K141793 · Pulmone Advanced Medical Devices, Ltd. · BZG · Mar 6, 2015 · Anesthesiology
Device Facts
Record ID
K141793
Device Name
PULMONE MINIBOXPFT 2.0
Applicant
Pulmone Advanced Medical Devices, Ltd.
Product Code
BZG · Anesthesiology
Decision Date
Mar 6, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1840
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The PulmOne MiniBoxPFT™ 2.0 is intended to measure lung function in adult and pediatric patients while at rest (including spirometry and lung volumes). The PulmOne MiniBoxPFT™ 2.0 is to be used by either a physician, respiratory therapist, or technician.
Device Story
The PulmOne MiniBoxPFT™ 2.0 is a diagnostic spirometer used by clinicians (physicians, respiratory therapists, technicians) in clinical settings to assess lung function. It measures spirometric parameters and absolute/relative lung volumes (TLC, TGV, RV, IC, ERV). The device utilizes a flow sensor, pressure sensor, and a computer-controlled solenoid valve within a sealed container. During operation, the patient breathes through a disposable viral-bacterial filter; the device performs brief flow interruptions (100 ms) to calculate lung volumes using a proprietary formula alongside spirometry data. The system integrates all measurements into a single unit, replacing the two-module design of the predicate. Output is displayed via a GUI, providing clinicians with quantitative lung function metrics to aid in respiratory diagnosis and clinical decision-making. The device benefits patients by providing non-invasive, rapid assessment of lung capacity and airflow.
Clinical Evidence
No clinical study was conducted. Substantial equivalence was supported by non-clinical bench testing, including electrical safety, environmental testing, software validation, and performance testing against ATS 2005 guidelines. Lung volume measurement (LVM) validation compared Total Lung Capacity (TLC) results between the subject and predicate devices, confirming equivalent performance.
Technological Characteristics
Device uses a bi-directional symmetric/averaging Pitot-tube flow sensor and piezo-resistive pressure sensor. Construction includes a 15L elliptic container and a detachable hand-held unit made of Makrolon 2825 polycarbonate (ISO 10993-1 compliant). Operates via computer-controlled solenoid valve (100 ms closure). Connectivity is standalone. Software is a C# .NET application. Complies with IEC 60601-1, IEC 60601-1-2, and ISO 26782.
Indications for Use
Indicated for measuring lung function, including spirometry and lung volumes, in adult and pediatric patients at rest. No specific contraindications listed.
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around the perimeter. In the center of the seal is a stylized image of three human profiles facing to the right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 6, 2015
PulmOne Advanced Medical Devices, Ltd. Dr. Susan Alpert Principal 200 Park Ave, Unit 403 Minneapolis, MN 55415
Re: K141793
Trade/Device Name: PulmOne MiniBoxPFT™ 2.0 Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic spirometer Regulatory Class: Class II Product Code: BZG, BZC Dated: January 27, 2015 Received: January 30, 2015
Dear Dr. Alpert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
Page 2 - Dr. Susan Alpert
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin Keith Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141793 pending
Device Name PulmOne MiniBoxPFT™ 2.0
Indications for Use (Describe)
The PulmOne MiniBoxPFT™ 2.0 is intended to measure lung function in adult and pediatric patients while at rest (including spirometry and lung volumes). The PulmOne MiniBoxPFT™ 2.0 is to be used by either a physician, respiratory therapist, or technician.
Type of Use (Select one or both, as applicable)
즈 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### SECTION 5 - 510(K) SUMMARY 5
#### 5.1 ADMINISTRATIVE INFORMATION
| Date: | November 12, 2014 |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | PulmOne Advanced Medical Devices, Ltd.<br>14 HaCharoshet St., Ra'ananna 4365707 , ISRAEL<br>Tel: +972 77 5100 938<br>Fax: +972 77 5100 939 |
| Official Correspondent: | Avi Lazar, CEO |
| Trade Name: | PulmOne MiniBoxPFT TM 2.0 |
| Classification Name: | Diagnostic spirometer |
| Classification Number: | 21 CFR 868.1840 |
| Product Code: | BZG |
| Additional Code: | BZC |
| Device Class: | Class II |
| Predicate Devices: | PulmOne MiniBoxPFTTM<br>510(k) Number – K133051 |
#### 5.2 DEVICE DESCRIPTION
The PulmOne MiniBoxPFT™ 2.0 is intended to measure lung function in adult and pediatric patients while at rest (including spirometry and lung volumes). The PulmOne MiniBoxPFT™ 2.0 is to be used by either a physician, respiratory therapist, or technician. A single-use, disposable viral-bacterial filter separates the patient from the internal components of the device.
