The MRI Patient Monitor (Model 865353) is intended to monitor vital signs for patients undergoing MRI procedures. The MRI Patient Monitor (Model 865353) is intended for use by healthcare professionals.
Device Story
MRI Patient Monitor (Model 865353) provides continuous SpO2, pulse rate, and perfusion index monitoring for patients in MRI environments. System comprises a battery-operated display unit and a wireless SpO2 module with fiber optic sensors. Wireless SpO2 module collects physiological data via fiber optic sensor and transmits data to display unit via 2.4 GHz telemetry. Display unit features 5.7-inch touch screen for visualization of vital signs and alarm management. Device used by healthcare professionals in clinical settings; provides audible, visual, and text alarms for patient status changes. System is MR Conditional per ASTM F2503. Benefits include real-time patient monitoring during MRI procedures without interference or safety risks associated with standard monitoring equipment in high-field magnetic environments.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including verification of performance specifications, MR conditions of use (3.0T and 1.5T), environmental testing, and wireless technology evaluation. Compliance with IEC 60601-1, 60601-1-2, 60601-1-4, 60601-1-6, 60601-1-8, ISO 9919, and ASTM F2503/F2052 demonstrated.
Indicated for monitoring vital signs (SpO2, pulse rate, perfusion index) in patients undergoing MRI procedures. Intended for use by healthcare professionals.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
MRI Patient Monitoring System (Model 865214) (K090785)
{0}------------------------------------------------
# K103700
## MAR 1 5 2011
## 510(K) SUMMARY
#### 2.1 SUBMITTER INFORMATION
vivo"
(1) Ir
| Establishment / Sponsor Name:<br>Establishment / Sponsor Address: | Invivo Corporation<br>12151 Research Parkway<br>Orlando, FL 32826 USA |
|-------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| Manufacturer Name: | Sanmina- SCI Systems Singapore<br>PTE.LTD. |
| Manufacturer Address: | 2 Chai Chee Dr<br>Singapore, SG-NOTA 469044, Singapore |
| Company Phone: | (407) 275-3220 |
| Company Fax: | (407) 249-2022 |
| Person to contact<br>regarding questions: | Rusty Kelly<br>Quality & Regulatory Manager, Invivo Corporation<br>(407) 455-6166<br>Rusty.Kelly@philips.com |
| Establishment<br>Registration Number: | 1051786 (Sponsor)<br>3006606827 (Manufacturer) |
| Date Summary Prepared: | December 17, 2010 |
## 2.2 DEVICE IDENTIFICATION
| Trade name: | Essential MRI Patient Monitor (Model 865353) |
|----------------------|----------------------------------------------|
| Common name: | MRI patient monitor |
| Classification name: | Oximeter (21 CFR 870.2700, Product Code DQA) |
{1}------------------------------------------------
#### 2.3 IDENTIFICATION OF LEGALLY MARKETED CLEARED DEVICE
The MRI Patient Monitor (Model 865353) is substantially equivalent to the following cleared device:
| Cleared Device | Manufacturer | 510(k) No. | Clearance Date |
|-------------------------------------------------|--------------------|------------|----------------|
| MRI Patient Monitoring System<br>(Model 865214) | Invivo Corporation | K090785 | Dec 15, 2009 |
#### 2.4 MODIFIED DEVICE DESCRIPTION
The MRI Patient Monitor (Model 865353) is substantially equivalent to the cleared device.
Invivo has marketed a stand-alone pulse oximeter for the last 20 years as cleared to market in K864730 (Invivo Pulse Oximeter, Model 4500). This technology is over 20 years old and has not changed in design or fundamental technology since cleared to market. Invivo has identified the need to market an improved standalone pulse oximeter as a replacement for this device.
