The MD300M122/MD300M222/MD300M322 handheld pulse oximeter is intended for continuous monitoring, spot-checking of functional pulse oxygen saturation (SpO2) and pulse rate (PR) of single adult and pediatric patients in hospitals and home care.
Device Story
Handheld pulse oximeter (models MD300M122, MD300M222, MD300M322) measures SpO2 and pulse rate via external sensor applied to pulsating arteriolar vascular bed. Sensor utilizes dual light source (660nm red, 940nm infrared) and photodetector. Device processes signals to calculate SpO2 and PR; displays values, waveforms, battery status, and alarm states. Features include audible/visual alarms (three-level priority), data storage (72 hours), and transmission via Bluetooth or USB. Used in hospitals and home care by clinicians or patients. Output assists in patient monitoring and clinical decision-making regarding oxygenation status. Powered by AA or rechargeable batteries; includes AC/DC adapter.
Clinical Evidence
Clinical testing conducted at Beijing Military General Hospital per ISO 9919:2005, Annex EE.2. Accuracy of SpO2 and PR measurements compared against 'Golden Standard' Co-Oximeter. Results demonstrated compliance with manufacturer specifications.
Technological Characteristics
Handheld form factor; dual-wavelength (660nm/940nm) optical sensing; Bluetooth and USB connectivity; powered by 3 AA or rechargeable batteries; AC/DC adapter (100-240V input, 5V output). Standards: IEC 60601-1, IEC 60601-1-2, ISO 9919:2005, FCC Part 15 Subpart C.
Indications for Use
Indicated for continuous monitoring or spot-checking of functional SpO2 and pulse rate in adult and pediatric patients in hospital and home care settings. Prescription use only.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{1}------------------------------------------------
Additional Information K111494
## Exhibit I 510(K) Summary
As required by 21 CFR 807.92 The Assigned 510(K) Number is: K111494
1. Date Prepared: December 16, 2011
## 2. Sponsor Information
Beijing Choice Electronic Technology Co., Ltd. Bailangyuan Building B 1127-1128, Fuxing Road, A36 Beijing 100039, China
Mr. Lei Chen North Building 3F, No.9 Shuangyuan Road, Bdachu Hi-tech Zone, Shijingshan District Beijing China 100041 Phone:+86-10-88790480-6018 Email:cc@choicemmed.com
## 3. Submission Correspondent
Ms. Sunny Wang Beijing Choice Electronic Technology Co., Ltd. Floor 4, Jingyang Building No. 15, Xijing Rd., Shijingshan District, Beijing 100041, P.R. China Phone: +86-10-88204630-8176 Fax: +86-10-88204632 Email: wangsuqin@choicemmed.com & cc@choicemmed.com
## 4. US Agent
Wei Huang Mid-Link International INC 307 Eagle Heights Apt E, Madison, Wisconsin, 53705, United States Phone: 608-3355979 Fax: 760-4665084 Email:info@mid-link.net
{2}------------------------------------------------
## 5. Proposed Device Information
Device Common or Usual Name: Pulse Oximeter Device Trade or Proprietary Name: Handheld Pulse Oximeter Model: MD300M122, MD300M222, MD300M322 Classification Name: Oximeter Regulation Number: 21 CFR 870.2700 Product Code: DQA Panel: Anesthesiology
## 6. Predicate Device
Fingertip Pulse Oximeter MD300C318 (K092620) Beijing Choice Electronic Technology Co., Ltd. Room 1127-1128 Building B, Bailangyuan Fuxing Road, No.A36 Beijing, CHINA 10039
## 7. Device Description
#### ● MD300M122 Handheld Pulse Oximeter
The applicant device MD300M122 Handheld Pulse Oximeter is integrated with Bluetooth® technology allowing the user to transfer measurement data any time and anywhere. The oximeter is designed with the measurement, storage, review, visible and audible alarms and data transmission function.
The MD300M122 Handheld Pulse Oximeter works by applying a external sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660nm, which is red light; the other is 940nm, which is infrared light.
The applicant device of MD300M122 Handheld Pulse Oximeter is the handheld equipment, which mainly function are measurement, display, alarm, data storage & replay etc.
## Measurement function
That's the mainly function of the device, which use the method described above to measure the SpO2 value and Pulse Rate value of user.
## Display function
The display function of the device display the SpO2&PR value, waveform synchronously, battery capacity indicator, alarm state indication and time.
