The STIMPOD ST2-3010 is a nerve stimulation device designed to be used for the purpose of nerve locating using invasive electrodes or needles (not supplied)
Device Story
Battery-powered peripheral nerve stimulator; used by anesthetists for nerve localization. Device connects to conducting needles (not supplied) to deliver electrical stimulation to specific neural pathways. Constant current source generates monophasic square wave pulses (0.1, 0.2, 0.3 ms pulse widths; 20, 50, 100 Hz frequencies). Unit continuously monitors circuit impedance; delivers stimulus only when closed circuit detected. Visual (LED) and audible (buzzer) feedback provided to clinician. Device operates via tactile membrane keypad and capacitive sensing wheel. Output helps clinician identify nerve proximity during regional anesthesia procedures, potentially improving needle placement accuracy and patient outcomes.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1 (3rd ed), IEC 60601-1-2 (EMC), and IEC 60601-2-10 (nerve/muscle stimulator safety) demonstrated. No clinical data provided.
Technological Characteristics
Battery-powered (4x AAA) peripheral nerve stimulator. ABS enclosure; 145mm x 90mm x 30mm; 130g. Constant current monophasic square wave output (0-10mA, 0.1-0.3ms pulse width). Class IIa, Type BF. Connectivity: wired stimulation cable. Software: C-language, microprocessor-controlled. Cleaning: soap/water or ethyl alcohol-based disinfectant.
Indications for Use
Indicated for nerve locating using invasive electrodes or needles in patients requiring regional anesthesia.
Regulatory Classification
Identification
An electrical peripheral nerve stimulator (neuromuscular blockade monitor) is a device used to apply an electrical current to a patient to test the level of pharmacological effect of anesthetic drugs and gases.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of three human profiles facing right, stacked on top of each other. The profiles are rendered in a simple, line-art style.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 6, 2014
Xavant Technology (Pty) Limited Mr. Roche van Rensburg Chairman Ravello 1st Floor, Delmondo Office Park 169 Garsfontein Road Ashlea Gardens Pretoria, Gauteng 0081 SOUTH AFRICA
Re: K140853
Trade/Device Name: Stimpod ST2-3010 Nerve Stimulator Regulation Number: 21 CFR 868.2775 Regulation Name: Electrical Peripheral Nerve Stimulator Regulatory Class: II Product Code: BXN Dated: October 1, 2014 Received: October 10, 2014
Dear Mr. Rothman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Rothman
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
# ATTACHMENT 2
# INDICATIONS FOR USE
510(k) Number (if known):
K140853
Device Name: STIMPOD ST2-3010 nerve stimulator
Indications for Use:
The STIMPOD ST2-3010 is a nerve stimulation device designed to be used for the purpose of nerve locating using invasive electrodes or needles (not supplied)
X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
# ATTACHMENT 4
# 510(k) SUMMARY
[As required by 21CFR807.92]
Image /page/3/Picture/3 description: The image is a logo for Xavant Technology. The logo features a stylized blue "X" with a gray orbital path and dot, followed by the word "AVANT" in red, block letters. Below "AVANT" is the word "TECHNOLOGY" in a smaller, gray, sans-serif font.
