Neurora is intended for use by healthcare professionals as an aid in diagnosis and assessment of Autism Spectrum Disorder (ASD).
Device Story
Neurora is a cloud-based SaMD platform used by healthcare professionals in clinical settings to aid in ASD diagnosis. The device takes video and audio recordings of semi-structured behavioral assessments as input; these are captured using off-the-shelf microphones and wearable glasses with forward-facing cameras. The Neurora Engine, an AI/ML module, processes these inputs to detect and quantify biometrics including vocal patterns, facial expressions, eye movements, and gaze. The system calculates temporal event durations and frequencies, then applies ML algorithms to compare these patterns against diagnostic criteria. The output is a binary classification ('Consistent with ASD' or 'Not Consistent with ASD') provided via a web portal. Clinicians review these results alongside other clinical data to inform their diagnostic decision-making. The device benefits patients by providing objective, automated behavioral metrics to support the diagnostic process, potentially improving the efficiency and consistency of ASD evaluations.
Clinical Evidence
Multicenter, blinded, non-interventional study (n=173) across five US sites. Population: children 18 months–12 years undergoing ASD evaluation. Reference standard: DSM-5 criteria integrated by a single clinician using ADOS-2, history, and observation. Results: 70.59% sensitivity (95% CI: 63.08-78.13%), 83.78% specificity (95% CI: 70.27-94.87%), 94.12% PPV, 43.66% NPV. 100% of subjects received a determinate result. No device-related adverse events reported.
Technological Characteristics
SaMD/SaaS platform; cloud-hosted. Utilizes ML algorithms for multi-modal analysis (audio prosody, facial/eye movement). Inputs: video/audio from off-the-shelf cameras/microphones. Connectivity: web-based portal/API. Cybersecurity: Tier 2 classification, encryption, authentication, and secure communication controls. No hardware components; no electrical safety testing required.
Indications for Use
Indicated for non-intrusive quantification and reporting of biometrics derived from video and audio recordings collected during semi-structured behavioral assessment sessions to aid healthcare professionals in the diagnosis of ASD in patients aged 18 months through 12 years; intended for use in conjunction with the diagnostic process and is not a stand-alone diagnostic tool.
Regulatory Classification
Identification
A pediatric Autism Spectrum Disorder diagnosis aid is a prescription device that is intended for use as an aid in the diagnosis of Autism Spectrum Disorder in pediatric patients.
Special Controls
In combination with the general controls of the FD&C Act, the pediatric Autism Spectrum Disorder diagnosis aid is subject to the following special controls:
- (1) Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including an evaluation of sensitivity, specificity, positive predictive value, and negative predictive value using a reference method of diagnosis and assessment of patient behavioral symptomology.
- (2) Software verification, validation, and hazard analysis must be provided. Software documentation must include a detailed, technical description of the algorithm(s) used to generate device output(s), and a cybersecurity assessment of the impact of threats and vulnerabilities on device functionality and user(s).
- (3) Usability assessment must demonstrate that the intended user(s) can safely and correctly use the device.
- (4) Labeling must include:
- (i) Instructions for use, including a detailed description of the device, compatibility information, and information to facilitate clinical interpretation of all device outputs:
- (ii) A summary of any clinical testing conducted to demonstrate how the device functions as an interpretation of patient behavioral symptomology associated with Autism Spectrum Disorder. The summary must include the following:
- (A) A description of each device output and clinical interpretation;
- (B) Any performance measures, including sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV);
- (C) A description of how the cut-off values used for categorical classification of diagnoses were determined; and
- (D) Any expected or observed adverse events and complications.
(iii)A statement that the device is not intended for use as a stand-alone diagnostic.
*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including an evaluation of sensitivity, specificity, positive predictive value, and negative predictive value using a reference method of diagnosis and assessment of patient behavioral symptomology.
(2) Software verification, validation, and hazard analysis must be provided. Software documentation must include a detailed, technical description of the algorithm(s) used to generate device output(s), and a cybersecurity assessment of the impact of threats and vulnerabilities on device functionality and user(s).
(3) Usability assessment must demonstrate that the intended user(s) can safely and correctly use the device.