{3}------------------------------------------------
## Additional Information K111494
## Alarm function
The applicant device has physiological alarm function as SpO2 and PR parameter and technology alarm function as Low Battery Voltage, probe off and Finger out.
Each alarm of device contains VISUAL and AUDIBLE alarm. And you can turn on or off the alarm and set the alarm limits.
There are three-level alarm priorities in Oximeter. All the three priorities divided by built-in module and can't be changed by user.()High priority: "Di- Di ----- Di - Di" indicates the patient is in the very dangerous situation. @Medium priority: "Di - Di" indicates the warning-attention should be paid.(3)Low priority: "Di" indicates the technical alarm caused by the device itself.
## Data storage, replay & transmission function
The measured record is stored automatically every four seconds. The monitor can store 72 hours records.
The device includes the function as "record review" on the Oximeter equipment. And you can review "Data Browse", "SpO2 Trend", "PR Trend" and "Data Clear" by this function. The data stored in the device can be transmitted through Bluetooth or USB cable.
### Power
The applicant device uses three AA alkaline batteries or rechargeable batteries for power supply. The device has charge function by AC/DC power adapter, the specification of which is: 100-240 (VAC), 47/63 (Hz) input; 5Vd.c. output.
The device can measure normally during the process of charging, but we won't recommend to do so.
The applicant device is not for life-supporting or life-sustaining, not for implant. The device is not sterile and does not need sterilization or re-sterilization. The device is for prescription. The device does not contain drug or biological product.
The device is software-driven and the software validation is provided in Section 10 Software.
#### . MD300M222 Handheld Pulse Oximeter
MD300M222 is the same as MD300M122 except the appearance.
#### ● MD300M222 Handheld Pulse Oximeter
MD300M322 is the same as MD300M122 except the appearance.
## 8. Intended use
{4}------------------------------------------------
## Additional Information K111494
The MD300M122/MD300M222/MD300M322 handheld pulse oximeter is intended for continuous monitoring, spot-checking of functional pulse oxygen saturation (SpO2) and pulse rate (PR) of single adult and pediatric patients in hospitals and home care.
## 9. Substantial Equivalence
The applicant devices MD300M222/MD300M222/MD300M322 Handheld Pulse Oximeter have the same classification, same intended use, same design principle, similar specifications and same safety performance as the predicate device. These are no obvious differences to influence the effectiveness and safety of the device.
The proposed device is Substantially Equivalent (SE) to the predicate device which is US legally market device. Therefore, the subject device is determined as safe and effectiveness.
## 10. Test Conclusion
Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The proposed device was performed the tests according to the following standards:
IEC 60601-1:1988+A1:1991+A2:1995, Medical Electrical Equipment - Partl: General requirements for safety.
IEC 60601-1-2:2007, Medical Electrical Equipment - Part 1: General requirements for safety-2, Collateral Standard: Electromagnetic compatibility - Requirements and tests.
## FCC PART 15, SUBPART C: 2008 (Senction 15.249)
The Clinical Tests following ISO 9919:2005, Annex EE.2 are conducted in the lab of Beijing Military General Hospital. The study protocol is subjected to ISO 9919:2005 Annex EE. Procedures of testing required in EE.2 are adopted. It can be determined from the result of the study that the accuracy of the proposed device is compliance to the specification claimed by the manufacturer compared with "Golden Standard" Co-Oximeter.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Sunny Wang Correspondent Beijing Choice Electronic Technology Company, Limited Floor 4. Jungyang Bldg, No.15 Xiiing Road Shijingshan District, Beijing CHINA 100041
JAN 1 8 2012
Re: K111494
Trade/Device Name: Handheld Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: December 16, 2011 Received: December 19, 2011
## Dear Ms. Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm 1 1 5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Anthony D. Watson, B.S., M.S., M.B.A.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
## Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): _pending_
Device Name: Handheld Pulse Oximeter_
Indications for Use:
The MD300M122/MD300M222/MD300M322 handheld pulse oximeter is intended The MD300M122/MD300MZ22/MD300M222 fitunctional pulse oxygen saturation
for continuous monitoring, spot-checking of functional pulse oxygens in hospitals and for continuous monitoring, spot-checking of functions. Partin and provins and home care.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(Please do not write below this line-continue on another page IF NEEDED)
Moll J. Ward br Lester Schultheis
(Division Sign-Off) Division of Anesthesiology, General Hospital nfection Control, Dental Devices
510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.