MASIMO RAINBOW SET PRONTO -7 PULSE CO- OXIMETER AND ACCESSORIES, MODEL PRONTO-7
Applicant
Masimo Corporation
Product Code
DQA · Cardiovascular
Decision Date
Jun 23, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Masimo Rainbow SET Pronto-7 Pulse CO-Oximeter and Accessories are indicated for noninvasive spot checking of functional saturation of arterial oxygen hemoglobin (SpO2), pulse rate, and total hemoglobin concentration (SpHb). The Masimo Rainbow SET Pronto-7 Pulse CO-Oximeter and Accessories are indicated for use by trained personnel, with adult and pediatric individuals, in clinical and non-clinical settings (e.g., hospital-type facilities, home, clinics, physician offices, blood donation facilities, and ambulatory surgery centers).
Device Story
The Masimo Rainbow SET Pronto-7 is a noninvasive pulse CO-oximeter for spot-checking SpO2, pulse rate, and total hemoglobin (SpHb). It utilizes optical sensors to measure light absorption, processing signals to derive physiological parameters. The device features a touch-screen interface, pleth waveform display, and wireless connectivity (Bluetooth/WiFi). It is intended for use by trained personnel in diverse settings including hospitals, clinics, and home environments. Output is displayed on the device screen, allowing clinicians to perform rapid assessments of patient oxygenation and hemoglobin status, facilitating immediate clinical decision-making and patient monitoring.
Clinical Evidence
Clinical validation performed per ISO 9919:2005. SpO2 accuracy validated in human blood studies on healthy adult volunteers (light to dark skin) during induced hypoxia (70-100% range) vs. laboratory CO-oximeter and ECG. SpHb accuracy validated in healthy adults and patients (6-18 g/dl range) vs. laboratory CO-oximeter. Pulse rate accuracy validated in healthy adults and patients (40-110 bpm). Results: SpO2 accuracy (rms) < 2%, pulse rate accuracy (rms) < 3 bpm, SpHb accuracy (rms) < 1 g/dl. No device-related adverse events reported.
Technological Characteristics
Noninvasive optical pulse CO-oximeter. Features: touch-screen, pleth waveform display, rechargeable lithium polymer battery, USB, Bluetooth 2.0, WiFi b/g, Micro SD slot. Standards: EN 60601-1-2 (EMC), IEC 60601-1/UL 60601-1 (Electrical Safety), Type BF-Applied Part, IPX1. Operates via spot-check mode. Connectivity: Wireless and wired data output.
Indications for Use
Indicated for noninvasive spot checking of SpO2, pulse rate, and total hemoglobin (SpHb) in adult and pediatric patients. For use by trained personnel in clinical and non-clinical settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Masimo Rainbow SET® Pronto Pulse CO-Oximeter and Accessories (K091057)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a handwritten text that appears to be a combination of letters and numbers. The text reads 'Klo 403' in a bold, slightly irregular font. The characters are closely spaced, and the overall impression is that of a quick, informal note.
## 510(k) SUMMARY
| Submitted by: | Masimo Corporation<br>40 Parker<br>Irvine, CA 92618<br>949-297-7000<br>FAX 949-297-7592 |
|------------------------------------|------------------------------------------------------------------------------------------|
| Company Contact: | Marguerite Thomlinson, Manager of Regulatory Affairs |
| Date Summary Prepared: | April 23, 2010 |
| Trade Name | Masimo Rainbow SET Pronto-7 Pulse CO-Oximeter and<br>Accessories |
| Common Name | Oximeter |
| Classification Name/ Product Code: | Oximeter Section 870.2700/ Product Code DQA |
| Predicate Device: | Masimo Rainbow SET® Pronto Pulse CO-Oximeter and<br>Accessories, 510(k) Number - K091057 |
#### Device Description
The Masimo Rainbow SET Pronto-7 Pulse CO-Oximeter and Accessories (Pronto-7) include the noninvasive technology of measurement for functional saturation of arterial oxygen hemoglobin (SpO2), pulse rate and total hemoglobin concentration (SpHb), which is based on the same technology of the predicate device, the Masimo Rainbow SET® Pronto Pulse CO-Oximeter and Accessories (Pronto), 510(k) no. K091057.
### Intended Use/ Indications for Use
The Masimo Rainbow SET Pronto-7 Pulse CO-Oximeter and Accessories are indicated for noninvasive spot checking of functional saturation of arterial oxygen hemoglobin (SpO3), pulse rate, and total hemoglobin concentration (SpHb). The Masimo Rainbow SET Pronto-7 Pulse CO-Oximeter and Accessories are indicated for use by trained personnel, with adult and pediatric individuals, in clinical and non-clinical settings (e.g., hospital-type facilities, home, clinics, physician offices, blood donation facilities, and ambulatory surgery centers).
