K091732 · Deep Breeze , Ltd. · OCR · Mar 4, 2010 · Cardiovascular
Device Facts
Record ID
K091732
Device Name
VRLXP, MODEL XP
Applicant
Deep Breeze , Ltd.
Product Code
OCR · Cardiovascular
Decision Date
Mar 4, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1875
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The VRIxp is intended for monitoring and recording lung sounds and automatic detection of crackles and wheezes. When interpreted by physicians with general medical training and experience, the VRIxp aids in diagnosis and patient management. The VRIxp is intended to be used in healthcare facilities on adults, adolescents, and/or children over the height of 2 feet 9 inches.
Device Story
VRIxp is a non-invasive electronic stethoscope system for lung sound monitoring. Inputs: acoustic energy collected via dermal sound sensors. Processing: Digital Collection Module (DCM) converts analog signals to digital; mobile computer workstation processes, displays, and stores data. Output: visual grayscale lung sound map and audible playback per sensor; automated detection of crackles and wheezes. Usage: healthcare facilities; operated by physicians. Clinical utility: aids diagnosis and patient management by providing visual/audible lung data for physician interpretation, allowing validation against conventional auscultation. Benefits: objective recording and automated detection of adventitious lung sounds to support clinical decision-making.
Clinical Evidence
Bench testing only. Biocompatibility testing and software verification/validation were performed. Accuracy of automated crackle and wheeze detection was compared to the STG Monitor Multichannel Lung Sound Analysis System, demonstrating similar performance.
Technological Characteristics
Electronic stethoscope system; non-energy emitting. Components: dermal sound sensors, Digital Collection Module (DCM), mobile computer workstation. Connectivity: PC-based display/storage. Software: automated detection of crackles and wheezes. Biocompatible materials used.
Indications for Use
Indicated for monitoring/recording lung sounds and automatic detection of crackles and wheezes in adults, adolescents, and children over 2 feet 9 inches tall in healthcare facilities. Aids physicians in diagnosis and patient management.
Regulatory Classification
Identification
Manual stethoscope: A mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs. Electronic stethoscope: An electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
STG Monitor Multichannel Lung Sound Analysis System (K012387)
Submission Summary (Full Text)
{0}------------------------------------------------
K091732
page 1 of 3
# 510(K) SUMMARY
| Submission Date: | June 11, 2009 |
|------------------|---------------|
|------------------|---------------|
February 26, 2010 Preparation Date:
MAR - 4 2010
# Submitter Information:
| Company Name: | | | Deep Breeze Ltd. |
|---------------|--|--|------------------|
|---------------|--|--|------------------|
Applicant Address: Deep Breeze, Ltd. Attn: Alon Kushnir 2 Hailan Street, North Industrial Park PO Box 140, Or-Akiva, 30600, Israel Telephone: +972 (4) 626-6650 Fax: +972 (4) 626-6649
| Contact Person: | Jack A. Kent |
|-----------------|----------------------------------------|
| | Project Manager, Strategic Consulting |
| | Becker & Associates Consulting, Inc. |
| | 2001 Pennsylvania Avenue NW, Suite 950 |
| | Washington, DC 20006 |
| | Telephone: (202) 822-1850 |
| | Fax: (202) 822-1859 |
| | Email: jack.kent@becker-consult.com |
| | |
# Device Information:
| Trade Name: | VRIXP |
|----------------------|------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Electronic Stethoscope |
| Classification Name: | Electronic Stethoscope |
| Device Class: | Class II, 21 CFR 870.1875(b)<br>Product Code OCR |
| Predicate Devices: | VRIxv (K073582)<br>Deep Breeze, Ltd.<br>Class II<br><br>VR Lung Electrosonograph (K06 1495)<br>Deep Breeze, Ltd.<br>Class II |
{1}------------------------------------------------
Meditron Stethoscope System (K991367) Meditron AS Class II
STG Monitor Multichannel Lung Sound Analysis System (K012387) Stethographics, Inc. Class II
#### Device Description:
The VRIxp is a non-invasive, non-energy emitting device indicated for monitoring and recording lung sounds and automatic detection of crackles and wheezes. When interpreted by physicians with general medical training and experience, the VRIxP aids in diagnosis, monitoring, and patient management.
