AI/ML, Software as a Medical Device, PCCP, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K260217 · Feb 24, 2026
AI Platform 2.2 (AIP002)
Exo Imaging
Retrospective clinical ultrasound images from 184 patients
Retrospective clinical data was used to validate the performance of the Quality AI module for abdominal views by comparing AI-generated quality scores against ratings provided by experienced sonographers.
96.6% of clips rated ACEP 3+ by experts received at least 'Minimum criteria met' by AI; 96.1% of AI-rated 'Minimal criteria met' or 'good' clips were deemed diagnostic by experts (ACEP 3+).
AI Platform 2.2 is intended for noninvasive processing of ultrasound images to detect, measure, and calculate relevant medical parameters of structures and function of patients with suspected disease. In addition, it can provide Quality Score feedback to assist healthcare professionals, trained and qualified to conduct echocardiography, abdominal, and lung ultrasound scans in the current standard of care while acquiring ultrasound images. The device is intended to be used on images of adult patients.
Device Story
SaMD accessory for diagnostic ultrasound systems; processes DICOM ultrasound images to detect, measure, and calculate medical parameters; provides real-time frame/clip quality scores for cardiac (A4C, PLAX), lung (A-lines, B-lines), and abdominal (upper quadrant, pelvic) views. Used in point-of-care (POC) settings by trained healthcare professionals; assists clinicians in evaluating scan quality and generating reports. AI modules (Deep Convolutional Neural Networks) perform segmentation, landmark detection, and classification. Output aids clinical decision-making but does not replace professional interpretation; clinicians remain responsible for diagnosis and patient management. Benefits include standardized image quality assessment and workflow simplification.
Clinical Evidence
Bench testing and validation on 184 patients (200 clips, 29,371 frames) for abdominal quality AI. Compared AI quality ratings against expert sonographer ratings. Results: ICC 0.94 for frames, ICC 0.95 for clips. Real-time validation with 12 users (including 8 novices) on 186 scans showed 96.6% agreement with expert-rated diagnostic quality (ACEP score ≥3) and 96.1% positive predictive value for diagnostic clips. Cardiac and lung modules unchanged from predicate.
Technological Characteristics
Software as a Medical Device (SaMD); Deep Convolutional Neural Networks for segmentation, landmark detection, and classification. Processes single/multi-frame DICOM ultrasound images. Integrated as software components into third-party diagnostic ultrasound systems. Validated per IEC 62304:2006/AC:2015.
Indications for Use
Indicated for adult patients with suspected disease requiring noninvasive ultrasound image processing for detection, measurement, and calculation of medical parameters, and quality score feedback during echocardiography, abdominal, and lung ultrasound scans.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 24, 2026
Exo Imaging
Dornoosh Zonoobi
VP of AI
4201 Burton Dr.
Santa Clara, California 95054
Re: K260217
Trade/Device Name: AI Platform 2.2 (AIP002)
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical image management and processing system
Regulatory Class: Class II
Product Code: QIH
Dated: January 23, 2026
Received: January 23, 2026
Dear Dornoosh Zonoobi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260217 - Dornoosh Zonoobi
Page 2
premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) regulation (21 CFR Parts 4 and 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part
{2}
K260217 - Dornoosh Zonoobi
Page 3
803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Jessica Lamb, Ph.D.
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260217 | ? |
| Please provide the device trade name(s). | | ? |
| AI Platform 2.2 (AIP002) | | |
| Please provide your Indications for Use below. | | ? |
| AI Platform 2.2 is intended for noninvasive processing of ultrasound images to detect, measure, and calculate relevant medical parameters of structures and function of patients with suspected disease. In addition, it can provide Quality Score feedback to assist healthcare professionals, trained and qualified to conduct echocardiography, abdominal, and lung ultrasound scans in the current standard of care while acquiring ultrasound images. The device is intended to be used on images of adult patients. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
EXO
510(k) Summary - AI Platform 2.2
# 510(k) Summary
K260217
# Purpose of the Submission
This Special 510(k) submission is intended to support a limited modification to the legally marketed Al Platform 2.0 (K240953). The modification extends the existing Quality AI functionality to support additional abdominal ultrasound views (upper quadrants and pelvic). The modification does not change the intended use, clinical role, core technology, software architecture, user workflow, or risk profile of the device.
