The spirometer (LA104, LA105) is a diagnostic tool to measure the maximal volume and flow of air that can be moved in and out of a patient's lungs. The system is intended for use with pediatric (5 to 21 years) and adult (22 years and older) patients in hospitals, physician's offices, laboratories, and occupational health environments.
Device Story
Spirometer (models LA104, LA105) measures pulmonary function via differential pressure flow sensor. Patient exhales/inhales through disposable filter/mouthpiece; sensor detects pressure differential across probe; converts to electrical then digital signals. Computer system processes signals to calculate FVC, SVC, MVV, and MV; displays volume-time and flow-volume curves. LA104 includes incentive animations for pediatric patients. Used in hospitals, clinics, labs by clinicians. Output assists in diagnosing ventilatory conditions and assessing medication response (bronchial diastolic/provocation tests). Benefits include standardized pulmonary function assessment.
Clinical Evidence
No clinical studies were conducted. Substantial equivalence is supported by bench testing, including performance testing per ATS/ERS 2005 and ISO 26782:2009, biocompatibility (ISO 10993-1), gas pathway safety (ISO 18562-2/3), electrical safety (ANSI AAMI ES60601-1), and EMC (IEC 60601-1-2).
Technological Characteristics
Differential pressure flow sensing; portable form factor; 8.4-inch touch screen; USB connectivity; power adapter/battery operated. Complies with ATS/ERS 2005, ISO 26782, ISO 10993-1, ISO 18562, ANSI AAMI ES60601-1, and IEC 60601-1-2. Software includes incentive animations (LA104).
Indications for Use
Indicated for pediatric (5-21 years) and adult (22+ years) patients requiring diagnostic measurement of maximal air volume and flow in lungs. Used in hospitals, clinics, labs, and occupational health settings.
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
Predicate Devices
CardioTech Spirometry Model System, GT-105 (K090646)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 10, 2021
MeHow Innovative Ltd % Olivia Meng Regulatory Affairs Manager Guangzhou Osmunda Medical Device Technical Service Co., Ltd 8-9th Floor, R&D Building, No.26 Qinglan Street, Panyu District Guangzhou, Guangdong 510006 China
Re: K201493
Trade/Device Name: Spirometer Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: Class II Product Code: BZG Dated: June 3, 2021 Received: June 7, 2021
Dear Olivia Meng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201493
Device Name Spirometer
## Indications for Use (Describe)
The spirometer (LA104, LA105) is a diagnostic tool to measure the maximal volume and flow of air that can be moved in and out of a patient's lungs. The system is intended for use with pediatric (5 to 21 years and older) patients in hospitals, physician's offices, laboratories, and occupational health environments.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
l. SUBMITTER
MeHow Innovative Ltd
Block A&B&C, NCBC Industrial Park, 6th Baolong Road, Longgang District, Shenzhen
518116, RP China
Phone: 086-0755-83051518
Fax: 086-0755-83051789
| Primary Contact Person: | Olivia Meng |
|---------------------------|----------------------------------------------------|
| | Regulatory Affairs Manager |
| | Guangzhou Osmunda Medical Device Technical Service |
| | Co., Ltd |
| | Tel: (+86) 20-6231 6262 |
| | Fax: (+86) 20-8633 0253 |
| Secondary Contact Person: | Gloria Sun |
| | Regulatory registration supervisor |
| | MeHow Innovative Ltd |
| | Tel: 86-0755-83051518 |
| | Fax: 086-0755-83051789 |
| Date Prepared: | 06/2/2021 |
II. DEVICE
| Name of Device: | Spirometer |
|-----------------------|------------------------------------------|
| Common or Usual Name: | Diagnostic Spirometer |
| Classification Names: | Spirometer, Diagnostic (21 CFR 868.1840) |
| Regulation Class: | II |
| Product Code: | BZG |
{4}------------------------------------------------
#### III. PREDICATE DEVICE
Predicate device
GT-105: CardioTech Spirometry Model System, K090646
#### DEVICE DESCRIPTION IV.
The spirometer is used to detect the ventilatory conditions of patients using a flow sensor. Basic test items include: Forced Vital Capacity (FVC), Slow Vital Capacity (SVC), Maximum Ventilator Volume (MVV), and Minute ventilation (MV). The device also provides bronchial diastolic and bronchial provocation tests comparison before and after medication along with time-volume and time-flow curves of the above tests.
The device comes in two models: LA104 and LA105. There are no differences between the two models apart from a minor software function. LA104 includes software incentive animations to encourage children to follow breathing instructions. The spirometer (model: LA104, LA105) consists of the main body, handle, power adapter and a single-use flow sensor. In order to conduct simple spirometry testing, the spirometer is used with a commercially available single-use disposable filter with integrated mouthpiece. This device is compatible with 30mm diameter filters.
The fundamental technology to measure flow is differential pressure. While the patient breathes, the air flows through both ends of the flow probe and produces different pressures. Then the sensor detects the pressure gap between both ends and converts it to electrical signals. The electrical signals are converted into digital signals of the pressure gap. Then digital signals are input into the computer system, which outputs values of pulmonary function related parameters after digital signal processing and data analysis.
### V. INDICATION FOR USE
The spirometer (LA104, LA105) is a diagnostic tool to measure the maximal volume and flow of air that can be moved in and out of a patient's lungs. The system is intended for use with pediatric (5 to 21 years) and adult (22 years and older) patients in hospitals, physician's offices, laboratories, and occupational health environments.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The spirometer (LA104, LA105) is substantially equivalent to the cleared predicate device (K090646). The spirometer (LA104, LA105) has same indications for use,
{5}------------------------------------------------
patient population and technological characteristics with the predicate.