(4) Labeling must include:
(i) Instructions for use, including a detailed description of the device, compatibility information, and information to facilitate clinical interpretation of all device outputs; and
(ii) A summary of any clinical testing conducted to demonstrate how the device functions as an interpretation of patient behavioral symptomology associated with Autism Spectrum Disorder. The summary must include the following:
(A) A description of each device output and clinical interpretation;
(B) Any performance measures, including sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV);
(C) A description of how the cutoff values used for categorical classification of diagnoses were determined; and
(D) Any expected or observed adverse events and complications.
(iii) A statement that the device is not intended for use as a stand-alone diagnostic.
{1}
[LOGO]
FDA
U.S. FOOD & DRUG
ADMINISTRATION
October 2, 2026
Argus Cognitive, Inc.
Andrea Marosan
Director of Quality-Regulatory Affairs
35 S. Main St.
Hanover, New Hampshire 03755
Re: K260064
Trade/Device Name: Neurora
Regulation Number: 21 CFR 882.1491
Regulation Name: Pediatric Autism Spectrum Disorder Diagnosis Aid
Regulatory Class: Class II
Product Code: QPF
Dated: September 4, 2026
Received: September 4, 2026
Dear Andrea Marosan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download)
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K260064 - Andrea Marosan
Page 2
and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}
K260064 - Andrea Marosan
Page 3
Sincerely,
# Patrick Antkowiak -S
Patrick Antkowiak, PhD
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K260064
Device Name
Neurora
Indications for Use (Describe)
Neurora is indicated for the non-intrusive quantification and reporting of biometrics derived from video and audio recordings collected during semi-structured behavioral assessment sessions to aid healthcare professionals in the diagnosis of ASD in patients aged 18 months through 12 years; Neurora is intended for use in conjunction with the diagnostic process and is not a stand-alone diagnostic tool.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{5}
Argus Cognitive, Inc
Neurora - K260064
[LOGO]
ARGUS Cognitive
# 510(k) SUMMARY
Neurora Software as a Medical Device
## General Information [807.92(a)(1)]
**Submitter:**
Argus Cognitive, Inc.
35 South Main street, Hanover, NH 03755
www.arguscognitive.com
**Contact person:**
Andrea Marosan
Director, Quality-Regulatory Affairs
andi@arguscognitive.com
**Submission date:** January 8, 2026 - *updated on September 04, 2026*
**Submission type:** 510(k) - New Device
**510(k) number:** k260064
**Establishment Registration Number:** 3036502204
## Device Information [807.92(a)(2)]
**Trade Name:** Neurora
**Generic Name:** Pediatric Autism Spectrum Disorder diagnosis aid
**Classification:** 21 CFR 882.1491
**Product Code:** QPF
**Device Class:** Class II
## Predicate Device(s) [807.92(a)(3)]
DEN200069 - Cognoa ASD Diagnosis Aid
1
{6}
Argus Cognitive, Inc
Neurora - K260064
[LOGO]
ARGUS Cognitive
# Device Description [807.92(a)(4)]
## Device characteristics
Neurora is a standalone Software as a Medical Device (SaMD) operating as a web-based Software as a Service (SaaS) platform. It is not part of a hardware medical device and operates independently as cloud-based diagnostic support software.
## Environment of use
Healthcare facility/Hospital: Neurora is intended for use by healthcare professionals who are qualified to conduct semi-structured behavioral assessments, interpret behavioral assessment results, and diagnose ASD.
## General description
Neurora is a software as a medical device (SaMD) designed to aid in the diagnosis of Autism Spectrum Disorder (ASD) by utilizing multiple machine-learning algorithms to measure behavioral outcomes in a post-assessment environment. To measure biometrics during ASD assessments, Neurora relies on a pair of glasses with a central forward-facing camera to record the clinician's field of view, along with off-the-shelf microphones to record audio. Users will then upload audio and video recordings to a web portal as input data. Neurora analyzes this data to detect vocal patterns, body movements, facial expressions, and gaze. The system processes temporal event data to calculate duration metrics and occurrence frequencies, then uses AI/ML algorithms to analyze these patterns against validated diagnostic criteria to generate an autism spectrum disorder output for healthcare professionals. The analysis results are visually displayed and stored in a patient manager along with other patient data.
The device is intended to provide information for use by the care professional, in conjunction with other clinically relevant information, to assist in the diagnosis of ASDs.
