The WiCap™ patient monitor is intended for the continuous or spot-check monitoring of carbon dioxide concentration of the expired (EtCO2) and inspired (FiCO2) breath and respiration rate (RR), and functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) measurements. Intended patient populations include: Adult, Pediatric, and Neonate patients. The device is for use in hospitals, healthcare facilities and clinics, nursing home facilities, and other healthcare environments. The WiCap™ patient monitor is to be used by trained healthcare providers.
Device Story
WiCap Patient Monitor performs continuous or spot-check monitoring of EtCO2, FiCO2, respiration rate, SpO2, and pulse rate. Capnography uses side-stream NDIR spectroscopy; a DC-driven pump aspirates breath samples via nasal cannula to an OEM module. Pulse oximetry uses spectrophotometry; LEDs in patient-applied sensors interrogate tissue to measure light absorption differences between oxyhemoglobin and deoxyhemoglobin. Device is used in hospitals, clinics, and nursing homes by trained healthcare providers. Output is displayed for provider review to assist in physiological monitoring and clinical decision-making. Benefits include non-invasive, real-time assessment of respiratory and circulatory status.
Clinical Evidence
Bench testing only. No clinical studies on patients performed. SpO2 accuracy validated via induced hypoxia studies on healthy subjects (motion/no-motion) compared to co-oximeter blood samples per ISO 80601-2-61. OEM module compliance statements and functional testing using pulse oximeter simulators provided.
Technological Characteristics
Enclosure: plastic. Sensing: NDIR spectroscopy (CO2), spectrophotometry (SpO2). Energy: rechargeable lithium batteries. Connectivity: OEM modules for CO2 and SpO2. Standards: IEC 60601-1, IEC 60601-1-2 (Class B), IEC 60601-1-8, ISO 80601-2-55, ISO 80601-2-61. Software: embedded firmware, moderate level of concern.
Indications for Use
Indicated for continuous or spot-check monitoring of EtCO2, FiCO2, respiration rate, SpO2, and pulse rate in adult, pediatric, and neonate patients in clinical and healthcare environments.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Oridion Medical 1987 Limited, Capnostream 20 (K060065)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, layered on top of each other to create a sense of depth and unity.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 2, 2016
Athena GTX Sean Mahoney V.P. Regulatory Affairs 5900 NW 86th Street, Suite 300 Johnston, IA 50131
Re: K160582
Trade/Device Name: WiCap™ Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA, CCK Dated: August 4, 2016 Received: August 5, 2016
Dear Sean Mahoney:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below
510(k) Number (if known)
Device Name
WiCap™ Patient Monitor
#### Indications for Use (Describe)
The WiCap™ patient monitor is intended for the continuous or spot-check monitoring of carbon dioxide concentration of the expired (EtCO2) and inspired (FiCO2) breath and respiration rate (RR), and functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) measurements. Intended patient populations include: Adult, Pediatric, and Neonate patients. The device is for use in hospitals, healthcare facilities and clinics, nursing home facilities, and other healthcare environments. The WiCap™ patient monitor is to be used by trained healthcare providers.
Type of Use (Select one or both, as applicable)> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
PSC Publishing Services (301) 443-6740 EF
4 -
{4}------------------------------------------------
# 510(k) Summary
#### I. SUBMITTER 807.92(a)(1)]:
Athena GTX 5900 NW 86th Street, Suite 300 Johnston, lowa 50131
Phone: 515.288.3360 Fax: 515.288.3394
Company Contact: Sean Mahoney (VP of Regulatory Affairs) 515.288.3360 x103 smahoney@athenagtx.com
Date Prepared: August 31, 2016
II. DEVICE [807.92(a)(2)]:
| Trade Name: | WiCapTM |
|----------------------|----------------------------|
| Common Name: | Patient Monitor |
| Classification Name: | Oximeter (21 CFR 870.2700) |
| Device Class: | Class II |
|-----------------------|------------|
| Product Code: | CCK, DQA |
| Basis for Submission: | New Device |
#### III. PREDICATE DEVICE [807.92(a)(3)]:
Legally Marketed Oridion Medical 1987 Limited, Capnostream 20 (K060065) (Predicate) Device:
#### IV. DEVICE DESCRIPTION [807.92(a)(4)]:
#### Device Identification
The WiCap Patient Monitor is a device that monitors physiological parameters associated with carbon dioxide gas and pulse oximetry.
