K161295 · Pulmone Advanced Medical Devices, Ltd. · BZC · Oct 21, 2016 · Anesthesiology
Device Facts
Record ID
K161295
Device Name
MiniBox +
Applicant
Pulmone Advanced Medical Devices, Ltd.
Product Code
BZC · Anesthesiology
Decision Date
Oct 21, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1880
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The PulmOne MiniBox+ is intended to measure lung function in adult and pediatric patients, 5 years and older, while at rest (including spirometry, lung volumes and diffusing capacity). The PulmOne MiniBox+ is to be used by a physician, respiratory therapist, or technician in a hospital or clinic setting.
Device Story
MiniBox+ measures lung function, including spirometry, lung volumes, and diffusing capacity (DLco). Device inputs include patient breath samples (inspired/expired gas concentrations of CO and CH4) and mouth pressure/flow measurements. It utilizes an embedded infrared absorption gas analyzer and a single-breath maneuver to calculate DLco, following ATS/ETS guidelines. Lung volumes are calculated using proprietary equations based on mouth pressure and flow. Used in hospitals or clinics by physicians, respiratory therapists, or technicians. Output is displayed on an integrated LCD touch panel for clinical assessment of pulmonary function. Benefits include non-invasive, resting-state lung function diagnostics.
Clinical Evidence
No clinical study was conducted. Evidence consists of bench testing, including DLco performance verification against a Hans Rudolf simulator, human comparative testing, and human reproducibility testing.
Technological Characteristics
Class II pulmonary-function data calculator. Features an embedded infrared absorption gas analyzer for CO/CH4. Uses a single-use, disposable viral-bacterial filter. Complies with IEC 60601-1 (3.1 Ed) and IEC 60601-1-2 (4th Ed). Dimensions: 21 x 10.5 x 4.5 cm; Weight: ~10 kg. Software/firmware updated to support the DLco module.
Indications for Use
Indicated for measuring lung function (spirometry, lung volumes, diffusing capacity) in adult and pediatric patients aged 5 years and older at rest. Used by physicians, respiratory therapists, or technicians in clinical settings.
Regulatory Classification
Identification
A pulmonary-function data calculator is a device used to calculate pulmonary-function values based on actual physical data obtained during pulmonary-function testing.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 21, 2016
PulmOne Advanced Medical Devices, Ltd % Dr. Susan Alpert Principal SFA Consulting, LLC 200 Park Avenue, Unit 111 Minneapolis, Minnesota 55415
Re: K161295
Trade/Device Name: MiniBox + Regulation Number: 21 CFR 868.1880 Regulation Name: Pulmonary-Function Data Calculator Regulatory Class: Class II Product Code: BZC, BZG Dated: September 20, 2016 Received: September 22, 2016
Dear Dr. Susan Alpert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K161295
Device Name MiniBox+
Indications for Use (Describe)
The PulmOne MiniBox+ is intended to measure lung function in adult and pediatric patients, 5 years and older, while at rest (including spirometry, lung volumes and diffusing capacity). The PulmOne MiniBox+ is to be used by a physician, respiratory therapist, or technician in a hospital or clinic setting
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
______________________________________________________________________________________________________________________________________________________________________________
## 5 SECTION 5 – 510(K) SUMMARY
#### 5.1 Administrative Information
| Date: | 15-October-2016 |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | PulmOne Advanced Medical Devices, Ltd.<br>14 HaMelacha Street, Ra'ananna 4366107 , ISRAEL<br>Tel: +972 77 5100 938<br>Fax: +972 77 5100 939 |
| Official Correspondent: | Avi Lazar, CEO |
| Trade Name: | MiniBox+ |
| Classification Name:<br>Classification Number:<br>Product Code:<br>Device Class: | Pulmonary-function data calculator<br>21 CFR 868.1880<br>BZC, BZG<br>Class II |
| Predicate Devices: | Primary Predicate:<br>PulmOne MiniBoxPFT™ 2.0<br>510(k) Number -K141793<br><br>Secondary Predicate:<br>ZAN LUNG-FUNCTION LAB<br>510(k) Number - K052328 |
| Application Correspondent: | Dr. Susan Alpert<br>SFA Consulting, LLC<br>200 Park Ave., unit 111<br>Minneapolis, MN 55415<br>Phone- 612-202-7019<br>drsusanalpert@gmail.com |
{4}------------------------------------------------
#### 5.2 DEVICE DESCRIPTION
The PulmOne MiniBox+ is intended to measure lung function in adult and pediatric patients while at rest (including spirometry, lung volumes and diffusing capacity). The PulmOne MiniBox+ is to be used by either a physician, respiratory therapist, or technician. A cleared single-use, disposable viral-bacterial filter separates the patient from the internal components of the device.