### 4.1 Date Prepared [21CFR807.92(a)(1)]
July 28, 2014
# 4.2 Submitter's Information [21CFR807.92(a)(1)]
| Company Name: | XAVANT Technology (Pty) LTD |
|-----------------|------------------------------------------|
| Street Address: | Ravello, 1st Floor. Delmondo Office Park |
| | 169 Garstfontein Road |
| | Ashlea Gardens |
| City: | Pretoria |
| State/Province: | Gauteng |
| Country: | South Africa |
| Telephone: | +27(0) 12 755 9491 |
| Facsimile: | +27(0) 12 346 7271 |
| Contact Person: | Brian Rothman |
| Contact Title: | Quality Assurance and Regulatory |
| | Compliance Officer |
| Contact Email: | brian@xavant.com |
#### 4.3 Trade Name, Common Name, Classification [21CFR807.92(a)(2)]
| Trade Name: | STIMPOD ST2-3010 nerve stimulator |
|----------------------|---------------------------------------------|
| Common Name: | Battery Powered Peripheral Nerve Stimulator |
| Classification Name: | Anesthesiology<br>per 21 CFR § 868.2775 |
| Device Class: | Class II |
| Product Code: | BXN |
{4}------------------------------------------------
## 4.4 Identification of Predicate Device(s) [21CFR807.92(a)(3)]
#### PREDICATE DEVICES
Xavant Technology, STIMPOD NMS450 (K102084)
There are no significant differences between the STIMPOD ST2-3010 Nerve Stimulator and the predicate devices which would adversely affect the use of the product. It is substantially equivalent to the device in design, function, materials, operational principles and intended use.
The STIMPOD NMS450 is a battery powered peripheral nerve stimulator that can be used for
- general anesthesia, for the purpose of establishing the efficacy of a � Neuromuscular Blocking Agent using non-invasive surface electrodes (not supplied)
- nerve mapping using the non-invasive Nerve Mapping Probe (supplied) �
- nerve locating using invasive electrodes/needles (not supplied) �
The stimulus is generated by a constant current source. The waveform is a square wave with various pulse width options.
The units will continuously check for a closed circuit. Once a closed circuit is detected, the unit will deliver a stimulus until an open circuit is detected.
Once operational the unit will flash an LED when the circuit with the patient is a closed circuit and the stimulation was successfully delivered. In the case that the circuit with the patient is an open circuit, a stimulus will not be delivered.
The anode comprises of an ECG electrode (not supplied). The cathode comprises a permanently attached nerve mapping probe (supplied), and/or a separate nerve locating needle (not supplied), and/or an ECG electrode (not supplied), depending on the mode of the unit
This product is a nerve stimulation device designed to be used by an anaesthetist. durina
- 1. General anaesthesia, for the purpose of establishing the efficacy of a Neuromuscular Blocking Agent using non-invasive surface electrodes (not supplied)
- 2. Regional anaesthesia for the purpose of
- a) nerve mapping using the non-invasive Nerve Mapping Probe (supplied) and
- b) nerve locating using invasive electrodes/needles (not supplied)
### 4.5 Description of the Device [21CFR807.92(a)(4)]
{5}------------------------------------------------
The STIMPOD ST2-3010 device is a battery powered peripheral nerve stimulator that can be used for
- the purpose of nerve locating using invasive electrodes or needles (not supplied) .
The STIMPOD ST2-3010 is a precision nerve locating tool used for localizing specific neural pathways. Localization of nerves by electrical stimulation involves connecting the nerve stimulator to a conducting needle.
The stimulus is generated by a constant current source. The waveform is a monophasic square wave with 3 options for pulse width. These are: 0.1, 0.2 and 0.3 milliseconds.
The unit dimensions are 145mm x 90mm x 30mm with a weight of 130 g
### 4.6 Intended Use [21CFR807.92(a)(5)]
This product is a nerve stimulation device designed to be used for the purpose of nerve locating using invasive electrodes or needles (not supplied).
The difference in intended use is not critical to its use and does not affect the safety and efficacy of the subject device. Both subject and predicate device are nerve stimulators. Where as the predicate device has three (3) indications for use, the subject device has only one. This is to be used for the purpose of nerve locating using invasive electrodes or needles (not supplied). The predicate device has this use listed and already has clearance.
The intended use of the subject device is limited to within the currently cleared indications of the predicate device.
