The Micro I spirometer is intended to measure the maximal volume and flow of air that can be moved out of a patient's lungs. The system is intended for use with pediatric and adult patients over the age of 3 years in hospitals, physician offices, laboratories and occupational health testing environments.
Device Story
Micro I is a handheld, portable, microprocessor-based spirometer for measuring expiratory lung flows and volumes. Input: patient exhales into a disposable mouthpiece connected to a uni-directional digital volume transducer. Operation: transducer consists of an acrylic tube with a low-inertia vane on jeweled bearings; airflow creates a vortex causing vane rotation; rotation detected by infrared beam interruption producing electrical pulses. Software: onboard processor converts pulses to volume and flow rate; calculates indices including FEV1, FVC, FEV6, PEF, and FEF metrics. Usage: hospitals, clinics, laboratories, occupational health settings; operated by clinicians or trained personnel. Output: graphic monochrome LCD displays results. Benefit: provides objective pulmonary function data to assist in diagnosing and monitoring respiratory conditions.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including biocompatibility testing per ISO 10993-1, safety and EMC testing per IEC 60601-1 and IEC 60601-2, and performance verification against ATS/ERS standardization of lung function testing.
Technological Characteristics
Handheld spirometer; uni-directional digital volume transducer (acrylic tube, polyester vane, stainless steel pivot). Housing: PC/ABS Cycoloy C2800 with soft coating. Power: 2x AA NiMH rechargeable batteries. Display: 128x128 pixel backlit monochrome LCD. Connectivity: USB. Software: microprocessor-based, rule-based calculation of expiratory indices. Sterilization: turbine is cleanable/sterilizable.
Indications for Use
Indicated for pediatric and adult patients over 3 years of age requiring pulmonary function testing to measure lung volume and flow parameters.
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for CareFusion. The logo consists of a circular icon on the left and the word "CareFusion" on the right. The icon appears to be a stylized shield within a circle. The text is in a simple, sans-serif font.
# 510(k) Summary
i13096
## GENERAL INFORMATION
#### 5.1 Type of Submission
Traditional 510(k) Submission
Submission date: 10/10/2011
#### 5.2 Submitter
Name:
CareFusion Germany 234 GmbH
Address:
Phone:
FAX:
E-mail
Phone: Fax:
E-mail:
Leibnizstrasse 7 D-97204 Hoechberg Germany
### Contact person in Germany:
(Regulatory Affairs Specialist) Address:
Elmar Niedermeyer CareFusion Germany 234 GmbH Leibnizstrasse 7, 97204 Hoechberg Germany +49 931 49 72 - 361 +49 931 49 72 - 62361 elmar.niedermeyer@carefusion.com
Contact person in the U.S .: (Official Correspondent) Address
Carol Emerson
CareFusion 22745 Savi Ranch Parkway Yorba Linda, CA 92887 714-919-3342 714-283-8420 carol.emerson@carefusion.com
{1}------------------------------------------------
- 5.3 Establishment Registration Number 9615102
- 5.4 Common Name or Classification Name Spirometer, Diagnostic (CFR 868.1840, Product Code BZG)
- 5.5 Trade Name Micro I
- 5.6 Device Classification This is a Class II device
- 5.7 Classification Panel 73 Anesthesiology Part 868 Code BZG
- 5.8 Reason for Premarket Notification --- new device ---
### 5.9 Legally predicate marketed devices
- Microlab Spirometer . K031102 Code BZG
- . Micro Diary K100928 Code BZG
- 5.10 Predicate Device Company CareFusion
### 5.11 Device Description
Description & function:
The Micro I spirometer is a hand held portable microprocessor based device designed to measure expiratory flows and volumes. To perform a spirometry test the user first inserts a mouthpiece into the mouthpiece holder of the spirometer, which aligns it with the volume transducer. The unit is then turned on and displays instructions and prompts according to the device type. When prompted to do so by the operator, the patient inhales as deeply as possible, seals his/her lips around the mouthpiece and exhales as hard and as fast as possible until no more air can be exhaled. The device converts the airflow to an electrical signal, and the onboard software calculates the required values.