General Information
| 510(k) Sponsor | Exo Imaging |
| --- | --- |
| Address | 4201 Burton Drive Santa Clara, CA 95054 |
| Correspondence Person | Dornoosh Zonoobi |
| Contact Information | dornoosh@exo.inc Cell: +1 780 717 4817 |
| Date Prepared | January 23, 2026 |
Proposed Device
| Proprietary Name | AI Platform 2.2 (AIP002) |
| --- | --- |
| Common Name | AI Platform 2.2 |
| Premarket Notification | K260217 |
| Classification Name | Automated Radiological Image Processing Software |
| Regulation Number | 21 CFR 892.2050 |
| Product Code | QIH |
| Regulatory Class | II |
Predicate Device
| Proprietary Name | AI Platform 2.0 (AIP002) |
| --- | --- |
| Premarket Notification | K240953 |
| Classification Name | Automated Radiological Image Processing Software |
| Regulation Number | 21 CFR 892.2050 |
| Product Code | QIH |
| Regulatory Class | II |
{5}
EXO
510(k) Summary - AI Platform 2.2
# Device Description
Exo AI Platform 2.2 (AIP 2.2) is a software as a medical device (SaMD) that helps qualified users with image-based assessment of ultrasound examinations in adult patients. It is designed to simplify workflow by helping trained healthcare providers evaluate, quantify, and generate reports for ultrasound images. AIP 2.2 takes as an input in the Digital Imaging and Communications in Medicine (DICOM) format from ultrasound scanners of a specific range and allows users to detect, measure, and calculate relevant medical parameters of structures and function of patients with suspected disease. In addition, it provides frame and clip quality score in real-time for the Left Ventricle from the four-chamber apical and parasternal long axis views of the heart, Abdominal Upper Quadrant and Pelvic views, and lung scans. The AI modules are provided as software components to be integrated by another computer programmer into their legally marketed ultrasound imaging device. Essentially, the Algorithm and API, which are modules, are medical device accessories.
Key features of the software are:
- Lung AI: An AI-assisted tool for suggesting the presence of lung structures and artifacts on ultrasound images, namely A-lines and B-lines.
- Cardiac AI: An AI-assisted tool for the quantification of Left Ventricular Ejection Fraction (LVEF), Myocardium wall thickness (Interventricular Septum (IVSd), Posterior wall (PWd)), and IVC diameter on cardiac ultrasound images.
- Quality AI: An AI tool designed to assess ultrasound per frame and per clip quality across Cardiac (A4C and PLAX), Lung, and Abdominal (Upper Quadrants and Pelvic) views.
# Warning:
It is important to note that patient management decisions should not be made solely on the AI Platform 2.2 analysis results. Users are fully responsible for ensuring scan quality and making diagnoses. Images obtained through the use of Quality AI should be interpreted exclusively by certified healthcare professionals. It is also imperative that a qualified healthcare professional reviews the data, ensuring its adequacy and appropriateness for the intended diagnosis or management.
# Elements Unchanged from the Predicate Device:
The following aspects remain unchanged from the predicate device (AI Platform 2.0, K240953)
- Intended use and patient population
- AI architecture and system design
Data inputs, outputs, and user workflow
- Labeling, warnings, and clinician responsibility
- Risk classification and mitigation strategies
{6}
EXO
510(k) Summary - AI Platform 2.2
# Indications for Use
AI Platform 2.2 is intended for noninvasive processing of ultrasound images to detect, measure, and calculate relevant medical parameters of structures and function of patients with suspected disease. In addition, it can provide Quality Score feedback to assist healthcare professionals, trained and qualified to conduct echocardiography, abdominal, and lung ultrasound scans in the current standard of care while acquiring ultrasound images. The device is intended to be used on images of adult patients.
# Description of the Modification
The modification extends the existing Quality AI module to accept and assess two additional abdominal ultrasound views: upper-quadrant and pelvic. No changes were made to the intended use, core technology, system architecture, data inputs, outputs, or user workflow.
The predicate device (AI Platform 2.0, K240953) includes a Predetermined Change Control Plan (PCCP). The device continues to contain this PCCP; however, the subject modifications described in this submission are outside the scope of the PCCP.
