BEDSIDE RESPIRATORY PATIENT MONITORING SYSTEM, BEDSIDE RESPIRATORY PATIENT MONITORING SYSTEM WITH RESPIRATION RATE SOFTW
Applicant
Covidien
Product Code
DQA · Cardiovascular
Decision Date
Feb 4, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Nellcor™ Bedside Respiratory Patient Monitoring System is a portable pulse oximeter intended for prescription use only as a continuous non-invasive monitor of arterial oxygen saturation (SpO2) and pulse rate of adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused. The monitoring system is intended for use in hospitals, hospital-type facilities, and during intra-hospital transport. The OxiMax SPD™ Alert (SPD) feature is intended only for facility-use care of adults to detect patterns of desaturation indicative of repetitive reductions in airflow through the upper airway and into the lungs. The Respiration Rate parameter, when used in conjunction with the Nellcor™ Bedside Respiratory Patient Monitoring System and Nellcor™ Respiratory Sensor, is intended for the continuous, non-invasive monitoring of respiration rate in adult patients who are well perfused during no motion conditions, in hospitals and hospital-type facilities.
Device Story
Portable pulse oximeter; monitors arterial oxygen saturation (SpO2), pulse rate, and respiration rate. Inputs: optical signals from patient-attached sensors. Processing: internal PCBA performs signal analysis; Respiration Rate Software calculates rate. Outputs: digital values for SpO2, pulse, and respiration rate; plethysmographic waveform or 'blip bar' for pulse amplitude. Used in hospitals and during intra-hospital transport by clinicians. Provides continuous monitoring to assist in clinical assessment of respiratory status; aids in detecting desaturation patterns via OxiMax SPD Alert. Powered by AC or internal rechargeable lithium-ion battery.
Clinical Evidence
Clinical validation conducted on healthy, well-perfused adult volunteers with varying skin pigmentation. Compared SpO2 and pulse rate against CO-Oximeter reference measurements during motion and non-motion conditions. Respiration Rate Software (V 1.0) validated against predicate; accuracy demonstrated within ±1 breath per minute. Results confirm performance equivalence to predicate. No device-related adverse events observed.
Technological Characteristics
Class II pulse oximeter. Materials: molded plastic exterior. Energy: AC (80-264VAC) or 7.2V, 83W/hr rechargeable lithium-ion battery. Sensing: optical transducers. Connectivity: standalone. Software: Respiration Rate Software V 1.0. Standards: EN 60601-1, EN 60601-1-2, EN 60601-1-6, EN 60601-1-8, EN 80601-2-61.
Indications for Use
Indicated for adult, pediatric, and neonatal patients requiring continuous non-invasive monitoring of SpO2 and pulse rate; includes patients who are well or poorly perfused, under motion or no-motion conditions. Respiration rate monitoring indicated for adult patients who are well-perfused under no-motion conditions. OxiMax SPD Alert feature indicated for adult facility-use to detect desaturation patterns indicative of repetitive airflow reduction.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Bedside Respiratory Patient Monitoring System (K121806)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the word "COVIDIEN" in bold, black letters. To the left of the word is a black square with a white cross in the center. The cross is made up of four smaller squares, and there are also some other white shapes in the black square.
# 510(k) Summary
Date: 24-Jan-2014
# 510(k) Submitter/Holder
Covidien IIc 6135 Gunbarrel Avenue Boulder, CO 80301
# Contact
Elizabeth Malo Sr. Regulatory Affairs Specialist Office 303.305.2763 Mobile 720.539.9634 Fax: 303-305-2212 Email: elizabeth malo@covidien.com
# Name of Device
| Device Name: | Bedside Respiratory Patient Monitoring System |
|-------------------------|-----------------------------------------------|
| Catalog Numbers: | GR101704, GR101704-RR, 10099029 |
| Common Name: | Pulse oximeter |
| Classification: | Class II |
| Regulation: | 21 CFR § 870.2700 Oximeter |
| Primary Product Code: | DQA |
| Secondary Product Code: | BZQ |
# Purpose of Submission
Bedside Respiratory Patient Monitoring System has undergone an engineering design change associated with the replacement of the printed circuit board assembly that performs oximetry functions for the device. Futhermore, the indications for use have been expanded to specifically claim oximetry accuracy for "both no motion and motion conditions". The purpose of this submission is to demonstrate substantial equivalence of the Covidien Mellor Bedride Respiratory Patient Monitoring System to the predicate devices as related to these specific changes.
Summaries of clinical and non-clinical testing were provided to support 1) accuracy of oximetry performance, and 2) the expansion in the indications for use to include accuracy in the presence of motion.