Differential pressure is the principle of spirometry testing for both the subject and predicate devices, in particular the same sensor technology is used for spirometry testing. Both the proposed device and predicated device comply with the ATS (American Thoracic Society) standards for measuring FVC, SVC, MVV, and MV (accuracy and repeatability).
| Specification | Proposed Device | Predicate Device | Comparison |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Device name | Spirometer (Model: LA 104, LA 105) | CardioTech Spirometry System, Model GT-105 | |
| K number | K201493 | K090646 | |
| Indications for Use | The spirometer (LA104, LA105) is a diagnostic tool to measure the maximal volume and flow of air that can be moved in and out of a patient's lungs. The system is intended for use with pediatric (5 to 21 years) and adult (22 years and older) patients in hospitals, physician's offices, laboratories, and occupational health environments. | The CardioTech Spirometry System, Model GT-105 is a diagnostic tool to measure the maximal volume and flow of air that can be moved in and out of a patient's lungs. The system is intended for use with pediatric (4 to 17 years) and adult patients in hospitals, physician's offices, laboratories, and occupational health environments. | Same |
| Product code | BZG | BZG | Same |
| Classification regulation | 21 CFR 868.1840 | 21 CFR 868.1840 | Same |
| Patient population | Pediatric (5 to 21 years) and adult (22 years and older) patients | Pediatric (4 to 17 years) and adult patients | Same |
| Environment of use | hospitals, physician's offices, laboratories, and occupational health environments | hospitals, physician's offices, laboratories, and occupational health environments | Same |
| Prescription use or not | Prescription use | Prescription use | Same |
| Technology for measure flow and volume | Differential pressure | Differential pressure | Same |
| Device type | Portable | Portable | Same |
| Flow range | ±16 L/s | ±14 L/s | Similar |
| Flow accuracy | Comply with ATS/ERS 2005: ATS/ERS Task Force | Comply with ATS/ERS 2005: ATS/ERS Task Force | Same |
| Flow resistance | <0.15 kPa/l/s | <0.15 kPa/l/s | Same |
| Volume range | 0-16 L | 0 -10 L | Similar |
| Volume accuracy | ± 3% or ±50 ml, whichever is greater | ± 3% or ±50 ml, whichever is greater | Same |
| Data Storage<br>capacity | ≥ 3000 patients data | More than 300 patient's data (HD) | Same |
| Communication | USB | USB and RS-232C | Same |
| Coaching | Display of volume-time and<br>flow-volume curves | Display of volume-time and<br>flow-volume curves | Same |
| Measurement<br>parameters | SVC, FVC, MVV, MV + BD + Challenge | SVC, FVC, MVV, MV | Similar |
| Power source | Power adapter, battery | Power adapter, battery | Same |
| Display | 8.4 inch<br>touch<br>(resolution<br>screen<br>800x600) | 8 inch TFT color LCD (640x480dots) | Similar |
| Printer | 104 mm width, thermo-sensitive printer | 112mm width, Thermal dot printer | Similar |
| Patient contact | Indirect contact with tissue less than 24h | Indirect contact with tissue less than 24h | Same |
| Biocompatibility<br>contact duration | Less than 24 hours | Less than 24 hours | Same |
{6}------------------------------------------------
#### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
# Performance testing
The Spirometer meets the spirometry recommendations for accuracy and precision published by the American Thoracic Society (ATS). Testing demonstrated equivalence to the predicate device with regards to performance of forced vital capacity (FVC), slow vital capacity (SVC), maximum ventilator volume (MVV) and minute ventilation (MV) spirometry tests. Spirometer performance was tested according to:
- । ATS/ERS 2005: ATS/ERS Task Force: Standardization of Lung Function Testing
- ISO 26782:2009: Anaesthetic and respiratory equipment — Spirometers intended for the measurement of time forced expired volumes in humans
# Biocompatibility testing
The device is an external communicating, non-sterile, with single use patient contacting
{7}------------------------------------------------
components (Flow sensor/Flow sensor head). The biocompatibility evaluation for the Spirometer was conducted in accordance with the FDA guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", June 16, 2016, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The biocompatibility of testing included the following tests:
- Cytotoxicity
- Sensitization
- Irritation
Also, the gas pathway testing has been conducted in accordance with ISO 18562-2 and ISO 18562-3.
- Particulate Matter
- Volatile Organic Compounds
Additionally, condensate testing was performed to show that no condensate is formed in the single-use flow sensor and mouthpiece. The flow sensor is considered to be tissue indirect contacting type for duration of less than 24 hours.
# Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the Spirometer. The device complies with the ANSI AAMI ES60601-1:2005/(R) 2012 and A1:2012, C1:2009/(R) 2012 and A2:2010/(R) 2012 (Consolidated Text) for safety and the IEC 60601-1-2:2014 standard for EMC.
# Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, since a failure of latent design flaw could directly result in minor injury to the patient or operator.
# Clinical Testing
Based on the similarities of the device specifications, intended use, indications for use between the Spirometer and its predicate devices, no clinical studies were needed to support this 510(k) Premarket Notification.
VIII. CONCLUSION
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the word "MeHow" in a bold, blue font. The letters are connected and slightly slanted to the right. The word is centered on a white background.
The Spirometer and its application comply with standards as detailed the premarket notification. The non-clinical tests conducted on the device determined the Spirometer to be substantially equivalent to its predicate in terms of safety, effectiveness and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.