The modified device is predicated from the MRI Patient Monitoring System (Model 865214) which received clearance to market under 510(k) K090785 on December 15, 2009. The cleared device consists of a processing unit, display, and wireless SpO2 module which is directly connected to the patient by a fiber optic sensor, clip, and grip. All oximetry data collection and processing is completed by the wireless SpO2 module and fiber optic sensor. The module transmits the patient data to the processing unit using telemetry, and the processing unit transmits the data to the display using telemetry. The modified device incorporates the same wireless SpO2 module and fiber optic sensor that is currently used with the cleared device. In the modified device, the module transmits the data directly to the display using telemetry.
The following modifications have been incorporated to create the modified device:
- . The display was reduced in size and weight.
- The operator interface is a touch screen (instead of a rotary knob and � keypad).
- . SpO2, pulse rate (derived from SpO2), and perfusion index parameters only are measured and displayed. The following parameters are no longer measured and displayed: ECG, non-invasive blood pressure, invasive blood pressure, respiration, temperature, end-tidal CO2, and anesthetic agents concentrations.
{2}------------------------------------------------
- . The device does not provide signals for synchronization for the MRI scanner ("gating"). "Gating", or lack thereof, has no effect on patient monitoring ability or quality.
- . The display is battery-operated only. The display does not offer connections to AC mains.
- The radio printed circuit board layout was reduced in size to accommodate . the smaller display size.
- The display radio antenna was reduced in size to accommodate the smaller . display size and the antenna was concealed in the device handle.
- The module used with the device also communicates with the existing . cleared MRI Patient Monitoring Systems, Models 865214 and 3160.
- . The operating platform was modified to the MicroBlaze soft-core microprocessor running on the Spartan 6LX45 FPGA. The operational system was modified to the uC ("micro C") from company Micrium.
- . The display was modified to provide unlatched alarms only. Latched alarms are no longer provided.
- The intelligent battery charger is located in the battery (instead of in the . device).
The modified device has the same performance specifications (SpO2 operating range and accuracy) and uses the same patient applied parts (wireless SpO2 module, fiber optic sensor, clips, and grips) as the cleared device.
The MRI Patient Monitor consists of the following primary components:
- Display ("Vital Signs Viewer") .
- Wireless SpO2 Module .
- SpO2 Sensor .
- SpO2 Grips and Clips .
The display and SpO2 module enclose the transceiver and antennas that support bi-directional 2.4 GHz wireless communication. The modified device's operator interface consists of a color 5.7-inch LCD display and touch screen. The modified device's display provides the same vital sign information regarding SpO2, generates the same types of alarms (audible, lights, and text), and controls the same functions (alarm limits, patient type, display setup, and network selection) as the cleared device.
#### 2.5 INTENDED USE
The intended use of the modified device as described in its labeling has not changed from that of the cleared device as a result of the modification.
The MRI Patient Monitor (Model 865353) is intended to monitor vital signs for patients undergoing MRI procedures. The MRI Patient Monitor (Model 865353) is intended for use by healthcare professionals.
