SOLARIS COMPATIBLE REUSABLE ADULT SPO2 FINGER SENSORS, MODELS S100A-090103 AND 300103, S200A-090101 AND 300101
K061931 · Solaris Medical Technology, Inc. · DQA · Jan 31, 2007 · Cardiovascular
Device Facts
Record ID
K061931
Device Name
SOLARIS COMPATIBLE REUSABLE ADULT SPO2 FINGER SENSORS, MODELS S100A-090103 AND 300103, S200A-090101 AND 300101
Applicant
Solaris Medical Technology, Inc.
Product Code
DQA · Cardiovascular
Decision Date
Jan 31, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Indications for Use
Solaris Sensors are indicated for use in continuous, non-invasive monitoring of arterial oxygen saturation and pulse rate for patients weighing more than 40 kg.
Device Story
Solaris Compatible Reusable Adult SpO2 Finger Sensors are electro-optical sensors for use with patient monitors/oximeters. Sensors utilize two LEDs (light sources) and one photodiode (detector) housed in silicon rubber pads. Device operates by non-invasively illuminating arterially perfused tissue with two wavelengths of light; photodetector measures time-varying absorbance to determine arterial oxygen saturation and pulse rate. Sensors connect to external patient monitors which perform signal processing and display results to clinicians. Used in clinical settings for continuous monitoring; provides real-time data to assist healthcare providers in assessing patient oxygenation status. Benefits include non-invasive, continuous monitoring without skin penetration, electrical contact, or heat transfer.
Clinical Evidence
Clinical testing conducted under institutional protocol with informed consent. Accuracy validated against co-oximetry in controlled hypoxia. Results confirm sensors meet published accuracy specifications for SpO2 range 70% - 100%. Biocompatibility testing per ISO 10993-1:2003 confirmed materials are non-toxic, non-sensitizing, and non-irritating.
Technological Characteristics
Electro-optical sensor; two LEDs and one photodiode in silicon rubber pads. No skin penetration, electrical contact, or heat transfer. Biocompatibility per ISO 10993-1:2003. Electrical safety per IEC 60601-1:1988 (A1:1991, A2:1995). EMC per IEC 60601-1-2:2001. Reusable form factor.
Indications for Use
Indicated for continuous, non-invasive monitoring of arterial oxygen saturation and pulse rate in patients weighing >40 kg.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
## K061931
Image /page/0/Picture/1 description: The image shows the logo for Solaris Medical Technology, Inc. The logo features a stylized mountain range to the left of the word "SOLARIS" in bold, uppercase letters. Below the logo is the text "Solaris Medical Technology, Inc." in a smaller font.
JAN 3 1 2007
## 510(k) Summary
| Submission Date: | 23 June 2006 |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Solaris Medical Technology, Inc.<br>427 26th Avenue, Suite 101<br>San Francisco, CA 94121 USA |
| Submitter Contact: | Rachel Cheng, Director, Regulatory Affairs<br>415 221 2350<br>415 221 2360 (fax)<br>rcheng@solarismedtech.com |
| Manufacturing Site: | Solaris Medical Technology, Inc.<br>6/F Building 9, 30th District, Keji Road<br>Science Park, Nanshan District<br>Shenzhen, Guangdong, 518057 CHINA |
| Official Contact: | Thomas Kroenke<br>Principal Consultant<br>Speed To Market, Inc.<br>2355 East Flamingo Road, Suite 201 G<br>Las Vegas, NV 89119 USA<br>tkroenke@speedtomarket.net<br>303 956 4232 |
| Trade Name: | Solaris Compatible Reusable Adult SpO2 Finger Sensors<br>• Model S100A-090103 SpO2 Finger Sensor<br>• Model S100A-300103 SpO2 Finger Sensor<br>• Model S200A-090101 SpO2 Finger Sensor<br>• Model S200A-300101 SpO2 Finger Sensor<br>• Model S300A-300046 SpO2 Finger Sensor |
| Common Name: | Pulse Oximeter Sensor |
| Classification Name: | Oximeter |
| Classification<br>Regulation: | 21 CFR §870.2700 |
| Product Code: | DOA |
427 26th Avenue, Ste. 101, San Francisco, CA 94121 USA Tel: 001-415-221-4350 Fax: 001-415-221-4360() ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( ) ( www.solarismedtech.com
ﺒﻪ .
