The VEO Multigas Monitor for Pocket PC is intended to provide monitoring of carbon dioxide and oxygen during anesthesia, recovery and respiratory care. It may be used in the operating suite, intensive care unit, patient room, and emergency medicine settings for adult and pediatric patients.
Device Story
VEO Multigas Monitor for Pocket PC performs mainstream infrared gas analysis for CO2 and uses an oxygen fuel cell for O2 monitoring. Device consists of a transducer attached to a breathing circuit via an airway adapter. It transmits data to an off-the-shelf Pocket PC for display of numeric and graphic information. Used in OR, ICU, patient rooms, and emergency settings by clinicians. Provides real-time gas concentration data to assist in patient monitoring during anesthesia and respiratory care. Benefits include rapid response, accuracy, and portability for patient management.
Clinical Evidence
Bench testing only. Non-clinical testing conducted in direct comparison to predicate devices across the full operating range using calibrated gas samples and legally marketed anesthesia and ventilation equipment. Results demonstrated performance, safety, and effectiveness equivalent or superior to predicates.
Technological Characteristics
Mainstream infrared gas analysis bench; oxygen fuel cell; airway adapter interface; off-the-shelf Pocket PC display; numeric and graphic output; disposable single-patient-use airway adapters.
Indications for Use
Indicated for monitoring carbon dioxide and oxygen levels in adult and pediatric patients during anesthesia, recovery, and respiratory care in operating suites, ICUs, patient rooms, and emergency settings.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
Predicate Devices
Tidal Wave Model 610, Novametrix Medical Systems Inc. (K963327)
{0}------------------------------------------------
K051857
## 510(k) SUMMARY
- 510(k) Owner Name and Address: 1. PHASEIN AB Svärdvägen 15 182 33 Danderyd Sweden Telephone: 46-8-544-98-150 Fax: 46-8-544-98-169
- Contact Person: 2. David Weissburg Weissburg Associates Madison, Wisconsin Telephone: 1-608-770-0223
- Date: 6 September 2005 ന്
- Trade Name: VEO Multigas Monitor for Pocket PC 4.
- Common Name: Multigas Monitor 5.
- Classification Names: 6.
- a. Carbon dioxide gas analyzer (21 CFR 868.1400, Product Code CCK)
- Oxygen gas analyzer (21 CFR 868.1720, Product Code CCL) ﺗﻘ
- Substantially equivalent to: 7.
- Tidal Wave Model 610, Novametrix Medical Systems Inc. (K963327) a.
- MX300 Portable Oxygen Monitor, Teledyne Analytical Instruments (K024155) ﻗ
- S/5 Multigas Monitor, Datex-Ohmeda-GE (K051092) C.
- Handi, Ceramatec, Inc (K973282) ರ.
- Device description: The VEO Multigas Monitor for Pocket PC combines a miniature 8. mainstream infrared gas analysis bench with an ultra-fast response oxygen fuel cell. The complete multigas analyzer is contained within a transducer that is attached to the breathing circuit via an airway adapter.
- Intended Use: The VEO Multigas Monitor for Pocket PC is intended to provide monitoring ത inflenced ooor the not oxygen during anesthesia, recovery and respiratory care. It may be used in the operating suite, intensive care unit, patient room, and emergency medicine settings for adult and pediatric patients.
- 10. Comparison to predicates: The VEO Multigas Monitor for Pocket PC combines the gas monitoring capabilities of two predicate devices into one device. The VEO Multigas Monitor for Pocket PC uses the same basic technology concepts used in the predicate devices, while adding improvements derived from advanced electronics and miniaturization. The intended uses of the VEO Multigas Monitor for Pocket PC and its predicates are the same. All the devices consume equivalent amounts of electric power produce disposable single-patient-use airway adapters to interface with gases in the breathing circuit. Labeling and materials used are equivalent, except that the VEO Multigas Monitor for Pocket PC displays numeric and graphic information on an off-theshelf Pocket PC. PHASEIN-approved Pocket PCs have been tested and validated as reliable components of the VEO Multigas Monitor device.
