K142303 · Medical Respiratory Devices S.L. · BZG · Nov 20, 2014 · Anesthesiology
Device Facts
Record ID
K142303
Device Name
NeumoFilt Ergo/Neumo Filt BiteOn
Applicant
Medical Respiratory Devices S.L.
Product Code
BZG · Anesthesiology
Decision Date
Nov 20, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1840
Device Class
Class 2
Indications for Use
For use with pulmonary function testing (PFT) equipment only, to filter air between the patient's exhaled air and the testing equipment. Single patient use, disposable. Duraton of use < 24 hours. Environment of use: Hospital, Sub-acute facilities, Clinics, Physician offices.
Device Story
NeumoFilt Ergo and BiteOn are disposable, single-patient-use electrostatic filters for pulmonary function testing (PFT) equipment. Device placed between patient and PFT machine to prevent cross-contamination of equipment by inhaled/exhaled breath. Two models: Ergo (mouthpiece-shaped end) and BiteOn (biteblock-shaped end). Used in hospitals, clinics, and physician offices. Filters air via electrostatic media; protects equipment from biological contaminants. Healthcare providers use the device as a standard accessory during routine PFT procedures. Benefits include equipment protection and patient safety through bacterial/viral filtration.
Clinical Evidence
Bench testing only. Testing included bacterial filtration efficiency (99.9998% per MIL-M-36954C), viral filtration efficiency (99.9939% per MIL-M-36954C), flow resistance at 1, 5, and 14 L/sec, dead space, housing burst strength, and age testing. Biocompatibility testing performed per ISO 10993-1 and G95-1, including cytotoxicity, sensitization, and intracutaneous reactivity/irritation.
Technological Characteristics
Electrostatic filtration media; plastic housing. Dimensions include various end fittings (30, 33, 35, 36, 48 mm). Dead space: 70 ml (Ergo), 68 ml (BiteOn). Burst pressure > 120 cm H2O. Biocompatibility per ISO 10993-1. Single-patient, disposable, non-sterile (implied).
Indications for Use
Indicated for use with pulmonary function testing (PFT) equipment to filter exhaled air. No specific patient population limitations; single-patient, disposable use for < 24 hours in clinical settings.
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of three faces in profile, stacked on top of each other, resembling a stylized human figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 20, 2014
Medical Respiratory Devices S.L. Mr. Paul Dryden President, PreMedic, Inc. Avda. de las Americas 4, Nave A7 Coslada (Madrid) 28823 SPAIN
Re: K142303
Trade/Device Name: NeumoFilt Ergo and NeumoFilt BiteOn Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: II Product Code: BZG Dated: October 21, 2014 Received: October 22, 2014
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Dryden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the
Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K142303
Device Name NeumoFilt Ergo
Indications for Use (Describe)
For use with pulmonary function testing (PFT) equipment only, to filter air between the patient's exhaled air and the testing equipment. Single patient use, disposable. Duraton of use < 24 hours. Environment of use: Hospital, Sub-acute facilities, Clinics, Physician offices.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K142303
Device Name NeumoFilt BiteOn
Indications for Use (Describe)
For use with pulmonary function testing (PFT) equipment only, to filter air between the patient's exhaled air and the testing equipment. Single patient use, disposable. Duraton of use < 24 hours. Environment of use: Hospital, Sub-acute facilities, Clinics, Physician offices.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{4}------------------------------------------------
### 510(k) Summary Page 1 of 4 8-Nov-14
| Medical Respiratory Devices S.L.<br>Avda. de las Americas 4, Nave A7<br>28823 Coslada (Madrid) SPAIN | Telephone: +34 91 6745003 |
|------------------------------------------------------------------------------------------------------|--------------------------------------------|
| Official Contact: | Jose Alberto Hernandez<br>General Manager |
| Proprietary or Trade Name: | NeumoFilt Ergo<br>NeumoFilt BiteOn |
| Common/Usual Name: | Diagnostic Spirometer (Accessory) |
| Regulation description: | Diagnostic Spirometer |
| Product code: | BZG |
| Regulation number: | 21 CFR 868.1840 |
| Device class: | Class II |
| Predicate Device: | Air Safety Model 2800 PFT filter (K051712) |
Device Description: The MRD NeumoFilt pulmonary function filter ("PFT filter") is a standard electrostatic filter that is placed between the patient and the pulmonary function testing equipment to keep a patient inhaled and exhaled breath from contaminating the equipment. There are 2 models with the only difference being the patient end piece which is either a biteblock shape (BiteOn) or the mouthpiece shape (Ergo).
Indications for Use: For use with pulmonary function testing (PFT) equipment only, to filter air between the patient's exhaled air and the testing equipment. Single patient use, disposable. Duration of use < 24 hours. Environment of use: Hospital, Sub-acute facilities, Clinics, Physician offices.
### Contraindications: None
## Substantial Equivalence Rationale:
The MRD NeumoFilt is viewed as substantially equivalent to the predicate devices because:
Indications - For use with pulmonary function testing (PFT) equipment only, to filter air between the patient's exhaled air and the testing equipment. Single patient use, disposable. Duration of use < 24 hours. Environment of use: Hospital, Sub-acute facilities, Clinics, Physician offices.
Discussion - This is identical to the predicate - Air Safety Model 2800 PFT Filter - K051712, which is intended for use with pulmonary function testing equipment, to filter air between the patient's exhaled air and the testing equipment.