The MiniBoxPFT™ 2.0 measures all common spirometric measurements as well as relative and absolute lung volumes, including the following (the full list is detailed in Section 11 - Device Description):
| Absolute Lung Volumes: | units |
|---------------------------------|-------|
| Total lung capacity (TLC) | L |
| Thoracic Gas Volume (TGV) | L |
| Residual volume (RV) | L |
| Relative Lung Volumes: | |
| Inspiratory capacity (IC) | L |
| Expiratory reserve volume (ERV) | L |
Spirometry:
{4}------------------------------------------------
| Forced vital capacity (FVC) | L |
|----------------------------------------------|-------|
| Forced inspiratory vital capacity (FIVC) | L |
| Slow vital capacity (SVC) | L |
| Slow inspiratory vital capacity (IVC) | L |
| Forced expiratory volume in 1 second (FEV1) | L |
| Forced inspiratory volume in 1 second (FIV1) | L |
| Ratio of FEV1 to SVC (FEV1/SVC) | % |
| Ratio of FEV1 to FVC (FEV1/FVC) | % |
| Forced expiratory volume in 6 seconds (FEV6) | L |
| Ratio of FEV1 to FEV6 (FEV1/FEV6) | % |
| Peak expiratory flow (PEF) | L/s |
| Peak inspiratory flow (PIF) | L/s |
| Forced Expiratory Flow at 50% of FVC (FEF50) | L/s |
| Maximum Voluntary Ventilation (MVV) | L/min |
#### 5.3 -INTENDED USE AND INDICATIONS FOR USE
The PulmOne MiniBoxPFT™ 2.0 is intended to measure lung function in adult and pediatric patients while at rest (including spirometry and lung volumes). The PulmOne MiniBoxPFT™ 2.0 is to be used by either a physician, respiratory therapist, or technician.
#### 5.4 SUMMARY OF TECHNICAL CHARACTERISTICS
The MiniBoxPFT" 2.0 provides both lung volume and spirometry measurements in a single device. The MiniBoxPFT™ 2.0's predicate device is the MiniBoxPFT™ which also provides both lung volume and spirometry measurements however the predicate is comprised of two modules.
The MiniBoxPFT™ 2.0 is substantially equivalent to the MiniBoxPFT™ and both utilize the same technologies and methods of operations, with the main difference being that the MiniBoxPFT™ uses an OEM spirometer and the MiniBoxPFT™ 2.0 integrates all measurements in one integrated device. The lab and bench testing demonstrates that the differences between the devices do not raise any new questions and that the proposed device is substantially equivalent to the predicate device.
Device Construction: Both devices utilize well-recognized flow sensors, the same pressure sensor, a valve and a sealed container in their operation. The MiniBoxPFT110 2.0 has a detachable hand-held unit for use with the spirometric measurements as opposed to an OEM spirometer in MiniBoxPFT™
Methods of Operation: The methods of operation for both devices require the subject to perform breathing maneuvers – breathing via a disposable viral-bacterial filter during flow interruptions. The MiniBoxPFT™ 2.0 requires less time to perform some of the breathing maneuvers, and has shorter duration flow interruptions.
{5}------------------------------------------------
Principles of Operation: The principles of operation for both devices are the same. Both devices uses data from spirometry measurements and data obtained during flow interruptions with a proprietary formula to calculate Total Lung Capacity.