Comparison of Technological Characteristics with the Predicate Device
| Feature/Function | Subject Device Exo AI Platform 2.2 | Predicate Device: Exo AI platform 2.0 (K240953) |
| --- | --- | --- |
| Indications for Use | AI Platform 2.2 is intended for noninvasive processing of ultrasound images to detect, measure, and calculate relevant medical parameters of structures and function of patients with suspected disease. In addition, it can provide Quality Score feedback to assist healthcare professionals, trained and qualified to conduct echocardiography, abdominal, and lung ultrasound scans in the current standard of care while acquiring ultrasound images. The device is intended to be used on images of adult patients. | AI Platform 2.0 is intended for noninvasive processing of ultrasound images to detect, measure, and calculate relevant medical parameters of structures and function of patients with suspected disease. In addition, it can provide Quality Score feedback to assist healthcare professionals, trained and qualified to conduct echocardiography and lung ultrasound scans in the current standard of care while acquiring ultrasound images. The device is intended to be used on images of adult patients. |
| Scan type | Single and Multi-frame ultrasound images | Same as subject device. No changes were made. |
{7}
EXO
510(k) Summary - AI Platform 2.2
| Feature/Function | Subject Device Exo AI Platform 2.2 | Predicate Device: Exo AI platform 2.0 (K240953) |
| --- | --- | --- |
| Principle of Operation and Technology | Ultrasound image processing software implementing artificial intelligence, including non-adaptive machine learning algorithms trained with clinical data intended for non-invasive analysis of ultrasound data | Same as subject device. No changes were made. |
| AI Algorithm | Deep Convolutional Neural Networks for Segmentation, Landmark Detection and Classification | Same as subject device. No changes were made. |
| Cardiac Measurements | LVEF IVC Minimum diameter on inspiration and Maximum diameter on expiration Myocardium wall thickness (Interventricular Septum and Posterior wall) from Plax view | Same as subject device. No changes were made. |
| Non Cardiac AI modules | Absence/presence of A-lines B-lines count | Same as subject device. No changes were made. |
| Real-time feedback on quality | Yes | Same as subject device. No changes were made. |
| Anatomical Sites for Quality AI | Heart, Lungs, and Abdomen | Heart, Lungs |
| Retrospectively recording of Diagnostic quality clip | Yes | Same as subject device. No changes were made. |
| AI modules are an accessory to compatible general purpose diagnostic ultrasound systems | Yes | Same as subject device. No changes were made. |
{8}
EXO
510(k) Summary - AI Platform 2.2
# Risk Analysis Summary
A risk analysis was conducted for the AI Platform 2.2 device in accordance with Exo's existing risk management process and ISO 14971, to assess the potential impact of adding support for abdominal views. The primary identified hazard for the Quality AI functionality - misclassification of image quality - remains unchanged from that identified for the predicate device.
New risk control measures were implemented to address potential risks associated with the added abdominal views. Following implementation and verification of these controls, residual (post-mitigation) risk levels remain consistent with those of the predicate device. The modification does not introduce new hazard types beyond those previously identified, and does not raise new questions of safety or effectiveness. The overall benefit-risk profile remains consistent with that of the predicate device.
# Performance Data
Safety and performance of the AI Platform 2.2 has been evaluated and verified in accordance with software specifications and applicable performance standards through software verification and validation testing. Software validation activities were performed in accordance with IEC 62304:2006/AC:2015 - Medical device software - Software life cycle processes, FDA's "Content of Premarket Submissions for Device Software Functions" Guidance for Industry and Food and Drug Administration Staff Document issued on June 14, 2023 and FDA Guidance (June 2022) "Technical performance assessment of quantitative imaging in radiological device premarket submissions". All verification and validation activities confirmed that the modification did not impact the core functionality, performance, or safety of the system.
# Validation Performance Testing and Results
Validation of the AI Platform 2.2 followed the same methodology, acceptance criteria, and expert-reference framework used for the predicate device and was applied to the additional abdominal views. The performance was successfully evaluated on test data encompassing diverse demographic variables, including gender, age, and ethnicity from multiple clinical sites in metropolitan cities with diverse racial patient populations. The test data was entirely separated from the training/tuning datasets and was not used for any part of the training/tuning. We also established auditability measures, by assigning a unique identification number to each study and its corresponding images.
The validation of Abdominal Quality AI consists of validating on:
a. Data previously acquired from 184 patients from various ultrasound devices and various abdominal pathologies, compared to quality rating by experienced sonographers on each frame and the entire clip. A total of 200 clips (29,371 frames) were used in this test. The overall agreement between the Quality AI and quality rated by the experienced sonographers was ICC = 0.94 (.94 - .95) for frames and ICC = 0.95 (.94 - .96) for clips.
Page 5 of 6
{9}
EXO
510(k) Summary - AI Platform 2.2
b. Data acquired using the image and clip Quality AI in real time while scanning the pelvic and upper quadrant views of the abdomen. In total, 186 abdomen scans were acquired by 12 users with a wide range of ultrasound experience in POC settings, including 8 novice users who received two hours of training. 96.6% of the clips rated as ACEP quality of 3 or above by expert readers, also received at least "Minimum criteria met for diagnosis" image quality by Clip Quality AI. Additionally, 96.1% of scans that were considered as "Minimal criteria met for diagnosis" or "good" by Quality AI were also deemed diagnostic by experts (ACEP score of 3 or higher).
# Conclusions
Exo's AI Platform 2.2 is substantially equivalent in intended use, design, principles of operation, technological characteristics, and safety features to the predicate device. The modification implemented in version 2.2 - an enhancement to the Quality AI module with expanded view support - does not introduce new questions of safety or effectiveness when the device is used as intended.
The performance of the AI Platform 2.2 product was successfully evaluated on test data encompassing diverse demographic variables, including gender, age and ethnicity from multiple clinical sites in metropolitan cities with diverse racial patient populations.
The lung and cardiac functionalities did not change during the development of the AI Platform 2.2 product and remain as per the cleared algorithms in K240953.
Page 6 of 6
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.