DualGuard™ is intended for use in Adult patients who require supplemental oxygen delivery and CO2 monitoring during endoscopy procedures and recovery. The complete device with bite block is to be used during endoscopy procedures. The DualGuard™ Bite Block is to be removed after endoscopy, leaving the O2 delivering/CO2 sampling cannula in place for patient recovery period.
Device Story
DualGuard is an endoscopic bite block with integrated nasal oxygen delivery and CO2 monitoring cannula; used during endoscopy procedures and patient recovery. Device inputs: supplemental oxygen source (via tubing) and patient respiratory gases (via nasal/oral sampling lines). Device outputs: oxygen delivery to patient; CO2 gas samples for external capnograph monitoring. Operated by clinicians in clinical settings. Bite block protects teeth during endoscopy; removable after procedure while cannula remains for continued monitoring/oxygenation. Features soft TPE over-molded bite block, self-sealing luer port for oral CO2 line, and adjustable head strap. Benefits include simultaneous oral/nasal CO2 sampling and secure, comfortable patient interface. Healthcare providers connect device to standard oxygen cylinders and capnograph machines; output data (CO2 levels) informs clinical respiratory management.
Clinical Evidence
Bench testing only. Evaluated against pre-determined acceptance criteria and recognized standards (BS EN 13544-2, ISO 594-1/2, BS EN ISO 10993-5/10, ASTM F1980). Tests included visual inspection, flow resistance, tensile strength, kinking resistance, dimensional gauging, leakage (air/liquid), luer performance, CO2 monitoring function, and accelerated aging. Biocompatibility testing (cytotoxicity, irritation, sensitization) confirmed safety. All samples passed.
Technological Characteristics
Materials: Medical grade PVC (tubing/cannula), Polypropylene (bite block), ABS (connectors), TPE (over-mold), ABS/Nylon (filter). Dimensions: 2.0m lines. Sensing: Passive gas sampling for external capnograph. Connectivity: Standard luer/oxygen connectors. Sterilization: Non-sterile. Software: None. Standards: ISO 594-1/2, BS EN 1707, BS EN 13544-2, BS EN ISO 10993, ASTM F1980.
Indications for Use
Indicated for adult patients requiring supplemental oxygen and CO2 monitoring during and after endoscopy procedures.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
{0}------------------------------------------------
Page 2 of 7
Cynon Valley Business Park, Mountain Mid Glamorgan. CF45 4ER UK. Tel: +44 (0) 1443 474647 APR 1 0 2014 Fax: +44 (0) 1443 474222 Customer Services: +44 (0) | 443 471580 Purchasing Fax: +44 (0) 1443 476332
· K140473
Email: enquiries@flexicare.com www.flexicare.com
# 510(k) Summary
510(k) Sponsor, Contact Person and Date Summary Prepared:
Flexicare Medical Limited Cynon Valley Business Park Mountain Ash, Mid. Glamorgan CF45 4ER. United Kingdom
Christopher Watkins Quality Regulatory & Technical Director Telephone: 00 44 1443 471 593 Fax: 00 44 1443 474 222
Summary prepared on: 10th April 2014
Device Name:
Trade Name: DualGuard™ Common/Usual Name: Endoscopic bite block and nasal oxygen cannula with CO2 monitoring accessory
Classification Name: Primary Classification: Carbon dioxide analyzer 21 CFR 868.1400 Primary Product Code: CCK Secondary Product Codes: Endoscopic bite block (MNK), Nasal oxygen cannula (CAT)
## Legally Marketed Equivalent Device:
The DualGuard™ is substantially equivalent to Trawax's TwinGuard® cleared under 510(k) K080527.
## Device Description:
DualGuard™ is an endoscopic bite block which is designed to fit into the nose and mouth of the patient during an endoscopy procedure.
DualGuard™ consists of a bite block with a removable oxygen delivering/ CO2 monitoring cannula, which when attached located under the patient nose. The bite block also has an Oral CO2 monitoring line attached at the mouth opening. DualGuard™ is able to deliver oxygen via the nose, whilst sampling CO2 from the nose and mouth.
When the Endoscopy is complete the bite block of DualGuard™ can be removed whilst leaving the cannula in place on patient, delivering oxygen and monitoring CO2.
{1}------------------------------------------------
Page 3 of 7
Image /page/1/Picture/2 description: The image shows the logo for "Flexicare". The logo is in black and white and features a stylized "F" followed by the word "lexicare" in a bold, sans-serif font. Below the word "lexicare" is the tagline "TOTAL QUALITY - TOTAL CARE" in a smaller font.
