NONIN PULSE OXIMETER AND CARBON DIOXIDE DETECTOR, MODELS 9843, 9845 AND 9847
K982969 · Nonin Medical, Inc. · CCK · Aug 12, 1999 · Anesthesiology
Device Facts
Record ID
K982969
Device Name
NONIN PULSE OXIMETER AND CARBON DIOXIDE DETECTOR, MODELS 9843, 9845 AND 9847
Applicant
Nonin Medical, Inc.
Product Code
CCK · Anesthesiology
Decision Date
Aug 12, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1400
Device Class
Class 2
Indications for Use
The NONIN® Pulse Oximeter and Carbon Dioxide Detector (Model 9840-Series) is indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, and approximate carbon dioxide (CO2) changes in the airway of intubated patients. These functions may be used separately or simultaneously.
Device Story
Hand-held, battery-operated, noninvasive monitor; tracks patient status via visual/audible indicators. Inputs: SpO2 and CO2 signals from patient airway. Operation: processes physiological signals to display SpO2, pulse rate, and CO2 levels; provides audible/visual alarms for apnea, low battery, or sensor conditions. Used by healthcare professionals in clinical settings. Output: real-time patient data and alarm notifications. Clinical utility: assists providers in assessing patient respiratory status and oxygenation; enables timely intervention for apnea or abnormal vital signs.
Clinical Evidence
Bench testing performed per FDA guidance and international standards. Includes hazard analysis, risk assessment, and software/hardware verification and validation. Field evaluations conducted in controlled clinical environments to confirm accuracy of the CO2 detector.
Technological Characteristics
Hand-held, battery-operated, noninvasive monitoring device. Features visual and audible indicators. Includes alarm logic for apnea, low battery, and sensor conditions. Software-controlled monitoring of SpO2, pulse rate, and CO2.
Indications for Use
Indicated for intubated patients requiring monitoring of functional oxygen saturation (SpO2), pulse rate, and airway CO2 changes.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
{0}------------------------------------------------
K982969
# 510(k) Summary
| SUBMITTER: | Nonin Medical, Inc. | |
|-----------------|---------------------|-----------------------------------------------------------------------------|
| | Address: | Nonin Medical, Inc.<br>2605 Fernbrook Lane North<br>Plymouth, MN 55447-4755 |
| | Telephone: | 612-553-9968 |
| CONTACT PERSON: | | Richard P. Bennett, Director of Regulatory Affairs |
| DATE PREPARED: | | August 21, 1998 |
| TRADE NAME: | | NONIN® Pulse Oximeter and Carbon Dioxide Detector |
| | COMMON NAME: | Pulse oximeter and carbon dioxide detector |
### SUBSTANTIALLY EQUIVALENT TO:
The NONIN Pulse Oximeter and Carbon Dioxide Detector are substantially equivalent to: (1) The pulse oximeter segment is substantially equivalent to the NONIN Models 8500/8500A Pulse Oximeters [510(k)s 893221/945290]. (2) The Carbon Dioxide [CO2] Detector segment is substantially equivalent to the Nellcor Stat Cap Model N-60B [510(k) 915494]. (3) The NONIN Model 9843, 9845, and 9847 Pulse Oximeter and CO2 Detector is substantially equivalent to the two-parameter Spegas MicroCap/NPB-75 [K964239] which likewise functions as a dual function instrument displaying CO2 and SpO2.
# DESCRIPTION of the DEVICE:
The NONIN® Pulse Oximeter and Carbon Dioxide Detector (Model 9840-Series), is a hand-held, battery-operated, noninvasive monitoring device that has visual and audible indicators for tracking patient status. Each Pulse Oximeter and Carbon Dioxide Detector model performs similar basic functions, but each model has been customized to meet the needs of different demands placed upon the health care professional.
Each model of the NONIN® Models 9840-Series has informational tones, (along with visible indicators) to alert healthcare providers to patient conditions. However only
510(k) Submission Pulse Oximeter & Carbon Dioxide Detector Model 9840-Series - 81098
{1}------------------------------------------------
# 510(k) Summary-Continued
Model 9845 and Model 9847 have an high priority "absence-of-breath" (apnea) audible alarm. The Model 9843 indicates that there is an "absence-of breath" by the cessation of the information tone, the breath-beep indicator.
Model 9845 and Model 9847 have medium priority (equipment) alarm that either indicates that the batteries have reached a critically low power state or that a CO2 sensor alarm condition is occurring. The Model 9843 indicates a low battery condition by an illuminated low battery indicator.
Model 9847 has a high priority (patient) alarm to alert the healthcare provider to "absence of breath" (apnea), a high or low oxygen saturation, a high or low pulse rate or inadequate pulse quality conditions.
Audible alarms can be "permanently" or "temporarily" disabled using the audible alarm disable switch on Model 9845 and Model 9847.
#### INDICATIONS FOR USE:
The NONIN® Pulse Oximeter and Carbon Dioxide Detector (Model 9840-Series) is indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, and approximate carbon dioxide (CO2) changes in the airway of intubated patients. These functions may be used separately or simultaneously.
#### SUMMARY of TESTING:
The Model 9840-Series testing has followed (where applicable), the Reviewer Guidance for Premarket Notification Submission of November 1993, from the Anesthesiology and Respiratory Branch, Division of Cardiovascular, Respiratory and neurological Devices. In addition, NONIN Medical, Inc. has evaluated these models for conformance with referenced voluntary international standards. NONIN Medical, Inc. has conducted an extensive Hazard Analysis and Risk Assessment, and developed extensive software/hardware verification and validation procedures to confirm the performance of the product to design requirements. Field evaluations have confirmed the accuracy of the Carbon Dioxide Detector in a controlled clinical environment.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features an abstract eagle design with three stylized lines representing the bird's body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The seal is presented in black and white.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 12 1999
Mr. Richard P. Bennett Nonin Medical, Inc. 2605 Fernbrook Lane North Plymouth, MN 55447-4755
Re: K982969 NONIN Pulse Oximeter and Carbon Dioxide Detectors: Model 9840-Series Regulatory Class: II (two) Product Code: 73 CCK and 74 DQA Dated: July 3, 1999 Received: July 8, 1999
Dear Mr. Bennett:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to leqally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such Failure to comply with the GMP regulation may result in assumptions. requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obliqation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Mr. Richard P. Bennett
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labelinq regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807,97), Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known)______________________________________________________________________________________________________________________________________________________
Device Name: NONIN® Pulse Oximeter and Carbon Dioxide Detector
# Indications for Use:
The NONIN® Pulse Oximeter and Carbon Dioxide Detector (Model 9840-Series) is indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, and approximate carbon dioxide (CO2) changes in the nemogrom (bpoz), paise rates, assess functions may be used separately or simultaneously.
# (Please Do Not Write Below This Line-Continue On Another Page If Needed) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use __
OR
Over -The-Counter Use
(Optional Format 1-2-96)
Joann A. Weathershen
Shipp, Sh. M.
Sion-Off Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number ***_***
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.