The Panoramic Oxygen Mask (POM) is a single patient, disposable device intended for delivering supplemental oxygen and monitoring expired gases from the patient, with ports to allow the clinician to insert scopes, probes, or tubes. It is for non-intubated, spontaneously breathing patients greater than 30 kg.
Device Story
The Panoramic Oxygen Mask (POM) is a multi-port, single-patient disposable mask for non-intubated, spontaneously breathing patients. It delivers supplemental oxygen via standard tubing and enables monitoring of exhaled gases (e.g., EtCO2) via a gas sampling line connected to a capnography monitor. The mask features integrated oral and nasal membranes (slit access ports) allowing clinicians to insert scopes, probes, or tubes while maintaining oxygen delivery and gas sampling. Used in hospitals, clinics, and pre-hospital settings. The device benefits patients by allowing airway access or diagnostic procedures without interrupting oxygen therapy or gas monitoring. Healthcare providers use the mask to facilitate clinical procedures while maintaining patient respiratory support and monitoring.
Clinical Evidence
Bench testing only. Comparative performance testing evaluated EtCO2 accuracy, respiration rate, and FiO2 across various simulated patient settings (BPM 12-20, Tidal Volume 300-500ml, CO2 concentrations 1-5%, O2 flow 1-12 Lpm) for both adult and pediatric sizes. Results demonstrated performance comparable to the predicate device. Biocompatibility testing (ISO 10993-5, -10, -11) and mechanical/aging/shelf-life testing were performed.
Technological Characteristics
Multi-port mask with oral/nasal membranes for instrument access. Materials: biocompatible, limited contact (<24 hours). Sensing: Luer slip fit for gas sampling. Connectivity: standard oxygen tubing and gas sampling line. Sterilization: not specified. Dimensions: Adult (198 ml) and Child (93 ml) internal volumes. Features one-way valves to prevent rebreathing.
Indications for Use
Indicated for non-intubated, spontaneously breathing patients >30 kg requiring supplemental oxygen, expired gas monitoring, and/or airway access via scopes, probes, or tubes.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
April 19, 2018
POM Medical, LLC % Paul Dryden Consultant ProMedic. LLC 131 Bay Point Dr NE St. Petersburg, Florida 33704
Re: K172365
Trade/Device Name: Panoramic Oxygen Mask (POM) Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: Class II Product Code: CCK Dated: March 17, 2018 Received: March 20, 2018
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
### K172365
Device Name
### Panoramic Oxvgen Mask (POM)
Indications for Use (Describe)
The Panoramic Oxygen Mask (POM) is a single patient, disposable device intended for delivering supplemental oxygen and monitoring expired gases from the patient, with ports to allow the clinician to insert scopes, probes, or tubes. It is for non-intubated, spontaneously breathing patients greater than 30 kg,
Type of Use (Select one or both, as applicable)
XX Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
EF PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
### 510(k) Summary Page 1 of 7 19-Apr-18
| Company: | POM Medical, LLC<br>1230 Madera Rd., Suite 5, Box 101<br>Simi Valley, CA 92065 |
|----------------------------|----------------------------------------------------------------------------------------------|
| Official Contact: | Jeff Voss, VP Operations<br>Tel - (855) 766-0202 |
| Proprietary or Trade Name: | Panoramic Oxygen Mask (POM) |
| Common/Usual Name: | Gas sampling oxygen mask |
| Classification Name: | 21 CFR 868.1400<br>Procode – CCK<br>Analyzer, gas, carbon-dioxide, gaseous phase<br>Class II |
| Predicate Device: | K133806 – Monitored Mask, M1 Capnography mask<br>CapnoVue Scope |
### Device Description:
The Panoramic Oxygen Mask (POM) is a multi-port mask that serves several functions:
- A standard oxygen mask for when a patient requires supplemental oxygen ●
- Sampling of exhaled gases for monitoring, typically end-tidal CO2 ●
- Additional ports (membranes) to allow for most types of scopes, probes, and tubes to be ● inserted while still delivering supplemental O2 and sampling exhaled gases.
The design of the POM's oral or nasal membranes allow access of a scope and are soft and pliable to help maintain the oxygen concentration to the patient while having scopes, etc. inserted through these oral or nasal membranes.
### Indications for Use:
The Panoramic Oxygen Mask (POM) is a single patient, disposable device intended for delivering supplemental oxygen and monitoring expired gases from the patient, with ports to allow the clinician to insert scopes, probes, or tubes. It is for non-intubated, spontaneously breathing patients greater than 30 kg.
### Patient Population:
For non-intubated, spontaneously breathing patients greater than 30 kg.
