K955234 · Monaghan Medical Corp. · BZH · Mar 18, 1996 · Anesthesiology
Device Facts
Record ID
K955234
Device Name
TRUPEAK PEAK FLOW METER
Applicant
Monaghan Medical Corp.
Product Code
BZH · Anesthesiology
Decision Date
Mar 18, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1860
Device Class
Class 2
Indications for Use
The TruPeak Peak Flow Meter is a simple mechanical device that responds to and indicates the Peak Expiratory Flow Rate (PEFR) generated by the user during a forced exhalation maneuver.
Device Story
Mechanical device; measures Peak Expiratory Flow Rate (PEFR). User performs forced exhalation into device inlet; internal piston extends spring via pressure spike; indicator moves to maximum displacement position; position corresponds to PEFR value on scale. Logarithmic piston displacement provides higher resolution at lower flow rates. Used by patients for self-monitoring; provides quantitative feedback to assist in respiratory management.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Mechanical peak flow meter; logarithmic piston-spring mechanism; manual indicator; plastic housing; no electronic components; no software; no energy source.
Indications for Use
Indicated for patients requiring measurement of Peak Expiratory Flow Rate (PEFR) via forced exhalation maneuver to monitor respiratory status.
Regulatory Classification
Identification
A peak-flow meter for spirometry is a device used to measure a patient's maximum ventilatory flow rate.
{0}
MAR 18 1996
K955235+
# SECTION 2 - SUMMARY AND CERTIFICATION
## 510(k) SUMMARY
Monaghan Medical Corporation
5 Latour Ave., Suite 1600
P.O. Box 2805
Plattsburgh, NY 12901-0299
Phone: 518-561-7330 Fax: 518-561-5660
Contact: James A. Cochie
QA/RA Manager
Date Prepared: October 30, 1995
Device Proprietary Name: TruPeak Peak Flow Meter
Device Classification name: Peak Flow Meter for Spirometry, 73 BZH
Legally Marketed Device Identity: TruZone Peak Flow Meter, 510(k) #926307
## Device Description:
The TruPeak Peak Flow Meter is a simple mechanical device that responds to and indicates the Peak Expiratory Flow Rate (PEFR) generated by the user during a forced exhalation maneuver.
The inlet of the device is placed into the user's mouth after which the maximum rate at exhalation is attempted from nearly full lungs. As this forced exhalation commences an internal piston extends a spring as reaction to an internal pressure spike. An indicator is pushed by the moving piston and remains at the maximum displacement position of the piston after the maneuver is terminated. This position corresponds to the PEFR and is quantified in liters per minute by reading a scale next to the indicator. The piston displacement function is logarithmic thereby providing greater resolution of scale at lower PEFR readings.
The TruPeak Peak Flow Meter is identical in form and function to the TruZone Peak Flow Meter as described in its 510(k) #926307 except for the use of zone management tapes. The essence of the modification to this 510(k) is changes in the device labeling to conform to the "Guidance for labeling Peak Flow Meters for Over the Counter Sale", dated June 1993.