AVIEW Aorta Metrics

K260391 · Coreline Soft Co., Ltd. · JAK · Oct 16, 2026 · Radiology

Device Facts

Record IDK260391
Device NameAVIEW Aorta Metrics
ApplicantCoreline Soft Co., Ltd.
Product CodeJAK · Radiology
Decision DateOct 16, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1750
Device ClassClass 2
AttributesAI/ML, Software as a Medical Device

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Aortic segmentationFCN (Full Convolutional Network) Deep Learning AI—All DSC results met established pass criteria.————
Aortic diameter measurementFCN (Full Convolutional Network) Deep Learning AI95% LoAs ±4.61 mm92.33% of measurements fell within LoA; MAE ≤ 1.95 mm; bias -1.58 mm.————

Indications for Use

AVIEW Aorta Metrics is software intended for use in adult patients 18 years of age or older that automatically segments the aorta and measures the aortic diameter by anatomical region using an AI-trained model from CT images.

Device Story

AVIEW Aorta Metrics is software that processes chest CT images to automatically segment the aorta and measure its diameter. It utilizes a Full Convolutional Network (FCN) deep learning AI model to perform segmentation and diameter quantification. The device is intended for use with a cleared AVIEW platform and integrates with PACS via DICOM standards. It is designed for use by clinicians to assist in the evaluation of aortic anatomy. By providing automated measurements and visualizations of aortic zones, the device supports clinical decision-making regarding aortic health. It is deployed in both cloud and on-premise environments.

Clinical Evidence

No clinical studies were conducted. Substantial equivalence was established via non-clinical standalone performance testing. Aortic segmentation accuracy was validated using Dice Similarity Coefficient (DSC) metrics meeting pass criteria. Maximum-diameter measurement performance showed an overall Mean Absolute Error (MAE) of ≤ 1.95 mm (95% CI: 1.78–2.12) and a bias of -1.58 mm (95% CI: -1.79 to -1.37), with 92.33% of measurements falling within 95% Limits of Agreement (±4.61 mm).

Technological Characteristics

Software-based device utilizing FCN (Full Convolutional Network) deep learning AI for image segmentation. Operates on DICOM-compliant chest CT images. Deployment supports cloud and on-premise configurations. Integrates with PACS. No physical materials or energy sources involved.

Indications for Use

Indicated for adult patients 18 years of age or older requiring automated segmentation of the aorta and measurement of aortic diameter by anatomical region from CT images.