## Scientific Concept:
The scientific concept which forms the basis of this device is the Carefusion uni-directional digital volume transducer. The transducer consists of an acrylic tube with a freely rotating vane supported on iewelled bearings positioned between a fixed swirl plate and a cross bar. As air is passed through the transducer, a vortex is created by the
{2}------------------------------------------------
swirl plate, which causes the low inertia vane to rotate. The rotation of the vane is detected by the interruption of an infra red beam which produces an electrical pulse train at the output of a phototransistor. The number of rotations is proportional to the volume of air passed through the turbine, and the rate of rotation is proportional to the flow rate. Using the integrated software the Micro I calculates a range of expiratory indices.
### Significant physical & performance characteristics:
Performance (measurements) Forced Expired Volume in 1 second (FEV1) Forced Expired Volume in 6 second (FEV6) Forced Vital Capacity (FVC) Forced Expiratory Ratio (FEV1/FEV6) Forced Expiratory Ratio (FEV1/FVC) Peak Expiratory Flow Rate (PEF) Mid Expiratory Flow (FEF25-75) Expiratory Flow at 75% of volume remaining (FEF75) Expiratory Flow at 25% of volume remaining (FEF75) Dimensions 162 x 61 x 30mm Weight 152g Display 128 x 128 pixel graphic backlit monochrome LDC Transducer type CareFusion Uni-Directional Digital Volume Power supply 2 x AA size NiMH rechargeable cells
## 5.12 Intended Use Statement
The Micro I spirometer is intended to measure the maximal volume and flow of air that can be moved out of a patient's lungs. The system is intended for use with pediatric and adult patients over the age of 3 years in hospitals, physician offices, laboratories and occupational health testing environments.
### 5.13 Required Components
- . Micro I spirometer
- Digital Volume Transducer .
- Cardboard Mouthpieces .
- USB/Charging cable .
- . Cardboard Mouthpiece Adapter
- Instruction for Use .
- ban .
- Accessories .
{3}------------------------------------------------
## 5.14 Summary Table of Comparison
.
| MicroLab K031102 | Micro Diary K100928 | Micro I<br>K113096 (new) | |
|-----------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | The Microlab spirometer is<br>intended, for prescription use only,<br>to measure the maximal volume<br>and flow of air that can be moved in<br>and out of a patient's lungs. The<br>system is intended for use with<br>pediatric (4 to 17 years) and adult<br>(18 to 99 years) patients in<br>hospitals, physician offices,<br>laboratories and occupational<br>health testing environments. | The Micro Diary Spirometer is used<br>in pulmonary function testing to<br>measure the volume of gas moving<br>in or out of a patient's lungs.<br>Specifically, the Micro Diary<br>Spirometer measures the following<br>lung function parameters: FEV1,<br>FVC, FEV6 and PEF. | The Micro I spirometer is intended<br>to measure the maximal volume<br>and flow of air that can be moved<br>out of a patient's lungs. The system<br>is intended for use with pediatric<br>and adult patients over the age of 3<br>years in hospitals, physician offices,<br>laboratories and occupational<br>health testing environments. |
| Target<br>population | Pediatric (4 to 17 years)<br>Adult (18 to 99 years) | The device can be used on patients<br>who require lung function<br>measurements. These patients are<br>usually suffering from diseases<br>such as asthma and<br>chronic obstructive pulmo-nary<br>disorder. It can be utilized for<br>patients from 4 years and older,<br>providing that they are able to<br>follow the medical practitioner's<br>instructions. | Over the age of 3 years |