The Neurora System consists of the following components:
- Neurora Web Portal: A user interface (UI) enabling healthcare professionals (HCPs) to upload video and audio recordings from structured behavioral assessment sessions. It interfaces with the Application Programming Interface (API) server for the secure transmission and management of patient data.
- Video/Audio Analyst Portal: A UI for HCPs to remotely review patient videos and audio analysis results, verify or correct observed behaviors, and access patient evaluation results.
- Outcome Report: Following the AI/ML analysis, the system generates a comprehensive report based on patterns identified by the Neurora Engine. This report provides:
- Autism Spectrum Disorder (ASD) Result:
- The system evaluates the behavioral patterns detected by the Neurora Engine during the semi-structured interview to generate the ASD Result, the system's estimation of the presence of ASD based on the model's output. The ASD Result should be interpreted
2
{7}
Argus Cognitive, Inc
Neurora - K260064
[LOGO]
ARGUS Cognitive
by qualified healthcare professionals in conjunction with other clinical measures and their professional judgment.
- Neurora API: This component shall act as an interface between the Neurora Engine and the Web Portal. It should expose the functionalities of the engine in such a way that they can be utilized in the web frontend.
- Neurora Engine: An artificial intelligence software module that processes audio data for prosodic vocal expressions and video data for facial and eye movement patterns. It quantifies these patterns to generate biometrics
## Intended use and Indications for Use [807.92(a)(5)]
Intended Use: Neurora is intended for use by healthcare professionals as an aid in diagnosis and assessment of Autism Spectrum Disorder (ASD).
Indications for Use: Neurora is indicated for the non-intrusive quantification and reporting of biometrics derived from video and audio recordings collected during semi-structured behavioral assessment sessions to aid healthcare professionals in the diagnosis of ASD in patients aged 18 months through 12 years; Neurora is intended for use in conjunction with the diagnostic process and is not a stand-alone diagnostic tool.
Neurora's intended use population (18 months through 12 years) is broader than the predicate's (18 months through 6 years). This difference is not critical to the diagnostic use of the device or affects its safety and effectiveness when used as labeled, as detailed under Comparison of Technological Characteristics [807.92(a)(6)] below.
## Comparison of Technological characteristics with the Predicate Devices [807.92(a)(6)]
The following table compares the technological characteristics of Neurora to the predicate device, Cognoa ASD Diagnosis Aid (DEN200069), and identifies where the devices differ.
3
{8}
Argus Cognitive, Inc
Neurora - K260064

ARGUS Cognitive
| Category | Predicate Device Cognoa ASD Diagnosis Aid | Subject Device Neurora | Comparison |
| --- | --- | --- | --- |
| General Comparison | | | |
| Submission Number | DEN200069 | [k260064] | |
| Regulation Number and Regulation Name | 21 CFR 882.1491 Pediatric Autism Spectrum Disorder diagnosis aid | 21 CFR 882.1491 Pediatric Autism Spectrum Disorder diagnosis aid | Equivalent |
| Classification Product Code | QPF | QPF | Equivalent |
| Class | Class II | Class II | Equivalent |
| Classification Panel | Neurology (882) | Neurology (882) | Equivalent |
| Intended Use | Intended for use as an aid in the diagnosis of ASD. | Neurora is intended for use by healthcare professionals as an aid in diagnosis and assessment of Autism Spectrum Disorder (ASD). | Similarities: Both the Predicate Device and the Subject device is intended for non-intrusive objective evaluation of ASD symptoms of a child by using software to analyze the child's behavior in response to external stimuli. Both are intended for use in the evaluation of young children. Differences: While the predicate device's indications for use statement specifies 'diagnosis,' |
4
{9}
Argus Cognitive, Inc
Neurora - K260064

ARGUS Cognitive
| Category | Predicate Device Cognoa ASD Diagnosis Aid | Subject Device Neurora | Comparison |
| --- | --- | --- | --- |
| | | | the subject device includes both diagnosis and assessment of ASD characteristics. Assessment refers to the same diagnostic evaluation process applied to patients who have already received an ASD diagnosis, consistent with clinical practice where the diagnostic tools are used to support ongoing clinical decision-making about ASD characteristics. This is a clarification, and not a significant change. |
| Indications for Use | The Cognoa ASD Diagnosis Aid is intended for use by healthcare providers as an aid in the diagnosis of ASD for patients ages 18 months through 72 months who are at risk for developmental delay based on concerns of a parent, caregiver, or healthcare Provider. The device is not intended for use as a stand-alone diagnostic device but as an adjunct to the diagnostic process. | Neurora is indicated as an aid healthcare professionals in the diagnosis of ASD in patients aged 18 months through 12 years; Neurora is intended for use in conjunction with the diagnostic process and is not a stand-alone diagnostic tool. | Differences: • Age range (Subject device offers a wider range) |
