K130570 · Neurovision Medical Products, Inc. · BZQ · Dec 26, 2013 · Anesthesiology
Device Facts
Record ID
K130570
Device Name
NEUROVISION MEDICAL MOTION SENSOR
Applicant
Neurovision Medical Products, Inc.
Product Code
BZQ · Anesthesiology
Decision Date
Dec 26, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.2375
Device Class
Class 2
Indications for Use
The Neurovision Medical Motion Sensor is intended to detect, monitor and/or record limb movements on a physiological recorder, in adult patient sleep disorder studies, in a clinical setting.
Device Story
Neurovision Medical Motion Sensor is a disposable, battery-powered piezo accelerometer sensor. Device attaches to patient limbs to detect gross motor movement. Input signal is physical limb motion; device transforms motion into electrical signal via piezo element. Output is transmitted via DIN connector to standard electrophysiologic monitors (EMG/ECG units) for recording. Used in clinical settings by healthcare professionals during sleep disorder studies. Provides continuous motion data to physiological recorders; assists clinicians in identifying limb movement patterns during sleep. Benefits include objective monitoring of patient movement without requiring dedicated monitor-powered sensors.
Clinical Evidence
Bench testing only. Demonstrated ability to signal gross motor movement using two different EMG units and one standard ECG unit. Compliance with IEC 60601-1, EN 60601-1, UL 60601-1, and CAN/CSA-C22.2 standards verified. Sensitivity compared favorably to the predicate device based on predicate's historical 510(k) data.
Technological Characteristics
Disposable piezo accelerometer sensor; battery-powered; non-sterile; DIN connector interface. Compatible with standard electrophysiologic monitors. Complies with IEC 60601-1:1988, EN 60601-1:1990, UL 60601-1:2003, and CAN/CSA-C22.2 standards.
Indications for Use
Indicated for adult patients undergoing sleep disorder studies in a clinical setting to detect, monitor, and/or record limb movements.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
Predicate Devices
Gereonics Ultra-Piezo Limb Movement Sensor (K062883)
Submission Summary (Full Text)
{0}------------------------------------------------
K120570
Image /page/0/Picture/1 description: The image shows the logo for Neurovision Medical. The word "neurovision" is in large, bold, black letters. Underneath, there is a line with a heartbeat symbol, and the word "MEDICAL" is in smaller, bold, black letters.
DEC 2 6 2013
#### 5. 510K Summary
In accordance with Title 21 of the Code of Federal Regulations, Part 807 and in particular 21 CFR §807.92, the following summary of information is provided:
## Applicant Information
Christine Vergely Regulatory Affairs Manager Neurovision Medical Products, Inc. 2225 Sperry Ave., Suite 1000 Ventura, CA 93003 Ph 866-815-6999 Fax: 413-410-4579 christie@neurovisionmedical.com
#### Date of Summary: 11/4/13
#### Device Identification:
Trade or Proprietary Name: Common or Usual Name: Classification Name: Device Class: Classification: Product Code:
Neurovision Medical Motion Sensor Medical Motion Sensor Breathing Frequency Monitor Class II \$868.2375 BZQ
#### Predicate Devices
The Neurovision Medical Motion Sensor is substantially equivalent to the following predicate devices currently distributed commercially in the U.S .:
- K062883; Gereonics Ultra-Piezo Limb Movement Sensor .
## Device Description
The Neurovision Medical Motion Sensor is a disposable, multipurpose piezo accelerometer sensor configured to be self-powered (by "watch" type battery), disposable, and capable of being read by a small signal electrophysiologic monitor.
#### Indication for Use
The Neurovision Medical Motion Sensor is intended to detect, monitor and/or record limb movements on a physiological recorder, in adult patient sleep disorder studies, in a clinical setting.
