SpO2 Sensors are indicated for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate(PR) for adult patients weighing greater than 40kg at hospital facilities.
Device Story
SpO2 sensors connect to compatible pulse oximeters to monitor arterial hemoglobin oxygen saturation and pulse rate. Device consists of a connector, cable, two LEDs (red 660-666nm, infrared 880-950nm), and a photodetector housed in finger clip, soft tip, or textile adhesive formats. Operation involves transmitting red and infrared light through patient tissue; photodetector captures modulated light, converting it into electrical signals. External oximeter unit processes signals to calculate SpO2 and pulse rate, displaying plethysmographic curves and numerical values. Used in hospital facilities by clinicians. Provides real-time physiological data to assist in patient monitoring and clinical decision-making.
Clinical Evidence
Clinical hypoxia testing performed on 13 adult human volunteers. Accuracy of SpO2 sensors validated against arterial oxygen saturation (SaO2) measured via co-oximetry. Results support accuracy claims of ±3% for SpO2 (70-100%) and ±3 bpm for pulse rate (30-250 bpm).
Technological Characteristics
Materials: ABS, PVC, TPU, Silicone, 3M. Sensing: 2-wavelength optical absorption (Red 660-666nm, IR 880-950nm). Photodetector: 20V breakdown, 50ns response time. Connectivity: Wired (DB9/Round-head connectors) to compatible monitors. Standards: IEC 60601-1, IEC 60601-1-2, ISO 80601-2-61, ISO 10993-5/10.
Indications for Use
Indicated for continuous non-invasive monitoring of functional oxygen saturation (SpO2) and pulse rate (PR) in adult patients weighing >40kg in hospital settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Unimed Disposable And Reusable Spo2 Sensors (K142832)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right of it is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in a larger, bolder font, and "ADMINISTRATION" in a smaller font below.
June 5, 2018
Shenzhen Greatmade Tech limited Mei Mei QA Manager 3th floor, B building, BaiFuli Industrial Zone ShangHengLang, HuaHui Road Shenzhen, CN
Re: K173045
Trade/Device Name: SpO2 Sensor Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: April 26, 2018 Received: April 30, 2018
Dear Mei Mei:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K173045
Device Name SpO2 Sensor
Indications for Use (Describe)
SpO2 Sensors are indicated for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate(PR) for adult patients weighing greater than40kg at hospital facilities.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Section 6
# 510(K) Summary
#### 1. Prepared Date: 2018/2/24
### 2. Submitter Information
| Name | Shenzhen Greatmade Tech limited |
|---------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address | 3th floor , B building , BaiFuli Industrial Zone ,<br>ShangHengLang, HuaHui Road , DaLang Street , LongHua new<br>district , Shenzhen City , GuangDong Province ,China |
| Tel | 86-755-28144774 |
### 3. Contact Person
| Contact person | Mei |
|----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Title | QA manager |
| Address | 3th floor , B building , BaiFuli Industrial Zone ,<br>ShangHengLang, HuaHui Road , DaLang Street ,<br>LongHua new district , Shenzhen City , GuangDong<br>Province ,China |
| Tel | 0086-755-23463462 |
| Fax | 0086-755-29553084 |
| E-mail | Mei_FDA@foxmail.com |
#### 4. Proposed Device Information
| Trade Name | SpO2 Sensor |
|------------------|--------------------------------------------------------|
| Model | AF009, AS009, AF063, AS063, AF024-1, AS024-1, DS024-NB |
| Common name | Oximeter |
| Regulatory class | 디 |
| Production | |
| requlation | 21 CFR §870.2700 |
| Product code | DQA |
| Panel | Cardiovascular |
### 5. Predicate Device Information
| 510(K)No. | Trade Name/model | Submitter |
|-----------|------------------------------------------------|------------------------------|
| K142832 | Unimed Disposable And Reusable<br>Spo2 Sensors | Unimed Medical Supplies Inc. |
### 6. Device description
{4}------------------------------------------------
The SpO2 Sensors are a family of oximeter sensors designed for compatibility with listed predicate oximeter manufacturers/monitors. The sensors are made up of connector, cable, two specific wavelength LEDs & a photo detector assembled into the sensor housing. The sensors use optical means to determine the light absorption of functional arterial hemoglobin by being connected between the patient and the oximeter. The Sensors contain finger clip type, soft tip and textile adhesive type. All the sensors will be labeled for compatibility for a specific monitor.