| Intended User Group | The Device is intended for use by healthcare professionals trained and | Neurora is intended for use by healthcare professionals who are | Identical to predicate device |
5
{10}
Argus Cognitive, Inc
Neurora - K260064

ARGUS Cognitive
| Category | Predicate Device Cognoa ASD Diagnosis Aid | Subject Device Neurora | Comparison |
| --- | --- | --- | --- |
| | qualified to interpret the results of a behavioral assessment examination and to diagnose ASD. | qualified to conduct semi-structured behavioral assessments, interpret behavioral assessment results, and diagnose ASD. | |
| Intended Use Population | Patients ages 18 months through 72 months | Infant (from 18 months to 2 years of age) Child (from 2 years to 12 years of age) | Differences: • Age Range (Neurora offers a wider range) |
| Clinical decision | Made by healthcare provider | Made by healthcare provider | Identical to predicate device |
| Technical Comparison | | | |
| Device description | The Cognoa ASD Diagnosis Aid is a software as a medical device (SaMD) that utilizes a machine-learning algorithm that receives independent information from caregivers or parents, trained analysts, and healthcare professionals (HCPs) to aid in the diagnosis of ASD. | Neurora is a software as a medical device (SaMD) that utilizes a machine-learning algorithm to detect and analyze biometrics by processing video and audio data collected by healthcare professionals (HCPs) to aid in the diagnosis of ASD during a semi-structured behavioral assessment. The system operates non-intrusively, without requiring the patient to remain stationary or interact directly with the equipment. | Differences: The subject device does not receive information from caregivers or parents. The subject device's description includes additional technical detail about specific biometrics analyzed (e.g., facial landmarks, gaze, speech patterns, etc) but this represents elaboration rather than fundamental functional difference. |
6
{11}
Argus Cognitive, Inc
Neurora - K260064

ARGUS Cognitive
| Category | Predicate Device Cognoa ASD Diagnosis Aid | Subject Device Neurora | Comparison |
| --- | --- | --- | --- |
| Device architecture | It consists of multiple software applications and hardware platforms. Input data is acquired via a Mobile App, a Video Analyst Portal, and a HCP Portal. | System components: - Neurora Web Portal: A user interface (UI) enabling healthcare professionals (HCPs) to upload video and audio recordings - Video/Audio Analyst Portal: A UI for HCPs to remotely review patient videos and audio analysis results, verify or correct observed behaviors, and access patient evaluation results. - Outcome Report: Following the analysis, the system generates a comprehensive report based on patterns identified by the Neurora Engine. | Equivalent. Both are multi-component software systems with HCP portals for data input, review interfaces, and automated reporting. Architectural differences are minor implementation variations that do not affect safety or effectiveness. |
| Data collection | Data collected using normal commercially available camera to videotaped behavior. | Data is collected using commercially available cameras, including wearable glasses with integrated forward-facing cameras (worn by the clinician) to record the session from the clinician's field of view. | Equivalent |
7
{12}
Argus Cognitive, Inc
Neurora - K260064

ARGUS Cognitive
| Category | Predicate Device Cognoa ASD Diagnosis Aid | Subject Device Neurora | Comparison |
| --- | --- | --- | --- |
| Technology | The Cognoa ASD Diagnosis Aid is a software as a medical device (SaMD) that utilizes a machine-learning algorithm. | Neurora is a software as a medical device (SaMD) that utilizes a machine-learning algorithm. | Equivalent |
| Data output | If the value is greater than the upper threshold, then the device output is 'Positive for ASD.' If the value is less than the lower threshold, then the device output is 'Negative for ASD.' If the available information does not allow the algorithm to render a reliable result, the device output is 'No Result.' | The system evaluates the biometrics detected by the Neurora Engine during the semi-structured interview to provide a binary classification of the subject's behavior: • ○ "Consistent with ASD" ○ "Not consistent with ASD" | Difference: Neurora does not include a 'No result' output. Every assessed patient receives a binary classification. |
| Design | SaMD | SaMD | Identical to predicate device |
| Software / Cybersecurity | MODERATE level of concern. Documentation includes a threat model, cybersecurity mitigation information, an upgrade and maintenance plan, and other | MODERATE level of concern Documentation includes all relevant information as required by the FDA Guidance "Content of Premarket Submissions for | Identical to predicate device |
8
{13}
Argus Cognitive, Inc
Neurora - K260064

ARGUS Cognitive
| Category | Predicate Device Cognoa ASD Diagnosis Aid | Subject Device Neurora | Comparison |
| --- | --- | --- | --- |
