CO2 MONITORING LINE, CO2 MONITORING LINE WITH FILTER
Applicant
Flexicare
Product Code
CCK · Anesthesiology
Decision Date
Apr 30, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1400
Device Class
Class 2
Indications for Use
CO2 Monitoring lines are intended to connect from a patient interface CO2 sampling port to the expired gas monitor.
Device Story
Single-patient-use, small-diameter PVC tubing with luer lock connectors; used to transport exhaled breath from a patient interface to a gas analyzer (e.g., capnograph). Optional 0.45µm hydrophobic filter prevents water vapor transfer to the analyzer. Operated in hospital settings by clinicians. Device relies on the external gas analyzer's pump to draw gas through the line. Output is the delivery of gas samples to the analyzer for CO2 monitoring. Benefits include facilitating continuous respiratory gas monitoring for patient safety.
Clinical Evidence
Bench testing only. Testing included visual/dimensional inspection, resistance to flow (at 100ml/min and 300ml/min), leakage, CO2 monitoring performance, tensile testing (40N), and accelerated aging (ASTM F1980). Biocompatibility testing performed per ISO 10993-3, 5, and 10. All samples passed pre-determined acceptance criteria.
Technological Characteristics
Flexible PVC tubing; ABS luer connectors; 2.0m or 3.0m lengths; 1.2mm ID / 2.7mm OD. Optional 0.45µm hydrophobic filter. Non-sterile, single-use. Compliant with ISO 594-2 (BS EN 1707:1997) and ISO 10993-3&5. No software, electronics, or energy source.
Indications for Use
Indicated for any patient requiring gas monitoring. Used to connect a patient interface CO2 sampling port to an expired gas monitor in a hospital setting.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
Predicate Devices
Intersurgical - CO2 Monitoring line and CO2 monitoring line with filter (K122075)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for "Flexicare". The logo is in black and white and features a stylized "F" with two curved lines above it. The word "Flexicare" is written in a bold, sans-serif font to the right of the "F". Below the word "Flexicare" is the tagline "TOTAL QUALITY - TOTAL CARE" in a smaller font.
#### 510(k) Summary
Trade name
C.F.R Section: Device tvpe: Product Code:
Manufacturer:
Regulatory Affairs Contact:
Christopher Watkins Quality Requlatory & Technical Director Flexicare Medical Limited Cynon Valley Business Park Mountain Ash, Mid. Glam. CF45 4ER. United Kingdom
CO2 monitoring line & CO2 monitoring line with filter
Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase
Telephone:
Date Summary Prepared:
Common Name:
Classification:
Predicate Device
Description:
00 44 1443 471 593
22nd April 2014
CO2 monitoring line
Class 2
K133516
сск
21 CFR 868.1400.
Flexicare Medical Limited
Cynon Valley Business Park Mountain Ash, Mid. Glam. CF45 4ER. United Kingdom
Intersurgical - CO2 Monitoring line and CO2 monitoring line with filter, which was cleared for marketing by 510(k) No K122075
Flexicare's CO2 monitoring line is single patient use, small diameter tubing intended to be connected to a luer lock connector on a patient interface to allow for gas sampling from a patients exhaled breath by gas sampling equipment. The Gas monitoring/analysing device will feature a pump, which draws expired breath from the patient through the monitoring line and into the gas analyser.
Flexicare's CO2 monitoring line is available with or without a 0.45um hydrophobic filter, which prevents the transfer of water vapour down the monitoring line and into the gas sampling equipment.
Flexicare's CO2 monitoring lines feature clear construction to aid visual inspection of the device.
The CO2 monitoring line is a single patient use device, designed for use with one patient over a single course of treatment. The CO2 monitoring line is supplied in a sealed poly bag in a non sterile state and is not to be sterilized.
The Flexicare CO2 monitoring lines feature 2M & 3M length tubes.