ج Page 1 of 3
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Solaris Medical Technology, Inc. The logo features a stylized "S" that resembles a mountain range. The word "SOLARIS" is written in bold, uppercase letters to the right of the "S", and the company name is written in a smaller font below the logo.
| Substantially<br>Equivalent Devices: | Solaris Model | Predicate 510(k)<br>Number | Predicate<br>Manufacturer / Model |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|-------------------------------------------------------|
| | S100A-090103 SpO2<br>Finger Sensor | K993637* | Nellcor DS-100A SpO2<br>Sensor, 0.9 m cable |
| | S100A-300103 SpO2<br>Finger Sensor | K993637* | Nellcor DS-100A SpO2<br>Sensor, 3.0 m cable |
| | S200A-090101 SpO2<br>Finger Sensor | K030407* | BCI 3044 SpO2 Sensor,<br>0.9 m cable |
| | S200A-300101 SpO2<br>Finger Sensor | K030407* | BCI 3044 SpO2 Sensor,<br>3.0 m cable |
| | S300A-300046 SpO2<br>Finger Sensor | K962127* | Datex-Ohmeda OXY-<br>F4-H SpO2 Sensor, 3.0<br>m cable |
| | * Please note these 510(k) numbers represent monitors containing a pulse oximetry<br>module in which the predicate sensors were included as accessories. | | |
| Device Description: | Solaris Compatible Reusable Adult SpO2 Finger Sensors (Solaris<br>Sensors) are compatible reusable sensors for use with major brands of<br>patient monitors and oximeter devices. | | |
| | Solaris Sensors are electro-optical sensors which function without skin<br>penetration, electrical contact, or heat transfer. The sensors use optical<br>means to determine the light absorption of functional arterial<br>hemoglobin by being connected between the patient and the patient<br>monitor or oximeter device. The sensor contains three optical<br>components: two light emitting diodes (LED) that serve as light<br>sources and one photodiode that acts as a light detector. The LED and<br>photodiode are contained in silicon rubber pads. | | |
| Intended Use: | Solaris Sensors are indicated for use in continuous, non-invasive<br>monitoring of arterial oxygen saturation and pulse rate for patients<br>weighing more than 40 kg. | | |
| Technology<br>Comparison: | Solaris Sensors employ the same technological characteristics as the<br>predicate devices to determine arterial oxygen saturation: arterially<br>perfused tissue is illuminated sequentially by two wavelengths of<br>LEDs, and the time varying absorbance of the tissue is measured by a<br>photodetector.<br>This method is characteristic of all reusable sensors that are the subject of this submission as well as the predicate devices. | | |
:
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image contains the logo for Solaris Medical Technology, Inc. The logo features a stylized mountain peak symbol to the left of the word "SOLARIS" in a bold, sans-serif font. Below the word "SOLARIS" is the text "Solaris Medical Technology, Inc." in a smaller font size.
## Performance Testing:
Cleaning Instruction Solaris Sensors were tested in accordance with internal protocols to Testing ensure that the cleaning instructions do not damage sensor labeling or degrade the material. Test results indicated that the cleaning instructions do not damage sensor labeling or degrade the material. Biocompatibility Patient contact materials used in Solaris Sensors were tested in accordance with ISO 10993-1: 2003, Biological evaluation of medical Testing devices - Part 1: Evaluation and testing for skin surface-contact, limited-duration devices. Test results indicated that the patient contact materials were non-toxic, non-sensitizing and non-irritating. Electrical Safety Solaris Sensors were tested in accordance with IEC 60601-1:1988: Testing Aml: 1991; A2: 1995, Medical electrical equipment -- Part 1: General requirements for safety, Clauses 3, 5, 6, 48 and 56. Test results indicated that the sensors comply with the stated clauses. Solaris Sensors were tested in accordance with IEC 60601-1-2:2001, Electromagnetic Compatibility Medical electrical equipment - Part 1-2: General requirements for Testing safety - Collateral standard: Electromagnetic compatibility -Requirements and tests, Clauses 5.1.1, 6.2.1 and 6.3.1. Test results indicated that the sensors comply with the stated clauses of the Standard. Clinical Testing Solaris Sensors were clinically tested to validate the performance and accuracy of the sensors under controlled hypoxia versus arterial oxygen saturation as determined by co-oximetry. All testing was performed under an institutionally approved protocol with subject informed consent. Test results indicated that the sensors meet the published specifications for accuracy over the 70 % - 100 % SpO2 range. Conclusion Based upon a comparison of devices and performance testing results, Solaris Sensors are substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Solaris Medical Technology, Incorporated C/O Mr. Thomas Kroenke Principal Consultant Speed to Market, Incorporated 2355 East Flamingo Road, Suite 201G Las Vegas, Nevada 89119
JAN 3 1 2007
Re: K061931
Trade/Device Name: Solaris Compatible Reusable Adult SpO2 Finger Sensors Including: Model S100A-090103; Model S100A-300103; Model S200A-090101; Model S200A-300101; and Model S300A-300046
Regulation Number: 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: December 31, 2006 Received: January 3, 2007
Dear Mr. Kroenke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Kroenke
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
510(k) Number (if known): K Solaris Compatible Reusable Adult SpO2 Finger Sensors Device Name: including: Model S100A-090103; . Model S100A-300103; . Model S200A-090101; ● . Model S200A-300101; and Model S300A-300046. � Solaris Medical Technology, Inc. SpO2 sensors are indicated for Indications for Use: use in continuous, non-invasive monitoring of arterial oxygen saturation and pulse rate for patients weighing more than 40 kg.
Prescription Use
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Chr. Kuhl
KCGVM31
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.