| Perfor-<br>mance<br>specifica-<br>tion<br>(measure-<br>ment para-<br>meter) | VC, FEV75, FEV1, FEV3, FEV6,<br>FVC, PEF, FEV75/VC,<br>FEV75/FVC, FEV1/VC,<br>FEV1/FVC, FEV3/VC, FEV3/FVC,<br>FEV75/FEV6, FEV1/FEV6, FEF75,<br>FEF50, FEF25, FEF25-75,<br>FEF50/VC, FEF50/FVC, MVV,<br>FIV1, FIVC, PIF, FIV1/FIVC,<br>FIF25, FIF50, FIF75,<br>FEF50/FIF50, MET2575, FET, TV,<br>ERV, IRV, IC, EVC, IVC, FR, Ti,<br>Te, Ti/Ttot, TV/TI | FEV1, FVC, FEV6, PEF | FEV1, FVC, FEV6,<br>FEV1/FVC, FEV1/FEV6,<br>PEF, FEF25, FEF75,<br>FEF25-75 |
| Transducer<br>material<br>(breathing<br>path<br>contacting) | Turbine swirl plate<br>(Polycarbonate)<br>Turbine van<br>(Polyester type 427)<br>Turbine tube<br>(Clear Acrylic)<br>Pivot<br>(Nivapoint stainless steel) | Turbine swirl plate<br>(Makrolon 2607)<br>Turbine van<br>(Polyester type 427)<br>Turbine tube<br>(Makrolon 2607)<br>Pivot<br>(Nivapoint stainless steel) | Identical<br>(to K100928) |
| | Turbine Flow deflector<br>(stainless steel)<br>Mouthpiece holder<br>(ABS Plastic) | N/A | Turbine Flow deflector<br>(stainless steel)<br>Mouthpiece holder<br>(PC/ABS Cycoloy C2800) |
| Transducer<br>(type) | Type 36-TDX 1048<br>CareFusion Bi-Directional Digital<br>Volume | Type 36-TDX 1050<br>CareFusion Uni-Directional Digital<br>Volume | Type 36-TDX 1051<br>CareFusion Uni-Directional Digital<br>Volume |
| Resolution | 0,01 liter | 0,01 liter | Identical<br>(to K100928 & K031102) |
| | MicroLab K031102 | Micro Diary K100928 | Micro I<br>K113096 (new) |
| Accuracy<br>(FEV1/FVC/<br>PEF) | +/- 3% to ATS<br>Recommendations | to the requirements of the<br>ATS/ERS Taskforce:<br>Standardization of<br>Spirometry 2005 | Identical<br>(to K100928) |
| Volume<br>Range | 0.1 - 8 Litres | 0 - 8 l as per ATS/ERS | Identical<br>(to K100928) |
| Flow Range | 0.2 - 15 Litres/Second | 0.2 - 15 Litres/Second | 0 - 14 l as per ATS/ERS |
| Sterilization | Turbine can be sterilized or<br>cleaned | Turbine can be sterilized or<br>cleaned | Identical<br>(to K100928 & K031102) |
| User input<br>type | Touchscreen<br>(PC/ABS) | Keypad 5 key<br>(Silicon rubber) | Keypad 4 key<br>(Silicon rubber) |
| Anatomical<br>sites<br>(Housing) | 255 x 120 x 35 mm<br>Weight: 630g<br>Handle (transducer):<br>50 x 60 x 90 | 130 x 57 x 37 mm<br>Weight: 100g | 162 x 61 x 30 mm<br>Weight: 152g |
| Housing<br>material<br>(patient<br>contacting) | PC/ABS Cycoloy C2950<br>(Housing)<br>Polyurethane "rubber feel"<br>& black ABS<br>(Transducer) | PC/ABS Babyblend T65<br>(Housing) | PC/ABS Cycoloy C2800<br>(upper & lower casing & MLD1621<br>Mouthpiece adaptor)<br>PC/ABS LEXAN 141 Resin<br>(case tube)<br>Alexit 401-75 soft coating<br>(coating for upper & lower case) |
| Display | Colour 1/4 VGA LCD | Graphic LCD monochrome<br>size 44,0 x 27,0<br>100 x 64 dot matrix | Graphic LCD monochrome<br>size 33,90 x 33,90<br>128 x 128 dot matrix |
| Display &<br>keypad foil<br>material<br>(patient<br>contacting) | PC/ABS<br>(Cycoloy C 2950)<br>no patient contact | Clarex Precision Optical<br>Sheet<br>(Cast Acrylic sheet) | Identical<br>(to K100928) |
| Energy<br>used<br>(battery) | 8,4V NiMH<br>1000mAh | 3V Lithium Coin Cell<br>(CR2450), 600mAh | 2.4V NiMH AA, 1600mAh |
| Environ-<br>mental<br>specifi-<br>cations | Operating:<br>0 to +40 °C<br>30% to 90% RH<br>Storage:<br>-20 to +70 °C<br>10% to 90% RH | Operating:<br>0 to +40 °C<br>30% to 90% RH<br>Storage:<br>-20 to +70 °C<br>10% to 90% RH | Operating:<br>10 to +35 °C<br>20% to 80% RH<br>Storage:<br>-20 to +70 °C<br>30% to 90% RH |
| | MicroLab K031102 | Micro Diary K100928 | Micro I<br>K113096 (new) |
| Accessory<br>(patient<br>contacting) | VOL2104 Nose Clips (pack<br>of 5)<br>Body material: Polyacetal<br>Pad material:<br>Ethylene Vinyl Acetate | VOL2104 Nose Clips (pack<br>of 5)<br>Body material: Polyacetal<br>Pad material:<br>Ethylene Vinyl Acetate | Identical<br>(to K100928 & K031102) |