# Predicate Devices
Bedside Respiratory Patient Monitoring System was compared and found to be substantially equivalent to the following products of comparable type in commercial distribution:
| Trade Name: | Bedside Respiratory Patient Monitoring System |
|---------------------|-----------------------------------------------------------|
| Device Common Name: | Oximeter, monitor, breathing frequency |
| 510(k) Number: | K121806 (cleared 9/2012) |
| Manufacturer: | Covidien, formerly Nellcor, a division of Tvco Healthcare |
| Trade Name: | Nellcor Puritan Bennett OxiMAX N-595 Pulse Oximeter |
| Device Common Name: | Monitor, breathing frequency |
| 510(k) Number: | K012891 (cleared 3/2002) |
| Manufacturer: | Covidien, formerly Nellcor, a division of Tyco Healthcare |
{1}------------------------------------------------
#### Device Description
The Covidien Nellcor Bedside Respiratory Patient Monitoring System is designed for continuous, non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate by use of a range monioning of functional onlygon transducers (sensors). The Bedside Respiratory Patient Monitoring System displays digital values of SpO2 and pulse and respiratory rate. Pulse amplitude is displayed by means of a "blip ber" presentation or plethysmographic waveform. The Bedside Respiratory Patient Monitoring System can be bur "proomation" on plannyamosphy operating on AC from a standard electrical utility receptacle (from 80VAC to 264VAC) or alternatively by an integral sealed 7.2V, 83W/hr rechargeable lithium-ion battery.
#### Intended Use
The Nellcor™ Bedside Respiratory Patient Monitoring System is a portable pulse oximeter intended for prescription use only as a continuous non-invasive monitor of arterial oxygen saturation (SpO2) and pulse rate of adult, pediatric, and neonatal patients during both no motion conditions, and for patients who are well or poorly perfused. The monitoring system is intended for use in hospital-type facilities, and during intrahospital transport. The OxiMax SPD™ Alert (SPD) feature is intended only for facility-use care of adults to detect natterns of desaturation indicative of repettive reductions in airflow through the upper airway and into the lungs.
The Respiration Rate parameter, when used in conjunction with the Nellcor™ Bedside Respiratory Patient Monitoring System and Nellcor™ Respiratory Sensor, is intended for the continuous, non-invasive monitoring of respiration rate in adult patients who are well perfused during no motion conditions, in hospital-type facilities.
# Technological and Performance Characteristics
The Bedside Respiratory Patient Monitoring System features the same performance characteristics and design, chemical composition, and energy source as the predicate K121806 (9/2012).
The external appearance and hardware of the subject device is exactly the same as that of the primary predicate, utilizing a similar molded plastic exterior case, display, and interface cables.
#### Design Change
The only significant difference between the subject and primary predicate device is the internal printed circuit board assembly (PCBA). The key element of the design change to the subject device is the difference in the analog hardware that was introduced through integration of the new PCBA. These hardware changes are specifically related to the Sample Rate/Interval and T-Mux frequency which were the main drivers for clinical validation of the design change to demonstrate that these differences produced equivalent results in the clinical setting. Results of performance testing has demonstrated that the performance of the Bedside Respiratory Patient Monitoring System has not changed due to the integration of the new PCBA and that the performance of the oximeter is equivalent to that of the primary predicate device.
There are no significant changes to the function or performance of the device other than the expansion in the indications for use to include accuracy in the presence of motion as based upon test methods cleared through K012891 (3/2002).
#### Usability / Human Factors
Usability of the Bedside Respiratory Patient Monitoring System was evaluated with users in simulated operating environments during original design and testing. These studies consisted of formative studies, which demonstrate the device provides adequate assurance of safety and performance (in regards to human factors/usability aspects) for the patient and operator. No features associated with device usability have changed through design change to the subject device and thus no additional usability testing was performed.
# Performance Data
Clinical and non-clinical tests were performed to support the determination of substantial equivalence. Clinical studies were conducted following regulations under Title 21 of the Code of Federal Regulations (21 CFR), Part 812 - Investigational Device Exemptions, Part 50 - Protection of Human Subjects and Part 56 Institutional Review Boards.
{2}------------------------------------------------
# Non-clinical Summary
Appropriate safety, environmental, performance and functional tests were conducted to ensure that the specifications of the Bedside Respiratory Patient Monitoring System were met.
Verification testing of the performance of the Respiration Rate parameter, when used in conjunction with the Nellcor™ Bedside Respiratory Patient Monitoring System and Nellcor™ Respiratory Sensor has been performed. Noneor - Docent data has been applied to closely mimic clinical free breathing studies on healthy volunteer subjects. Respiration rate accuracy was tested to demonstrate accuracy from 4-40 breaths per minute (BrPM) in the range of 40 -170 beats per minute (BPM) at a fixed O2 saturation of 97%.
Testing was conducted according to the FDA Guidance documents and consensus standards shown below:
- "Pulse Oximeters Premarket Notification Submissions [510(k)s]: Guidance for Industry and Food and Drug . Administration Staff" - July 19, 2007
- "Use of Standards in Substantial Equivalence Determinations" March 12, 2000 .