## 7. Operation principle& Mechanism of action
The SpO2 Sensors will connect with a compatible patient monitor or a pulse oximeter to continuously, non-invasively monitor the functional arterial oxygen saturation (SpO2) of the patient under stationary state. The SpO2 measurement is based on the absorption of pulse blood oxygen to red and infrared light by means of sensor and SpO2 measuring unit(Pulse Oximeter). The light-electronic transducer in sensor converts the pulse red and infrared light modulated by pulse blood oxygen into electrical signal, the signal is processed by hardware and software of SpO2 measuring unit(Pulse Oximeter). The pleth curve or(and) numeral value of Sp02 will be obtained.
### 8. Key electrical components description
Spo2 sensor is made up of connector, cable, two specific wavelength LEDs & a photo detector. LED & photo detector are the key electrical components to the subject device. The LED is with wavelength Red: 660-666/Infrared: 880-950m. The minimum breakdown voltage photo detector is 20V and its response time is 50ns.
### 9. Indications for use
SpO2 Sensors are indicated for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate(PR) for adult patients weighing greater than 40kg at hospital facilities.
#### Comparison to predicate device 10.
A comparison of key similarities and differences between the subject devices and the predicate devices (K142832) is provided below
| Comparison<br>item | Subject Device | Predicate Device<br>K142832 | Note |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Intended<br>use&<br>Indications<br>for Use | Spo2 sensors are indicated<br>for continuous non-invasive<br>monitoring of functional<br>oxygen saturation of arterial<br>hemoglobin (SpO2) and<br>pulse rate(PR) for adult | Unimed Disposable and<br>Reusable SPO2 Sensors<br>are indicated for continuous<br>non-invasive monitoring of<br>functional oxygen saturation<br>of arterial hemoglobin | Similar |
{5}------------------------------------------------
#### K173045
| Comparison<br>item | Subject Device | Predicate Device<br>K142832 | Note |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|---------|
| patients weighing greater<br>than 40kg. | (SpO2) and pulse rate(PR)<br>for adult patients weighing<br>greater than40kg, pediatric<br>patients weighing 10 -50 kg,<br>and neonatal patients<br>weighing no less than 3 Kg. | | |
| Measuremen<br>t Method | 2-wavelength Relative<br>Optical Absorption | 2-wavelength Relative<br>Optical Absorption | Same |
| Light<br>Emitting | Red: 660-666nm,<br>Ired: 880-950nm | Red: 660-666nm,<br>Ired: 880-950nm | same |
| Signal<br>Detection<br>Method | Photodetector | Photodetector | Same |
| SPO2<br>Accuracy | ±3%(70-100%) | ±3%(70-100%) | Same |
| Pulse Rate<br>Accuracy | ±3(30-250bpm) | ±3(30-250bpm) | Same |
| Applied<br>population | Adult(≥40Kg) | Adult(≥30Kg)&Pediatric(10-<br>50Kg) | Similar |
| Measuremen<br>t part | Fingers | Fingers or toes | Similar |
| compatible<br>monitor | Nellcor(N395)<br>Ohmeda3800 | Nellcor(N395)<br>Ohmeda3700 | Similar |
| Sterility | No | No | Same |
| Usage | Reusable&disposable | Reusable&disposable | Same |
| Material | ABS,PVC,TPU,Silicone,3M | ABS,PVC,TPU,Silicone,3M | Same |
| Cable Length | 3.0/1.0 m | 3.0/1.1m | Similar |
| Proximal<br>connector<br>Design | DB9 9pin&Round-head<br>7pin/8p | DB9 9pin&Round-head<br>7pin/8p | Same |
| Distal<br>connector<br>Design | finger clip , soft tip, textile<br>adhesive | finger clip , soft tip, textile<br>adhesive and sponge<br>adhesive | Similar |
| Conformance<br>standard | IEC 60601-1 ,IEC<br>60601-1-2, ISO<br>80601-2-61, ISO<br>10993-5/10 | IEC 60601-1 ,IEC<br>60601-1-2, ISO<br>80601-2-61, ISO<br>10993-5/10 | Same |
{6}------------------------------------------------
From the comparison form above, both devices have the same Measurement Method, Light Emitting, Signal Detection Method, Sterility, Usage & Conformance standard. There are slightly differences between the subject devices and predicate devices as follows.
