The OxiprobE is indicated for non-invasive, continuus, beat-by-beat monitoring of oxygen saturation of functional arterial hemoglobin, pulse rate and pulse amplitude. Individual models are labeled specifically for either adult or pediatric use. It is for prescription use only.
Device Story
OxiprobE consists of pulse oximetry sensors and cables; inputs include optical signals from patient tissue; device measures blood oxygen saturation, pulse rate, and pulse amplitude; disposable and reusable sensor configurations available; disposable sensors use adhesive for secure positioning; reusable sensors intended for short-term/intermittent monitoring of relatively immobile patients; device connects to compatible pulse oximetry monitors; healthcare providers use output for continuous patient monitoring; facilitates clinical decision-making regarding patient oxygenation status; benefits include non-invasive, real-time monitoring and reduced cross-contamination risk with single-use options.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and design similarities to predicate devices.
Technological Characteristics
Pulse oximetry sensor system consisting of connector, cable, and sensor. Available in disposable (adhesive) and reusable configurations. Sensing principle based on optical light source and photo detector for non-invasive measurement of arterial hemoglobin saturation. Designed for specific patient populations (adult/pediatric) based on model. Non-sterile.
Indications for Use
Indicated for non-invasive, continuous, beat-by-beat monitoring of arterial hemoglobin oxygen saturation, pulse rate, and pulse amplitude in adult and pediatric patients. Specific models are designated for adult use (BM-100, BM-200, BM-300, BM-300s) or pediatric use (BM-400, BM-600p). Pediatric models are weight-restricted (BM-600p for 10-50 kg). Prescription use only.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Model D-25, N-25, D-20, Nellcor Puritan Bennett Inc. (K993637)
Philips Reusable SpO2 Sensors-M1191T, M1192T, Philips Medical Systems (K063783)
Submission Summary (Full Text)
{0}------------------------------------------------
K092549
Image /page/0/Picture/1 description: The image shows a diagram of a molecule. The molecule is made up of three atoms, labeled as M, B, and M. The atoms are connected by lines, indicating that they are bonded together. The text below the diagram says 'Bimetallic homogeneous molecular catalyst'.
Medical Technologies Co., Ltd 1 Bldg. #431-17. Shindang-dong, Jung-gu,Seoul, 100-452. Korea cl.822.2234.0780~2 Fax: 822.2234.0748
## NOV 1 6 2009
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92(c).
Date: Aunistrator 6, 2009
1. Company and Correspondent making the submission:
| | Company |
|--------------|-----------------------------------------------------------------------------------|
| Name | Bio Medical Technologies Co., Ltd. |
| Address | Megu Bldg. #431-17. Shindang-dong, Jung-<br>gu, Seoul, 100-452, Republic of Korea |
| Phone<br>Fax | +82 2 2234-0780-2<br>+82 2 2234-0748 |
| Contact | K.N. Kim |
#### 2. Device:
Proprietary Name - OxiprobE Common Name - Oximetry Sensor Classification Name - Oximeter
#### 3. Predicate Device:
Model D-25, N-25, D-20, Nellcor Puritan Bennett Inc., K993637 Philips Reusable SpO2 Sensors-M1191T, M1192T, Philips Medical Systems, K063783
## 4. Classifications Names & Citations:
Regulation Number: 21 CFR 870.2700 Regulatory Class: Class II
Regulation Name: Oximeter Product Code: DQA
5. Description:
The Oxiproble is Pulse Oximetry sensors and cables used for non-invasive measurement of blood oxygen saturation and pulse rute measurement. It consists of Connector, Cable and sensor. Bio Medical Technologies Co., Ltd. supplies both disposable and reusable sensors. Disposable sensors are designed to enable the sensor's light source and photo detector to be securely and properly positioned on the patient. The adhesive stabilizes
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo with three circles connected by lines. The top circle contains the letters 'm' and 'm', the left circle contains the letters 'b' and 'b', and the right circle contains the letters 't' and 't'. Below the circles, the text 'An Medical Technologies' is written, followed by 'www.am-tech.co.kr'.
### Bio Medical Technologies Co.,Ltd legu Bldg. #431-17. Shindang-dong, Jung-gu,Seoul, 100-452, Korea Tel.822.2234.0780~2 Fax.822.2234.0748
these important optical components and provides a comfortable fit. Adhesive sensors are patient-dedicated and can travel with your patients. Single-use sensors do not present the risk of cross-contamination caused by products that are reused from patient up patient. Reusable sensors are designed to monitor various sizes of relatively immobile patients. When short-term or intermittent monitoring is necessary, these reusable, nonsterile sensors are an effective monitoring alternative.
The Oxiproble models BM-100, BM-200 and BM-300 are indicated for adult patients. The Oxiproble model BM-400 is indicated for pediatric patients.
The Oxiproble models BM-300s is indicated for adult patients over 30 kg. The OxiproblE models BM -600p is indicated for pediatric patients over 10 kg and less than 50 kg.
- ర్. Indication for use:
The OxiprobE is indicated for non-invasive, continuus, beat-by-beat monitoring of oxygen saturation of functional arterial hemoglobin, pulse rate and pulse amplitude. Individual models are labeled specifically for either adult or pediatric use. It is for prescription use only.
#### 7. Review:
The OXIPROBE has the same device characteristics as the predicate device, the intended use, materials, design and use concept are similar.
Based on the comparison of intended use and technical features, the OXIPROBE is substantially equivalent to the predicate device.
## 8. Conclusions:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification Bio Medical Technologics Co., Ltd. concludes that the OXIPROBE are safe and effective and substantially equivalent to predicate devices as described herein.
- 9. Bio Medical Technologies Co., Ltd. will update and include in this summary any other information deemed reasonably necessary by the FDA.
END
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird, with its wings forming a flowing, abstract design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
Bio Medical Technologies Company, Limited C/O Ms. Cathryn Cambria Arkin Consulting Group, L.L.C. 5536 Trowbridge Drive Dunwoody, Georgia 30338
NOV 1 6 2009
Re: K092549
Trade/Device Name: OxiprobE Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II . Product Code: DQA Dated: August 6. 2009 Received: August 19, 2009
Dear Ms. Cambria:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2- Ms. Cambria
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
for
Susan Runner, D.D.S., M.A. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
# Indications for Use
510(k) Number K
Device Name: OxiprobE
Indication for use:
The OxiprobE is indicated for non-invasive, continuus, beat-by-beat monitoring of oxygen saturation of functional arterial hemoglobin, pulse rate and pulse amplitude. Individual models are labeled specifically for either adult or pediatric use. It is for prescription use only.
Prescription Use N ાર (Per 21CFR801 Subpart D)
Over-The-Counter Use (Per 21CFR807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schultheis
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: k 092549
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.