| | information for safeguarding the device algorithms. | Management of Cybersecurity in Medical Devices": threat modeling, cybersecurity controls, security assessment using CVSS scoring system, and other information for safeguard and data protection. | |
| Performance Characteristics | | | |
| Safety | Adverse events (AEs) and serious adverse events (SAEs) were collected and reported from enrollment through completion of the reference diagnosis clinic visit. | Safety monitoring protocol was implemented as part of the clinical studies. All adverse events and serious adverse events were collected, documented, and assessed from enrollment through study completion per FDA guidance and IRB requirements, including the external US cohort collected across the five additional sites.. | Identical to predicate device |
| Electrical Safety Testing | N/A as the device is a SaMD. | Not applicable (SaMD on general computing platform) | Identical to predicate device |
| Test data | Data available to show safety and effectiveness | Data available to show safety and effectiveness - See NEU-V&V - Populations and databases document. | Equivalent |
9
{14}
Argus Cognitive, Inc
Neurora - K260064

ARGUS Cognitive
| Category | Predicate Device Cognoa ASD Diagnosis Aid | Subject Device Neurora | Comparison |
| --- | --- | --- | --- |
| Overall ASD classification quality | When measured on all clinical study completers: Sensitivity: 51.6% [95% CI 42.4 - 60.8] Specificity: 18.5% [95% CI 14.3 - 23.3] For the 31.8% of participants who received a determinate output (ASD positive or negative): Sensitivity 98.4% [95% CI 91.6%-99.96%] Specificity 78.9% [95% CI 67.6%-87.7%] | New external US test cohort (Multisite US Independent Test set, n=173): Sensitivity 70.59% [63.08-78.13%], Specificity 83.78% [70.27-94.87%], PPV 94.12% [89.11-98.15%], NPV 43.66% [32.05-55.26%]. 100% of assessed subjects received a determinate binary classification (0% no-result rate) | Neurora's new-cohort performance (sensitivity 70.59%, specificity 83.78%, 100% of assessed subjects receiving a determinate result) is compared against two predicate benchmarks. FDA's De Novo summary for the predicate reports performance only on the 31.76% of subjects who received a determinate output (sensitivity 98.44% [95% CI 91.60%-99.96%], specificity 78.87% [95% CI 67.56%-87.67%]); no all-completers figure is reported there directly. An all-completers estimate (sensitivity 51.64%, specificity 18.48%) can be derived from the confusion matrix in FDA's own De Novo summary (FDA, 2020) by scoring "No Result" as negative; the same estimate, with confidence intervals, is also reported in Megerian et al. (2022). |
| No response rate | 68.24% | 0% no response rate - system always provides a result | Differences: Neurora provides an output for every subject as opposed to the predicate device that gives a "No result" output for 68% of the subjects. |
10
{15}
Argus Cognitive, Inc
Neurora - K260064

ARGUS Cognitive
| Category | Predicate Device Cognoa ASD Diagnosis Aid | Subject Device Neurora | Comparison |
| --- | --- | --- | --- |
| Clinical reference standard | DSM-5 criteria, determined by majority assessment of up to three specialists: diagnosis by a site-specific specialist, followed by independent review by one or two central specialist clinicians, with majority rule resolving disagreement. | DSM-5 criteria, established through clinical evaluation — administration of an age- and language-appropriate ADOS-2, caregiver/guardian interview, and review of developmental and medical history — integrated by a single qualified diagnosing clinician. Assessment components may be administered by trained assisting personnel under the diagnosing clinician's supervision. The reference standard diagnosis was finalized independently of, and prior to, any Neurora output. | Differences: The predicate requires independent central-reviewer confirmation for every case, with majority rule among up to three specialists. Neurora's reference standard relies on a single diagnosing clinician integrating ADOS-2, caregiver interview, and developmental history. There is no routine second-reviewer consensus step, and inter-rater agreement was not calculated across the study population. Both reference standards apply DSM-5 criteria through structured, validated assessment processes and were established independently of, and prior to, exposure to the device output. |
| Other | | | |
| Labeling | The labeling is sufficient and satisfies the requirements of 21 CFR 801.109 for prescription devices. | Meets the requirements of 21 CFR 801.109 for prescription devices. | Equivalent to predicate device |
| Bench testing | Non-clinical or bench testing was generally not needed to evaluate the hardware that the software is intended to be run on. | Bench testing was generally not needed to evaluate the hardware that the software is intended to be run on. | Equivalent to predicate device |
11
{16}
Argus Cognitive, Inc
Neurora - K260064
[LOGO]
ARGUS Cognitive
Neurora's software design leverages and expands upon the technological capabilities of the predicate device, Cognoa ASD Diagnosis Aid (DEN200069), while maintaining substantial equivalence.