# APR 3 0 2014
1
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Flexicare. The logo is black and white and features a stylized "F" followed by the word "lexicare" in a bold, sans-serif font. Below the word "lexicare" is the tagline "TOTAL QUALITY - TOTAL CARE" in a smaller font.
#### Part Numbers:
| Part Number | Product description |
|-------------|----------------------------------------------------------------------|
| 010-700U | CO2 MONITORING LINE WITH MALE/MALE CONNECTORS - 2M |
| 010-705U | CO2 MONITORING LINE WITH FEMALE/FEMALE CONNECTORS - 2M |
| 010-710U | CO2 MONITORING LINE WITH MALE/FEMALE CONNECTORS - 2M |
| 010-700XU | CO2 MONITORING LINE WITH MALE/MALE CONNECTORS - 3M |
| 010-700AU | CO2 MONITORING LINE WITH MALE/MALE CONNECTORS - 2M - WITH FILTER |
| 010-705AU | CO2 MONITORING LINE WITH FEMALE/FEMALE CONNECTORS - 2M - WITH FILTER |
| 010-710AU | CO2 MONITORING LINE WITH MALE/FEMALE CONNECTORS - 2M - WITH FILTER |
| 010-700XAU | CO2 MONITORING LINE WITH MALE/MALE CONNECTORS - 3M - WITH FILTER |
#### Sizing
| Part Number | Length (mm) | Tube OD (mm) | Tube ID (mm) |
|-------------|-------------|--------------|--------------|
| 010-700U | 2000 | 2.7 | 1.2 |
| 010-705U | 2000 | 2.7 | 1.2 |
| 010-710U | 2000 | 2.7 | 1.2 |
| 010-700XU | 3000 | 2.7 | 1.2 |
| 010-700AU | 2000 | 2.7 | 1.2 |
| 010-705AU | 2000 | 2.7 | 1.2 |
| 010-710AU | 2000 | 2.7 | 1.2 |
| 010-700XAU | 3000 | 2.7 | 1.2 |
Intended Use:
CO2 Monitoring lines are intended to connect from a patient interface CO2 sampling port to the expired gas monitor.
Substantial Equivalence
Flexicare's CO2 monitoring line & CO2 monitoring line with filter have the same intended use as predicate device:
Intersurgical - CO2 Monitoring line and CO2 monitoring line with filter, which was cleared for marketing by 510(k) No K122075
Both Flexicare's CO2 monitoring lines and Intersurgical CO2 monitoring lines are single patient use devices.
Neither Flexicare's CO2 monitoring lines or Intersurgical CO2 monitoring lines are life supporting or life sustaining.
Neither Flexicare's CO2 monitoring lines or Intersurgical CO2 monitoring lines use software, and are not electronically driven.
Both Flexicare's CO2 monitoring lines and Intersurgical CO2 monitoring lines are supplied non- sterile in individual poly bags.
Both Flexicare's CO2 monitoring lines and Intersurgical CO2 monitoring lines are designed for the same intended use in the same intended conditions.
Both Flexicare's CO2 monitoring lines and Intersurgical CO2 monitoring lines consist of 3-4 components made from injection molded & extruded plastics.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for "Flexicare". The logo is black and white and features a stylized "F" with two curved lines. The word "Flexicare" is written in a bold, sans-serif font to the right of the "F". Below the word "Flexicare" is the phrase "TOTAL QUALITY - TOTAL CARE" in a smaller font.
Both Flexicare's CO2 monitoring lines and Intersurgical CO2 monitoring lines are supplied either with or without 0.45µm hydrophobic filters. Flexicare's hydrophobic filter appeared to be identical to the Intersurgical filters in appearance and product marking. It is likely they are sourced from the same supplier.
From the performance testing it was discovered both Flexicare's and the predicate device tubes are manufactured from flexible PVC, and that both devices feature male and female luers with the option of a hydrophobic filter.