### 4.7 Technological Characteristics [21CFR807.92(a)(6)]
#### Operating Modes:
| Current Range: | 0.00 – 10.00mA ±5%<br>adjustable in 0.1mA increments. |
|------------------------|-------------------------------------------------------|
| Pulse Width Options: | 100µs, 200µs, 300µs |
| Stimulus: | Monophasic square wave |
| Stimulating Frequency: | 20Hz, 50Hz and 100Hz |
{6}------------------------------------------------
## Waveform:
Constant Current Monophasic Square wave
#### Technical Specifications
| Device Classification: | Class Ila, Type BF |
|------------------------------------|------------------------------|
| Power Supply: | 4x AAA alkaline batteries |
| Power Consumption: | 18mA |
| Maximum Stimulation Voltage: | 70V |
| Weight: | 130g |
| Dimensions: | 145mm x 90mm x 30mm |
| Load Impedance: | 0 kOhm – 7 kOhm |
| Operating Temperature: | 10 - 40 Degrees Celsius (°C) |
| Storage and Transport Temperature: | 0 - 50 Degrees Celsius (°C) |
| Operating Humidity: | 90% Relative Humidity |
Summary of technical characteristics between predicate device (STIMPOD NMS450) and new device (STIMPOD ST2-3010)
| Device Name: | STIMPOD ST2-3010 | STIMPOD NMS450 | Affects<br>Indications<br>for Use /<br>Technology | Comments |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Product<br>Components: | 1. Nerve Stimulator<br>2. Stimulation Cable<br>3. IFU | 1. Nerve Stimulator<br>2. Nerve Locating Cable<br>3. Nerve Mapping/Locating Cable<br>4. NMBA Monitoring Cable<br>5. IFU | NO (e) | Both contain a nerve stimulator and a cable.<br>The NMS450 has additional cables for additional functions |
| Indications for<br>Use: | This product is a nerve stimulation device designed to be used for the purpose of nerve locating using invasive electrodes or needles (not supplied) | This product is a nerve stimulation device designed to be used by an anesthetist during<br>1. general anesthesia, for the purpose of establishing the efficacy of a Neuromuscular Blocking Agent using non-invasive surface electrodes | NO - In this case, the indications for use are limited to within the cleared indications for use of the predicate device (a) (c) | Essentially, both devices are for nerve localization using invasive electrodes/needles (not supplied) |
| Device Name: | STIMPOD ST2-3010 | STIMPOD NMS450 | Affects Indications for Use / Technology | Comments |
| | | (not supplied<br>2. regional anesthesia<br>for the purpose of<br>a. nerve mapping<br>using the non-<br>invasive Nerve<br>Mapping Probe<br>(supplied) and<br>b. nerve locating<br>using invasive<br>electrodes/need<br>les (not<br>supplied) | | |
| Current Ranges | 0 -10mA | Nerve Locating: 0-5mA<br>Nerve mapping: 0-20mA<br>NMBA Mode: 0-80mA | NO (d) | The current<br>range is<br>covered |
| Energy Type<br>(1) | 4 X AAA alkaline batteries, 6V | 4 X AAA alkaline batteries, 6V | NO (e) | Both units<br>are battery<br>powered |
| Activation<br>Labeling: | ON/OFF Power button<br>According to ISO 15223 | ON/OFF Power button<br>According to ISO 15223 | NO (e)<br>NO (d) | |
| Anatomical<br>site of use<br>Testing | Location of peripheral<br>nerves<br>SAFETY TESTING | Location of peripheral<br>nerves<br>SAFETY TESTING | NO (a) , (b),<br>(c) | |
| | Passed<br>(IEC 60601-1) 3rd edition | Passed<br>(IEC 60601-1) 3rd edition | NO (b) , (h) | |
| | RADIATED RADIO<br>FREQUENCY<br>ELECTROMAGNETIC<br>FIELD 80 MHz TO 2500<br>MHz<br>IEC61000-4-3<br>Passed | RADIATED RADIO<br>FREQUENCY<br>ELECTROMAGNETIC<br>FIELD 80 MHz TO 2500<br>MHz<br>IEC61000-4-3<br>Passed | NO (b) , (h) | |