{3}------------------------------------------------
#### 2.6 SUBSTANTIAL EQUIVALENCE COMPARISON
The modified device, the MRI Patient Monitor (Model 865353), and the cleared device, the MRI Patient Monitoring System (Model 865214) which received clearance to market under 510(k) K090785 on December 15, 2009, are identical with respect to indications for use, intended use, and fundamental scientific technology. Both devices are multi-parameter patient monitors intended to monitor vital signs for patients undergoing MRI procedures. Details of all modifications are listed in the following table:
{4}------------------------------------------------
| Table 2-1: Substantial Equivalence Comparison | | |
|-----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | Cleared Device<br>MRI Patient Monitoring System<br>(Model 865214) | Modified Device<br>MRI Patient Monitor<br>(Model 865353) |
| 510(k) Number | K090785, Cleared on December 15, 2009 | Pending |
| General Safety | Compliant to IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-1-6, and IEC 60601-1-8<br>Defibrillator protection up to 5kV | IDENTICAL TO CLEARED DEVICE |
| Power<br>Requirements | Battery Type: Lithium-ion<br>Battery Operation Time: At least 8 hours<br>Also uses AC mains to AC-DC power converter | Battery Type: IDENTICAL TO CLEARED DEVICE<br>Battery Operation Time: IDENTICAL TO CLEARED DEVICE<br>No AC power available |
| Environmental | Operating Temperature: 10-44°C<br>Relative Humidity: 0-80% non-condensing<br>Storage Temperature: -40 – 75°C | IDENTICAL TO CLEARED DEVICE |
| Display | Type: Color LCD with 800 x 600 resolution<br>Size: 12.25-inch diameter<br>Weight: 17 lbs (7.7 Kg)<br>Dimensions:<br>Height: 11.5 inches (29.2 cm)<br>Width: 15.4 inches (39.1 cm)<br>Depth: 5.5 inches (14.0 cm) | Type: Color LCD with 640 x 480 resolution<br>Size: 5.7-inch diameter<br>Weight: 3.3 lbs (1.5 Kg)<br>Dimensions:<br>Height: 6.1 inches (15.5 cm)<br>Width: 6.9 inches (17.5 cm)<br>Depth: 3.7 inches (9.4 cm) |
| Operator<br>Interface | Rotary knob and keypad | Touch screen<br>The layout of menus and patient data is IDENTICAL TO<br>CLEARED DEVICE |
| Mounting<br>Configurations | Traditional roll around cart<br>Mounting options able to be mounted permanently to<br>various surfaces | IDENTICAL TO CLEARED DEVICE |
| | Cleared Device<br>MRI Patient Monitoring System<br>(Model 865214) | Modified Device<br>MRI Patient Monitor<br>(Model 865353) |
| | RF Output Power: +20 dBm<br>Frequency Range: 2.4GHz band | The functionality, technology, and operating performance of the<br>wireless communication is IDENTICAL TO CLEARED DEVICE. |
| | | The wireless SpO2 module including its radio is IDENTICAL TO<br>CLEARED DEVICE. |
| Wireless<br>Communication | Radios have FCC approval under the following<br>identification numbers:<br><br>SpO2 Module: S6W2GMOD Processing Unit: S6WDR3160BAS Display: HSW-2410NF | The cleared device radio in the Processing Unit was modified to<br>create the radio in the display. The only modifications are as<br>follows:<br>The radio printed circuit board layout was reduced in size to<br>accommodate the smaller device size. The radio antenna was reduced in size to accommodate the<br>smaller device size and is concealed in the device handle. For these reasons, a new FCC evaluation was required for the radio<br>in the display. |
| Vital Signs<br>Monitored | SpO2, ECG, NIBP, pulse rate (derived from SpO2, ECG,<br>or NIBP), perfusion index, IBP, ETCO2, respiration,<br>agents, and temperature | Radios have FCC approval under the following identification<br>numbers:<br>SpO2 Module: IDENTICAL TO CLEARED DEVICE Display: S6WESSENTIAL SpO2, pulse rate (derived from SpO2), and perfusion index |
| Pulse Oximetry<br>(SpO2) | Saturation Range: 1-100%<br>Saturation Accuracy: ±3% at 70-100%<br>Pulse Rate Range: 30-250 BPM<br>Pulse Rate Accuracy: ±2% or 1 BPM, whichever is greater | IDENTICAL TO CLEARED DEVICE |
| Patient Applied<br>Parts | SpO2 Fiber optic sensor, SpO2 fiber optic grips and clips | IDENTICAL TO CLEARED DEVICE |
| | Cleared Device<br>MRI Patient Monitoring System<br>(Model 865214) | Modified Device<br>MRI Patient Monitor<br>(Model 865353) |
| | Device is MR Conditional according to ASTM F2503 | Device is MR Conditional according to ASTM F2503 |
| Labeling | | Labeling was modified as follows:<br>Added photographs and specifications showing dimensional modifications to display Cited use of the touch screen display Removed references to ECG, non-invasive blood pressure, invasive blood pressure, temperature, respiration, end-tidal CO2, and anesthetic agents monitoring, and gating which were present on the cleared device Cited that device is battery operated only Added instructions for battery charging Added description of communication between the modified device and existing cleared devices, MRI Patient Monitoring System Models 865214 and 3160 |
| Battery<br>Charger | Integral to patient monitor and separate charger | Integral to battery |
| Operating<br>Platform | STPC 486 Processor | MicroBlaze soft-core microprocessor running on the Spartan 6LX45 FPGA |
| Operating<br>System | Multitask | uC ("micro C") |
| Alarms | Provides latched and unlatched alarms.<br>Alarm system compliant to IEC 60601-1-8. | Provides unlatched alarms only.<br>Alarm system compliant to IEC 60601-1-8. |
r
:
Page 2-5
۔
{5}------------------------------------------------
:
Page 2-6
{6}------------------------------------------------
.