DualGuard™ has a soft TPE over molded Bite block to reduce the risk of damage to teeth, a tube slide for securing, and an adjustable head strap for a secure comfortable fit.
The CO2 line from the DualGuard™ has a self sealing "luersafe" luer port which allows the oral CO2 line to be removed with the bite block without any leaking of CO2 or ingress of air that may cause an inaccurate reading.
DualGuard™ is available in adult size only.
Intended Use:
DualGuard™ is intended for use in Adult patients who require supplemental oxygen delivery and CO2 monitoring during endoscopy procedures and recovery.
The complete device with bite block is to be used during endoscopy procedures. The DualGuard™ Bite Block is to be removed after endoscopy, leaving the O2 delivering/CO2 sampling cannula in place for patient recovery period.
Substantial Equivalence:
DualGuard™ has the same intended use as the predicate device.
Both the DualGuard and TwinGuard are single patient use devices. Supplied in Adult size onlv.
Neither device is life supporting or life sustaining.
Patient Contact -Bite block - Invasive - Mucosal membrane - Limited duration (>24hrs) Cannula & tubing - skin contact (intact) and externally communicating - Limited duration (>24hrs)
Neither the DualGuard nor the TwinGuard uses software/ neither are electronically driven. Both devices are supplied non-sterile in individual poly bags.
Both devices are able to be used with industry standard devices such as oxygen cylinders/connectors and Capnograph machines.
Both the DualGuard and TwinGuard devices are designed for the same intended use in the same intended conditions.
Both designs consist of 10+ components made from injection molded & extruded plastics.
From comparison testing it was determined that the only invasive components of the two devices were the bite blocks.
The DualGuard bite blocks are manufactured from Polypropylene with a soft TPE section over molded.
TwinGuard bite blocks are also manufactured from Polypropylene but do not have an over molded section.
The Oxygen tubing, cannula, CO2 line and oxygen connector of both the DualGuard and TwinGuard devices are manufactured from PVC.
The Trawax "TwinGuard" bite block is Green in color with clear tubing. The Flexicare Duale, Luard TM in Dlun 9 aranan in antar uith aloar fuhina Thia diffaran
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for "Flexicare". The logo has a stylized "F" on the left, followed by the text "lexicare" in a serif font. Below the name is the text "TOTAL QUALITY - TOTAL CARE" in a smaller, sans-serif font.
manufacturer's choice/ branding, and is not related to sizing, intended use, gender of patient or performance of device.
Both the DualGuard and Predicate devices have 2M oxygen & CO2 lines.
| | Flexicare DualGuard™ | Trawax "TwinGuard" |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | K: Unknown | K080527 |
| Components | O2 tube | O2 tube |
| | O2 connector | O2 connector |
| | CO2 line | CO2 line |
| | CO2 luer | CO2 luer |
| | Bite block | Bite block |
| | TPE over mold | Nasal cannula |
| | Nasal cannula | Head strap |
| | Head strap | Tube slide |
| | Tube slide | Oral CO2 tube link |
| | Oral CO2 tube link | |
| | Self sealing CO2 Y-piece | |
| | Oral CO2 luer | |
| | 0.45µm Hydrophobic filter | |
| | Medical grade PVC | |
| Materials | | Medical grade PVC |
| | Medical grade Polypropylene | Medical grade Polypropylene |
| | Medical grade ABS | Medical grade ABS |
| | Medical grade TPE | |
| | Medical grade ABS/Nylon (filter) | |
| Target | Adult | Adult |
| population | | |
| Home use | No | No |
| Emergency use | Yes | Yes |
| Assembly | Bonded - Solvent adhesive | Bonded - Solvent adhesive |
| method | | |
| Connectable to | Yes | Yes |
| Capnograph? | | |
| Monitors CO2? | Yes | Yes |
| Nasal CO2 | Yes | Yes |
| monitoring? | | |
| Oral CO2 | Yes | Yes |
| monitoring? | | |
| Use | Single use | Single use |
| Supplied sterile | No | No |
| | | |
| Soft molded bite | Yes | No |
