The Welch Allyn Spot Vital Signs is intended for measurement of blood pressure, pulse rate, temperature and oxygen saturation (SpO2) of adult and pediatric patients. The device is not designed, sold or intended for use except as indicated. The Welch Allyn Spot Check Device is not designed for use with neonates. To ensure pediatric blood pressure accuracy and safety, note that the Welch Allyn small cuff (5200-03) and the small One Piece cuff (5200-13) are the smallest cuffs approved for use with young children and infants. The circumference of the child's arm must fit within the range markings on the cuff. The Welch Allyn Spot Check Device should not be used on patients who are linked to heart/lung machines. The Welch Allyn Spot Check Device is not designed for use of axillary temperature option above three years of age in children. The Welch Allyn Spot Check Device is not intended to monitor patient's vital signs. The Welch Allyn Spot Check Device is not defibrillator proof.
Device Story
Spot vital signs measurement device; provides one-time readings for blood pressure, pulse, temperature, and SpO2. Inputs: non-invasive BP cuff, thermometry probe, pulse oximetry sensor. Outputs: LCD display of vital signs; optional thermal printer output via infrared port. Used in clinical settings by physicians, physician assistants, or nurses for high-traffic patient assessment. Device lacks continuous monitoring, programmable alarms, or memory. Benefits: cost-effective, portable, rapid spot-check capability for adult and pediatric patients.
Clinical Evidence
Bench testing only. Compliance with recognized standards including EN60601-1, EN60601-1-1, EN60601-1-2, EN60601-1-4, AAMI SP10, ASTM 1112-86, and EN 865 demonstrated for the MP506 pulse oximetry module and base unit.
Technological Characteristics
Non-invasive BP measurement, SureTemp thermometry, Nellcor SpO2 pulse oximetry. Infrared port for external printer. Rechargeable battery. Standards: EN60601-1, EN60601-1-1, EN60601-1-2, EN60601-1-4, AAMI SP10, ASTM 1112-86, EN 865. LCD display. Portable form factor with mounting accessories.
Indications for Use
Indicated for adult and pediatric patients (excluding neonates) for non-invasive measurement of blood pressure, pulse rate, temperature, and SpO2. Contraindicated for patients on heart/lung machines. Axillary temperature measurement restricted to children age 3 and under. Not for continuous monitoring or defibrillator use.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
OxiMax Pulse Oximetry System w/N-595 Pulse Oximeter (K012891)
Submission Summary (Full Text)
{0}------------------------------------------------
K 022163
SEP 1 7 2002
# V. 510(k) Summary
[As described in CFR 807.92]
| Submitted by: | Welch Allyn Inc.<br>4341 State Street Road<br>Skaneateles Falls, NY 13153 |
|----------------------|----------------------------------------------------------------------------------------------|
| Contact Person: | David Klementowski<br>Corporate Regulatory Affairs Manager |
| Date Prepared: | 20 May 2002 |
| Proprietary Name: | Welch Allyn Spot Vital Sign Monitor with MP506 |
| Common Name: | Vital Signs measurement device |
| Classification Name: | Class II 870.1130 Noninvasive Blood Pressure<br>Measurement System |
| Predicate Devices: | Welch Allyn Spot Vital Signs<br>Welch Allyn, Inc.<br>510(k) Document Control Number K002530 |
| | OxiMax Pulse Oximetry System w/N-595 Pulse Oximeter<br>Nellcor Puritan Bennett, Incorporated |
Noneor Parkan Delhott, Incorporator
510(k) Document Control Number K012891
{1}------------------------------------------------
## Description of the Device:
The Welch Allyn Spot Vital Signs is not a monitor, but a one-time vital signs measurement device. This product will not have continuous monitoring capability with timed cycle intervals, memory, or any various programmable alarm features. The device is intended to provide the physician, physician's assistant, or nurse, facing high patient traffic or multiple tasks, a cost effective method to determine a one-time vital signs reading on the spot. The base unit will have non-invasive blood pressure (BP) measurement. Options will also be offered such as SureTemp® thermometry, Nellcor® pulse oximetry (SpO2), mounting bracket, and rolling stand. The device may be interfaced with an external printer via an infrared port.
