The Chartis System is indicated for use by bronchoscopists during a bronchoscopy in adult patients with emplysema, a form of Chronic Obstructive Pulmonary Disease (COPD), in a bronchoscopy suite. The system, composed of the Clartis Catheter and Chartis Console, is designed to measure pressure and flow in order to calculate resistance to airflow and quantify collateral ventilation in isolated lung compartments. The Chartis Catheter is used through the working channel of a bronchoscope and connects to the Chartis Console. The Chartis Console is capital equipment that is reusable and displays the patient information.
Device Story
System measures pressure and flow in isolated lung compartments to quantify collateral ventilation; used by bronchoscopists in bronchoscopy suites. Input: airflow and pressure signals from Chartis Catheter (inserted via bronchoscope working channel). Processing: Tablet Console hardware translates catheter signals into electrical data; software analyzes and displays real-time airflow, pressure, resistance, and volume. Output: real-time graphs and metrics on touchscreen display. Two modes: Standard (spontaneous breathing) and Ventilator (mechanical ventilation). Clinical decision-making: assists physicians in assessing lung compartments for collateral ventilation. Benefits: enables targeted patient selection for lung procedures.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including software verification and validation, electrical safety (AAMI/ANSI ES60601-1, IEC 60601-1-6), electromagnetic compatibility (IEC 60601-1-2), hardware verification, stress cycling, cleaning/cycling testing, usability evaluation (IEC 62366), and packaging validation (ASTM D4332, ASTM 4169).
Technological Characteristics
System comprises touchscreen computer and sensor enclosure. Connectivity: wired connection to Chartis Catheter. Materials: polymer tube with filter. Sensing: pressure and flow transducers. Standards: AAMI/ANSI ES60601-1, IEC 60601-1-6, IEC 60601-1-2, IEC 62366, ASTM D4332, ASTM 4169. Software: real-time signal processing and display. Reusable capital equipment.
Indications for Use
Indicated for adult patients with emphysema (COPD) undergoing bronchoscopy in a bronchoscopy suite to quantify collateral ventilation in isolated lung compartments.
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 13, 2018
Pulmonx Corporation Ms. Sherry Kim Sr. Regulatory Affairs Specialist 700 Chesapeake Drive Redwood City, California 94063
Re: K180011
Trade/Device Name: Pulmonx Chartis Tablet Console Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic spirometer Regulatory Class: Class II Product Code: BZG Dated: June 12, 2018 Received: June 13, 2018
Dear Ms. Kim:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# James J. Lee -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180011
Device Name Pulmonx Chartis Tablet Console
#### Indications for Use (Describe)
The Chartis System is indicated for use by bronchoscopists during a bronchoscopy in adult patients with emplysema, a form of Chronic Obstructive Pulmonary Disease (COPD), in a bronchoscopy suite. The system, composed of the Clartis Catheter and Chartis Console, is designed to measure pressure and flow in order to calculate resistance to airflow and quantify collateral ventilation in isolated lung compartments. The Chartis Catheter is used through the working channel of a bronchoscope and connects to the Chartis Console. The Chartis Console is capital equipment that is reusable and displays the patient information.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------|--------------------------------------------------------------|
| <div> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
This summary of 510(k)-safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
#### Applicant Information:
Pulmonx Corporation 700 Chesapeake Drive Redwood City, California 94063
#### Contact Person:
Sherry Kim Email: skim@pulmonx.com Phone: (650) 216-0189
#### Device Information:
Trade Name: Regulation Name: Regulation Number: Device Class: Product Code:
Pulmonx Chartis Tablet Console Diagnostic Spirometer 21 CFR 868.1840 II BZG
### Predicate Device:
Chartis Console, K111764
#### Date Prepared:
June 7, 2018
#### Device Description:
The Chartis Tablet Console is a two-part spirometry system comprised of a touchscreen computer and sensor enclosure designed for use in the bronchoscopy suite in conjunction with the Chartis Catheter. The Chartis Catheter is a previously cleared device (under K111522) and the subject 510(k) K180011 is solely for the Chartis Tablet Console. The proximal end of the Chartis Catheter is attached to a polymer tube with a filter whose opposite end is attached to an input fitting on the Chartis Console. The balloon on the Chartis Catheter isolates the lung compartment of interest. The hardware components of the Tablet Console translate airflow and pressure detected through the Chartis Catheter into electrical signals. The Tablet Console analyzes and displays airflow and pressure from the isolated lung compartment in real time.