| area? | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| Indications for<br>use | DualGuard™ is intended for use in Adult<br>patients who require supplemental oxygen<br>delivery and CO2 monitoring during<br>endoscopy procedures and recovery.<br><br>The complete device with bite block is to be<br>used during endoscopy procedures.<br>The DualGuard™ Bite Block is to be removed<br>after endoscopy, leaving the 02<br>delivering/CO2 sampling cannula in place for<br>patient recovery period. | TwinGuard®, a bite block with a nasal oxygen<br>cannula, is used during endoscopy<br>procedures to administer continuous oxygen<br>to the nose and mouth.<br><br>After endoscopy the bite block is removed,<br>leaving the cannula in place during recovery,<br>to continue delivering oxygen to the nose<br>and mouth.<br><br>In addition, the TwinGuard® contains an<br>optional accessory which collects samples of<br>the patient's breathing at the nose and<br>mouth to measure CO2 with a capnograph<br>during endoscopy procedures. It is intended<br>for adult patients. |
| Prescription Use? | Yes | Yes |
| Processes carried<br>out | Injection moulding<br>Tube extrusion<br>Solvent bonding<br>Packaging | Injection moulding<br>Tube extrusion<br>Solvent bonding<br>Packaging |
| Component<br>interactions | O2 & CO2 Tube - non-Invasive, skin contact<br>(intact) and Externally communicating.<br>Limited use (<24hr)<br><br>Connectors - skin contact (hands), limited<br>use (<24h)<br><br>Bite Block - invasive (natural orifice)<br>mucosal membrane, limited use(<24h) | O2 & CO2 Tube - non-Invasive, skin contact<br>(intact) and Externally communicating.<br>Limited use (<24hr)<br><br>Connectors - skin contact (hands), limited<br>use (<24h)<br><br>Bite Block - invasive (natural orifice)<br>mucosal membrane, limited use(<24h) |
| Storage<br>conditions | Store in a cool, dry place out of direct<br>sunlight | Store in a cool, dry place out of direct<br>sunlight |
| Available sizes | Adult | Adult |
| Available lengths | 2.0m | 2.0m |
| O2 tube<br>dimensions | ID: 3.9mm<br>OD: 5.0mm | ID: 3.9mm<br>OD: 5.6mm |
| CO2 tube<br>dimensions | ID: 1.2mm<br>OD: 2.7mm | ID: 1.4mm<br>OD: 2.7mm |
| Colour | Blue, Orange, Clear/Colorless | Green, Clear/ Colorless |
{3}------------------------------------------------
Page 5 of 7
Image /page/3/Picture/2 description: The image shows the logo for Flexicare. The logo consists of a stylized "F" followed by the word "lexicare" in a sans-serif font. Below the word "lexicare" is the tagline "TOTAL QUALITY - TOTAL CARE" in a smaller font.
.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Flexicare. The logo consists of a stylized "F" symbol followed by the word "lexicare" in a bold, sans-serif font. Below the word "lexicare" is the tagline "TOTAL QUALITY - TOTAL CARE" in a smaller font. The logo is simple and professional, and it conveys a sense of quality and care.
Although very similar in design and function, there are some differences (described below) between the Flexicare DualGuard™ and the predicate device. However, these differences do not raise and new safety or effectiveness issues.
The Flexicare DualGuard™ has an over molded soft TPE section around the bite area of the bite block whilst the predicate Trawax "TwinGuard" does not have this feature.
The Flexicare DualGuard TM also differs from the Trawax "TwinGuard" predicate as DualGuard™ includes a nasal CO2 sampling cannula. This cannula is bonded side-by-side with the Oxygen delivering cannula and samples CO2 through its prongs that locate beneath the nostrils. This nasal sampling cannula is not present on the predicate device, which obtains its nasal CO2 samples via a plug & monitoring line that is inserted into one nostril. This does however restrict oxygen delivery to one nostril.
Image /page/4/Picture/5 description: The image shows two different medical devices, labeled as "DualGuard™M" and "TwinGuard". The DualGuard device has three tubes connected to a central piece, with arrows indicating the direction of flow. The TwinGuard device has a knob on top and a black, circular base with a hole in the center.
DualGuard™ also features a self sealing Y-Piece, allowing CO2 to be sampled from the nose and mouth simultaneously or from the nose only, without the need to cap a disconnected port.
Another difference between DualGuard™ and its predicate device "TwinGuard", although not critical, is strap securing locations. Flexicare's device features flexible molded wings which the strap secures onto, which keeps the bite block secure in the mouth. The "TwinGuard" features a hook on location each side of the bite block to secure head strap.