#### Device Characteristics
The WiCap Patient Monitor is a multi-patient use non-sterile device. It utilizes embedded firmware. Patient applied parts are needed for physiological measurement and are provided via FDA cleared OEM accessories to the WiCap monitor.
#### Environment of Use
WiCap is intended to be used in hospitals, healthcare facilities and clinics, nursing home facilities, and other healthcare environments.
\$\qquad 5-1\$
{5}------------------------------------------------
# Principle of Operation
WiCap uses non-dispersive infrared (NDIR) infrared spectroscopy to perform the capnography function (measuring carbon dioxide in exhaled breath (etCO2)). It is a side stream monitoring technique that utilizes a sample of the inhaled/exhaled patient's breath aspirated via a sample line (nasal cannula) to the CO2 OEM Module. The sample is drawn into the measurement module via DC driven pump.
Pulse oximetry measures the functional oxygen saturation of arterial hemoglobin. The patient's tissue is interrogated optically and the difference in the absorption of red and infrared light (spectrophotometry) by oxyhemoqlobin and deoxyhemoglobin is analyzed. This measurement is also done non-invasively via light emitting diodes (LEDs) in the sensors applied to the patient.
## Materials
The WiCap Patient Monitor enclosure is primarily plastic and does not contact the patient. The applied parts are OEM accessories that are FDA cleared and meet the biocompatibility requirements for intact skin contact.
## Key Performance Specification
Key performance specifications are listed in the table in section VI below under capnography and pulse oximeter. Specific accuracy specifications for sensors that are compatible with the XPod pulse oximeter OEM module, are listed in the table below.
| | Accuracy (± X digits) | | |
|------------------|-----------------------|--------|---------------|
| Sensor Model | No Motion | Motion | Low Perfusion |
| Adult/Pediatric | | | |
| 8000AA-1, -2, -3 | ± 2 | ± 3 | ± 2 |
| 8000AP-1, -3 | ± 2 | ± 3 | ± 2 |
| 8000SS-1, -3 | | | |
| 8000SM-1, -3 | ± 2 | ± 3 | ± 2 |
| 8000SL-1, -3 | | | |
| 8000J-1, -3 | ± 2 | --- | ± 2 |
| 8001J | ± 2 | --- | ± 2 |
| 8008J | ± 2 | --- | ± 2 |
| 8000Q2 | ± 3 | --- | ± 2 |
| 8000R | ± 2 | --- | ± 2 |
| 6000CA, CP, CI | ± 2 | ± 3 | ± 2 |
| 7000A, P, I | ± 2 | ± 3 | ± 2 |
| 6500MA, SA | ± 2 | ± 2 | ± 3 |
| Neonate | | | |
| 6000CN | ± 3 | --- | ± 2 |
| 7000N | ± 3 | --- | ± 2 |
{6}------------------------------------------------
# V. INDICATIONS FOR USE [807.92(a)(5)]:
The WiCap patient monitor is intended for the continuous or spot-check monitoring of carbon dioxide concentration of the expired (EtCO2) and inspired (FiCO2) breath and respiration rate (RR), and functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) measurements. Intended patient populations include: Adult, Pediatric, and Neonate patients. The device is for use in hospitals, healthcare facilities and clinics, nursing home facilities, and other healthcare environments. The WiCap patient monitor is to be used by trained healthcare providers.