The technological characteristics of Flexicare's CO2 monitoring line and Intersurgical's CO2 monitoring line are very similar. These similarities are evident in the comparison table below.
| | Flexicare - CO2 monitoring line<br>K133516 | Intersurgical - CO2 Monitoring Line<br>K122075 |
|-------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Intended use | | |
| Target<br>Population | Any patient from which gas monitoring is<br>required | Any patient from which gas monitoring is required |
| Indications for,<br>use | CO2 Monitoring lines are intended to<br>connect from a patient interface CO2<br>sampling port to the expired gas monitor. | The CO2 monitoring lines are intended to connect<br>from a CO2 sampling port to the expired gas<br>monitor |
| Place of use | Hospitals | Hospitals |
| Product labeling | CO2 monitoring line | CO2 monitoring line |
| Single use or re-<br>usable? | Single use | Single use |
| Design and<br>performance | | |
| Resistance to<br>flow | 8.8 mbar at 100ml/min flow<br>30.2 mbar at 300ml/min flow | 15.2 mbar at 100ml/min flow<br>51.2 mbar at 300ml/min flow |
| Leakage | <1.0ml/min | <1.0ml/min |
| Connectors | 2x luer lock connectors - male or female | 2x luer lock connectors- male or female |
| Outer diameter | 2.7mm | 3.05mm |
| Inner diameter | 1.2mm | 1.47mm |
| Lengths available | 2.0m, 3.0m | 0.7m, 1.8m, 2.45m, 3.0m |
| Standards | ISO 594-2 (BS EN 1707:1997) , ISO 10993-<br>3&5 | ISO 594-2 (BS EN 1707:1997) , ISO 10993-3&5 |
| Processes | Extrusion, injection moulding, adhesive<br>bonding | Extrusion, injection moulding, adhesive bonding |
| Colours | None - colourless | None - colourless |
| Energy<br>Used/delivered: | Gas is pulled away from the patient by a<br>pump in the gas sampling device | Gas is pulled from one end of the tube to the other<br>by a pump in the gas sampling device |
| Compatability : | Designed for use with gas monitoring<br>device (for example a Capnograph) with<br>luer connections to gas sampling tubing | Designed for use with gas monitoring device (for<br>example a Capnograph) with luer connections to<br>gas sampling tubing |
| Materials : | PVC (main line) and ABS (luer connectors) | PVC (main line) and Polycarbonate (luer<br>connectors) |
| Biocompatibility: | Compliant with ISO 10993-3&5 | Compliant with 10993-3&5 |
| Sterility: | Non sterile | Non sterile |
| Storage | Store in a cool, dry place out of direct sunlight | Store in a cool, dry place out of direct sunlight |
| Component<br>interactions | Tube – intact skin contact<br>Connector - intact skin contact ( whilst connecting to analyzing machine | Tube - intact skin contact<br>Connector - intact skin contact ( whilst connecting to analyzing machine |
| Intended use | Any patient from which gas monitoring is required | Any patient from which gas monitoring is required |
| Target<br>Population | Any patient from which gas monitoring is required | Any patient from which gas monitoring is required |
| Indications for<br>use | CO2 Monitoring lines are intended to connect from a patient interface CO2 sampling port to the expired gas monitor. | The CO2 monitoring lines are intended to connect from a CO2 sampling port to the expired gas monitor |
| Place of use | Hospitals | Hospitals |
| Product labeling | CO2 monitoring line + Hydrophobic filter | CO2 monitoring line + Hydrophobic filter |
| Single use or re-usable? | Single use | Single use |
| Design and<br>performance | | |
| Resistance to<br>flow | 30.0 mbar at 100ml/min flow<br>95.7 mbar at 300ml/min flow | 33.4 mbar at 100ml/min flow<br>114.5 mbar at 300ml/min flow |
| Leakage | <1.0ml/min | <1.0ml/min |
| Connectors | 2 x luer lock connectors - male or female (plus luer lock male to luer lock female connection between monitoring line and filter) | 2 x luer lock connectors - male or female (plus luer lock male to luer lock female connection between monitoring line and filter) |
| Outer diameter | 2.7mm | 3.05mm |
| Inner diameter | 1.2mm | 1.47mm |
| Length | 2.0m, 3.0m | 0.7m, 1.8m, 2.45m, 3:0m |
| Standards | ISO 594-2 (BS EN 1707:1997), ISO 10993-3&5 | ISO 594-2 (BS EN 1707:1997), ISO 10993-3&5 |
| Processes | Extrusion, injection moulding, adhesive bonding | Extrusion, injection moulding, adhesive bonding |