{5}------------------------------------------------
## 510(k) Summary Page 2 of 4 8-Nov-14
| Features | Proposed<br>NeumoFilt PFT Filter | Predicate<br>Air Safety Model 2800 PFT Filter<br>(K051712) |
|---------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | For use with pulmonary function testing<br>(PFT) equipment only, to filter air<br>between the patient's exhaled air and the<br>testing equipment. Single patient use,<br>disposable. Duration of use < 24 hours.<br>Environment of use: Hospital, Sub-acute<br>facilities, Clinics, Physician offices. | Model 2800 is indicated for use with<br>pulmonary function testing equipment,<br>to filter air between the patient's<br>exhaled air and the testing equipment. |
| Patient Use / Duration if<br>use | Single patient use, disposable,<br>< 24 hours | Single patient use, disposable,<br>< 24 hours |
| Environment of Use | Hospital, Sub-acute facilities, Clinics,<br>Physician offices | Hospital, Sub-acute facilities,<br>Physician offices |
| Patient Population | No specific limitations | No specific limitations |
| Contraindications | None | None |
| Features, Specifications and Performance | | |
| Can be used with several<br>different PFT machines | Yes | Yes |
| Various sizes of machine<br>side end fittings – 30, 33,<br>35, 36, 48 mm | Yes | Yes |
| Patient uses a<br>mouthpiece or biteblock | Yes | Yes |
| Filtration method | Electrostatic | Electrostatic |
| Resistance to flow @ 1,<br>5, 14 l/sec per ATS<br>standard for spirometry | 0.56 cmH2O @ 1 l/sec<br>0.74 cmH2O @ 5 l/sec<br>1.47 cmH2O @ 14 l/sec | 0.5 cm H2O @ 1 l/sec<br>0.7 cm H2O @ 12 l/sec |
| Bacterial Filtration<br>Efficiency (BFE) | 99.9998% | 99.9999% |
| Viral Filtration<br>Efficiency (VFE) | 99.9939% | 99.999+% |
| Weight (gm) | 30 grams | 40 grams |
| Internal volume/<br>Dead space | 70 ml – Ergo<br>68 ml - BiteOn | 75 ml |
| Housing Burst Pressure | > 120 cm H2O | Not reported |
| Performance | None under section 514 | None under section 514 |
| Biocompatibility | Tested per ISO 10993<br>Cytotoxicity<br>Sensitization<br>Irritation | Not reported |
Technology – The MRD NeumoFilt is an electrostatic media that is equivalent in performance and the fundamental principle of operation as the predicate.
Discussion - There are no differences in technology between the proposed device and the predicate Air Safety Model 2800 PFT filter - K051712.
{6}------------------------------------------------
## 510(k) Summary Page 3 of 4 8-Nov-14
Environment of Use - Hospitals, Sub-acute facilities, clinics and physician offices. Discussion - The environment of use is substantially equivalent to the predicate Air Safety Model 2800 PFT filter – K051712.
Patient Population - There are no patient specific limitations. Discussion - This is identical to the predicate Air Safety Model 2800 PFT filter - K051712.
Performance Specifications - The performance specifications are substantially equivalent to the predicates - Air Safety Model 2800 PFT filter - K051712.
Non-clinical Testing - We performed a number of non-clinical tests to demonstrate the safety and efficacy of the MRD NeumoFilt. These tests included:
Materials – The materials have been evaluated and tested in accordance with ISO 10993-1 and G95-1. Based upon ISO 10993-1 and G95-1 the NeumoFilt filter would be considered as:
- Externally communicating (indirect gas pathway) ●
- Tissue contact ●
- Limited duration (< 24 hours) .
And the housing:
- Surface Contact ●
- Mucosa contact
- Limited duration (< 24 hours) ●
We have performed the following tests:
- Cytotoxicity ●
- Sensitization
- Intracutaneous reactivity or Irritation ●
Performance Testing - We have performed the following performance data and testing:
- Resistance to flow at 1, 5, 14 1/second ●
- Bacterial / Viral filtration efficiency per Nelson Laboratories ●
- o Pre - and Post - conditioning
- Bacterial Filtration Efficiency (BFE) 99.9998% per MIL-M-36954C, S. aureus o ATCC# 6538 of approximately 3.0 micron size, > 107 challenge.
- Viral Filtration efficiency (VFE) 99.9939% per MIL-M-36954C. X174 о bacteriophage of approximately 2.9 micron size, > 106 challenge.
- Dead space ●
- Housing burst strength ●
- Age testing
### Discussion of Differences and Similarities -
As the above table and rationale present there are no major differences between the proposed NeumoFilt and the predicate. Each filter is operates on the same filtration technology that is placed between two houses. They each connect to PFT testing equipment. The predicate may be used with a separate mouthpiece to make it easier for the user whereas the proposed Ergo design the patient end-fitting is shaped like a mouthpiece thus not requiring a separate piece. Their physical
{7}------------------------------------------------
### 510(k) Summary Page 4 of 4 8-Nov-14
size, design, dimensions are very similar and not considered significantly different as demonstrated by the comparative performance testing. The slight differences in BFE / VFE, weight, and Dead Space do not raise any new safety concern and thus can be considered substantially equivalent.
Substantial Equivalence Conclusion - The sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predicate have been found to be substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.