The LCI Option for Innocor is intended to measure the Lung Clearance Index (LCI), which is the cumulative expired volume required to clear an inert gas from the lungs during normal breathing in a multiple-breath washout (MBW) test divided by the Functional Residual Capacity (FRC). The specific parameters measured by the Innocor LCI Option include: LCI (Lung Clearance Index) and FRC (Functional Residual Capacity).
Device Story
LCI Option for Innocor is a software-based enhancement for the existing Innocor point-of-care pulmonary function system. It utilizes inert gas rebreathing (IGR) for initial wash-in of SF6 gas to determine Functional Residual Capacity (FRC), followed by multiple-breath washout (MBW) to measure cumulative expired volume required to clear the tracer gas. The device uses a pneumotachometer for flow measurement and a photoacoustic infrared gas analyzer for SF6 concentration. Operated by clinicians in point-of-care settings, the system processes respiratory data to calculate LCI, representing lung volume turnovers required for gas clearance. Output is used by healthcare providers to assess peripheral airway function and lung volume. The device benefits patients by providing non-invasive pulmonary diagnostics.
Clinical Evidence
Bench testing only. No clinical data. Performance evaluated using a calibration syringe to simulate lungs for FRC measurement and simulation models for LCI wash-out testing. Results demonstrate the device is as safe and effective as predicate devices.
Technological Characteristics
Compact, portable point-of-care device. Hardware identical to Innocor (K051907). Sensing: pneumotachometer for flow, photoacoustic infrared gas analyzer for SF6 concentration. Connectivity: standalone. Software: LCI calculation algorithm. Sterilization: N/A (non-invasive).
Indications for Use
Indicated for use in children and adults for pulmonary function testing to measure Lung Clearance Index (LCI) and Functional Residual Capacity (FRC) via multiple-breath inert gas washout (MBW).
Regulatory Classification
Identification
A pulmonary-function data calculator is a device used to calculate pulmonary-function values based on actual physical data obtained during pulmonary-function testing.
SensorMedics Vmax Series Pulmonary/Metabolic System (K942211)
Submission Summary (Full Text)
{0}------------------------------------------------
# 5. 510(k) Summary
Image /page/0/Picture/1 description: The image shows the word "innovision" in a stylized font. The letters are black and bold, and the word is slightly slanted upwards from left to right. There is a curved line above the "i" in "innovision", with a small circle at the end of the line.
NOV
1 2011
LCI Option
| Date of Summary | April 5, 2011 |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter/Contact<br>Person | H. Carl Jenkins<br>The Wood Burditt Group<br>10 E. Scranton Ave., Suite 201<br>Lake Bluff, IL 60044<br>(ph) (847) 234-7500 x 205<br>(fax) (847) 574-0728<br>(email) hcjenkins@woodburditt.com |
| Applicant | Innovision A/S<br>Lindvedvej 75<br>DK-5260 Odense S<br>Denmark |
| | Phone: +45 65 95 91 00<br>Fax: +45 65 95 78 00<br>info@innovision.dk<br>www.innovision.dk |
| Device Name | LCI Option for Innocor |
| Common Name | Pulmonary function test equipment |
| Classification | <b>LCI Option for Innocor - K102047</b><br>Calculator, pulmonary function data<br>Regulation Number: 21 CFR §868.1880<br>Product Code: BZC<br>Panel Code: Anesthesiology<br>Device Class: II |
| Legally Marketed<br>Predicate Devices | The LCI Option for Innocor is substantially equivalent in respect<br>to the intended use, design and method of operation to the<br>following legally marketed devices:<br><br>Predicate Device No. 1<br>Name: Innocor |
| | 510(k) number: K051907 |
| | Manufacturer: Innovision A/S, Denmark |
| | Predicate Device No. 2 |
| | Name: Cardiopulmonary Exercise Testing Option |
| | 510(k) number: K071911 |
| | Manufacturer: Innovision A/S, Denmark |
| | Predicate Device No. 3 |
| | Name: Spirometry Option |
| | 510(k) number: K083879 |
| | Manufacturer: Innovision A/S, Denmark |
| | Predicate Device No. 4 |
| | Name: COSMED QUARK |
| | 510(k) number: K001174 |
| | Manufacturer: COSMED S.r.l., Italy |
| | Predicate Device No. 5 |
| | Name: SensorMedics Vmax Series Pulmonary/Metabolic System |
| | 510(k) number: K942211 |
| | Manufacturer: SensorMedics |
| Device Description | Innocor is a compact point-of-care device intended to be used for<br>non-invasive measurement of a) cardiac output (CO) and other<br>hemodynamic parameters utilizing inert gas rebreathing (IGR)<br>technology, b) metabolic parameters including oxygen uptake by<br>means of a breath-by-breath gas exchange method, c) spirometry<br>parameters by means of forced respiratory maneuvers, and<br>d) peripheral airway function by means of multiple-breath inert<br>gas washout. |
{1}------------------------------------------------
{2}------------------------------------------------
The basic Innocor provides cardiac output (CO) as the principal measured parameter. Utilizing inert gas rebreathing, Innocor measures the relative levels of two inhaled gases of differing blood solubility over approximately 3-4 respirations. The disappearance curve for the blood soluble gas is used to calculate pulmonary blood flow (PBF), which in the absence of a significant intrapulmonary shunt is equal to cardiac output. The functional residual capacity (FRC) is determined from the dilution of the relatively insoluble gas during the same maneuver.