| | PSA1000 Adult Disposable<br>Mouthpiece<br>(material: Polyethylene coated<br>bleached kraft paper)<br>SST1250 One-way Safety<br>Mouthpiece (250 per box)<br>(material: Polyethylene coated<br>bleached kraft paper / Safety<br>wheel & valve: Polystyrene &<br>Rectaleen 8/170)<br>SST1000 One-way Safety<br>Mouthpiece (500 per box)<br>(material: Polyethylene coated<br>bleached kraft paper / Safety<br>wheel & valve: Polystyrene &<br>Rectaleen 8/170)<br>PSA1200 Paediatric<br>Disposable Mouthpiece<br>(material: Polyethylene coated<br>bleached kraft paper) | PSA2200 Mouthpiece<br>(material: ABS) | Identical<br>(to K031102 & also used in<br>Micro Diary Card K965042) |
| | PSA1100 Paediatric<br>adapter<br>(Polypropylene co-polymer, natural<br>Stamylan P512MN10, colour<br>UN0001 white) | N/A | Identical<br>(to K031102) |
·
.
{4}------------------------------------------------
・
·
{5}------------------------------------------------
## Summary to the table above:
The Micro I and the main predicate device MicroLab have the same indication for use only the sentence for the patient population has been changed into "over the age of 3 years" instead of "4 years to 17 years" and "18 to 99 years". The Micro I has the same technological characteristics as the predicate devices. For the materials used in the Micro I spirometer which are not identical to the predicate devices a biocompatibility test has been done. The patient contacting accessories are the same as used in the predicate devices.
### 5.15 Summary of Device Testing
The following practices were followed and monitored for development of the Micro I Spirometer:
- . The modification for the above device was developed in accordance with the CareFusion Design and Development QP 0301.
- The modifications were developed according to IEC 62366 (Usability) . standard.
{6}------------------------------------------------
- The risk analysis method used to assess the impact of the modifications . was a Failure Modes and Effects Analysis (FMEA) according standard ISO 14971.
- Biocompatibility for the patient contacting material has been tested . according ISO 10993-1 standard at Nelson test laboratory in the USA.
- Safety and EMC testing to IEC 60601-1 and IEC 60601-2 .
- Packaging test .
- Environmental Testing (climatic chamber) .
- Testing according ATS/ERS standardization of lung function testing .
### 5.16 Conclusion
Based on the above, CareFusion 232 UK concludes that the Micro I Spirometer is substantially equivalent to the legally marketed predicate devices, the CareFusion Microlab Spirometer K031102 and Micro Diary K100928 and is safe and effective for its intended use, and performs at least as well as the predicate devices.
{7}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. The text is written in all capital letters.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Elmar Niedermeyer Regulatory Affairs Specialist CareFusion Germany 234 GmbH Leibnizstrasse 7 Hoechberg, Bavaria GERMANY 97204
Re: K113096
Trade/Device Name: Micro I Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: II Product Code: BZG Dated: February 27, 2012 Received: February 29, 2012
### Dear Mr. Niedermeyer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
APR 1 3 2012
{8}------------------------------------------------
### Page 2 - Mr. Niedermeyer
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Nh for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
### Enclosure
{9}------------------------------------------------
## Indications for Use
510(k) Number (if known):
K113096
Device Name:
Micro I
Indications for Use:
The Micro I spirometer is intended to measure the maximal volume and flow of air that can be moved out of a patient's lungs. The system is intended for use with pediatic and addlt patients over the age of 3 years in hospitals, physician offices, laboratories and occupational health testing environments.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
റ
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.