### Technology Comparison
Features in the Pronto-7, which are not in the Pronto, include touch-screen user interface, pleth waveform display, rechargeable batteries or AC power, USB connection, flash memory, earphone jack, and wireless connection. The Pronto-7 has the following specifications:
| FEATURES | SPECIFICATIONS |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Display Range | Saturation (SpO2): 0-100%; Pulse Rate (bpm): 30-250 bpm<br>(Total Hemoglobin (SpHb): 0-25 g/dl; Perfusion Index (PI): 0.02-20% |
| Accuracy:<br>Adults and Pediatrics > 30kg | SpO2: 70-100±2%; Pulse Rate: 30-250±3 bpm; SpHb: 6-18 g/dl ±1 g/dl |
| Resolution | SpO2: 1%; Pulse Rate: 1 bpm; SpHb: 0.1 g/dl |
| AC Power | Voltage Input: 100-240 VAC, 50-60 Hz; Max Power Consumption: 15 VA |
| Batteries | Rechargeable lithium polymer |
{1}------------------------------------------------
# 510(k) SUMMARY
| FEATURES | SPECIFICATIONS |
|--------------------------------------|-------------------------------------------------------------------------|
| Temperature | Operating/ Storage: 41 to 104°F (5 to 40°C)/ -40 to 158°F (-40 to 70°C) |
| Relative Humidity | 5 to 95% noncondensing |
| Operating Altitude | 500 to 1,060 mbar pressure; -1,000 to 18,000 ft (-304 to 5,486m) |
| Alarm | System Failure; Low Battery Alarm |
| Display and Indicators | SpO2 (%); Pulse Rate (bpm); SpHb (g/dl); Perfusion Index (%) |
| Display and Indicators | Pleth Waveform; Sensor Status; Status Messages; Battery Status |
| Connection/ Output | Wireless: Bluetooth 2.0; WiFi b/g; Flash Memory: Micro SD card slot |
| Earphone jack | Connection for standard 3.5 mm earphone jack |
| EMC/ Electrical Safety<br>Compliance | EN 60601-1-2, Class B |
| EMC/ Electrical Safety<br>Compliance | IEC 60601-1, UL 60601-1, Internally Powered, AC Power Class 2 |
| EMC/ Electrical Safety<br>Compliance | Type BF-Applied Part, IPX1, Class 2 |
| Mode of Operation | Spot check |
### Clinical Data Summary
Clinical Studies: SpO2 accuracy has been validated in human blood studies on healthy adult male and female volunteers with light to dark skin pigmentation. Pronto-7 measurements on volunteers induced hypoxia in the range of 70-100% SpO2 are compared against a laboratory CO-oximeter and ECG monitor. SpHb accuracy has been validated with (arterial/ venous) blood from healthy adult male and female volunteers and on patients with light to dark skin pigmentation in the range of 6-18 g/dl SpHb, with Pronto-7 measurements compared against a laboratory CO-oximeter. Pulse rate accuracy has been validated on healthy adult male volunteers and on patients with light to dark skin pigmentation in the range of 40-110 bpm. The variation in these studies equals plus or minus one standard deviation encompasses 68% of the population.
#### Clinical Results: No device-related adverse events.
The clinical studies were performed in accordance with ISO 9919:2005. The studies resulted in SpO3 accuracy (rms) < 2%, pulse rate accuracy (rms) < 3bpm, and SpHb accuracy (rms) < 1g/d.
#### Non-Clinical Data Summary
The Pronto 7 complies with the voluntary standards as detailed in this submission. Laboratory testing for biocompatibility, safety and environmental was conducted to verify that the Pronto 7 met all design specifications and was substantially equivalent to the predicate device.
#### Conclusions
The information in this 510(k) submission demonstrates that the Masimo Rainbow SET Pronto 7 Pulse CO-Oximeter and Accessories are substantially equivalent to the predicate device, with respect to safety. effectiveness, and performance.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings and tail feathers. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA", which is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Marguerite Thomlinson Manager of Regulatory Affairs Masimo Corporation 40 Parker Irvine, California 92618
JUN 2 3 2010
Re: K100403
Trade/Device Name: Masimo Rainbow SET Pronto-7 Pulse CO-Oximeter and Accessories Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: June 11, 2010 Received: June 14, 2010
Dear Ms. Thomlinson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2- Ms. Thomlinson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Susan Rioner
Anthony D. Watson, BS, MS, MBA Director Division of Anesthesiology, General Hospital
Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure .
{4}------------------------------------------------
### Indications for Use
#### 510(k) Number (if known):
Masimo Rainbow SET Pronto-7 Pulse CO-Oximeter and Accessories Device Name:
#### Indications For Use:
The Masimo Rainbow SET Pronto-7 Pulse CO-Oximeter and Accessories are indicated for noninvasive spot checking of functional saturation of arterial oxygen hemoglobin (SpO₂), pulse rate, and total hemoglobin concentration (SpHb). The Masimo Rainbow SET Pronto-7 Pulse CO-Oximeter and Accessories are indicated for use by trained personnel, with adult and pediatric individuals, in clinical and non-clinical settings (e.g., hospital-type facilities, home, clinics, physician offices, blood donation facilities, and ambulatory surgery centers).
| Prescription Use _X_ | AND/OR | Over-The-Counter Use _ |
|--------------------------------|--------|--------------------------------|
| (Per 21 CFR 801.109 Subpart D) | | (Per 21 CFR 801.109 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
rrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K100403
510(k) Traditional, Masimo Pronto-7
Page 10 of 1254. 2/5/2010
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.