15 (American Board of Radiology and MQSA-certified radiologists)
Indications for Use
Lunit INSIGHT DBT is a computer-assisted detection and diagnosis (CADe/x) software intended to be used concurrently by interpreting physicians to aid in the detection and characterization of suspected lesions for breast cancer in digital breast tomosynthesis (DBT) exams from compatible DBT systems. Through the analysis, the regions of soft tissue lesions and calcifications are marked with an abnormality score indicating the likelihood of the presence of malignancy for each lesion. Lunit INSIGHT DBT uses screening mammograms of the female population. Lunit INSIGHT DBT is not intended as a replacement for a complete interpreting physician’s review or their clinical judgment that takes into account other relevant information from the image or patient history.
Device Story
Software as a Medical Device (SaMD) using deep learning AI to analyze digital breast tomosynthesis (DBT) slices; identifies and marks regions of soft tissue lesions and calcifications; provides abnormality scores for malignancy likelihood. Used in clinical settings by radiologists to augment diagnostic workflow. Outputs include lesion location, type, abnormality scores, and optional pre-populated reports. Features include current-prior comparison and volumetric breast density integration. Clinicians use output to assist in detection and characterization of breast cancer; device does not replace clinical judgment. Benefits include improved diagnostic accuracy and efficiency in screening.
Clinical Evidence
Standalone performance testing on 3,277 DBT exams from US institutions. Primary endpoint: ROC AUC 0.9388 (95% CI: 0.9304, 0.9472), p < 0.05. Secondary metrics: JAFROC AUC 0.9206; sensitivity 91.11% and specificity 77.62% at default threshold (0.1). Comparative analysis against 15 radiologists (258 cases) showed AI standalone AUROC (0.9430) significantly superior to average reader (0.8983, p=0.0083). Lesion type agreement 75.61%.
Technological Characteristics
AI/ML-based software algorithm using deep learning. Compatible with Hologic, GE, Siemens, and Fujifilm DBT systems. Operates on standard clinical hardware. Complies with IEC 62304 (software lifecycle) and IEC 62366-1 (usability).
Indications for Use
Indicated for female patients undergoing screening mammography to aid interpreting physicians in the detection and characterization of suspected breast cancer lesions in digital breast tomosynthesis (DBT) exams. Not intended to replace physician review or clinical judgment.
Regulatory Classification
Identification
A radiological computer-assisted detection and diagnostic software is an image processing device intended to aid in the detection, localization, and characterization of fracture, lesions, or other disease-specific findings on acquired medical images (e.g., radiography, magnetic resonance, computed tomography). The device detects, identifies, and characterizes findings based on features or information extracted from images, and provides information about the presence, location, and characteristics of the findings to the user. The analysis is intended to inform the primary diagnostic and patient management decisions that are made by the clinical user. The device is not intended as a replacement for a complete clinician's review or their clinical judgment that takes into account other relevant information from the image or patient history.
Special Controls
A radiological computer assisted detection and diagnosis software must comply with the following special controls: Design verification and validation must include: 1. i. A detailed description of the image analysis algorithm, including but not limited to a description of the algorithm inputs and outputs, each major component or block, how the algorithm and output affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide improved assisted-read detection and diagnostic performance as intended in the indicated user population(s), and to characterize the standalone device performance for labeling. Performance testing includes standalone test(s), side-by-side comparison(s), and/or a reader study, as applicable. iii. Results from standalone performance testing used to characterize the independent performance of the device separate from aided user performance. The performance assessment must be based on appropriate diagnostic accuracy measures (e.g., receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Devices with localization output must include localization accuracy testing as a component of standalone testing. The test dataset must be representative of the typical patient population with enrichment made only to ensure that the test dataset contain a sufficient number of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, concomitant disease, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment. iv. Results from performance testing that demonstrate that the device provides improved assisted-read detection and/or diagnostic performance as intended in the indicated user population(s) when used in accordance with the instructions for use. The reader population must be comprised of the intended user population in terms of but not limited to clinical training, certification, and years of experience. The performance assessment must be based on appropriate diagnostic accuracy measures (e.g., receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Test datasets must meet the requirements described in 1(iii) above. v. Appropriate software documentation, including device hazard analysis, software requirements specification document, software design specification document, traceability analysis, system level test protocol, pass/fail criteria, testing results, and cybersecurity measures. 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the device instructions for use, including the intended reading protocol and how the user should interpret the device output. iii. A detailed description of the intended user, and any user training materials as programs that addresses appropriate reading protocols for the device to ensure that the end user is fully aware of how to interpret and apply the device output. iv. A detailed description of the device inputs and outputs. v. A detailed description of compatible imaging hardware and imaging protocols. vi. Warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality or for certain subpopulations), as applicable. vii. A detailed summary of the performance testing, including: test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as anatomical characteristics, patient demographics and medical history, user experience, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the image analysis algorithm, including a description of the algorithm inputs and outputs, each major component or block, how the algorithm and output affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide improved assisted-read detection and diagnostic performance as intended in the indicated user population(s), and to characterize the standalone device performance for labeling. Performance testing includes standalone test(s), side-by-side comparison(s), and/or a reader study, as applicable.