### Environment of Use:
Locations where procedures are performed where the patient requires supplemental oxygen, monitoring exhaled gases, and scope access
Hospital, sub-acute, clinic, physician offices, pre-hospital.
### Contraindications:
None.
{4}------------------------------------------------
## 510(k) Summary
### K172365
## Page 2 of 7
19-Apr-18
## Predicate Device Comparison - Table 1 - Comparison to the Predicate
| Attributes | Proposed<br>POM Panoramic Oxygen Mask | Predicate<br>Monitor Mask – K133806<br>M1 Capnography mask (Scope) | Comments |
|------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The Panoramic Oxygen Mask (POM) is<br>a single patient, disposable device<br>intended for delivering supplemental<br>oxygen and monitoring expired gases<br>from the patient, with ports to allow the<br>clinician to insert scopes, probes, or<br>tubes. It is for non-intubated,<br>spontaneously breathing patients<br>greater than 30kg. | The M1 Capnography Mask is a<br>single-use device intended for<br>delivering supplemental oxygen and<br>monitoring exhaled carbon dioxide in<br>non-intubated spontaneously breathing<br>patients. Standard oxygen tubing and<br>two female luer ports for gas sample<br>line attachment are included. | Both provide<br>- supplemental oxygen<br>- monitoring of exhaled gases<br><br>Subject device provides<br>- access port(s) for scopes ( CapnoVue<br>Scope model) |
| Patient Population | non-intubated spontaneously breathing<br>patients | non-intubated spontaneously breathing<br>patients<br>Child and adult | Similar |
| Environment of Use | Locations where procedures are<br>performed where the patient requires<br>supplemental oxygen, monitoring<br>exhaled gases, and scope access | Locations where procedures are<br>performed where the patient requires<br>supplemental oxygen, monitoring<br>exhaled gases | Scope access for the subject device |
| | Hospital, sub-acute, clinic, physician<br>offices, pre-hospital | Hospital, sub-acute, clinic, physician<br>offices, pre-hospital | |
| Duration of Use | Single patient, disposable | Single patient, disposable | Similar |
| Prescriptive | Yes | Yes | Similar |
| Mode of Operation | O2 delivery through standard oxygen<br>supply tubing and simultaneous<br>exhaled gas monitoring via a gas<br>sampling line connected from mask to<br>capnography or oxygen only delivery<br><br>Slit access ports allowing for<br>introduction of a scope | O2 delivery through standard oxygen<br>supply tubing and simultaneous<br>exhaled gas monitoring via a gas<br>sampling line connected from mask to<br>capnography or oxygen only delivery | Slit access ports allowing for introduction<br>of a scope are provided in the CapnoVue<br>Scope model |
| Attributes | Proposed<br>POM Panoramic Oxygen Mask | Predicate<br>Monitor Mask – K133806<br>M1 Capnography mask (Scope) | Comments |
| Components which may be<br>supplied or packaged with the<br>mask | Standard oxygen tubing<br>Gas sampling line | Standard oxygen tubing<br>Gas sampling line | Similar |
| Gas sampling connection | Luer slip fit | Luer lock | Similar<br>Both connector type accommodate<br>standard gas sampling luer fittings |
| Sizes | Child<br>Adult | Child<br>Adult | Similar |
| Profile | Over the nose / mouth | Over the nose / mouth | Similar |
| Face strap | Yes | Yes | Similar |
| Performance | | | |
| Internal volume | Child - 93 ml<br>Adult - 198 ml | Child - 73 ml<br>Adult - 159 ml | The difference in internal volume does not<br>affect performance of oxygen delivery or<br>monitoring of exhaled gases |
| Access Ports and Maximum<br>instrument size | Oral port – 60 mm ID / 20 mm OD<br>Nasal – 36 mm ID / 12 mm OD | Not available | Access port allows introduction of<br>instruments while delivering supplemental<br>oxygen and sampling expired gases |
| Entrainment Vents | One-way valves to prevent rebreathing | Open vents allowing room entrainment | Without the one-way valves the exhaled<br>gas monitoring performance is reduced.<br>For the intended use the designs are<br>similar.<br>There are oxygen masks marketed as non-<br>rebreathing mask with one-way valves to<br>prevent rebreathing. (Teleflex Non-<br>rebreathing mask)<br>This does not raise different questions of<br>safety or effectiveness. |
| Attributes | Proposed<br>POM Panoramic Oxygen Mask | Predicate<br>Monitor Mask – K133806<br>M1 Capnography mask (Scope) | Comments |
| %CO2 accuracy and<br>Respiration rate | Testing was done at different simulated<br>patient settings for breath rate, Tidal<br>Volume at different CO2 concentrations<br>with waveforms | | The results showed similar performance<br>with the proposed device. |