## Indications for Use:
The Chartis System is indicated for use by bronchoscopists during a bronchoscopy in adult patients with emphysema, a form of Chronic Obstructive Pulmonary Disease (COPD), in a bronchoscopy suite. The system, composed of the Chartis Catheter and Chartis Console, is designed to measure pressure and flow in order to calculate resistance to airflow and quantify collateral ventilation in isolated lung
{4}------------------------------------------------
compartments. The Chartis Catheter is used through the working channel of a bronchoscope and connects to the Chartis Tablet Console. The Chartis Console is capital equipment that is reusable and displays the patient information.
#### Comparison of Intended Use and Technological Characteristics with the Predicate Device:
The subject and the predicate devices have the same intended use as spirometry systems in measuring pressure and/or flow to detect the presence of collateral ventilation in isolated lung compartments.
The subject and predicate device are based on the following same technological elements:
- Both devices operate with the Chartis Catheter as the Chartis System and have the same principles ● of operation.
- . Both devices have the same general device description. Both devices are comprised of a touchscreen computer, patient interface board and software.
- . Both devices display air pressure, airflow, resistance and total volume of expired air.
Whereas the predicate device is indicated for all adult patients, the indications for use for the subject device specify the intended patient population as adult patients with emphysema, a form of Chronic Obstructive Pulmonary Disease (COPD). The computer and patient interface board of the predicate device are integrated into a single unit; in the subject device they are two physically separate components. The subject device allows segmental level assessments.
The subject device additionally provides a second display option called the Ventilator Mode. The Ventilator Mode provides a display option that may be preferred when the physician selects to perform the bronchoscopy and Chartis assessment on a patient under general anesthesia with mechanical ventilation. For clarification, the term "Ventilator Mode" does not signify it is intended for use in patients that are ventilator dependent outside of the procedure. The Standard Mode, which is available in both the predicate device and the subject device, provides a display option that may be preferred when the physician selects to perform the bronchoscopy and Chartis assessment on a patient under conscious sedation with spontaneous breathing. Both modes display the flow graphs, and both modes can be used to detect the presence of collateral ventilation in isolated lung compartments. The main difference in the display of the two modes is that the Ventilator Mode does not display inspiratory pressure and resistance since these measurements are not generated when mechanical ventilation is used during the procedure.
#### Performance Data:
The following performance data were provided to demonstrate safety and efficacy in support of substantial equivalence determination:
| Test | Results and Conclusions |
|--------------------------------------------------------------------------------------------|---------------------------------------------------|
| Software verification and validation | Test samples passed all acceptance criteria. PASS |
| Electrical safety per applicable requirements of AAMI<br>ANSI ES60601-1 and IEC 60601-1-6. | Test samples passed all acceptance criteria. PASS |
| Electromagnetic compatibility per applicable<br>requirements of IEC 60601-1-2 | Test samples passed all acceptance criteria. PASS |
| Hardware Verification Testing | Test samples passed all acceptance criteria. PASS |
{5}------------------------------------------------
| Hardware Review | Test samples passed all acceptance criteria. PASS |
|------------------------------------------------------------------------------|---------------------------------------------------|
| Operational Temperature and Humidity Conditioning | Test samples passed all acceptance criteria. PASS |
| Stress Cycling Testing | Test samples passed all acceptance criteria. PASS |
| Repeated Cleaning and Tablet Stand Cycling Testing | Test samples passed all acceptance criteria. PASS |
| Usability Evaluation per IEC 62366 | Test samples passed all acceptance criteria. PASS |
| Packaging Validation per applicable requirements of<br>ASTM D4332, ASTM 4169 | Test samples passed all acceptance criteria. PASS |
#### Summary:
The Pulmonx Chartis Tablet Console has the same intended use as the predicate device. In addition, it has similar technological characteristics; performance data demonstrates that any differences in technological characteristics do not raise different questions of safety or effectiveness. Therefore, the Pulmonx Chartis Tablet Console is substantially equivalent to the cleared predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.