# **Key Software Features and Comparisons:**
- **Expanded Age Range:** Neurora is designed to support ASD assessment in children aged 18 months to 12 years. This expands upon the age range of the predicate device: Cognoa is limited to children up to 6 years old. Clinical performance was evaluated across chronological age subgroups spanning the full intended use range, including the 18-36-month subgroup (n=30 ASD, n=4 non-ASD; sensitivity 83.33% [95% CI: 68.75-96.43%], specificity 100%), as detailed in the **Clinical Validation Report (Table 1.)**. These results do not indicate that the expanded age range introduces a safety or effectiveness concern distinct from the predicate's.
- **Data Acquisition and Analysis:** Neurora utilizes video and audio recordings, captured using readily available, off-the-shelf devices, to collect data during semi-structured clinician-child interactions. This approach aligns with Cognoa in utilizing digital data for analysis but expands the modalities captured. The recorded data is then analyzed automatically to derive objective behavioral measurements. This automated analysis contrasts with Cognoa's reliance on manual data entry and annotation by parents, caregivers, and HCPs.
- **Output Reporting and Clinical Interpretation:** Neurora provides a classification to support clinician assessment of "Consistent with ASD" or "Not Consistent with ASD."
- This output format differs from the predicate, Cognoa, which provides a binary categorical output ("Positive for ASD" / "Negative for ASD") with a "No Result" category. Both devices, however, provide quantitative decision support intended to assist — but not replace — clinical judgment. Neither output directly determines or controls a diagnostic or treatment decision. Neurora maintains the same intended use and clinical function. Therefore, this difference in output reporting does not raise new questions of safety or effectiveness and is considered substantially equivalent.
The subject device, Neurora, has the same intended use and equivalent technological characteristics, performance, and safety profile as the legally marketed predicate device. Minor differences in implementation details (e.g., data collection methods, user interface design, age range extension) do not raise new questions of safety or effectiveness.
Therefore, Neurora is substantially equivalent to Cognoa ASD Diagnosis Aid (DEN200069) and may be marketed for its intended use.
### Performance Data [807.92(b)]
Performance testing conducted to support a determination of substantial equivalence included nonclinical testing — software verification and validation, analytical validation, and cybersecurity assessment — and clinical testing, summarized below.
12
{17}
Argus Cognitive, Inc
Neurora - K260064
[LOGO]
ARGUS Cognitive
# Nonclinical Testing Summary [807.92(b)(1)]
## Software verification
Software verification was performed at the unit, integration, and system levels, in accordance with the FDA's General Principles of Software Validation and ISO 13485:2016. Unit-level testing evaluated individual software modules using black-box and white-box test strategies; integration testing evaluated the complete system with all components operating concurrently. System-level testing assessed analytical engine processing time, user interface response time, availability, and behavior under concurrent load: both audio and video analysis workflows completed well within the applicable 24-hour turnaround requirement; average response time (0.7 seconds) met the 5-second success criterion; the system achieved 100% uptime over a 30-day monitoring period; and the system remained stable under concurrent-load testing, with response times increasing under high-concurrency conditions.