## 5.4.1 Summary table of Comparison
| | MiniBoxPFTTM<br>[K133051] | MiniBoxPFTTM2.0<br>[Proposed Device] |
|-------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use and<br>Indications for Use | The PulmOne MiniBoxPFTTM is<br>intended to measure lung function in<br>adult and pediatric patients while at<br>rest (including spirometry and lung<br>volumes).<br>The PulmOne MiniBoxPFTTM is to<br>be used by either a physician,<br>respiratory therapist, or technician. | Identical |
| Target Population | Adult and pediatric subjects | Identical |
| Biocompatibility | - Mouthpiece:<br>Single-use FDA-cleared viral<br>bacterial filter (K051712)<br>- Hand-Held sub-system:<br>Skin Contact - Using Cleared OEM<br>handle MIR , [K122384] | - Mouthpiece:<br>Identical<br>- Hand-Held sub-system:<br>Skin Contact , Using Biocompatible<br>Makrolon 2825 Polycarbonate material<br>in compliance with ISO10993-1 |
| Standards /<br>Performance<br>guideline<br>compliance | -General medical device safety: IEC<br>60601-1:2005 (3rd Ed)<br>-Electromagnetic compatibility:<br>IEC 60601-1-2:2007 (3rd Ed)<br>-Performance per ATS 2005<br>Standards | Same with additional of ISO 26782<br>compliance to section 7.2, 7.4-7.6 &<br>7.8-7.9 |
| Environmental<br>Operating<br>Conditions | Temperature: 0 to 40 °C<br>Relative Humidity: 20 to 90 %<br>Atmospheric Pressure: 900 to 1060<br>cmH2O | Identical |
| Flow Sensor | 1. Bi-directional hot-wire mass<br>airflow sensor<br>Range: +/- 5.0 L/s<br>2. Bi-directional digital turbine<br>Range: +/- 16.0 L/s<br>[Part of the incorporated<br>OEM cleared MIR Minispir,<br>K122384] | Bi-directional Symmetric and<br>averaging Pitot-Tube<br>Range: +/- 16.0 L/s |
| Pressure Sensor | Bi-directional piezo-resistive<br><br>Mouth Sensor:<br>Range: ±70 cmH2O (±1 PSI)<br>Accuracy: ±0.25% Full Scale | Identical |
| Valve | Computer-controlled solenoid valve<br>Closing response time: <30 ms<br>Closing duration: less than 300 ms | Computer-controlled solenoid valve<br>Closing response time:<30 ms<br>Closing duration: ~100 ms |
| Software | PulmOne 4.0, Matlab based | PulmOne C# .NET application |
| LUNG VOLUME<br>MEASUREMENT | | |
| Measurements | Mouth pressure and mouth flow | Identical |
| Calculations | TLC, TGV (FRC), RV, VC, IC | Same with addition of ERV |
| Principle of Operation | 1) Pressure and flow are measured at the mouth<br>2) Spirometry is measured with the handheld spirometer.<br>3) Proprietary equation is used to calculate total lung capacity | Identical |
| Method of Operation | 1) Patient seated at desk/table<br>2) 1-2 minutes normal breathing<br>3) 300 ms shutter closure per breath<br>4) Inspiration to TLC<br>5) Slow expiration to RV | Same, except shutter closure per breath is 100 ms |
| Container | 16.3 round container | 15L elliptic container |
| SPIROMETRY<br>MEASUREMENT | Cleared OEM MIR Minispir<br>spirometer [K122384] incorporated | MiniBox spirometer design incorporated |
| Measurements | Mouth flow | Identical |
| Calculated<br>Parameters | Used a cleared OEM MIR Minispir [K122384] incorporated :<br>FEV1, FVC, FEV6, FEV1/FVC,<br>FEV1/SVC, FEV1/VC, EFV1/FEV6,<br>PEF, FEF25-75, FEV3, FEV3/FVC,<br>FET, FEF 25, FEF50, FEF75, Evol,<br>FIVC, FIV1, FIVC1/FIVC, PIF, IC,<br>SVC, ERV, TV, IVC, EVC, MVV | same calculated parameters with addition of MVV Rate |
| Displayed<br>Parameters by<br>MiniBox GUI | FVC, FEV1, FEV1/SVC,<br>FEV1/FVC, FEV6, FEV1/FEV6,<br>PEF, SVC, IC, ERV | FEV1, FVC, FEV6, FEV1/FVC,<br>FEV1/SVC, FEV1/VC, EFV1/FEV6,<br>PEF, FEF25-75, FEV3, FEV3/FVC,<br>FET, FEF 25, FEF50, FEF75, Evol,<br>FIVC, FIV1, FIVC1/FIVC, PIF, IC,<br>SVC, ERV, TV, IVC, EVC, MVV and MVV Rate |