| | ELECTROMAGNETIC<br>COMPATIBILITY<br>Passed<br>(IEC 60601-1-2) | ELECTROMAGNETIC<br>COMPATIBILITY<br>Passed<br>(IEC 60601-1-2) | NO (b) , (h) | |
| | Medical electrical<br>equipment. Particular<br>requirements for the | Medical electrical<br>equipment. Particular<br>requirements for the | NO(b) , (h) | |
| Device Name: | STIMPOD ST2-3010 | STIMPOD NMS450 | Affects<br>Indications<br>for Use /<br>Technology | Comments |
| | safety of nerve and<br>muscle stimulators | safety of nerve and<br>muscle stimulators | | |
| | PASSED<br>(IEC 60601-2-10) | PASSED<br>(IEC 60601-2-10) | | |
| Weight (5) | 130g | 130g | NO (e) | Same weight |
| Size (unit) (5) | 145mm x 90mm x 30mm | 145mm x 90mm x 30mm | NO (e) | Identical<br>enclosures |
| Enclosure (4) | Manufactured from ABS | Manufactured from ABS | NO (i) | Identical<br>enclosures |
| Dial (4) | Tactile membrane with<br>capacitive sensing wheel.<br>Manufactured from<br>polycarbonate | Tactile membrane with<br>capacitive sensing wheel.<br>Manufactured from<br>polycarbonate | NO (e) | Same<br>keypads,<br>different<br>coloring and<br>function |
| LCD (4) | Not covered | Not covered | NO (e) | |
| Cable (4) | KE-PVC moulding<br>Connector – Gold plated | KE-PVC moulding<br>Connector – Gold plated | NO (e), (l) | Similar cable<br>construction, |
| Packaging (7) | Non-sterile<br>Multiple Use<br>Packaging in a<br>polypropylene carry case | Non-Sterile<br>Multiple Use<br>Packaging in a<br>polypropylene carry case | NO (e), (f) | Packaged in<br>the same<br>way |
| Software (6) | Language: C<br>Architecture | Language: C<br>Architecture | | |
| | - User Inputs<br>• Keypad<br>• Dial | - User Inputs<br>• Keypad<br>• Dial | | |
| | - External Inputs<br>• Open circuit<br>detection Low<br>voltage<br>• No Open circuit<br>detection High<br>voltage<br>• Battery Voltage<br>• Cable ID x 1<br>• No accelerometer | - External Inputs<br>• Open circuit<br>detection Low<br>voltage<br>• Open circuit<br>detection High<br>voltage<br>• Battery Voltage<br>• Cable ID x 3<br>• Accelerometer | NO (d), (g) | |
| | - Outputs<br>• Display Menu<br>• Display Main<br>Screen | - Outputs<br>• Display Menu<br>• Display Main<br>Screen | | |
| Device Name: | STIMPOD ST2-3010 | STIMPOD NMS450 | Affects<br>Indications<br>for Use /<br>Technology | Comments |
| | (Closed Circuit)<br>○ Red LED (Open<br>Circuit)<br>○ Buzzer Alarm<br>○ Square Wave<br>Stimulation Low<br>Voltage<br>○ No Square Wave<br>Stimulation High<br>voltage | (Closed Circuit)<br>○ Red LED (Open<br>Circuit)<br>○ Buzzer Alarm<br>○ Square Wave<br>Stimulation Low<br>voltage<br>○ Square Wave<br>Stimulation High<br>voltage | | |
| Waveform<br>Pulse Width<br>(3) | Monophasic Square Wave<br>0.1, 0.2, 0.3ms | Monophasic Square Wave<br>0.05, 0.1, 0.3, 0.5, 1ms | NO (h) | Similar and<br>within the<br>range of the<br>predicate |
| Shelf Life (7) | 5 Years | 5 Years | NO (e) | |
| Control<br>Mechanism<br>(3) | Digitally controlled by a<br>microprocessor | Digitally controlled by a<br>microprocessor | NO (g) | |
| Electronics | PCB:2 layer board<br>POWER SOURCE:: 4 x<br>AAA<br>POWER SUPPLY: Switch<br>Mode<br>MAX POWER<br>CONSUMPTION: ± 40mA<br>MAX VOLTAGE: 100V | PCB:2 layer board<br>POWER SOURCE:: 4 x<br>AAA<br>POWER SUPPLY: Switch<br>Mode<br>MAX POWER<br>CONSUMPTION: 79mA<br>MAX VOLTAGE: 400V | NO (h) | |
