Coreray Reusable SPO2 Sensor is intended to be used in hospital settings where patient care is offered by qualified healthcare personnel. The Reusable SPO2 Sensor is intended for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult patients weighing greater than 40kg.
Device Story
Reusable SpO2 sensor; consists of connector, cable, and finger clip sensor housing. Contains two-wavelength LEDs (red 660nm, infrared 905nm) and photodetector. Operates by measuring optical absorption of functional arterial hemoglobin. Used in hospital settings by qualified healthcare personnel. Connects between patient and compatible pulse oximeter monitor. Provides continuous, non-invasive SpO2 and pulse rate data to monitor. Output assists clinicians in assessing patient oxygenation status and heart rate, facilitating timely clinical intervention.
Clinical Evidence
Clinical validation performed per ISO 80601-2-61:2011 requirements. Testing conducted in the 70%-100% SaO2 range to validate SpO2 accuracy. No animal testing performed.
Technological Characteristics
Materials: ABS, TPU, silicone. Sensing: 2-wavelength optical absorption (Red 660nm±10nm, IR 905nm±10nm). Form factor: Reusable finger clip. Connectivity: Wired connection to compatible pulse oximeter. Standards: IEC 60601-1, IEC 60601-1-2, ISO 80601-2-61, ISO 10993-5/10.
Indications for Use
Indicated for continuous non-invasive monitoring of functional oxygen saturation (SpO2) and pulse rate (PR) in adult patients weighing >40kg in hospital settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Solaris Reusable & Disposable SPO2 Sensors (K100077)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Shenzhen Coreray Technology, Ltd. % Field Fu Consultant Shenzhen Joyantech Consulting Co., Ltd. Room 1122, International Mayor Communication Center No. 55 Shi Zhou Zhong Road Nanshan, Shenzhen 518100 CHINA
Re: K173368
Trade/Device Name: Reusable SpO2 Sensor Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: July 24, 2018 Received: August 2, 2018
Dear Field Fu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/6 description: The image shows the name "Todd D. Courtney -S" in black text. The text is positioned in front of a light blue FDA logo. The text is clear and easy to read, and the logo is a simple and recognizable design. The image is likely a nameplate or a document header.
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173368
Device Name Reusable SPO2 Sensor
#### Indications for Use (Describe)
Coreray Reusable SPO2 Sensor is intended to be used in hospital settings where patient care is offered by qualified healthcare personnel. The Reusable SPO2 Sensor is intended for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate(PR) for adult patients weighing greater than 40kg.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Version: A/0
# SECTION 05 510(k) Summary
# 5.1 Administrative Information
| Date of Summary<br>prepared | Aug. 24, 2017 |
|----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer<br>information | Company title: Shenzhen Coreray Technology<br>Co., Ltd<br>Company address:<br>ChuangYe Technology Park, 1th Dong Huan<br>Road, Longhua New District, Shenzhen, 510000<br>P.R China. P.C.:518109<br>Contact person: Simon Fan<br>Phone: +86-755-28839229<br>Fax: +86-755-28839229<br>E-mail: manager@core-ray.com |
| Submission<br>Correspondent<br>Image: Logo<br>卓远天成 | Shenzhen Joyantech Consulting Co., Ltd.<br>Address: Room 1122, International Mayors<br>Communication Centre, NO. 55 Shizhou middle<br>road , Nanshan District, Shenzhen<br>Contact person: Mr. Field Fu<br>E-Mail: cefda13485@163.com |
# 5.2 Device Information
| Type of 510(k)<br>submission: | Traditional |
|-------------------------------|-----------------------|
| Trade Name: | Reusable SpO2 Sensor |
| Model: | CR001-3106A |
| Classification name: | Pulse Oximeter Sensor |
| Review Panel: | Cardiovascular |
| Product Code: | DQA |
| Device Class: | II |
{4}------------------------------------------------
Regulation Number: | 870.2700
## 5.3 Predicate Device Information
| Sponsor: | Shenzhen Caremed Medical Technology Co., Ltd. |
|----------------|-----------------------------------------------|
| Device: | Caremed Reusable & Disposable SpO2 Sensors |
| 510(K) Number: | K153184 |
| Sponsor: | Solaris Medical Technology Inc |
Medical Technology, Inc. Sponsor: Solaris Reusable & Disposable SPO2 Sensors Device: K100077 510(K) Number: |
### 5.4 Device Description
The Reusable SPO2 Sensor is comprised of a connector and a cable which terminates into sensor housing.