- "General principles of Software Validation, Final Guidance for Industry and FDA Staff" January 11, 2002 .
- International Conference on Harmonization of Technical Requirements for Registration of Pharmaceuticals for Human . Use: ICH Harmonized Tripartite Guideline: Guideline for Good Clinical Practice E6(R1)10 - June 1996
- EN 60601-1:2006- Medical Electrical Equipment Part 1: General Requirements for Safety and Essential Performance -.
- EN 60601-1-4:1999- Medical electrical equipment Part 1-4 : general requirements for safety Collateral standard : . programmable electrical medical systems
- EN 60601-1-6:2010- Medical electrical equipment Part 1-6: General requirements for safety Collateral standard. . Usability
- EN 60601-1-8:2006- Medical electrical equipment Part 1-8: General requirements for basic safety and essential . EN 0000 F 7 0.2000 Thoursel equirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems -
- EN 60601-1-9- Medical electrical equipment Part 1-9: General requirements for basic safety and essential performance . - Collateral Standard: requirements for environmentally conscious design
- EN 80601-2-61:2011- Medical electrical equipment Part 2-61: Particular requirements for basic safety and essential . performance of pulse oximeter equipment
- IEC 60601-1-2: 2007- Medical electrical equipment Part 1-2: General requirements for basic safety and essential . performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
# Clinical Summary
Clinical testing has been performed on healthy, well perfused adults ranging in skin pigmentation from light to dark, and has demonstrated that the Bedside Respiratory Patient Monitoring System meets the Accuracy Root Mean Square (ARMS), in the 70%-100% SpO2 range acceptance criteria for both SpO2 and Pulse Rate in comparison to reference-standard measurements of blood SaO2 by a CO-Oximeter during motion and non-motion conditions. Testing has also demonstrated accuracy of SpO2 and Pulse Rate in low saturation conditions and equivalent performance of the Respiration Rate V 1.0 software to that of the cleared predicate.
Clinical validation studies on healthy adult volunteers were conducted to assess the accuracy of the Respiration Rate Software residing on the oximeter. The studies have demonstrated that the accuracy of the Respiration Rate Software algorithm is equivalent to that of the predicate. The studies demonstrated the Respiration Rate Software algorithm calculates respiration rate within a stated accuracy of ±1 breath per minute, and is shown to be substantially equivalent to the predicate device with respect to Respiration Rate (for adults).
No device related adverse event were observed during the clinical studies.
# Substantial Equivalence
In establishing substantial equivalence of the Bedside Respiratory Patient Monitoring System to the predicate devices, Covidien has evaluated the intended use, indications for use, technological characteristics, and elements of risk analysis and risk management. The use of the Bedside Respiratory Patient Monitoring System in patient monitoring environments does not raise any new questions of safety and effectiveness when compared with the predicate devices currently in use.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing wings or feathers. The symbol is enclosed within a circular border, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the border of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 04, 2014
Covidien, LLC Ms. Elizabeth Malo Senior Regulatory Affairs Specialist 6135 Gunbarrel Avenue Boulder, CO 80301
Re: K130320
Trade/Device Name: Bedside Respiratory Patient Monitoring System Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA, BZQ Dated: November 6, 2013 Received: November 7, 2013
Dear Ms. Malo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Malo
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Image /page/4/Picture/8 description: The image shows a logo or seal with text and graphic elements. The text includes "Tejashri Purohit-Sheth, M.D." and "Clinical Deputy Director", indicating a medical professional's name and title. The acronym "DAGRID" is also present, possibly representing an organization or department. The word "FOR" is visible at the bottom right, suggesting the purpose or recipient of the document or seal.
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use Statement
510(k) Number (if known): KI30320
Covidien Nellcor™ Bedside Respiratory Patient Monitoring System and the Covidien Device Name: Nellcor™ Bedside Respiratory Patient Monitoring System Respiration Rate Software
Indications for Use:
For Covidien Nellcor Bedside Respiratory Patient Monitoring System with Respiration Rate Software:
The Nellcor™ Bedside Respiratory Patient Monitoring System is a portable pulse oximeter intended for prescription use only as a continuous non-invasive monitor of arterial oxygen saturation (SpO2) and pulse rate of adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused. The monitoring system is intended for use in hospitals, hospitaltype facilities, and during intra-hosport. The OxiMax SPD™ Alert (SPD) feature is intended only for facility-use care of adults to detect patterns of desaturation indicative of repetitive reductions in airflow through the upper airway and into the lungs.
The Respiration Rate parameter, when used in conjunction with the Nellcor™ Bedside Respiratory Patient Monitoring System and Nellcor™ Respiratory Sensor, is intended for the continuous, non-invasive monitoring of respiration rate in adult patients who are well perfused during no motion conditions, in hospitals and hospital-type facilities.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
—
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
K130320
Navan J
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.