The reprocessed AllFit Oximetry sensor, OXY-AFR, is a single-patient-use, adhesive sensor for short-term or long-term monitoring of functional oxygen saturation and pulse rate. The device is indicated for use by qualified medical personnel only.
Device Story
Reprocessed pulse oximetry sensor; utilizes red and infrared LEDs to measure functional oxygen saturation and pulse rate; consists of sensor head, 1-meter cable, and connector. Used in hospital, ICU, anesthesia, respiratory therapy, and transport settings; operated by qualified medical personnel (nurses/doctors). Reprocessing involves testing, cleaning, inspection, decontamination, and application of new patient adhesive over optical windows. Output displayed on compatible oximeter monitor; assists clinicians in patient physiological monitoring.
Clinical Evidence
Bench testing only. Includes verification and validation data, safety testing, accuracy testing, decontamination process validation, and biocompatibility reports. No clinical data presented.
Technological Characteristics
Adhesive pulse oximetry sensor; red and two infrared LEDs; optical windows; 1-meter cable with custom connector. Reprocessed device; materials identical to predicate except for patient window adhesive. Operates via optical transmission/detection. Single-patient-use.
Indications for Use
Indicated for short-term or long-term monitoring of functional oxygen saturation and pulse rate in small adult and pediatric patients in hospital, ICU, anesthesia, respiratory therapy, and transport settings. For use by qualified medical personnel only.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
K042891
Page 1 of 3
Image /page/0/Picture/2 description: The image shows a black and white logo. The logo consists of a large circle with a white letter 'D' inside. There is a small square in the upper right corner of the circle, resembling a camera flash.
JUN 2 9 2005
# Premarket Notification 510(k) Summary As required by section 807.92 Datex-Ohmeda OXY-AFR Sensor
### GENERAL COMPANY INFORMATION as required by 807.92(a)(1)
### COMPANY NAME/ADDRESS/PHONE/FAX:
Datex-Ohmeda, Inc 1315 W Century Dr Louisville, Colorado 80021 USA FDA Registration No. 1719176. Tel: 303-666-7001 Fax: 303-665-9176
### NAME OF CONTACT:
Mr. Joel Kent Datex-Ohmeda 86 Pilgrim Road Needham, MA 02492 USA Tel: 781-449-8685 Fax: 781-433-1344
### DATE:
October 17, 2004
### DEVICE NAME as required by 807.92(a)(2)
### TRADE NAME:
Datex-Ohmeda OXY-AFR Sensor
### COMMON NAME:
Reprocessed Pulse Oximetry Sensor
### CLASSIFICATION NAME:
The following Class II classifications appear applicable:
73 NLF (Reprocessed) Oximeter 21 CFR 870.2700
{1}------------------------------------------------
### NAME OF LEGALLY MARKETED DEVICE FOR WHICH A CLAIM OF SUBSTANTIAL EQUIVALENCE IS MADE as required by 807.92(a)(3)
The Datex-Ohmeda OXY-AFR Sensor is substantially equivalent in safety and effectiveness to the legally marketed (predicate) Datex-Ohmeda OXY-AF SENSOR (K021265).
### DEVICE DESCRIPTION as required by 807.92(a)(4)
This OXY-AFR sensor has a sensor head, consisting of a red LED and two infrared LEDS mounting on a leadframe, wired to a detector with a 1 meter cable and custom "happisnap" connector. There are two optical windows where the LED light is transmitted between the detector and LEDS. These components are packaged in a heat sealed package. The OXY-AFR is identical to the OXY-AF sensor with the exception of the new patient window adhesive. The ICU, a FR Sensor is for use in small adult and pediatric patients in the hospital, ICU, anesthesia, respiratory therapy, and during transport.
The Datex-Ohmeda OXY-AFR Sensor has the following differences when compared to the Datex-Ohmeda OXY-AF predicate device. The OXY-AFR sensor is a reprocessed OXY-AF sensor During reprocessing the used OXY-AF sensor is tested, cleaned, inspected and decontaminated. After decontamination new patient adhesive is installed over the optical windows and the reprocessed sensor is repacked with new removable tape. The OXY-AFR Sensor and the predicate OXY-AF SENSOR have the same performance specifications. The changes in the labels and instructions for use were made to add the reprocessed part numbers and name of the new device to the IFU. No changes in the warnings, cautions or contraindications have been made
#### INTENDED USE as required by 807.92(a)(5)
#### Intended use:
The reprocessed AllFit Oximetry sensor, OXY-AFR, is a single-patient-use, adhesive sensor for short-term or long-term monitoring of functional oxygen saturation and pulse rate
#### Indications for use:
The reprocessed AllFit Oximetry sensor, OXY-AFR, is a single-patient-use, adhesive sensor for short-term or long-term monitoring of functional oxygen saturation and pulse rate. The device is indicated for use by qualified medical personnel only.
{2}------------------------------------------------
## SUMMARY OF TECHNOLOGICAL CHARACTERITICS OF DEVICE COMPARED TO THE PREDICATE DEVICE as required by 807.92(a)(6)
The Datex-Ohmeda OXY-AFR Sensor is substantially equivalent in safety and Fffectiveness to the legally marketed (predicate) Datex-Ohmeda OXY-AF SENSOR (K040831).
The Datex-Ohmeda OXY-AFR Sensor has the following similarities to the Datex-Ohmeda OXY-AF SENSOR predicate device:
-have the same indicated use
-have the same fundamental scientific technology and use the same operating principle -are manufactured using the same processes
-constructed of identical materials except for the patient window adhesive which is equivalent to the original window adhesive
The Datex-Ohmeda OXY-AFR Sensor has the following differences when compared to the Datex-Ohmeda OXY-AF SENSOR predicate device:
-The OXY-AFR sensor is a reprocessed OXY-AF sensor
-The patient window adhesives are different but equivalent.
In summary, the Datex-Ohmeda OXY-AFR Sensor, described in this submission is substantially equivalent to the predicate OXY-AF SENSOR (K040831).
### SUMMARY OF NONCLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(1)(3)
Datex-Ohmeda OXY-AFR Sensors are normally used and installed by a trained nurse or doctor. Necessary precautions and warnings are stated on the instructions for use. The Verification and Validation data, safety testing report and Accuracy testing is included.
The decontamination process validation was conducted and this report is included.
Biocompatibility summary and testing reports are included as well.
#### CONCLUSION:
The summary above shows that there are no new questions of safety and effectiveness for the Datex-Ohmeda OXY-AFR Sensor as compared to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" is arranged in a circular pattern around the caduceus symbol.
SEP - 6 2006
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Joel C. Kent Manager, Quality and Regulatory Affairs GE Healthcare 86 Pilgrim Road Needham, Massachusetts 02492
Re: K042891
Trade/Device Name: Datex-Ohmeda OXY-AFR Sensor Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: NLF Dated: May 17, 2005 Received: May 18, 2005
Dear Mr. Kent:
This letter corrects our substantially equivalent letter of June 29, 2005
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Kent
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Clive, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
List of Models: Datex Ohmeda Reprocessed Sensors
1
Datex Ohmeda OxyTip+ OXY-AFR
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: Datex-Ohmeda OXY-AFR Sensor
Indications for Use:
The reprocessed AllFit Oximetry sensor, OXY-AFR, is a single-patient-use, adhesive sensor for short-term or long-term monitoring of functional oxygen saturation and pulse rate. The device is indicated for use by qualified medical personnel only.
Over-The-Counter Use _ Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Aingulion
on Siyn-Off) (Di-Divin of Anesthesiology, General Hospital, Infection Collast, Deman De vices
510(k) Number _
Page _ of _
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.