## Analytical validation
Analytical validation assessed Neurora's individual algorithmic modules (vocalization/speaker diarization, facial action unit detection, and person/face detection) against benchmark datasets and an independently annotated human ground-truth dataset, using module-specific, pre-specified success criteria grounded in published literature. Inter-rater reliability of the ground-truth annotations was confirmed using Cohen's Kappa (≥0.7 across annotated event types). All evaluated modules met their pre-specified success criteria.
## Cybersecurity
All necessary cybersecurity testing was conducted on Neurora to support safe and secure operation as a web-based Software as a Medical Device. Security assessment activities were performed in accordance with FDA guidance "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions" (April 2022) and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" (October 2018).
Security testing demonstrated that Neurora implements appropriate security controls for a Tier 2 medical device. No critical or high-severity vulnerabilities were identified during testing. All security-critical functions including authentication, authorization, data encryption, and secure communications operate as intended.
Low-severity findings identified during testing were evaluated and determined to be acceptable within the device's Tier 2 classification and risk profile.
13
{18}
Argus Cognitive, Inc
Neurora - K260064
[LOGO]
ARGUS Cognitive
# Clinical Testing Summary [807.92(b)(2)]
## Clinical Studies
Clinical validation of Neurora's output consisted of a multicenter clinical validation study of the binary ASD classification and a supportive reader study evaluating clinician interaction with the Neurora report. The final user-facing report provides one of two binary results: "Consistent with ASD" or "Not Consistent with ASD."
## Multisite Clinical Validation Study
### Study Design
The primary clinical validation study was a multicenter, blinded, non-interventional study conducted at five clinical sites in the United States: Denver, Colorado; Dallas, Texas; Houston, Texas; Baton Rouge, Louisiana; and Phoenix, Arizona. The study evaluated Neurora in children aged 18 months through 12 years who were undergoing evaluation for ASD or developmental and behavioral concerns, together with typically developing comparison subjects.
The clinical validation population was independent of the data used for model development at both the subject and clinical-site levels. There was no overlap in subjects, clinical sites, or healthcare providers between the development datasets and the independent test set.
A total of 202 subjects were enrolled. Of these, 173 met the predefined clinical, recording-quality, assessment-completeness, reference-standard requirements, did not withdraw consent and were included in the primary diagnostic-performance analysis. The analyzed population included 136 subjects with an ASD reference diagnosis and 37 subjects in the combined Non-ASD reference group, comprising 22 subjects with a Non-ASD clinical condition and 15 typically developing subjects.
The analyzed population included 59 female and 114 male subjects. Mean chronological age was 60.60 months (SD 28.43), with a range of 18 to 144 months. 80 subjects were evaluated using the ADOS-2 Toddler Module or Module 1, 37 using Module 2, and 56 using Module 3.
### Clinical Reference Standard
The clinical reference diagnosis was established independently of Neurora by licensed psychologists with expertise in the evaluation and diagnosis of ASD, each with at least five years of experience in diagnosing autism.
The reference-standard assessment incorporated the Autism Diagnostic Observation Schedule, Second Edition (ADOS-2), developmental and medical history, caregiver information, direct clinical observation, and application of DSM-5 diagnostic criteria. The ADOS-2 classification alone did not determine the reference diagnosis. One final diagnosing clinician integrated the available clinical information and assigned the reference-standard outcome as ASD, Non-ASD clinical condition, or Typically developing.
14
{19}
Argus Cognitive, Inc
Neurora - K260064
[Non-Text]
ARGUS Cognitive
When the individual diagnostic inputs were not concordant, the diagnosing clinician reviewed all available information, obtained additional information when clinically indicated, and applied DSM-5 criteria to determine the final diagnosis. The reference diagnosis was finalized and locked before the clinician or assisting personnel had access to any Neurora output, and the evaluation was conducted under double-blind conditions, such that Neurora also had no knowledge of the reference diagnosis. For evaluation of Neurora's binary classification, the Non-ASD clinical and Typically developing categories were combined into a single Non-ASD reference group.
### Primary Performance Endpoint
The primary performance endpoint evaluated the sensitivity and specificity of Neurora's binary output relative to the locked clinical reference diagnosis. The analysis population consisted of 173 subjects with evaluable Neurora results and an established clinical reference diagnosis. Neurora generated a valid binary classification for all 173 subjects in the analysis population, corresponding to a no-result rate of 0%.
