The EarlySense (EverOn) System is intended for continuous measurement of respiration rate, heart rate and movement, in an automatic contact-less manner, at home, hospital or clinic setting. The system is indicated for use in children, adolescents and adults. The operation of the EarlySense has been studied in children (weight ≥ 10 Kg) and adults (weight <111 Kg) during sleep and resting condition. In addition, EarlySense (EverOn) System can continuously monitor oxygen saturation of arterial hemoglobin (SPO2) using pulse oximetry in pediatric (ages 2 years and older), adolescents, and adults at home, hospital, or clinical settings.
Device Story
System provides continuous, contact-less monitoring of respiration rate, heart rate, and movement via under-mattress piezoelectric sensor; optional OEM pulse oximetry module monitors SpO2 via finger sensor. Bedside unit processes signals, displays parameters, and generates alerts based on set thresholds. Used in home, hospital, or clinic settings by clinicians or patients. Data communicated via wired/wireless LAN to Central Display Station for real-time or off-line analysis. Output assists healthcare providers in patient monitoring; enables timely clinical intervention; supports patient safety during sleep and rest.
Clinical Evidence
Bench testing only. Included electrical safety and electromagnetic compatibility testing (IEC 60601-1, IEC 60601-1-2). Software verification and validation performed per SRS. Bench testing verified accuracy of oximetry data display and SpO2 alerting features.
Technological Characteristics
Piezoelectric sensor for mechanical movement detection; UART (RS232/TTL) interface for optional OEM oximetry module; Linux-based software platform; wired/wireless LAN connectivity. Complies with IEC 60601-1, IEC 60601-1-2, IEC 62304, and ISO 14971.
Indications for Use
Indicated for children (≥10 kg), adolescents, and adults for continuous monitoring of respiration rate, heart rate, and movement during sleep and rest. Additionally indicated for pediatric (≥2 years), adolescent, and adult patients for continuous SpO2 monitoring.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}------------------------------------------------
### 510(K) SUMMARY
# 510(K) Number K_12 Q 465
5.1 Applicant's Name: EarlySense Ltd. 12 Tzvi Street, Ramat Gan, Israel
5.2 Contact Person: Dalia Argaman EarlySense Ltd. 12 Tzvi Street, Ramat Gan, Israel Tel: +972 (3) 752 - 2330 Fax: +972 (3) 752 - 2340 Email: Dalia. Argaman(@)Earlysense.com
April 2012 5.3 Date Prepared:
EarlySense (EverOn™) System 5.4 Trade Name:
5.5 Classification Name: Breathing frequency monitor (868.2375) monitor, cardiac (incl. cardiotachometer & rate alarm) (870.2300) Oximeter (870.2700)
5.6 Product Code:
BZQ, DRT, and DQA
Class II 5.7 Device Class:
5.8 Predicate Devices:
- EverOn 1.0 (EarlySense Ltd.), cleared under K092062. .
- Avant™ 9600 Pulse Oximeter (Nonin Medical Inc.), cleared under K040589 .
- LifeShirt System with VivoLogic Analysis Software (VivoMeterics Inc.), cleared under ● K011903
### 5.9 Intended Use / Indication for Use:
The EarlySense (EverOn) System is intended for continuous measurement of respiration rate, The Darry other (1) an automatic contact-less manner, at home, hospital or clinic setting. The system is indicated for use in children, adolescents and adults. The operation of the The System 15 marcated 16th children (weight ≥ 10 Kg) and adults (weight <111 Kg) during
{1}------------------------------------------------
sleep and resting condition, In addition, EarlySense (EverOn) System can continuously monitor oxygen saturation of arterial hemoglobin (SPO2) using pulse oximetry in pediatric (ages 2 years and older), adolescents, and adults at home, hospital, or clinical settings.
#### Device Description 5.10
The EarlySense (EverOn) System consists of the following main components:
- A piezoelectric Sensing Unit placed under the mattress or mattress pad. .
- Bedside Unit with Proprietary recording and data analysis software .
- OEM Oximetry Module (optional) .
The EarlySense (EverOn) System is designed for continuous and contact-less monitoring of respiration rate, heart rate and movement. The under mattress Sensing Unit includes a piezoelectric sensor, which converts mechanical movements into an electric signal. The principles and the mode of operation for the capability of contactless monitoring of the EarlySense (EverOn) is identical to the cleared EverOn 1.0 (K092062). The EarlySense interfaces with a pulse oximeter OEM module (optional feature) and can therefore also monitor oxygen saturation. The oximetry module is connected externally via UART (RS232/TTL) to the Bedside unit. A compatible oximetry sensor is then attached to the patient's finger and the monitor begins to continuously display oximetry data (e.g. SpO2).