The Indications for Use statement for the WiCap device is not identical to the predicate device: however, the differences do not alter the key intended use of the device nor affect the safety and effectiveness of the device relative to the predicate. Both the subject and predicate devices measure the same physiological parameters, are to be used on the same patient populations and in the same environments and are for use by trained healthcare providers.
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The WiCap Patient Monitor for monitoring end tidal CO2 and Pulse Oximetry employs the same technology principles as the predicate device. For CO2 measurements non-dispersive infrared (NDIR) infrared spectroscopy is used and for pulse oximetry functional oxygen saturation of arterial hemoglobin by analyzing the difference in the absorption of red and infrared light (spectrophotometry) by oxyhemoglobin and deoxyhemoglobin is used.
At a high level, the subject and predicate devices are based on the following same technological elements:
- . Uses OEM modules for Pulse Oximetry and Capnography (CO2 gas monitor)
- Same measurement techniques are used for both oximetry and capnography .
- Both have alarms
- Both comply to the same industry and FDA recognized standards ●
- Both use rechargeable lithium based batteries .
The following technological differences exist between the subject and predicate devices:
- Different OEM modules are used for Pulse Oximetry and Capnography (CO2 gas ● monitor)
- . WiCap imparts a lower optical power to the patient
- WiCap is smaller and lighter
- WiCap has a lower battery/system voltage
- . WiCap has been tested to more severe EMC requirements (Class B vs. Class A)
{7}------------------------------------------------
The following is provided as a summary of how the technological characteristics of the device compare to the predicate device:
| Product Element | WiCap | Capnostream20 | Comments |
|-----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------|
| Manufacturer | Athena GTX | Oridion Capnography Inc. | - - - |
| Model Number(s) | WiCap | Capnostream20 | - - - |
| 510(k) Number | Pending - new device | K060065 | - - - |
| Physiological Parameters | carbon dioxide analyzer (EtCO2)<br>and Pulse-Ox (SpO2 and Pulse<br>Rate) | carbon dioxide analyzer<br>(EtCO2) and Pulse-Ox (SpO2<br>and Pulse Rate) | Same |
| Patient Population | Adult, pediatric and neonate<br>patients | Adult, pediatric and neonate<br>patients | Same |
| Indications | The WiCap patient monitor is intended<br>for the continuous or spot-check<br>monitoring of carbon dioxide<br>concentration of the expired (EtCO2)<br>and inspired (FiCO2) breath and<br>respiration rate (RR), and functional<br>oxygen saturation of arterial<br>hemoglobin (SpO2) and pulse rate (PR)<br>measurements. Intended patient<br>populations include: Adult, Pediatric,<br>and Neonate patients. The device is for<br>use in hospitals, healthcare facilities and<br>clinics, nursing home facilities, and other<br>healthcare environments. The WiCap<br>patient monitor is to be used by trained<br>healthcare providers. | The Capnostream20 is intended for<br>CO2 and SPO2 indications. The<br>Capnostream20 combined<br>capnograph/pulse oximeter monitor<br>is intended to provide professionally<br>trained health care providers the<br>continuous, non-invasive<br>measurement and monitoring of<br>carbon dioxide concentration of the<br>expired and inspired breath and<br>respiration rate, and for the<br>continuous non-invasive monitoring<br>of functional oxygen saturation of<br>arterial hemoglobin (SpO2 and pulse<br>rate). It is intended for use with<br>neonatal, pediatric and adult patients<br>in hospitals, hospital type facilities,<br>intra hospital moves and home<br>environments. | Substantially<br>Equivalent |