| Colours | None - colourless | None - colourless |
| Energy<br>Used/delivered: | Gas is pulled away from the patient by a pump in the gas sampling device | Gas is pulled from one end of the tube to the other by a pump in the gas sampling device |
| Compatability : | Designed for use with gas monitoring device (for example a Capnograph) with luer connections to gas sampling tubing | Designed for use with gas monitoring device (for example a Capnograph) with luer connections to gas sampling tubing |
| Materials : | PVC (main line) and ABS (luer connectors) | PVC (main line) and Polycarbonate (luer connectors) |
| Biocompatibility: | Compliant with ISO 10993-3&5 | Compliant with ISO 10993-3&5 |
| Sterility: | Non sterile | Non sterile |
| Storage | Store in a cool, dry place out of direct sunlight | Store in a cool, dry place out of direct sunlight |
| Component<br>interactions | Tube – intact skin contact<br>Connector - intact skin contact (whilst connecting to analyzing machine | Tube - intact skin contact<br>Connector - intact skin contact (whilst connecting to analyzing machine |
| Filter - intact skin contact (whilst<br>connecting to analyzing machine | Filter - intact skin contact (whilst connecting to<br>analyzing machine | |
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for "Flexicare". The logo has a stylized "F" on the left side, followed by the word "lexicare" in a sans-serif font. Below the word "lexicare" is the text "TOTAL QUALITY - TOTAL CARE" in a smaller font.
:
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Flexicare. The logo consists of a stylized letter "f" followed by the word "lexicare" in a serif font. Below the word "lexicare" is the phrase "TOTAL QUALITY : TOTAL CARE" in a smaller, sans-serif font.
From the table above it is evident that the CO2 monitoring line devices from both manufacturers have very similar technological characteristics. Both devices feature the same number of components, have been made using the same manufacturing methods, have been manufactured from the same (or very similar) materials and critical dimensions (tube ID, Luer dimensions) are identical. Any differences between the Flexicare and Intersurgical devices are stated below with justifications.
#### Differences
Both devices are simple products, comprising of 3 components each (4 each if with filter). There are very little differences between the two devices. Below are the only differences discovered.
The luer connectors of the Intersurgical predicate device are Polycarbonate, while Flexicare's luers are ABS. This difference in material is not crucial as both materials are amorphous polymers and have similar properties including high clarity, hardness, high impact strength and very low mould shrinkage.
The only other difference between the devices is that the Intersurgical sample has a frosted finish tube, which is purely a surface finish difference.
| Test | Standard / Pre-determined<br>Acceptance Criteria | Outcome |
|------------------------------------------------------------------------|--------------------------------------------------------------------------------|------------------|
| Visual inspection | Pre-determined Acceptance Criteria | All samples pass |
| Resistance to flow | Pre-determined Acceptance Criteria | All samples pass |
| Dimensional inspection. Length, ID, OD | Pre-determined Acceptance Criteria | All samples pass |
| Leakage | Pre-determined Acceptance Criteria | All samples pass |
| Performance - CO2 monitoring | Pre-determined Acceptance Criteria | All samples pass |
| Tensile testing | Pre-determined Acceptance Criteria | All samples pass |
| Luer connectors - Dimensional,<br>leakage, connection, stress cracking | ISO 594-1 / BS EN 20594-1:1994<br>ISO 594-2 (superseded by BS EN<br>1707:1997) | All samples pass |
| Accelerated age testing | ASTM F1980 | All samples pass |
| Cytotoxicity, Irritation, sensitization | BS EN ISO 10993-10:2010<br>BS EN ISO 10993-5:2009 | All samples pass |
| Genotoxicity | BS EN ISO 10993-3:2009 | All samples pass |
Non Clinical Tests Performed:
All the tests that were carried out on Flexicare's CO2 monitoring line are listed in the table above.