- 11. Testing vs. predicates: Non-clinical testing in direct comparison to predicates throughout the operating range was conducted using calibrated gas samples and legally marketed anesthesia and ventilation devices.
- 12. Conclusions from testing: The VEO Multigas Monitor for Pocket PC demonstrated performance, safety and effectiveness equivalent or superior to its predicates in all characteristics. The VEO Multigas Monitor for Pocket PC demonstrated superior performance in response time, accuracy, precision, and reliability.
{1}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows the logo for the Department of Health and Human Services, USA. The logo is circular and contains the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized caduceus, which is a symbol often associated with healthcare. The symbol consists of three curved lines that converge at the top and bottom, with a wavy line at the bottom.
Public Health Service
SEP 1 5 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Phasein AB C/O Mr. David Weissburg Weissburg Associates 4213 Winnequah Drive Madison, Wisconsin 53716
Re: K051857
R021857
Trade/Device Name: VEO Multigas Monitor For Pocket PC, Model 400221 Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon dioxide gas analyzer Regulatory Class: II Product Code: CCK, CCL Dated: June 30, 2005 Received: July 8, 2005
Dear Mr. Weissburg:
We have reviewed your Section 510(k) premarket notification of intent to market the indication we nave reviewed your becaon 310(t) person is substantially equivalent (for the indications felerenced above and nave decemblicate devices marketed predicate devices marketed in interstate for use stated in the Chelosure) to regard cate of the Medical Device Amendments, or to
commerce prior to May 28, 1976, the enactment date of the Medical Food Drya commerce proof to May 20, 1976, the excordance with the provisions of the Federal Food. Drug, devices that have been recalised in access approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The You may, therefore, manse the devices, belyer to the .
general controls provisions of the Act include requirements for annual registration, listing of general controls provisions of vactice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (500 aborb) and existing major regulations affecting your device can Inay be subject to suen additions, Title 21, Parts 800 to 898. In addition, FDA may be found in the Ood of I cases on one one on the Federal Register.
{2}------------------------------------------------
## Page 2 -- Mr. Weissburg
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issualles of a substitive complies with other requirements of the Act
that FDA has made a determination that your device Force agemains. You mus that FDA has made a determination that your arres over Federal agencies. You must and listing or any Federal statutes and regulations administered of one of existing and listing (21
comply with all the Act's requirements, including, but not asseting securements as set comply with all the Act s requirements, menames caring practice requirements as set
CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as a CFR Part 807); labeling (21 CFR Part 001); good manager (200); and if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR 100); 21 CFR 1000 1050 forth in the quality systems (Sections 531-542 of the Act); 21 CFF 100-1050.
product radiation control provisions (Sections 531-542 of the Act); vous Section S product radiation confor provisions (Scetions 95 %) on and on the more of the may of the may of This letter will allow you to begin marketing your avries as walnec of your device to a legally premarket notification. The PDA intallig of substantial equively .
marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your as the sees, please not the regulation entitled, a contact the Office of Compliance at (210) 276 : 1265, Part 807.97). You may obtain "Misbranding by reference to premarket notified.com the Division of Small
other general information on your responsibilities under the Act from worker (800) 638-200 other general information on your responsion.res-ander at its toll-fire number (800) 638-2041 or Manufacturers, International and Consumer Assistance of the economic of consisted on the many of the world index. html.
Sincerely yours,
Clive, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number: _______________________________________________________________________________________________________________________________________________________________
Device Name: VEO Multigas Monitor for Pocket PC Indications for Use:
The VEO Multigas Monitor for Pocket PC is intended to provide monitoring of The VEO Multigas Monitor for Postor of Is interespiratory and respiratory care. It
carbon dioxide and oxygen during anesthesia, recovery and respiration one Carbon dioxide and oxygon daning intensive care unit, patient room and may be used in the operaings for adult and pediatric patients.
Prescription Use × AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Cure Julion
(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital,
Division Onetral Dental Devices Division of Anesthoolorships Devices
510(k) Number _
Page __ of ___
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.