The MiniBox+ measures all common spirometric measurements as well as relative and absolute lung volumes and diffusing capacity.
The MiniBox+ is equipped with a diffusing capacity module which is embedded into the MiniBox+ enclosure and located in the upper end of the device. This feature allows easy measurement of the volumes and concentrations (CO and CH4) of inspired and expired breath, of the test gas mixture and patient sample respectively, making possible the calculation of the diffusion capacity [DLco]. MiniBox + DLco [diffusing capacity factor of the lung for carbon monoxide (CO)] module uses the single breath concept and enables the measurements by setting and controlling the maneuver stages as defined in ATS/ETS DLco guidelines.
#### 5.3 INTENDED USE AND INDICATIONS FOR USE
The PulmOne MiniBox+ is intended to measure lung function in adult and pediatric patients, 5 years and older, while at rest (including spirometry, lung volumes and diffusing capacity). The PulmOne MiniBox+ is to be used by a physician, respiratory therapist, or technician in a hospital or clinic setting.
#### 5.4 Summary of Technical Characteristics
The MiniBox+ has been updated from the predicate version K141793. There have been no changes to the general intended use of the device to measure lung function or to the fundamental scientific technology. The Indications for use were updated to include the diffusing capacity module which is substantially equivalent to the one in ZAN 500 [K052328] and both utilize the same technologies and methods of operations. In addition, the software, firmware and hardware for the MiniBox+ in this submission has been updated to introduce minor software, firmware and hardware modifications to support this change.
{5}------------------------------------------------
# 5.4.1 Summary table of Comparison
| Specification | MiniBox+ | MiniBoxPFT™ 2.0 | ZAN LUNG-FUNCTION LAB<br>(ZAN 500 Body<br>Plethysmograph) | Differences discussion |
|--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | | | |
| 510(k) Number | Proposed Device | K141793 | K052328 | NA |
| Product Code / | BZC- Calculator, Pulmonary | BZC- Calculator, Pulmonary | BZC- Calculator, Pulmonary | Identical |
| Classification | Function Data | Function Data | Function Data | |
| | BZG- Spirometer, Diagnostic<br>Class II | BZG- Spirometer, Diagnostic<br>Class II | BZG- Spirometer, Diagnostic<br>Class II | |
| Intended Use and | The PulmOne MiniBox+ is | The PulmOne MiniBoxPFT™ | ZAN Lung-Function Lab may | Equivalent. |
| Indications for | intended to measure lung | 2.0 is intended to measure | measure or monitor pulmonary | MiniBox + shares the same |
| Use | function in adult and | lung function in adult and | function in adult and pediatric | intended use as the cleared |
| | pediatric patients, 5 years | pediatric patients while at | subjects during exercise | MiniBoxPFT™ 2.0 device with |
| | and older, while at rest | rest (including spirometry | (including diagnosis, training, | addition of diffusing capacity |
| | (including spirometry, lung | and lung volumes). The | and stress testing) or while at | module. diffusing capacity |
| | volumes and diffusing | PulmOne MiniBoxPFT™ 2.0 | rest | module added to Minibox + is |
| | capacity). The PulmOne | is to be used by either a | (including spirometry, airway | included with ZAN 500 |
| | MiniBox+ is to be used by a | physician, respiratory | strength and compliance, | intended use |
| | physician, respiratory | therapist, or technician. | diffusion capacity, body | |
| | therapist, or technician in a<br>hospital or clinic setting | | plethysmography, nutritional<br>assessment, and indirect | |
| | | | cardiac output) for diagnosis, | |
| | | | rehabilitation, performance | |
| | | | enhancement, and other | |
| | | | related activities | |
| Target | Adult and Pediatric, 5 years | Adult and Pediatric | Adult and Pediatric | Identical to MiniBoxPFT™ 2.0 |