ﺑ
Page 2-7
.
.
{7}------------------------------------------------
#### 2.7 SUMMARY OF NON-CLINICAL PERFORMANCE DATA
The performance data included in this notification establishes substantial equivalence of the modified device, the MRI Patient Monitor (Model 865353), to the cleared device which received market clearance in 510(k) K090785 on December 15, 2009. The modified device was evaluated to the following safety and performance tests:
- FDA Guidance Documents .
- . Voluntary standards
- Verification and validation of performance specifications .
- . Verification and validation of MR conditions of use
- . Environmental testing
- . Evaluation of wireless technology
In all testing, the device was verified using a worst-case environment.
Verification and validation of SpO2 accuracy in the modified device is not required because the SpO2 module, technology, patient-applied parts, algorithm, and software is the same as the cleared device.
## FDA Guidance Documents
The modified device, MRI Patient Monitor (Model 865353), was designed and evaluated in accordance with the following FDA Guidance Documents:
- · Use of Standards in Substantial Equivalence Determination (Issued March 12, 2000)
- Pulse Oximeters Premarket Notification Submission (Issued July 19, . 2007)
- · Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (Issued May 11, 2005)
- · Radio-Frequency Wireless Technology in Medical Devices (Issued January 3, 2007)
## Voluntary Standards
Standards Data Reports (Form FDA 3654 (09/07)) are provided in Section 3 of this notification. The MRI Patient Monitor (Model 865353) was evaluated to the following voluntary standards where applicable:
- IEC 60601-1, Medical Electrical Equipment Part 1: General . Requirements for Safety
- IEC 60601-1-2, Medical Electrical Equipment Part 1-2: General . Requirements for Safety - Collateral Standard: Electromagnetic Compatibility -- Requirements and Tests
- IEC 60601-1-4. Medical electrical equipment Part 1-4: General . requirements for safety -- Collateral standard: Programmable electrical medical systems
- . IEC 60601-1-6, Medical electrical equipment - Part 1-4: General requirements for safety -- Collateral standard: Usability
{8}------------------------------------------------
- IEC 60601-1-8, Medical electrical equipment Part 1-8: General . requirements for safety - Collateral Standard: Alarm Systems -Requirements, tests and guidance - General requirements and guidelines for alarm systems in medical equipment
- IEC 60601-2-33, Medical electrical equipment Part 2-33: Particular ● requirements for the safety of magnetic resonance equipment for medical diagnosis
- ISO 9919, Medical electrical equipment Particular requirements for . the basic safety and essential performance of pulse oximeter equipment for medical use
- ISO 10993-1, Biological evaluation of medical devices -- Part 1: . Evaluation and testing within a risk management process
- ISO 14971, Medical devices Application of risk management to . medical devices
- . ASTM F2503-08, Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment
- ASTM F2052-06, Standard Test Method for Measurement of ● Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment
The modified device, MRI Patient Monitor (Model 865353), was evaluated by a third party laboratory to IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-1-6, IEC 60601-1-8, and ISO 9919 and complies to the requirements in these standards.