The LCI Option for Innocor, used alone or in conjunction with the entire Innocor system, provides information on lung volume and peripheral airway function.
The LCI is calculated as the cumulative expired volume (VcE) required to clear the inert tracer gas from the lungs during normal breathing, minus the product of the number of wash-out breaths and the external dead space outside the lips. divided by the subject's Functional Residual Capacity (FRC). FRC is the amount of air that stays in the lungs (up to the lips) after a normal expiration. In other words, LCI represents the number of lung volume turnovers (i.e. FRCs) that the subject must breathe to clear the inert tracer gas from the lungs (by convention, to an end-tidal concentration of 1/40th of the starting concentration over three subsequent breaths).
Innocor uses a combination of two techniques to determine the LCI, using SF6 as the inert tracer gas:
Inert gas rebreathing (IGR) is used for rapid wash-in of a very small amount of SF% until an even concentration is obtained in the lungs before the wash-out can start. This allows accurate determination of the functional residual capacity (FRC) by gas
{3}------------------------------------------------
analysis alone.
This is followed by multiple-breath wash-out (MBW) for determination of the cumulative expired volume (VCF) required to clear the SF6 from the lungs. The Innocor device is a compact and portable point-of-care device. The hardware in the LCI Option for Innocor is identical to that used in the Innocor Diagnostic Programmable Computer (K051907) and the Cardiopulmonary Exercise Testing Option to Innocor (K071911). combined rebreathing/multiple breath wash-out test The performed with the LCI Option for Innocor requires accurate measurements of the respiratory flow and the concentration of SF6 at the mouth during inspiration (inhalation) and expiration (exhalation). The Innocor device measures the flow and SF6 concentration by means of of a pneumotachometer and a photoacoustic infrared gas analyzer, respectively, while the patient takes multiple breaths over a period of time through a pneumatically activated respiratory valve unit (RVU). The rebreathing gas mixture for wash-in of SF6 is filled into a rubber bag prior to testing. Intended Use and The LCI Option for Innocor is intended to measure the Lung Clearance Index (LCI), which is the cumulative expired volume Indications required to clear an inert gas from the lungs during normal breathing in a multiple-breath washout (MBW) test divided by the Functional Residual Capacity (FRC). The specific parameters measured by the Innocor LCI Option include:
{4}------------------------------------------------
| Abbreviation | Name | Unit |
|---------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|
| LCI | Lung Clearance Index | (none) |
| FRC | Functional Residual capacity | L [BTPS] |
| Summary of technological<br>characteristics<br>compared to predicate<br>devices | Comparison with predicate device Innocor with options<br>[K051907, K071911, and K083879]:<br><br>The hardware in the LCI Option for Innocor is identical to that<br>used in the predicate devices. The connection of the patient and<br>the operation of the device are similar in that the wash-in is<br>performed by rebreathing as in the basic Innocor (K051907), and<br>the wash-out is performed by breathing room air in an open-circuit<br>breath-by-breath measurement as in the Cardiopulmonary Exercise<br>Testing Option (K071911). The gas mixture is the same as that<br>used with Innocor and approved by the FDA as an integral part of<br>the device (K051907). Thus, only the software is different from<br>previous models.<br><br>Comparison with predicate device COSMED QUARK<br>[K001174]:<br><br>Both the LCI Option for Innocor and the COSMED QUARK<br>predicate device are indicated for use in pulmonary function<br>testing to measure the wash-out of a gas from the lungs during a<br>multiple-breath procedure in open-circuit mode. The LCI Option<br>for Innocor performs the same LCI functions as the COSMED<br>QUARK, by providing lung function measurements of FRC and<br>LCI from analysis of a wash-out curve. Although there are some<br>minor descriptive characteristics differences between the LCI | |
.
{5}------------------------------------------------
Option for Innocor and the COSMED QUARK, performance data from bench testing demonstrates substantial equivalence. The performance of the LCI Option for Innocor has been evaluated against a calibration syringe for FRC determination and data from simulation for LCI determination. Performance data a demonstrates that the hardware and software of the LCI Option for Innocor is substantially equivalent to a legally marketed predicate diagnostic device for LCI determination.