Regulatory Classification

Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{1} **FDA** U.S. FOOD & DRUG ADMINISTRATION October 16, 2026 Coreline Soft Co., Ltd. Hyeyi Park Director 12, Donggyo-Ro 19-Gil, Mapo-Gu, Seoul, 04001 KOR Re: K260391 Trade/Device Name: AVIEW Aorta Metrics Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: JAK Dated: September 6, 2026 Received: September 8, 2026 Dear Hyeyi Park: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {2} K260391 - Hyeyi Park Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. {3} K260391 - Hyeyi Park Page 3 For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Jessica Lamb, PhD Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260391 | ? | | Please provide the device trade name(s). | | ? | | AVIEW Aorta Metrics | | | | Please provide your Indications for Use below. | | ? | | AVIEW Aorta Metrics is software intended for use in adult patients 18 years of age or older that automatically segments the aorta and measures the aortic diameter by anatomical region using an AI-trained model from CT images. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {5} K260391 Page 1 # 510(k) Summary # 1 SUBMITTER Coreline Soft Co., Ltd. 12, Donggyo-ro 19-gil, Mapo-gu, Seoul, Republic of Korea. Phone: 82.2.517.7321 Fax: 82.2.571.7324 Contact Person: hyeyi. Park # 2 DEVICE Name of Device: AVIEW Aorta Metrics Classification Name: Computed Tomography X-ray System (21CFR 892.1750) Regulatory Class: II Product Code: JAK # 3 PREDICATE DEVICE AI-Rad Companion (CardioVascular) by Siemens Medical Solutions U.S.A. Name of Device: AI-Rad Companion (CardioVascular) 510k Number: K222360 Classification Name: Computed Tomography X-Ray System (21CFR 892.1750) Regulatory Class: II Product Code: JAK This predicate has not been subject to a design-related recall. # 4 DEVICE DESCRIPTION AVIEW Aorta Metrics automatically segments the aorta using an aorta segmentation model learned with FCN (Full Convolutional Network) Deep Learning AI from chest CT images. Using the aorta segmentation result and CT images together, the aorta area segmentation model learned with FCN Deep Learning AI is used to automatically segment and measure the diameter. This product complies with the DICOM standard and can be linked to a picture storage and communication system (PACS). The device is intended to be used with a cleared AVIEW platform. # 5 INDICATIONS FOR USE AVIEW Aorta Metrics is software intended for use in adult patients 18 years of age or older that automatically segments the aorta and measures the aortic diameter by anatomical region using an AI-trained model from CT images. {6} K260391 Page 2 ## 6 COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVCIE | Characteristic | Subject Device | Predicate Device | | --- | --- | --- | | Device Name | AVIEW Aorta Metrics | AI-Rad Companion (CardioVascular) | | Classification Name | Computed Tomography X-ray System | Computed Tomography X-Ray System | | Regulatory Number | 892.1750 | 892.1750 | | Product Code | JAK | JAK | | Review Panel | Radiology | Radiology | | 510k Number | - | K222360 | | Indications for use | AVIEW Aorta Metrics is software intended for use in adult patients 18 years of age or older that automatically segments the aorta and measures the aortic diameter by anatomical region using an AI-trained model from CT images. | AI-Rad Companion (Cardiovascular) is image processing software that provides quantitative and qualitative analysis from previously acquired Computed Tomography DICOM images to support radiologists and physicians from emergency medicine, specialty care, urgent care, and general practice in the evaluation and assessment of cardiovascular diseases. It provides the following functionality: - Segmentation and volume measurement of the heart - Quantification of the total calcium volume in the coronary arteries - Segmentation of the aorta - Measurement of maximum diameters of the aorta at typical landmarks - Threshold-based highlighting of enlarged diameters The software has been validated for non-cardiac chest CT data with filtered backprojection reconstruction from Siemens Healthineers, GE Healthcare, Philips, and Toshiba/Canon. Additionally, the calcium detection feature has been validated on non-cardiac chest CT data with iterative reconstruction from Siemens Healthineers. Only DICOM images of adult patients are considered to be valid input. | | General Description | AVIEW Aorta Metrics automatically segments the aorta using an aorta segmentation model learned with FCN (Full Convolutional Network) Deep Learning AI from chest CT images. Using the aorta segmentation result and CT images together, the aorta area segmentation model learned with FCN Deep Learning AI is used to automatically segment and measure the diameter. This product complies with the DICOM standard and can be linked to a picture storage and communication system (PACS). The device is intended to be used with a cleared AVIEW platform. | AI-Rad Companion (Cardiovascular) SW version VA20 is an enhancement to the previously cleared device AI-Rad Companion (Cardiovascular) K183268 that utilizes machine and deep learning algorithms to provide quantitative and qualitative analysis to computed tomography DICOM images to support qualified clinicians in the evaluation and assessment of cardiovascular diseases. | | Aorta segmentation | Aorta Segmentation with deep learning-based algorithm. | Aorta Segmentation with deep learning-based algorithm. | {7} K260391 Page 3 | Landmark detection | Landmark Detection with deep learning-based algorithms, 3 AHA positions. | Landmark Detection with deep learning-based algorithms, 9 AHA positions. | | --- | --- | --- | | Aorta diameter measurement | Automatically measures the aorta diameter for each zone. | Aorta diameter measurements at nine predefined locations according to the AHA guidelines and at the locations of the maximum diameter of the ascending and descending aorta. | | Aorta categories | Categorization of diameter measurements at locations according to the AHA guidelines and at locations of the maximum diameter of ascending and descending aorta. | Categorization of diameter measurements at locations according to the AHA guidelines and at locations of the maximum diameter of ascending and descending aorta. | | Image format | DICOM | DICOM | | Outputs | - Visualization of aorta and aorta zone areas- Segmentation of aorta and aorta zone areas- Provides measured diameter values (max diameter) for each aorta zone | - Segmentation and volume measurement of the heart- Quantification of the total calcium volume in the coronary arteries- Segmentation of the aorta- Measurement of maximum diameters of the aorta at typical landmarks- Threshold-based highlighting of enlarged diameters | | Modality | Non-enhanced chest CT | Non-enhanced chest CT | | Deployment | Cloud and on-premise deployments | Cloud and Edge (on-premise) deployments | ## 7 PERFORMANCE DATA ### 7.1 Standalone Test To evaluate the performance of AVIEW Aorta Metrics, the results for aortic segmentation and diameter measurement were analyzed. All DSC results for segmentation accuracy met the established pass criteria. For maximum-diameter measurement performance, based on inter-reader variability assessment, 95% LoAs were established at ±4.61 mm, and 92.33% of the overall measurements from the AVIEW Aorta Metrics fell within the LoA. The subject device yielded an overall MAE of ≤ 1.95 mm (95% CI: 1.78–2.12) and a bias of -1.58 mm (95% CI: -1.79 to -1.37), demonstrating comparable performance to the predicate device (91.9% of the measurements provided by the predicate device were found to lie within the LoA, with a bias within ±1.5 mm, 95% CI: 1.2 mm – 1.8 mm / MAE ≤ 1.8 mm, 95% CI: 1.44 mm – 2.23 mm). ### 7.2 Software Verification and Validation #### 7.2.1 Unit Test - Perform unit tests according to software requirements. - If every unit test contents are 100% pass, unit test is done. - It was confirmed that Every unit test verification contests are 100% pass. #### 7.2.2 System Test - Total 100 tests were performed, if every test contets are 100% pass, system test is done. - It is confirmed that every system test verification contets were 100% pass. {8} K260391 Page 4 ### 7.2.3 Performance Test - System availability test - Image storage error test - Data deleting preventions on installation or uninstallation test - Importing time test - Ensure compatibility and reliability when running programs test ## 8 Clinical Test No clinical studies were conducted or necessary to establish substantial equivalence. Substantial equivalence was supported by nonclinical standalone performance testing of the device's automatic aorta segmentation and diameter measurement (see Nonclinical Performance Testing). ## 9 CONCLUSIONS AVIEW Aorta Metrics has the same intended use and the principle of operation and has similar features to the predicate devices. There might be slight differences in features and menu, but these differences between the predicate device and the proposed device are not so significant since they do not raise any new or potential safety risks to the user or patient and questions of safety or effectiveness. Based on the results of software validation and verification tests, we conclude that the proposed device is substantially equivalent to the predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...