Compliance of the modified device to ISO 14971 is demonstrated by risk management summary provided in Section 6.3.
Compliance of the modified device to ISO 10993-1 is demonstrated through risk management analysis and evaluation of the modified device against a history of cleared devices provided in Section 6.12.
- Compliance of the modified device to IEC 60601-2-33, ASTM F2503-08, and ASTM F2052-06 was demonstrated through validation testing performed by Invivo Corporation in the MR environment. The modified device complies to the applicable requirements of these standards.
A Standards Summary Report Table noting deviations, adaptations, or options used in demonstrating compliance of the modified device to the standards is provided in Section 6.10.
### Verification and Validation of Performance Specifications
All performance specifications of the modified device, MRI Patient Monitor (Model 865353), were defined by Invivo Corporation according to national standards, international standards, market needs, risk management, and intended
{9}------------------------------------------------
use. The verification and validation protocol for the specifications which are modified from the cleared device are provided in Section 6.4.
Results of the complete verification and validation indicate that the device operates as intended within the performance specifications. The results do not raise issues regarding the safety and effectiveness of the device and clinical data was not required to substantiate claims of safety and effectiveness.
### Verification and Validation of MR Conditions of Use
The MR conditions of use of the modified device, MRI Patient Monitor (Model 865353), were defined by Invivo Corporation according to national standards, international standards, intended use, risk management, and market needs. The modified device was evaluated for proton emissions, image artifact, magnetically induced displacement force, RF heating, Specific Absorption Rate, and Peripheral Nerve Stimulation (PNS) using 3.0T and 1.5T magnetic fields, simulators, and test equipment under the most extreme use conditions within the intended use. Details are provided in Sections 6.6, 6.7, and 6.8.
Test results demonstrate that the MRI Patient Monitor (Model 865353) met the MR conditions of use as defined in the modified device labeling. Test results are provided in Sections 6.6, 6.7, and 6.8.
### Environmental Testing
Environmental specifications for the modified device, MRI Patient Monitor (Model 865353) were defined by Invivo Corporation according to international standards, intended use, risk management, and market needs. Test results demonstrate conformity to customer requirement specifications over the device use life and ensure longevity of the modified device within the use model. Test data was not provided in this submission but is contained within the modified device's Design History File.
### Evaluation of Wireless Technology
The modified device display was evaluated to FCC Part 15 for Low Power Communication Device Transmitters. Test results are provided in Section 6.9.
The modified device incorporates the same wireless SpO2 module including radio that is currently used with the cleared device. Additional evaluation of the wireless SpO2 module was not required.
Integrity of the wireless communication between the modified device display and cleared wireless SpO2 module was validated to operate as intended. Results are provided in Section 6.9.
{10}------------------------------------------------
## Conclusion
The conclusion of all testing confirms that all identified risks have been mitigated, the device operates as designed and intended within the performance specifications, and the device meets the labeling claims.
The results do not raise issues regarding the safety and effectiveness of the device and clinical data was not required to substantiate claims of safety and effectiveness.
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circle around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Rusty Kelly Quality & Regulatory Manager Invivo Corporation 12150 Research Parkway Orlando, Florida 32826
MAR 1 5 2011
Re: K103700
Trade/Device Name: MRI Patient Monitor (Model 865353) Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DOA Dated: February 11, 2011 Received: February 14, 2011
Dear Mr. Kelly:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{12}------------------------------------------------
Page 2- Mr. Kelly
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{13}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name: MRI Patient Monitor (Model 865353)
Indications For Use: The MRI Patient Monitor (Model 865353) is intended to monitor vital signs for patients undergoing MRI procedures. The MRI Patient Monitor (Model 865353) is intended for use by healthcare professionals.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schuberth
(Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: _______________________________________________________________________________________________________________________________________________________________
Page 1 of _1 _________________________________________________________________________________________________________________________________________________________________
Page A-1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.