The sensor contains two specific wavelength LEDs and a photo detector assembled into the sensor housing which separate by one housing half and the other half. The sensor uses optical means to determine the light absorption of functional arterial hemoglobin by being connected between the patient and the oximeter.
One type of sensor housing is described in this submission:
Reusable finger clip sensor
The sensor has a labeling and specifications designed for compatibility with a specific monitor.
### 5.5 Intended Use/ Indications for Use
Coreray Reusable SPO2 Sensor is intended to be used in hospital settings where patient care is offered by qualified healthcare personnel. The Reusable SPO2 Sensor is intended for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult patients weighing greater than 40kg.
{5}------------------------------------------------
# 5.6 Technological characteristics of the subject device compared to the predicate device
# Predicate Device Information:
| 510(K) No.: | K153184 |
|------------------------|-------------------|
| Common name: | Oximeter |
| Classification name: | Oximeter |
| Production regulation: | 21 CFR § 870.2700 |
| Product code: | DQA |
| Panel: | Cardiovascular |
| 510(K) No.: | K100077 |
|------------------------|-----------------------|
| Common name: | Pulse Oximeter Sensor |
| Classification name: | Oximeter |
| Production regulation: | 21 CFR § 870.2700 |
| Product code: | DQA |
| Panel: | Cardiovascular |
# Comparison to predicate device:
| Comparison<br>item | Subject Device<br>Reusable and<br>disposable<br>SPO2 sensor | Predicate Device<br>K153184 | Predicate Device<br>K100077 | Remarks | |
|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|-----------------------------------------------------------|
| Product Code | DQA | DQA | DQA | same | |
| Regulation<br>Number | 870.2700 | 870.2700 | 870.2700 | same | |
| Classification | II | II | II | | |
| Intended<br>use&<br>Indication<br>s for Use | Coreray<br>Reusable<br>SPO2 Sensor is<br>intended to be | Caremed<br>Reusable &<br>Disposable SPO2<br>Sensors are | When used with a<br>compatible<br>patient monitor<br>or a pulse | SE as<br>K100077 | |
| | used in hospital<br>settings where<br>patient care is<br>offered by<br>qualified<br>healthcare<br>personnel. The<br>Reusable SPO2<br>Sensor is<br>intended for<br>continuous non-<br>invasive | indicated for<br>continuous<br>non-invasive<br>monitoring of<br>functional oxygen<br>saturation of<br>arterial<br>hemoglobin<br>(SpO2) and<br>pulse rate(PR) for<br>adult<br>patients weighing | oximeter device,<br>Solaris Medical<br>Technology,<br>Inc. reusable &<br>disposable<br>SpO2 sensors<br>are intended<br>to be used for<br>continuous,<br>non-invasive<br>functional<br>arterial oxygen | | |
| | monitoring of<br>functional<br>oxygen<br>saturation of<br>arterial<br>hemoglobin<br>(SpO2) and<br>pulse rate (PR)<br>for adult<br>patients<br>weighing<br>greater than<br>40kg.<br>Prescription<br>device. | greater<br>than40kg and<br>pediatric<br>patients weighing<br>10 -50 kg. | saturation<br>(SpO2)<br>and pulse rate<br>monitoring.<br>Solaris Medical<br>Technology,<br>Inc. reusable<br>multi-patient<br>use SpO2 Soft<br>Sensors,<br>reusable multi-<br>patient use<br>SpO2 Finger<br>Sensors, and<br>disposable single<br>patient<br>use SpO2 Soft-<br>finger<br>Sensors are for<br>use with<br>adult/pediatric<br>patients<br>weighing greater<br>than 40kg.<br>Solaris Medical<br>Technology,<br>Inc. disposable<br>single<br>patient use SpO2<br>Adhesive<br>Sensors are for<br>use with<br>adult patients | | |
| | | | adult patients | | |
| | | | greater than<br>40kg, pediatric<br>patients weighing<br>10 - 40<br>kg, and infant<br>(non-neonatal)<br>patients<br>weighing 3 -<br>15kg.<br>Prescription<br>device. | | |