| Biocompatibility | External Communicating (indirect)<br>Tissue contact<br>And<br>Surface Communicating (direct)<br>Skin contact<br>Limited duration of use (< 24 hours)<br><br>Testing included<br>Cytotoxicity (ISO 10993-5:2009)<br>Sensitization (ISO 10993-10:2010)<br>Intracutaneous / Irritation (ISO 10993-<br>10:2010)<br>Acute Systemic Toxicity (ISO 10993-<br>11:2017) | External Communicating (indirect)<br>Tissue contact<br>And<br>Surface Communicating (direct)<br>Skin contact<br>Limited duration of use (< 24 hours) | Similar |
| Storage | -20° to + 50°C | Not specified | Testing supports the claim |
| Shelf-life | 3 years real-time aging and shelf-life | | |
| Effects of Aging | No effects of aging on performance | | |
{5}------------------------------------------------
510(k) Summary
### K172365
## Page 3 of 7
19-Apr-18
{6}------------------------------------------------
510(k) Summary
### K172365
# Page 4 of 7
19-Apr-18
{7}------------------------------------------------
### Substantial Equivalence Discussion
The POM is viewed as substantially equivalent to the predicate device (K133806 - M1 Capnography mask) based on the following discussions:
### Indications -
- The intended use is to provide supplemental oxygen, monitor exhaled gases, and provide ● scope access is similar to the predicate.
Discussion - Both provide supplemental oxygen, monitoring of exhaled gases The differences of the subject device having access port(s) for scopes or instruments has been shown to not affect performance related to supplemental oxygen delivery and EtCO2 sampling.
### Patient Population and environment of Use -
- The intended population is non-intubated spontaneously breathing patients in locations ● where procedures are performed that the patient may require supplemental oxygen, monitoring of exhaled gases and scope access.
- There are 2 sizes Child and adult mask .
Discussion - The population is similar to the predicate Monitor Mask - K133806 - M1 Capnography mask.
### Technology –
- Identical technology of utilizing a luer fitting to connect gas sampling lines,
- This device modifies existing masks with ports for scope access. .
Discussion - The technology is similar to the predicate Monitor Mask - K133806 - M1 Capnography mask.
### Differences -
The subject device offers access ports for scopes while the mask sits on the patient face. This technological characteristic is a combination of supplemental O2 and CO2 sampling facilitated by cannulas, which still allows scopes to access the patient airway. While access ports are different technological characteristic, the risk is the inability to use scopes to access airway. Labeling is provided to ensure compatibility of scopes with this device for airway access.
The differences between the predicate and the proposed device do not impact the performance of the device for supplemental oxygen delivery and EtCO2 sampling. These differences do not raise different questions of safety or effectiveness.
### Non-clinical performance testing
We have performed non-clinical comparative performance testing that included:
- Internal volume ●
- Evaluation of ability to measure EtCO2 and Respiration Rate are various simulated patient settings typical of pediatric and adult users.
- . Storage, Drop and Aging
- . Biocompatibility
### Biocompatibility / Materials -
{8}------------------------------------------------
### 510(k) Summary Page 6 of 7 19-Apr-18
The materials in patient / drug contact have been tested they are characterized as:
- . External Communicating (Indirect gas pathway), Tissue / Bone / Dentin communicating, And
- Surface Contact, Skin,
- . Duration of Use - limited (< 24 hours)
We performed the following tests with guidance from ISO 10993-1 and the results were acceptable.
- Cytotoxicity ISO 10993-5:2009 .
- Sensitization ISO 10993-10:2010 ●
- . Irritation (for surface contact materials) - ISO 10993-10:2010
- Acute Systemic Toxicity ISO 10993-11:2017 .
Discussion - The materials used in the subject device are familiar and standard materials utilized in devices for this intended use.
### Bench / Performance testing -
We performed comparative performance testing the tests included:
- Internal volume ●
- . Evaluation of ability to measure EtCO2 and Respiration Rate are various simulated patient settings typical of pediatric and adult users.
- Biocompatibility of Materials
- Mechanical Drop test ●
- . Effects of Aging on Performance
- Environmental Testing / Shelf-life
- ISO 594 Luer connector .