| Population | and older | | | |
| Standards / | General medical device | General medical device | General medical device safety: | Equivalent. |
| Performance | safety: IEC 60601- | safety: | EN 60601-1 Medical electrical | MiniBox + complies with the |
| guideline | 1:2005+A1:2012 [3.1 Ed. ] | IEC 60601-1:2005 (3rd Ed) | safety: Electromagnetic | IEC 60601-1 [3.1Ed] and IEC<br>60601-1-2[4th Ed] |
| compliance | Medical electrical safety:<br>Electromagnetic | Electromagnetic | compatibility: EN 60601-1-2 | |
| | compatibility: IEC 60601-1- | compatibility: | | |
| | 2:2014 [4th Ed] | IEC 60601-1-2:2007 (3rd Ed) | | |
| | - ATS/ETS :2005, | -Performance per ATS 2005 | | |
| | Standardization of lung | Standards | | |
| | function testing | - ISO 26782 compliance to | | |
| | -- ISO 26782 compliance to | section 7.2, 7.4-7.6 & 7.8-7.9 | | |
| | section 7.2, 7.4-7.6 & 7.8-7.9 | | | |
| Patient | Patient direct contact with | Patient direct contact with | Single-use, disposable FDA- | Identical to MiniBoxPFT™ 2.0 |
| Contacting Parts | the cleared disposable viral | the cleared disposable viral | cleared viral bacterial filter or a | |
| | and bacterial filter, K051712 | and bacterial filter, K051712 | mouthpiece. Single-use, | |
| User Interface | LCD display with touch panel; | LCD display with touch panel; | disposable nose clips<br>LCD display with keyboard and | Identical to MiniBoxPFT™ 2.0 |
| and reports | top table device | top table device | mouse | |
| Physical | 21 x 10.5 x 4.5 cm | 21 x 10.5 x 4.5 cm | Dimensions: 71 x 87 x 174 cm | Equivalent. |
| Dimension | Weight: ~10 kg | Weight: ~300gr | Volume: 980 L Weight: 145 kg | Additional Weight due to<br>embedded DLco module |
| | | | | |
| | | | | |
| DIFFUSING CAPACITY [Gas Analyzer] | | | | |
| | -Technology: Infrared | Not Applicable | -Technology: Infrared | Identical |
| CO Analyzer | absorption | | absorption | |
| Specification | MiniBox+ | MiniBoxPFT™ 2.0 | ZAN LUNG-FUNCTION LAB<br>(ZAN 500 Body<br>Plethysmograph) | Differences discussion |
| | -Measurement range: 0 to<br>3000 ppm<br>-Accuracy: <1%<br>-Neutral point drift: <2% per<br>week<br>-Linearity: <1% | | -Measurement range: 0 to 3000<br>ppm<br>-Accuracy: <1%<br>-Neutral point drift: <2% per<br>week<br>-Linearity: <1% | |
| CH4 analyzer | Technology: Infrared<br>absorption<br>-Measurement range: 0 to<br>3000 ppm<br>-Accuracy: 2.5%<br>-Neutral point drift: <2% per<br>week<br>-Linearity: <1% | Not Applicable | -Technology: Infrared<br>absorption<br>-Measurement range: 0 to 3000<br>ppm<br>-Accuracy: 1%<br>-Neutral point drift: <2% per<br>week<br>-Linearity: <1% | Equivalent.<br>Since DLco is computed by<br>the ratio of CH4-in/CH4-out,<br>the more important Gas<br>Analyzer parameter is –<br>linearity. That linearity was<br>tested and verified by the<br>OEM manufacturer and<br>through extensive tests<br>versus the DLco gas simulator<br>[Hans Rudolf, Inc. The Gas<br>Analyzer accuracy was also<br>established by human<br>measurements, based on<br>internal repeatability<br>measurements as well as<br>versus the predicate device -<br>the ZAN Function Lab [ZAN-<br>500] concluded that MiniBox+<br>didn't adversely impact the<br>substantial equivalence to<br>ZAN 500 , the predicate<br>device |
| Test gas | mixture of 0.3% Carbon<br>monoxide (CO), 0.3%<br>Methane (CH4),<br>21% Oxygen (O2) with<br>balance Nitrogen (N2). | Not Applicable | mixture of 0.3% Carbon<br>monoxide (CO), 0.3% Methane<br>(CH4),<br>21% Oxygen (O2) with balance<br>Nitrogen (N2). | Identical |
| Demand Valve | OEM by MSA | Not Applicable | OEM by MSA | Equivalent<br>Same component as included<br>with ZAN 500 |
| Calculation and<br>Displayed<br>parameters | BHT , DLco, VA,VI, CO , CH4<br>and KCO | Not Applicable | BHT , DLco, VA,VI, CO , CH4 and<br>KCO | Identical |
| LUNG VOLUME MEASUREMENT | | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.