| Measurement<br>Method | 2-wavelength<br>Relative<br>Optical<br>Absorption | 2-wavelength<br>Relative<br>Optical<br>Absorption | 2-wavelength<br>Relative<br>Optical<br>Absorption | same | |
| Light<br>Emitting | Red:<br>660nm±10nm<br>Infrared:<br>905nm±10nm | Red:660-666nm,<br>Ired:880-950nm | Red:660-666nm,<br>Ired:880-950nm | similar<br>Within<br>the range<br>of the<br>frequency<br>of the<br>predicate<br>devices. | |
| Signal<br>Detection<br>Method | Photodetector | Photodetector | Photodetector | same | |
| SPO2<br>Accuracy | ±3% (70-100%) | ±3% (70-100%) | ±2% (70-100%) | same as<br>K153184,<br>and<br>compliance with<br>the<br>ISO80601<br>1-2-61. | |
| Pulse<br>Rate<br>Accuracy | ±3(30-250bpm) | ±3(30-250bpm) | ±3(30-250bpm) | same | |
| Applied<br>population | Adult (≥40Kg) | Adult(≥40Kg)<br>Pediatric(10-<br>50Kg) | Adult(≥40Kg)<br>Pediatric(10-<br>50Kg)<br>Neonatal ( 3- | SE as<br>K100077 | |
| | | | | 10Kg ) | |
| Measurement<br>part | Fingers | Fingers or toes | Fingers or toes | Fingers or toes | SE |
| Sterile | No | No | No | No | same |
| Material | ABS, TPU,<br>Silicone, | ABS, PVC, TPU,<br>Silicone,<br>sponge | ABS, PVC, TPU,<br>Silicone, sponge | ABS, PVC,<br>Silicone and 3M | similar<br>Passed<br>the<br>biocompa<br>tibility<br>test. |
| Biocompatibili<br>ty | Cytotoxicity<br>Irritation<br>Sensitization | Cytotoxicity<br>Irritation<br>Sensitization | Cytotoxicity<br>Irritation<br>Sensitization | Cytotoxicity<br>Irritation<br>Sensitization | same |
| Distal<br>connector<br>Design | clip | finger clip and<br>sponge<br>adhesive | finger clip and<br>sponge<br>adhesive | soft tip and textile<br>adhesive | clip<br>design<br>same as<br>K153184 |
| Electrical<br>Performance<br>and Safety | IEC60601-<br>1,IEC60601-1-<br>2,<br>ISO80601-2-<br>61,ISO10993-<br>5/10 | IEC60601-<br>1,IEC60601-1-2,<br>ISO80601-2-<br>61,ISO10993-<br>5/10 | IEC60601-<br>1,IEC60601-1-2,<br>ISO80601-2-<br>61,ISO10993-<br>5/10 | IEC60601-<br>1,IEC60601-1-2,<br>ISO80601-2-<br>61,ISO10993-<br>5/10 | same |
{6}------------------------------------------------
{7}------------------------------------------------
Version: A/0
{8}------------------------------------------------
Version: A/0
The subject device and the predicate device have the same intended use and similar technological characteristics; they both measure SpO2 values for the patients. Thus the subject device is substantially equivalent to the predicate devices.
# 5.7 Brief discussion of the nonclinical tests
Reusable SpO2 Sensor conforms to the following standards:
60601-1:2005+CORR.1:2006+CORR.2007+A1:2012 Medical IEC Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential
{9}------------------------------------------------
Performance.
IEC 60601-1-2: 2014 Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests
ISO 80601-2-61:2011 Medical Electrical Equipment - Part 2-61: Particular Requirements For Basic Safety And Essential Performance Of Pulse Oximeter Equipment
ISO 10993-5:2009 Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity.
ISO 10993-10:2010 Biological Evaluation of Medical Devices- Part 10: Tests for Irritation and Skin Sensitization.
# 5.8 Brief discussion of clinical tests
Clinical testing is conducted for validation of SoO2 the Coreray Reusable & Disposable SpO2 Sensor.
The Reusable SpO₂ Sensor was satisfied the requirements of FDA quidance and ISO 80601-2-61:2011 in the range of 70%-100%SaO2.
# 5.9 Other information (such as required by FDA guidance/Test)
Animal performance testing is not required and was not performed to demonstrate substantial equivalence of the Coreray Reusable SpO2 Sensor.
# 5.10 Conclusions
Based on the above information, we conclude the subject device, Reusable SpO2 Sensor, is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.