Further, except for the way the indicator gas is washed in and the way FRC is determined, the LCI Option for Innocor meets the relevant recommendations in the Official American Thoracic Society/European Respiratory Society Statement: Pulmonary Function Testing in Preschool Children, Am J Respir Crit Care Med Vol 175. pp 1304-1345, 2007, Section 7: The Multiple-Breath Inert Gas Washout Technique.
### Comparison with predicate device SensorMedics Vmax Series Pulmonary/Metabolic System [K942211]:
Both the LCI Option for Innocor and the SensorMedics Vmax Series Pulmonary/Metabolic System (K942211) are intended to measure LCI, and are for use in children and adults.
The SensorMedics Vmax Series Pulmonary/Metabolic System was granted premarket clearance based, in part, on the Vmax calculating and providing LCI as one of the Distribution Parameters in the device's Normal Predicted Equation Sets.
The numerator and denominator of the fraction which defines LCI in the Vmax are defined in the Vmax 510(k) submission by reference to the cited article titled "A NEW INDEX OF THE INTRAPULMONARY MIXTURE OF INSPIRED AIR" by Margaret R. Becklake, Thorax (1952), 7, 111. More specifically,
{6}------------------------------------------------
| | the numerator and denominator of the fraction which defines LCI<br>are defined in the Becklake article as: Litres ventilation required to<br>wash 90% F.R.A. free of N2 / 90% F.R.A.<br>The LCI Option for Innocor employs a substantially equivalent<br>method of measuring LCI, using essentially the same indexes and<br>providing the same information. |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance Testing / | The LCI Option for Innocor has been evaluated against a |
| Summary of | calibration syringe simulating the lungs for FRC measurement (in |
| Substantial | the same way as for the predicate device, Innocor), and data |
| Equivalence (SE) | obtained from a bench test using a calibration syringe to simulate<br>the lungs has been compared with data from a simulation model of<br>an LCI wash-out test. Performance data demonstrates that the<br>hardware and software of the LCI Option for Innocor is as safe, as<br>effective, and performs as well as or better than the predicate<br>devices and thus that the device is substantially equivalent to the<br>predicate devices. |
| Compliance with | Except for the way the indicator gas is washed in and the way<br>FRC is determined, the LCI Option for Innocor has been<br>developed in accordance with the Official American Thoracic<br>Society/European Respiratory Society Statement: Pulmonary<br>Function Testing in Preschool Children, Am J Respir Crit Care<br>Med Vol 175. pp 1304-1345, 2007, Section 7: The Multiple-<br>Breath Inert Gas Washout Technique. |
| | The Statement includes the following remarks: |
| | MBW is performed routinely in preschool children in only<br>- a limited number of laboratories, presumably because<br>suitable equipment is not commercially available.<br>Several different inert marker gases with low solubility in |
| | blood and tissues can be used for MBW. The most well<br>known is nitrogen (N2), which can be washed out from the<br>lungs by letting the patient breathe pure oxygen (100%<br>O2). Other gases, such as argon (Ar), helium (He), or sulfur<br>hexafluoride (SF6), may also be used, but measuring these<br>gases may require expensive equipment, such as a mass<br>spectrometer. |
{7}------------------------------------------------
# 5. 510(k) Summary
{8}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
NOV - 1 2011
Innovision A/V Regulatory Counsel C/O Mr. H. Carl Jenkins The Wood Burditt Group LLC 10 E. Scranton Avenue, Suite 201 Lake Bluff, Illinois 60044
Re: K102047
Trade/Device Name: LCI Option for Innocor Regulation Number: 21 CFR 868.1880 Regulation Name: Pulmonary-Function Data Calculator Regulatory Class: II Product Code: BZC Dated: April 5, 2011 Received: April 6, 2011
Dear Mr. Jenkins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{9}------------------------------------------------
Page 2 - Mr. Jenkins
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices
Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{10}------------------------------------------------
#
Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: LCI Option for Innocor
Indications for Use:
The LCI Option for Innocor is intended to measure the Lung Clearance Index (LCI), which is the cumulative expired volume required to clear an inert gas from the lungs during normal breathing in a multiple-breath washout (MBW) test divided by the Functional Residual Capacity (FRC).
The specific parameters measured by the Innocor LCI Option include:
| Abbreviation | Name | Unit |
|--------------|------------------------------|----------|
| LCI | Lung Clearance Index | (none) |
| FRC | Functional Residual Capacity | L [BTPS] |
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schult
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
4. Indications for Use Statement
:
510(k) Number: Kn 10204
Page 4.1 of 4.1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.