### Technological Characteristics of Device in Relation to Predicate Devices
The Neurovision Medical Motion Sensor is similar to the Gereonics Ultra-Piezo Limb Movement Sensor in terms of materials, design, performance and indications for use, differing only in being battery powered whereas the predicate is powered by a dedicated monitor.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Neurovision Medical. The word "neurovision" is in large, bold, black letters. Underneath the word "neurovision" are two horizontal lines with a heartbeat symbol in the middle, followed by the word "MEDICAL" in smaller, bold, black letters.
## Table 1 Device Comparison Table
| | K062883 | Subject Device |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Brand Name | Gereonics Ultra-Piezo Limb Motion Sensor | Medical Motion Sensor |
| Description | Piezo Sensor, attachable | Piezo Sensor, attachable |
| Indications for Use | For adult use. The Gereonics Ultra-Piezo<br>Limb Movement Sensor (LMS) is used to<br>detect<br>periodic limb movements for recording or<br>monitoring on a physiological amplifier.<br>The LMS is intended for use in sleep<br>disorders studies. | The Neurovision Medical Motion Sensor is intended<br>to detect, monitor and/or record limb movements on a<br>physiological recorder, in adult patient sleep disorder<br>studies, in a clinical setting. |
| Product Code | BZQ | BZQ |
| Materials and Design | Nonsterile attachment electrode with<br>unpowered Piezo Sensor | Nonsterile attachment device with battery powered<br>Piezo Accelerometer |
| Blo-compatibility | N/A | N/A |
| Compatible devices | Dedicator Monitor (Gereonics) | Electrophysiologic Monitors |
| Sterillty | Non-sterile | Non-sterile |
| Attachment to Monitor | DIN connector | DIN connector |
### Performance Testing
Compatibility with representative testing equipment was demonstrated.
Bench testing included in the submission includes:
- Demonstrated ability of the Neurovision Medical Motion Sensor to signal gross . motor movement with 2 different model EMG units.
- . Demonstrated ability of the Neurovision Medical Motion Sensor to signal gross motor movement with a standard ECG unit.
- . Compliance with IEC 60601-1:1988 + A1:1991 + A2:1995; EN 60601-1:1990+A1+A2+A13, UL 60601-1:2003 R4.06, CAN/CSA-C22.2
- . Comparable sensitivity of the Neurovision Medical Products Motion Sensor to typical patient movements expected under the conditions of intended use to the sensitivity of the Predicate Devices, Gereonics Ultra-Piezo Limb Movement Sensor, to similar movements, based on the testing submitted by the predicate for 510k review.
### Conclusion
The Neurovision Medical Motion Sensor is substantially equivalent to the Gereonics Ultra-Piezo Limb Movement Sensor in terms of materials, design, performance and indications for use, differing only in being battery powered whereas the predicate is powered by a dedicated monitor. There are no additional safety or effectiveness risks associated with this design difference. (End of 510(k) Summary)
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure embracing a globe, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
# December 26, 2013
Neurovision Medical Products Inc. Christine Vergely Regulatory Affairs Manager 2225 Sperry Ave, Suite 1000 VENTURA, CA 93003
Re: K130570
Trade/Device Name: Neurovision Medical Motion Sensor Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: Class II Product Code: BZQ Dated: November 25, 2013 Received: November 26, 2013
## Dear Ms. Vergely:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
Page 2 - Ms. Christie Vergely
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
**Kwame** **FOA**mer-S for
Erin I. Keith Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
510(k) Number (if known) KI30570
#### Device Name
******************************************************************************************************************************************************************************
Neurovision Medical Motion Sensor
Indications for Use (Describe)
The Newrovision Medical Motion Seasor is intended to detect, manitor and/or record limb movements on a physiological recorder, in actuat patient sleep disorder studies in a clinical setting.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
# Please Do not WRITE BELOW THIS LINE - Continue on a SEPARATE PAGE IF NEEDED.
Image /page/4/Figure/12 description: The image shows a document with the heading "FOR FDA USE ONLY". It includes the text "Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)". There is a digital signature by Richard C. Chapman, dated 2013.12.26 at 12:26:15 -05'00'. The form number "FORM FDA 3881 (9/13)" is at the bottom left.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.