The Welch Allyn Spot Vital Signs is designed to non-invasively measure systolic and diastolic blood pressure, pulse rate, temperature and oxygen saturation (SpO2) for adult and pediatric patients. The Welch Allyn Spot Vital Signs also calculates Mean Arterial Pressure (MAP). All blood pressure, pulse, temperature and SpO2 values are displayed on a large, easy-to-read LCD display, and may be printed via an external thermal printer, as desired. The rechargeable battery and wide variety of mounting accessories make the Welch Allyn Spot Vital Signs convenient for many locations.
## Intended Use
The Welch Allyn Spot Check Device Check Device is intended for measurement of blood pressure, pulse rate, temperature and oxygen saturation (SpO2) of adult and pediatric patients. The device is not designed, sold or intended for use except as indicated.
The Welch Allyn Spot Check Device is not designed for use with neonates. To ensure pediatric blood pressure accuracy and safety, note that the Welch Allyn small cuff (5200-03) and the small One Piece Cuff (5200-13) are the smallest cuffs approved for use with young children and infants. The circumference of the child's arm must fit within the range markings on the cuff.
## The Welch Allyn Spot Check Device should not be used on patients who are linked to heart/lung machines.
The Welch Allyn Spot Check Device is not designed for use of axillary temperature option above three years of age in children.
The Welch Allyn Spot Check Device is not intended to monitor patient's vital signs.
The Welch Allyn Spot Check Device is not defibrillator proof.
{2}------------------------------------------------
Action Taken to Comply with Section 514 of the Act
The agency has recognized the following standards:
- a) EN60601-1
- b) EN60601-1-1
- c) EN60601-1-2
- d) EN60601-1-4
- e) AAMI SP10
- f) ASTM 1112-86
- g) EN 865
The Welch Allyn Spot Vital Sign with MP506 Pulse Oximeter OEM module meets the requirements called out in these standards. Evidence of compliance is on file at Welch Allyn and is available for review upon demand.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle, there is a stylized symbol that resembles three overlapping human profiles or abstract shapes, possibly representing people or services provided by the department.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. David Klementowski Corporate Regulatory Affairs Manager Welch Allyn, Incorporated 4341 State Street Road Skaneateles Falls, New York 13153-0220
Re: K022163
Trade/Device Name: Welch Allyn Spot Vital Sign Monitor with MP506 Regulation Number: 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: August 13, 2002 Received: August 20, 2002
Dear Mr. Klementowski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{4}------------------------------------------------
Page 2 - Mr. Klementowski
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Timothy A. Ulatowski
Timothy Ulatowski Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## III. Indications for Use Statement
11 11 11 11
.
| 510(k) Number: | Unknown K022163 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Welch Allyn Spot Vital Signs |
| Indications for use: | The Welch Allyn Spot Vital Signs is intended for<br>measurement of blood pressure, pulse rate, temperature and<br>oxygen saturation (SpO2) of adult and pediatric patients.<br>The device is not designed, sold or intended for use except<br>as indicated.<br>* The Welch Allyn Spot Check Device is not designed for<br>use with neonates. To ensure pediatric blood pressure<br>accuracy and safety, note that the Welch Allyn small cuff<br>(5200-03) and the small One Piece cuff (5200-13) are the<br>smallest cuffs approved for use with young children and<br>infants. The circumference of the child's arm must fit<br>within the range markings on the cuff.<br>* The Welch Allyn Spot Check Device should not be used<br>on patients who are linked to heart/lung machines.<br>* The Welch Allyn Spot Check Device is not designed for<br>use of axillary temperature option above three years of age<br>in children.<br>* The Welch Allyn Spot Check Device is not intended to<br>monitor patient's vital signs.<br>* The Welch Allyn Spot Check Device is not defibrillator<br>proof. |
1. 1. 1.
## (Please Do Not Write Below This Line - Continue On Another Page If Needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | <div style="text-align:center;">✓</div> |
|----------------------|-----------------------------------------|
| (Per 21 CFR 801.109) | Or Over-The-Counter Use |
__________________________
(Division Sign-Off)
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices
510(k) Number: IL022163
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.