Tables 2 and 3 presents a summary of the testing performed.
| | | POM | | Predicate M1 | |
|-----------------|-----------------------|-------------------|-------------------|--------------|----|
| Adult | | EtCO2 | RR | EtCO2 | RR |
| | | Closed / Inserted | Closed / Inserted | | |
| Test conditions | BPM – 12, TV – 500 ml | | | | |
| 5% CO2 | O2 Flow – 1 Lpm | 7.37% / 6.57% | 12/12 | 6.43% | 12 |
| | O2 Flow – 8 Lpm | 5.17% / 4.78% | 12/12 | 2.50% | 12 |
| 1% CO2 | O2 Flow – 1 Lpm | 1.40% | 12 | 1.20% | 12 |
| | O2 Flow – 8 Lpm | 1.00% | 12 | 0.47% | 12 |
| | BPM – 20, TV – 300 ml | | | | |
| 5% CO2 | O2 Flow – 1 Lpm | 8.23% / 7.07% | 20/20 | 6.77% | 20 |
| | O2 Flow – 8 Lpm | 6.17% / 4.40% | 20/20 | 3.37% | 20 |
| 1% CO2 | O2 Flow – 1 Lpm | 1.63% | 20 | 1.27% | 20 |
| | O2 Flow – 8 Lpm | 1.13% | 20 | 0.60% | 20 |
| Internal Volume | | 198 ml | | 159 ml | |
### Table 2 - Summary of Comparative Testing - Adult Size
| Table 3 – Summary of FiO2 Testing – Adult Size | | | |
|------------------------------------------------|--|--|--|
|------------------------------------------------|--|--|--|
| | POM |
|-----------------|-----------------------|
| Adult | FiO2 |
| Test conditions | |
| | BPM - 12, TV - 400 ml |
| | Closed |
{9}------------------------------------------------
| | 510(k) Summary<br>Page 7 of 7<br>19-Apr-18 | |
|--------|--------------------------------------------|-------|
| 4% CO2 | O2 Flow – 8 Lpm | 77.6% |
| | O2 Flow – 10 Lpm | 83.4% |
| | O2 Flow – 12 Lpm | 87.0% |
### Table 4 - Summary of Comparative Testing - Pediatric size
| | | POM | | | | Predicate M1 | |
|-----------------|-----------------------|----------------------------|-------------------------|---------------------------|-------|--------------|--|
| Pediatric | | EtCO2<br>Closed / Inserted | RR<br>Closed / Inserted | FiO2<br>Closed / Inserted | EtCO2 | RR | |
| Test conditions | BPM - 12, TV – 500 ml | | | | | | |
| 5% CO2 | O2 Flow – 1 Lpm | 6.40% / 6.13% | 12 / 12 | 25.0% / 24.33% | 6.17% | 12 | |
| | O2 Flow – 8 Lpm | 3.40% / 3.17% | 12 / 12 | 48.67% / 49.0% | 4.07% | 12 | |
| 1% CO2 | O2 Flow – 1 Lpm | 1.10% | 12 | | 1.07% | 12 | |
| | O2 Flow - 8 Lpm | 0.73% | 12 | | 0.77% | 12 | |
| | BPM - 20, TV – 300 ml | | | | | | |
| 5% CO2 | O2 Flow - 1 Lpm | 6.40% / 6.23% | 20 / 20 | 23.67% / 24.67% | 6.03% | 20 | |
| | O2 Flow – 8 Lpm | 3.10% / 2.23% | 20 / 20 | 56.0% / 56.33% | 3.97% | 20 | |
| 1% CO2 | O2 Flow - 1 Lpm | 1.23% | 20 | | 1.17% | 20 | |
| | O2 Flow – 8 Lpm | 0.70% | 20 | | 0.77% | 20 | |
| Internal Volume | | 93 ml | | | 73 ml | | |
The results demonstrated that the device is substantially equivalent.
### Discussion of Differences and Substantial Equivalence Conclusion -
The differences in:
- Access ports
- While access ports are different technological characteristic, the risk is the o inability to use scopes to access airway. Labeling is provided to ensure compatibility of scopes with this device for airway access.
- Access ports impact on monitoring is characterized O
- Internal volume
- While the internal volume of the mask is different the measured results show that o the difference does not influence the measured performance.
- . One-way valves for non-rebreathing
- There are standard oxygen masks which contain one-way valves to prevent o rebreathing.
These differences do not raise different questions of safety or effectiveness.
In conclusion, the comparison to the predicate for features, indications for use, population, and comparative testing across the range of oxygen flow, simulated patient conditions for both adult and pediatrics, exposed to different CO2 levels demonstrated that both devices provide representative measurement of EtCO2 and respiratory rate and their respective waveforms were similar.
The results demonstrate equivalence in performance.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.