| Technical<br>Specifications | Flow range: +/- 16 L/s<br>Volume accuracy: +/- 3% or 50mL<br>Flow accuracy: +/- 5% or 200 mL/s<br>Dynamic resistance at 12 L/s: <0.5 cmH2O/(L/s)<br>Dimension: 142x49.7x26 mm | Flow range: +/- 16 L/s<br>Volume accuracy: +/- 3% or 50mL<br>Flow accuracy: +/- 5% or 200 mL/s<br>Dynamic resistance at 14 L/s: <0.7 cmH2O/(L/s)<br>Dimension: 210x105x45 mm |
| BTPS Correction | Manual | Automatic |
| ACCESSORIES | | |
| Bacterial Filters | FDA 510(K) Number K051712<br>Air Safety Spiroguard - Integral Mouthpiece<br>single-use, disposable viral-bacterial filter | Identical |
{6}------------------------------------------------
{7}------------------------------------------------
#### 5.5 UTILIZATION OF STANDARDS
The MiniBoxPFT™ 2.0 has met the following recognized standards:
IEC 60601-1:2005 - Medical electrical equipment - Part 1: General requirements for basic safety and essential performance, 3td edition
IEC 60601-1-2:2007 - Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests, 3td edition.
ASTM D4169-09, Standard Practice for Performance Testing of Shipping Containers and Systems
ASTM D4332-01, Standard practice for conditioning containers, packages, or packaging components for testing
ASTM D999-01 Standard Test Methods for Vibration Testing of Shipping Containers
ISO 26782:2009, to section 7.2, 7.4-7.6 & 7.8-7.9
#### SUMMARY OF NON-CLINICAL PERFORMANCE TESTING 5.6
Laboratory tests were performed to establish substantially equivalent performance of the MiniBoxPFTM 2.0 vs its predicate, and included electrical safety, software validation, and environmental testing. In addition, non-clinical bench testing was conducted to verify the performance of both the spirometry and lung volume measurements of the device.
{8}------------------------------------------------
The spirometry bench tests (FDA 510(k) sections 18-1 and 18-2) verified that the performance of the MiniBoxPFT TM 2.0 spirometry measurements met the ATS guidelines - requirements of accuracy and repeatability for spirometry equipment.
A Repeatability & Reproducibility test was conducted (FDA 510(k) sections 18-3 and 18-4) and validated that inter-device repeatability and intra-device reproducibility were within the accepted known range for lung volume measurement.
An LVM validation test was conducted (FDA 510(k) sections 18-5 and 18-6), comparing the Total Lung Capacity (TLC) measurement of the proposed device with that of the predicate. TLC measurements were performed using the proposed device, MiniBoxPFTM 2.0 and the predicate device and pre-defined success criteria were set which demonstrate substantial equivalence of Lung Volume Measurement of the proposed device and the predicate device. The study results demonstrate that the MiniBoxPFTTM 2.0 successfully met the study success criteria and the performances of the Lung Volume Measurements are substantially equivalent to those of the predicate device.
#### 5.7 SUMMARY OF CLINICAL PERFORMANCE DATA
No clinical study was conducted to support this application.
#### 5.8 CONCLUSIONS
Based on its underlying technology and the lab and bench tests performed, the PulmOne MiniBoxPFT™ 2.0 is substantially equivalent to the predicate device.
<sup>4</sup> M.R. Miller et al. Standardization of spirometry; Eur Respir J 2005; 26: 319-338
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.