The Bedside Unit, processes inputted signals, displays the patient's parameters, and generates alerts (respiration rate, heart rate, movement, and SpO2) as per set thresholds when needed. The antinuously accumulated data from the monitored period are displayed on the bedside unit and communicated, accammunicated, via wired or via wireless LAN communication to a dedicated may options of SerilySense Central Display Station (CDS) - cleared under K100376 and K110521). Analysis of the results may be performed either on-line during the monitoring session, or later, off-line. Data for each patient is recorded and can therefore also be retrieved and presented.
## 5.11 Performance Standards
The EarlySense System complies with voluntary standards such as:
- Medical electrical equipment- general requirements for safety. Part 1: General . Requirements for Safety. IEC 60601-1(1988): +A1(1991) +A2(1995);
- Electrical Equipment Part 1: General Requirements for Safety; l Medical Electromagnetic Compatibility -- Requirements and Tests. EN/IEC 60601-1-2 (2007)
- Medical device software Software life cycle processes. IEC 62304 (2006) ■
- Medical devices Application of risk management to medical devices. ISO 14971 1 (2009)
## 5.12 Substantial Equivalence
{2}------------------------------------------------
### Intended use and Technological Characteristics
The EarlySense (EverOn), subject to this submission, is similar to the EverOn 1.0 predicate device with the following changes: (1) optional interface with a pulse oximeter OEM module and therefore monitoring of oxygen saturation and (2) operation on a Linux platform. Bv incorporating an oximeter module into the EarlySense System, two cleared scientific technologies are combined into one System. The intended use and indications for use of the EarlySense is similar and encompassed within the intended use and indications for use of its Specifically, with the exclusion of monitoring oxygen saturation, the predicate devices. proposed intended use and indications for use of the EarlySense (EverOn) System is identical to the EverOn 1.0 (K092062) predicate device, while the monitoring of oxygen saturation is encompassed within the intended use and indications for use of the oximeter predicate device Avant 9600 Pulse Oximeter (K040589). Incorporation of an oximeter module into a physiological data monitor, a well understood technology, is shared by other cleared devices.
Testing was performed in order to demonstrate the safety and performance of the EarlySense System and to demonstrate that as a result of combining the EarlySense technology and a standard OEM, no new safety and effectiveness issues, in comparison to its predicate devices, are raised. In summary, the EarlySense System is as safe and effective as its predicate devices for its intended use and is substantially equivalent to its predicate devices without raising any new safety and/or effectiveness issues.
#### Performance Testing
A set of software and bench testing was performed in order to demonstrate the performance and accuracy of the EarlySense System and to verify that it does not raise any new safety and effectiveness issues in comparison to its predicate devices. The testing included the following:
- Electrical safety and electromagnetic compatibility testing according to IEC 60601-1 I (and amendments), and IEC 60601-1-2 (and amendment) standards.
- Software verification and validation testing was conducted to evaluate the B performance of the EarlySense System and to verify that it performs according to its specifications described in the Software Requirements Specifications (SRS).
- Bench testing including demonstrating the accuracy of the oximetry data display and . SpO2 alerting feature.
#### Summarv
The EarlySense System has the same or similar intended uses and indications, technological The Barrystics, and principles of operation as its predicate devices. Based on the performance testing results, including software verification and validation process and bench testing, the cooning roatio, including and differences, EarlySense Ltd. believes that the EarlySense System andryis of the simmariters o its predicates without raising new issues of safety or effectiveness.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the eagle. The eagle is black, and the text is also black.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Delia Argaman Vice President Clinical and Regulatory Affairs EarlySense Limited 12 Tzvi Street Ramat Gan Israel 52504
JUN - 8 2012
Re: K120465
Trade/Device Name: EarlySense (EverOn™) System Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: II Product Code: BZQ Dated: May 23, 2012 -Received: May 29, 2012
Dear Ms. Argaman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties: We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2- Ms. Argaman
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm1115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/ default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Wh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices
Office of Device Evaluation Center for Devices and Radiological Health ·
#### Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE
510(k) Number (if known): __
Device Name: EarlySense (EverOn™) System
The EarlySense (EverOn) System is intended for continuous measurement of respiration rate, heart rate and movement, in an automatic contact-less manner, at home, hospital or clinic setting. The system is indicated for use in children, adolescents and adults. The operation of the EarlySense has been studied in children (weight ≥ 10 Kg) and adults (weight <111 Kg) during sleep and resting condition. In addition, EarlySense (EverOn) System can continuously monitor oxygen saturation of arterial hemoglobin (SPO2) using pulse oximetry in pediatric (ages 2 years and older), adolescents, and adults at home, hospital, or clinical settings.
| Prescription Use | √ |
|-----------------------------|---|
| (Part 21 CFR 801 Subpart D) | |
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Division Sign-Off) Division Sign-On7
tection Control, Dental Devices
510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.