| Device<br>Cleaning | <b>Cleaning:</b> Soap and water,<br>applied with a damp cloth is<br>suitable to clean and<br>disinfect the STIMPOD. It is<br>imperative that no moisture<br>penetrates the STIMPOD.<br><b>Disinfecting:</b> Any<br>commercially available<br>methanol - free disinfectant<br>in an ethyl alcohol base<br>can be used for disinfection | <b>Cleaning:</b> Soap and water,<br>applied with a damp cloth is<br>suitable to clean and<br>disinfect the STIMPOD. It is<br>imperative that no moisture<br>penetrates the STIMPOD.<br><b>Disinfecting:</b> Any<br>commercially available<br>methanol - free disinfectant<br>in an ethyl alcohol base can<br>be used for disinfection. | NO (e) | |
| 510(k) | This submission. | K102084 | | |
Page 14-4
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
## A detailed comparison has provided the following conclusions:
The bolded letters reference the relevant section on the table above. This is indicated in the column "affects indications for use / Technology"
## Clinical Use
{10}------------------------------------------------
The target populations on which product usage would occur are the same as those of the cited predicate device.
The Stimpod ST2-3010 employs a subset of the indications for use of the Stimpod NMS450, the same contraindications for use, and the same warnings and precautions within labelling
Each of the above devices:
- used for the same clinical condition or purpose; (a) ●
- . have similar relevant critical performance according to expected clinical effect for specific intended use. (b)
#### Technical Characteristics
The STIMPOD ST2-3010 is equivalent to the identified predicate in design and utilized materials of construction of the currently marketed aforementioned predicate devices.
Also, the principles of operation of the subject device are directly equivalent to those of the cited predicate
Each of the above devices:
- are used under similar conditions of use; (c) ●
- have similar specifications and properties; (d)
- are of similar design; (e) ●
- use similar deployment methods; (f) ●
- have similar principles of operation; (g) and ●
- have similar electrical performance. (h) .
As indicated in the table above, the ST2-3010 and the NMS450 are exact matches on technical characteristics, with the following exceptions:
- 1. Current Range 0-10 compared to Current Range 0-5 (Nerve Locating)
- a. The ST2-3010 can adjust the current up to 10mA. The NMS450 however can only adjust the current to 5mA. However, the NMS450 can adjust pulse width up to 1ms, whereas the ST2-3010 can only adjust pulse width up to 0.3ms. That concludes that the maximum charge output of the ST2-3010 is 3uC, which is only 60% of the maximum charge output (5uC) of the NMS450. Thus patient safety is not affected by this change.
- 2. Pulse width options 0.1, 0.2, 0.3ms compared to 0.05, 0.1, 0.3, 0.5, 1.0ms
- a. The pulse width options of the ST2-3010 are a subset of the pulse width options of the NMS450, and therefor do not affect patient safety.
- 3. No High Voltage stimulation vs. High Voltage Stimulation
- a. As the ST2-3010 is only a nerve location device, it only employs the low voltage section of the electrical circuit, exactly the same as the Nerve Locating mode on the NMS450
{11}------------------------------------------------
#### Biological Characteristics
Each of the above devices has:
- no known biocompatibility issues; (i) and
- no known effect on the environment, or to other devices. .
#### Contents ST2-3010 Peripheral Nerve Stimulator Kit.
The Kit will include the following parts:
- Peripheral Nerve Stimulator a.
- b. Stimulation Adaptor Cable
- Carrying Case C.
- Instructions for Use d.
Note: Needles, ECG electrodes and batteries are not supplied
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.