### Clinical Performance Results
Overall sensitivity was 70.59% (95% CI: 63.08-78.13%), and specificity was 83.78% (95% CI: 70.27-94.87%), and balanced accuracy was 77.19% (95% CI: 69.62-84.03%).
Positive predictive value was 94.12% (95% CI: 89.11–98.15%), negative predictive value was 43.66% (95% CI: 32.05–55.26%). These predictive values reflect the 78.61% ASD prevalence of the validation population and should be interpreted in that context. At a prevalence of 70.9%, representative of children referred for ASD evaluation in U.S. clinical settings [Klaiman, C., White, S., Richardson, S. et al. Expert Clinician Certainty in Diagnosing Autism Spectrum Disorder in 16–30-Month-Olds: A Multi-site Trial Secondary Analysis. J Autism Dev Disord 54, 393–408 (2024). https://doi.org/10.1007/s10803-022-05812-8], prevalence-adjusted positive and negative predictive values are 91.38% and 53.90%, respectively, calculated from Neurora's observed sensitivity and specificity.
When compared to the predicate, Cognoa (CanvasDx), the sensitivity and specificity of the Neurora device are higher than of the predicate device.
Table 1. Comparison in sensitivity and specificity between Neurora and the predicative device.
| Population | SensitivityMean (n/N)95% CI | SpecificityMean (n/N)95% CI |
| --- | --- | --- |
| Cognoa (CanvasDx (mITD))N=425 | 51.6% (63/122)42.8% - 60.5% | 18.5% (56/303)14.3% - 23.3%) |
| NeuroraN=173 | 70.6% (96/136)63.1%-78.1% | 83.8% (31/37)70.3-94.9% |
## Safety
15
{20}
Argus Cognitive, Inc
Neurora - K260064
[LOGO]
ARGUS Cognitive
The clinical validation study was non-interventional. The Neurora output was not used to establish the clinical reference diagnosis or to direct treatment during the study. Neurora is intended to provide adjunctive information to qualified healthcare professionals and is not intended to establish or exclude an ASD diagnosis independently of a comprehensive clinical evaluation. No device-related adverse events or complications were reported during the clinical validation study.
## Summary
The independent multisite clinical validation study demonstrated an overall sensitivity of 70.59% and specificity of 83.78% for Neurora's binary ASD classification, with a valid result generated for every subject in the analysis population. The predefined binary-classification acceptance criterion based on comparison with the predicate device's whole-population performance estimates was met.
Taken together, the clinical validation results support use of Neurora's binary classification as an adjunct to, and not a replacement for, qualified clinical judgment in the evaluation of patients aged 18 months through 12 years who are being assessed for ASD.
## CONCLUSION AND SUBSTANTIAL EQUIVALENCE STATEMENT
Neurora is substantially equivalent to the predicate device based on:
1. Similar Intended Use: Both provide adjunctive diagnostic support for ASD in pediatric populations to healthcare providers in clinical settings
2. Similar Fundamental Scientific Technology: Both employ machine-learning algorithms to analyze behavioral information and generate a diagnostic classification for ASD, based on video-derived and/or clinician- and caregiver-reported inputs
3. Performance Characteristics: Neurora's specificity is comparable to the predicate's reported performance (83.78% vs. 78.87%). Sensitivity is lower than the predicate's determinate-subset figure (70.59% vs. 98.44%) but exceeds the predicate's performance when evaluated on an all-completers basis, consistent with Neurora providing a classification for 100% of assessed patients rather than the predicate's 31.8%. These differences do not raise new questions of safety or effectiveness.
4. Equivalent Safety Profile: Both operate as software-only devices with minimal risk profile due to adjunctive labeling and clinical oversight requirements
The technological enhancements in Neurora (multi-modal analysis including audio, extended age range, advanced ML architectures) do not raise different questions of safety and effectiveness. These enhancements provide improved clinical utility while maintaining the fundamental operating principles and risk profile of the predicate device.
16
{21}
Argus Cognitive, Inc
Neurora - K260064
[LOGO]
ARGUS Cognitive
Verification, validation, cybersecurity, and clinical testing demonstrate that Neurora is safe and effective for its intended use and is substantially equivalent to the predicate device.
17
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.