Module and Accessories can assist medical personnel to quantitatively determine the level of muscle relaxation. This information can be used to determine the dose of muscle relaxants and regional anesthetics when performing anesthesia in a clinical setting. It is intended for use by medical personnel in the operating room, recovery room, or intensive care unit.<br>NMT Module is a system comprised of NMT Module, Main Cable, Holder, and EMG Electrode. The NMT module is connected to an electrode via<br>Main Cable. The electrode is a single-use electrode array and each array includes two stimulating electrodes, two recording electrodes, and one ground electrode The NMT module can transmit an electrical | |
| | stimulation pulse to the patient and can receive EMG signals via the<br>electrode array. The captured data from the disposable electrode is sent<br>to the monitoring system via the Smart Cable interface connector. The<br>various stimulation settings are also sent to the monitoring system to<br>display.<br>The AF-201P NMT Module is used to control the electrical stimulation<br>and to measure the response. The operational setting is controlled via<br>buttons on the module or a touch screen. | |
| Indication for Us and<br>Intended Use | The Smart Cable NMT Module and Accessories are indicated for<br>monitoring the relaxation of the patient when neuromuscular blockades<br>are administered. | |
| | The Smart Cable NMT Module and Accessories are comprised of: | |
| | • AF-201P NMT Module with Smart Cable | |
| | • Disposable Electrodes | |
| | • Main cable | |
| | • Holder (optional) | |
| | The Smart Cable NMT Module and Accessories are intended to be<br>used as a system that requires Nihon Kohden compatible electrodes and<br>bedside monitoring systems. The Smart Cable NMT Module and<br>Accessories are intended for use by medical personnel in clinical<br>settings and are available by prescription only. | |
| Modification | There are no changes to the Smart Cable NMT Module and<br>Accessories as a result of the software update to the Life Scope® PT<br>BSM-1700 Series and or the AY Series input units. | |
| Summary of<br>Substantial<br>Equivalence | The device comparison and the results of the safety and performance<br>tests indicate that the Nihon Kohden Smart Cable NMT Module and<br>Accessories is substantially equivalent to the predicate devices. | |
{9}------------------------------------------------
- LIFE SCOPE® BSM-6000 SERIES BEDSIDE MONITORING SYSTEM 5.4
The intended Use for BSM-6000 was updated to include the statement "The system requires a BSM 6000 core unit with a compatible input unit: AY Series or Life Scope® PT BSM-1700 Series".
The Life Scope® BSM-6000 Series Bedside Monitoring System, the input unit software has been updated to recognize the connection to the Smart Cable NMT Module and Accessories. There are no changes to the specifications or performance from the predicate device under K201949.
The Intended Use further clarifies the addition of Life Scope® BSM-1700 Series and the AY Series Input Units to BSM-6000 Intended Use is substantially equivalent to the predicate K201949.
{10}------------------------------------------------
| Subject Device | | Life Scope BSM-6000 Series Bedside<br>Monitoring Systems |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| | Trade Name: | |
| | Common Name: | Bedside Monitor, Patient Monitor,<br>Cardiac Monitor, Vital Signs Monitor and<br>Anesthesia Monitor |
| | Classification Name | Monitor, Physiological Patient Monitor<br>with Arrhythmia Detection and Alarms |
| | Classification: | Class II |
| | Product Code: | MHX |
| | Regulation Number: | 21 CFR 870.1025 |
| Predicate Device | | Life Scope BSM-6000 Series Bedside<br>Monitoring System |
| | Trade Name:…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.