{5}------------------------------------------------
Page 7 of 7
Image /page/5/Picture/2 description: The image shows the logo for "flexicare". The logo has a stylized "f" on the left side of the word. Below the word "flexicare" is the phrase "TOTAL QUALITY - TOTAL CARE".
# Summary of Testing:
DualGuard™ has been evaluated in accordance with standards listed in table:
| Test | Standard / Pre-Determined<br>Acceptance Criteria | Outcome |
|------------------------------------------------------------------------|---------------------------------------------------|-------------------------------------------------------------|
| Visual inspection | Pre-Determined Acceptance<br>Criteria | All Samples pass |
| Tubing flow resistance | | All Samples pass |
| O2 Connector to tubing tensile strength | BS EN13544-2: 2002<br>+A1:2009. | All Samples pass |
| O2 Connector to nipple tensile strength | | All Samples pass |
| Tube resistance to kinking | | All Samples pass |
| Dimensional inspection of luer conical | ISO 594-1 / BS EN 20594-<br>1:1994. | All Samples pass |
| Dead Space | Pre-Determined Acceptance<br>Criteria | DualGuard Dead<br>space is less than<br>marketed predicate. |
| Dimensional inspection of luer conical | | All Samples pass |
| Gauging tests on luer | | All Samples pass |
| Liquid leakage from luer | | All Samples pass |
| Air leakage from luer | | All Samples pass |
| Luer separation force | | All Samples pass |
| Luer unscrewing torque | ISO 594-2 (BS EN 1707:1997) | All Samples pass |
| Luer ease of assembly | | All Samples pass |
| Luer resistance to overriding | | All Samples pass |
| Luer testing for stress cracking | | All Samples pass |
| Leak testing of all bonded components | Pre-Determined Acceptance<br>Criteria | All Samples pass |
| Co2 monitoring function test | Pre-Determined Acceptance<br>Criteria | All Samples pass |
| Tensile testing - CO2 & O2 lines to<br>connector and/or cannula prongs | Pre-Determined Acceptance<br>Criteria | All Samples pass |
| Dimensional characteristics | Pre-Determined Acceptance<br>Criteria | All Samples pass |
| Accelerated age testing | ASTM F1980 | All Samples pass |
| Cytotoxicity, Irritation, sensitization | BS EN ISO 10993-10:2010<br>BS EN ISO 10993-5:2009 | All Samples pass |
All Samples passed the performance testing when tested against methods and criteria from pre-determined acceptance criteria In-House test methods and relevant BS EN, ISO & ASTM FDA recognized standards.
The results of this testing show that DualGuard™ passes all performance tests and performs at least as well as the marketed predicate device "TwinGuard" by Trawax
The conclusions drawn from the nonclinical performance comparison testing demonstrate that DualGuard™ is as safe, as effective, and performs as well as or better than the legally marketed Predicate device "TwinGuard" by Trawax.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features a stylized caduceus, a symbol often associated with medicine and healthcare, with three parallel lines that curve and flow. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the symbol in a circular fashion.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 10, 2014 ,
Flexicare Medical Limited c/o Mr. Mark Job Regulatory Technology Services, LLC 1394 25th Street NW Buffalo, MN 55313
Re: K140473
Trade/Device Name: DualGuardTM Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: II Product Code: CCK, MNK, CAT Dated: March 25, 2014 Received: March 26, 2014
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 - Mr. Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Image /page/7/Picture/8 description: The image contains a logo with the text "Tejashri Purohit-Sheth, M.D." and "Clinical Deputy Director". Below this text is the word "DAGRID" in a stylized font. The word "FOR" is located in the bottom right corner of the image. The logo appears to be a stamp or seal.
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
## 510(k) Number (if known)
K140473
### Device Name DualGuardin
#### Indications for Use (Describe)
DualCourd™ is intended for use in Adult patients who require supplemental oxygen delivery and CO2 monitoring during endoscopy procedures and recovery.
The complete device with bite block is to be used during endoscopy procedures.
The DualGuard™ Bitc Block is to be removed after endoscopy, baving the O2 delivering/CO2 sampling cannula in place for palient recovery period.
Type of Use (Select one or both, as applicable)
🇿 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
| | Todd D-Courtney-S |
|--|-----------------------------|
| | 2014.04.09 12:58:53 -04'00' |
FORM FDA 3881 (9/13)
PSC Publishing Services ( 1015 641474) - EF
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.