The VRIxp consists of same fundamental components as its predecessors the VRLE and the VRIxy, which are I) sound sensors designed to collect lung sounds via dermal contact with human skin; 2) a Digital Collection Module (DCM) for the conversion of analog data to digital data; and 3) a mobile computer workstation to assist in processing, displaying, and/or storing recording information.
During the breathing process, the VRIxp detects lung sounds (i.e., acoustic energy) and converts them into a visual display, which can be viewed via a personal computer (PC) monitor and stored for future review. The device is designed to record breath sounds based on sensor location. Additionally, the VRIxp has an automated feature for detecting sounds consistent with crackles and wheezes for further clinical evaluation. Lung sounds can be viewed collectively as a grayscale image, as well as audibly by sensor. This latter feature provides greater flexibility for physicians to validate visual lung data by applying his or her previous clinical experience with conventional auscultation (gold standard).
### Indications for Use:
The VRIxp is intended for monitoring and recording lung sounds and automatic detection of crackles and wheezes. When interpreted by physicians with general medical training and experience, the VRIxp aids in diagnosis and patient management. The VRIxp is intended to be used in healthcare facilities on adults, adolescents, and/or children over the height of 2 feet 9 inches.
#### Comparison to Predicate Device:
The VRIxp is substantially equivalent to the VRLE (manufactured by Deep Breeze; K061495), the VRIxy (manufactured by Deep Breeze; K073582), and the Meditron stethoscope system (manufactured by Meditron AS; K991367). The VRIxp and these three predicate devices share the same intended use and fundamental technological
{2}------------------------------------------------
characteristics; the VRLE and VRIxy are indicated for monitoring and recording lung sounds, while the Meditron stethoscope system is indicated for use as an aid in diagnosis, treatment, and monitoring.
Additionally, the VRIxr has the same fundamental technological characteristics as the VRLE, VRIxy, and Meditron stethoscope system. Like the VRIxp, the VRLE and VRIxy use sound sensors to collect lung sounds via dermal contact, which is then converted to a visual display. The Meditron stethoscope system amplifies sound from the body's internal organs without introducing signals or energy into the body, and in fact, represents the electronic sensors contained within the VRLE and VRIxp. Additionally, the Meditron stethoscope system contains a software application that is designed to provide computer-aided recordings with the electronic stethoscope and to store these recordings along with other appropriate patient information, same as the VRIxp.
The STG Monitor Multichannel Lung Sounds Analysis System (STG) is a predicate device for the clearance of the VRIxp. It has the same intended use, namely: "automatic detection of crackles and wheezes." The VRIxp testing has demonstrated similar accuracy in crackles and wheezes detection as the reported accuracy of the predicate device.
Non-clinical testing has been conducted to demonstrate the performance of VRIxp and that it meets its intended use. The VRLxphas been evaluated for biocompatabiliuty, and appropriate software verification and validation testing was conducted. Pre-determined product specifications were met.
## Conclusion:
The VRIxp falls within the generic type of device as defined by 21 CFR 870.1875(b), Electronic Stethoscope. The VRIxp does not differ from the intended use or fundamental technological characteristics of the cited predicates, the VRLE, VRIxy, or the Meditron stethoscope system, and therefore is substantially equivalent to them.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three lines forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
MAR - 4 2010
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Deep Breeze, Limited C/O Mr. Jack Kent Project Manager, Strategic Consulting Becker & Associates Consulting, Incorporated 2001 Pennsylvania Avenue Northwest, Suite 950 Washington, DC 20006
Re: K091732
Trade Name: VRIxP Regulation Number: 21 CFR 870.1875(b) Regulation Name: Electronic Stethoscope Regulatory Class: Class II Product Code: OCR Dated: March 4, 2010 Received: March 4, 2010
Dear Mr. Kent:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
## Page 2 - Mr. Kent
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address. http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
flin for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{5}------------------------------------------------
### INDICATIONS FOR USE
K091732 510(k) Number (if known):
VRIxP Device Name:
Indications for Use:
The VRIxp is intended for monitoring and recording lung sounds and automatic detection of crackles and wheezes. When interpreted by physicians with general medical training and experience, the VRIxp aids in diagnosis and patient management. The VRIxp is intended to be used in healthcare facilities on adults, adolescents, and/or children over the height of 2 feet 9 inches.
Prescription Use X (Part 21 CFR 801 Subpart D)
・・
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schullitin
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of 1
510(k) Number: k091732
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.