The tests selected to be carried out in order to determine substantial equivalence would test the critical aspects of the each device's functionality.
The visual & dimensional inspections were carried out to ensure that the Flexicare device had the same number of components, the same general assembly, the same material type and was dimensionally the same as (or very similar to) the predicate device. Both manufacturers' devices were found to be very similar in appearance, assembly, materials and dimensions with critical dimensions such as tube ID being identical.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Flexicare. The logo is black and white and features a stylized "F" followed by the word "lexicare" in a bold, sans-serif font. Below the word "lexicare" is the tagline "TOTAL QUALITY - TOTAL CARE" in a smaller font.
6
Resistance to flow testing was carried out to compare the level of resistance between Flexicare's CO2 monitoring line with Intersurgical's at the same flow rates. This aspect is critical as a higher resistance would restrict CO2 reaching the gas analyzer. Testing was carried out at 100ml/min and again at 300ml/min.
CO2 monitoring testing was carried out in order to assess whether the devices would detect CO2 from beneath the nostrils as this is the main purpose of the device. Samples were not assessed for % of CO2 detected as this depends on the patient-end device that the CO2 monitoring line is attached to. Both manufacturers' devices detected CO2 consistently, proving substantial equivalence.
Leak and tensile testing were also carried out as a leak or disconnection from sampling line to luer could result in less or no CO2 being drawn to the gas sampling device. Both Flexicare's and Intersurgical's devices passed the leak testing and withstood the pre-determined acceptance criteria of 40N tensile force without disconnection. This also proves substantial equivalence of the devices.
#### Conclusion
Flexicare's & Intersurgical's CO2 monitoring line & CO2 monitoring line with filter passed the performance testing when tested against methods and criteria from both pre-determined acceptance criteria test methods and relevant standards. The results of this testing show that the CO2 monitoring line & CO2 monitoring line with filter passes all performance tests that are critical to its substantially equivalent operation. In all tests the Flexicare device performed as well as or better than the Predicate device.
Flexicare's CO2 monitoring line & CO2 monitoring line with filter is substantially equivalent to, and performs as well as the marketed predicate device manufactured by Intersurgical.
Signature of contact Person:
Chris Watkins Quality Regulatory & Technical Director
euashano
Dated: 22-04-2014
-------End of 510k Summary----------------------------------------------------------------------------------------------------------------------------------------------------
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 30, 2014
Flexicare Medical Limited Christopher Watkins Ouality Regulatory and Technical Director Cynon Vallev Business Park Mountain Ash, Mid. Glamorgan CF45 4ER. United Kingdom
Re: K133516
Trade Name: CO2 Monitoring Line & CO2 Monitoring Line with Filter Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: Class II Product Code: CCK Dated: March 26. 2014 Received: March 31, 2014
Dear Mr. Watkins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
### Page 2 - Mr. Watkins
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note
the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Ticket Park Shah M.D.
Tejashri Purohit Sheth, MiD.
Clinical Deputy Director,
DAGRID/ODE/CDRH FOR
Erin Keith, M.S. Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Flexicare. The logo consists of a stylized letter "f" with two curved lines above it, followed by the word "lexicare" in a bold, sans-serif font. Below the word "lexicare" is the tagline "TOTAL QUALITY - TOTAL CARE" in a smaller, sans-serif font.
## Indications for Use
510(k) Number (if known): _K133516
Page 1 of 1
CO2 monitoring line & CO2 monitoring line with filter Device Name:
Indications for Use:
CO2 Monitoring lines are intended to connect from a patient interface CO2 sampling port to the expired gas monitor.
Prescription Use
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Todd D. Cour
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.