| Difference clause | Discussion |
|-----------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use&<br>Indications for Use<br>& Applied<br>population | The subject device is intended use for adult<br>patients weighing greater than 40Kg. The<br>predicate device, which includes 12 models, is<br>intended for adult, pediatric and neonate patients.<br>but both devices are compatible with the same<br>monitor. This difference does not raise different<br>questions of safety and effectiveness. |
| Measurement part | The subject devices are intended to be used on<br>fingers not toes; the predicate devices will be<br>applied for fingers and toes. But the spo2 and PR<br>accuracy of subject device applied for fingers<br>meets the requirements of ISO 80601-2-61. This<br>difference does not raise different questions of<br>safety and effectiveness. |
| compatible monitor | The subject devices is compatible with<br>Nellcor(N395) and Ohmeda3800 monitors. The<br>predicate devices is compatible with Nellcor(N395)<br>and Ohmeda3700 monitors. This difference does<br>not raise different questions of safety and<br>effectiveness. |
| Cable Length | According to the clinical requirements, the subject<br>device's cable length were designed to be 1.0m &<br>3.0m. The predicate device's cable lengths are<br>1.1m & 3.0m. This difference does not raise<br>different questions of safety and effectiveness.<br>Meanwhile the subject device has passed the<br>IEC60601-1, IEC 60601-1-2 and ISO 80601-2-61. |
| Distal connector<br>Design | The spo2 sensor in this submission have 3 types<br>(finger clip , soft tip &3M). The predicate device<br>has an additional connector of sponge adhesive.<br>This difference does not raise different questions<br>of safety and effectiveness. This subject devices<br>have passed the IEC 60601-1, IEC 60601-1-2and<br>ISO 80601-2-61. |
#### Summary of Biocompatibility test 11.
{7}------------------------------------------------
According to IFU, the sensor end will contact with patient finger skin. This submission have three types of sensor housings (soft tip, finger clip & Non-woven) and its duration contact time is less than 2 hours.
The Nature of body contact is skin surface. And the contact duration is less than 24 hours. According to Table 1 - Initial evaluation tests for consideration in ISO 10993-1, the applicable tests are:
- Cytotoxicity ●
- Sensitization
- Irritation / Intracutaneous ●
#### 12. Summary of Non-clinical test data
Non clinical tests as follows were conducted to verify that the proposed device met all design specifications to demonstrate Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with requirements of the recognized standards and IFU.
- a) Biocompatibility test
- b) Testing/ANSI/AAMI/IEC60601-1
- c) Testing /ANSI/AAMI/IEC60601-1-2
- d) Testing /ISO80601-2-61
- e) Cleaning and Disinfection Validation test
- Shelf life test f)
- q) Performance test for sensors connecting with compatible monitors
- h) Pulse rate accuracy test
#### 13. Clinical test data
Clinical testing has been performed under an approved protocol with subject informed consent. Clinical hypoxia test results were obtained in 13 human adult volunteers to validate the accuracy of Spo2 Sensors versus arterial oxygen saturation (SaO2) as determined by co-oximetry. Clinical test results support device accuracy claims for the specified saturation range.
#### Substantial Equivalence Statement 14.
Based on the comparison, analysis, and the submitted performance data, Spo2 Sensors are as safe, as effective and performs as well as or better
{8}------------------------------------------------
than the predicate device and thus can be considered substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.