| Environment of Use | Hospitals, healthcare facilities<br>and clinics, nursing home<br>facilities, and other healthcare<br>environments | Hospitals, hospital type<br>facilities, intra hospital moves<br>and home environments. | Similar -<br>Capnostream<br>20 includes<br>home use. |
| Product Element | WiCap | Capnostream20 | Comments |
| Compliance | IEC 60601-1:2012 Basic safety<br>and essential performance | IEC/EN60601-1<br>UL 60601-1<br>CSA C22.2 No 601.1-M90 | |
| | IEC 60601-1-2:2007-03 and<br>2014-02 EMC | IEC/EN60601-1-2 Class A<br>Radiated and Conducted<br>Emission | |
| | IEC 60601-1-8:2012-11 Alarms | IEC 60601-1-8 (Audible and<br>Visual Alarms) | Same |
| | ISO 80601-2-55:2011-12-15<br>Respiratory gas monitors | ISO 80601-2-55<br>(Capnography) | |
| | ISO 80601-2-61:2011-04-01<br>Pulse oximeters | ISO 80601-2-61 (Pulse<br>Oximetry) | |
| | IEC 60601-2-49:2011<br>Multifunction patient<br>monitoring equipment | IEC 60601-2-49 Multifunction<br>patient monitoring<br>equipment | |
| Capnography | | | |
| CO2, EtCO2, FiCO2 Range | 0 to 150 mmHg, 0 to 19.7% | 0-150 mmHg | Same |
| CO2 Accuracy | 0 - 40 mmHg ± 2 mmHg<br>41 - 70 mmHg ± 5% of reading<br>71 - 100 mmHg ± 8% of reading<br>101 - 150 mmHg ± 10% of<br>reading<br>Above 80 breath per minute ±<br>12% of reading<br>* NOTE: Gas temperature at<br>25°C | 0-38 mmHg: ± 2 mmHg<br>39-150 mmHg: ± (5% of<br>reading + 0.08 x (expected<br>reading in mmHg -39mmHg)<br>Accuracy applies for breath<br>rates of up to 80 bpm. For<br>breath rates above 80 bpm,<br>accuracy is 4 mmHg or ±12 %<br>of reading whichever is<br>greater, for etCO2 values<br>exceeding 18 mmHg. | Substantially<br>Equivalent |
| Respiration Rate Range | 2 to 150 breaths per minute<br>(bpm) | 0-150 bpm | Substantially<br>Equivalent |
| Respiration Rate Accuracy | ± 1 breaths per minute (bpm) | 0-70 bpm: ±1 bpm<br>71-120 bpm: ±2 bpm<br>121-150 bpm: ±3 bpm | Substantially<br>Equivalent |
| CO2 Alarms | No breath, etCO2 high, etCO2<br>low, FiCO2 high, FiCO2 low, RR<br>high, RR low | No breath, etCO2 high, etCO2<br>low, RR high, RR low, IPI low | Same |
| Flow Rate | 50 ml/min ±10 ml/min | 50 (42.5 ≤ flow ≤ 65) ml/min,<br>flow measured by volume | Substantially<br>Equivalent |
| Product Element | WiCap | Capnostream20 | Comments |
| Pulse-Oximeter | | | |
| SpO2 Measurement Range | 0-100% | 0-100% | Same |
| SpO2 Accuracy | Varies per sensor type.<br>Typical specs | | |
| 70%-100%<br>Adult & Pediatric<br>No Motion | ± 2 digits | ± 2 digits | Substantially<br>Equivalent |
| Motion<br>Low Perfusion | ± 3 digits<br>± 2 digits | | |
| Neonatal<br>No Motion | ± 3 digits | ± 3 digits | Equivalent |
| Motion | --- | | |
| Low Perfusion | ± 2 digits | | |
| 0% - 69% | Unspecified | Unspecified | |
| Pulse Rate Range | 30 to 250 beats per minute<br>(bpm) | 20-250 bpm | Substantially<br>Equivalent |
| Pulse Rate Accuracy | Varies per sensor type.<br>Typical specs | ± 3bpm | Substantially<br>Equivalent |
| No Motion/Low Perfusion | ± 3 digits | | |
| Motion | ± 5 digits | | |
| Alarms | SpO2 high, SpO2 low, Pulse Rate<br>high, Pulse Rate low | Adjustable Alarm Limits<br>SpO2 high, SpO2 low, Pulse<br>Rate high, Pulse Rate low | Same |
| Measurement Technology | | | |
| CO2 Gas Detection OEM<br>Module | Yes | Yes | Same |
| CO2 Measurement<br>Technique | Non-dispersive infrared (NDIR)<br>Infrared spectroscopy | Non-dispersive infrared<br>(NDIR) Infrared spectroscopy | Same |
| Pulse-Ox OEM Module | Yes | Yes | Same |
| Sp02 Measurement<br>Technique | Functional oxygen saturation of<br>arterial hemoglobin | Functional oxygen saturation<br>of arterial hemoglobin | Same |
| Pulse-Ox Sensors<br>Red Wavelength<br>IR Wavelength | LED<br>660 nanometers<br>910 nanometers | LED<br>660 nanometers<br>900 nanometers | Substantially<br>Equivalent<br>wavelengths. |
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
# VII. PERFORMANCE DATA [807.92(b)(1)]:
## Biocompatibility
The WiCap patient monitor does not come in contact with the patient and was not tested for biocompatibility.
OEM cleared accessories are specified for use that contact the patient. The 510(k) clearance information was provided as part of this submission.
OEM in-cable oximeter has been tested for Biocompatibility since it may come in contact with the patient.
These products comply with ISO 10993 for intact skin contact for greater than 30 days and were tested for compliance to requirements for
- Cytotoxicity ●
- . Sensitization
- . Irritation
# Industry Standards for Electrical Safety, EMC and Essential Performance
Testing of the WiCap Patient Monitor has been completed to verify compliance with FDA guidance, recognized national and international standards for electrical safety and performance for medical devices, and particular requirements applicable to this device including:
- Pulse Oximeters Premarket Notification Submissions [510(k)s] Guidance for Industry . and Food and Drug Administration Staff. Document issued on: March 4, 2013.
- IEC 60601-1:2012 Basic safety and essential performance ●
- IEC 60601-1-2:2007-03 and 2014-02 EMC ●
- IEC 60601-1-8:2012-11 Alarms ●
- IEC 60601-2-49:2011 Multifunction patient monitoring equipment
- ISO 80601-2-55:2011-12-15 Respiratory gas monitors ●
- ISO 80601-2-61:2011-04-01 Pulse oximeters
- IEC 60601-1-6: 2010 + IEC 62366:2007 Usability ●
#### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern.
#### Animal Study Data
None.
# Clinical Study Data [807.92(b)(2)]:
OEM module and sensor clinical accuracy reports and manufacturer compliance statements were provided for the pulse oximeter parameter of the WiCap patient monitor that demonstrated compliance with the FDA Pulse Oximeter Guidance and ISO 80601-2-61 requirements. No further clinical testing was performed. Successful integration of the OEM module and sensor(s) was tested via a functional tester (pulse oximeter simulator) spanning the monitor's claimed range of pulse rate and saturation values.
5 - 7
{11}------------------------------------------------
SpO2 accuracy testing was conducted during induced hypoxia studies on healthy, non-smoking, light-to-dark-skinned subjects during motion and no-motion conditions in an independent research laboratory. The measured arterial hemoglobin saturation value (SpO2) of the sensors were compared to arterial hemoglobin oxygen (SaO2) value, determined from blood samples with a laboratory co-oximeter. The accuracy of the sensors in comparison to the co-oximeter samples was measured over the SpO2 range of 70 - 100%. Accuracy data was calculated using the root-mean-squared (Arms value) for all subjects, per ISO 80601-02-61:2011, Standard Specification for Pulse Oximeters for Accuracy.
# VIII. CONCLUSION [807.92(b)(3)]:
The results for all safety, compliance, and non-clinical performance testing demonstrates that the Athena GTX WiCap Patient Monitor is as safe and as effective as the above listed predicate device, has the same performance characteristics and is substantially equivalent to the above listed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.