(iii) Results from standalone performance testing used to characterize the independent performance of the device separate from aided user performance. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Devices with localization output must include localization accuracy testing as a component of standalone testing. The test dataset must be representative of the typical patient population with enrichment made only to ensure that the test dataset contains a sufficient number of cases from important cohorts (*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, concomitant disease, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment.(iv) Results from performance testing that demonstrate that the device provides improved assisted-read detection and/or diagnostic performance as intended in the indicated user population(s) when used in accordance with the instructions for use. The reader population must be comprised of the intended user population in terms of clinical training, certification, and years of experience. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Test datasets must meet the requirements described in paragraph (b)(1)(iii) of this section.(v) Appropriate software documentation, including device hazard analysis, software requirements specification document, software design specification document, traceability analysis, system level test protocol, pass/fail criteria, testing results, and cybersecurity measures.
(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use.
(ii) A detailed description of the device instructions for use, including the intended reading protocol and how the user should interpret the device output.
(iii) A detailed description of the intended user, and any user training materials or programs that address appropriate reading protocols for the device, to ensure that the end user is fully aware of how to interpret and apply the device output.
(iv) A detailed description of the device inputs and outputs.
(v) A detailed description of compatible imaging hardware and imaging protocols.
(vi) Warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality or for certain subpopulations), as applicable.(vii) A detailed summary of the performance testing, including test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as anatomical characteristics, patient demographics and medical history, user experience, and imaging equipment.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 26, 2026
Lunit, Inc.
Hyeseung Yoo
Sr. RA Specialist, RAC-Devices
4-8f, 374, Gangnam-Daero, Gangnam-Gu
Seoul, 06241
Republic Of Korea
Re: K253796
Trade/Device Name: Lunit INSIGHT DBT (V1.2)
Regulation Number: 21 CFR 892.2090
Regulation Name: Radiological Computer-Assisted Detection And Diagnosis Software
Regulatory Class: Class II
Product Code: QDQ
Dated: November 28, 2025
Received: November 28, 2025
Dear Hyeseung Yoo:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K253796 - Hyeseung Yoo
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253796 - Hyeseung Yoo
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# YANNA S. KANG -S
Yanna Kang, Ph.D.
Assistant Director
Mammography and Ultrasound Team
DHT8C: Division of Radiological Imaging
and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253796 | |
| Device Name Lunit INSIGHT DBT (V1.2) | |
| Indications for Use (Describe) Lunit INSIGHT DBT is a computer-assisted detection and diagnosis (CADe/x) software intended to be used concurrently by interpreting physicians to aid in the detection and characterization of suspected lesions for breast cancer in digital breast tomosynthesis (DBT) exams from compatible DBT systems. Through the analysis, the regions of soft tissue lesions and calcifications are marked with an abnormality score indicating the likelihood of the presence of malignancy for each lesion. Lunit INSIGHT DBT uses screening mammograms of the female population. Lunit INSIGHT DBT is not intended as a replacement for a complete interpreting physician’s review or their clinical judgment that takes into account other relevant information from the image or patient history. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Page 1/8
# K253796
# 510(k) Summary
# Lunit INSIGHT DBT v.1.2
This 510(k) summary of safety and effectiveness information is prepared in accordance with the requirements of 21 CFR §807.92.
# 1. Submitter
| Applicant Information | Lunit Inc.
4-8 F, 374, Gangnam-daero, Gangnam-gu,
Seoul, 06241, Republic of Korea
Tel: +82-2-2138-0827
Fax: +82-2-6919-2702 |
| --- | --- |
| Primary Correspondent | Hyeseung Yoo,
Sr. Regulatory Affairs Specialist
Email: hsyoo@lunit.io |
| Secondary Correspondents | Juyoung Jung,
Regulatory Affairs Specialist
Email: jjyoung@lunit.io
Sulgue Choi
Regulatory Affairs Team Leader
Email: sulgue@lunit.io |
| Date Prepared | November 28, 2025 |
# 2. Device Names and Classifications
Subject Device
| Name of Device | Lunit INSIGHT DBT (V1.2) |
| --- | --- |
| Version | 1.2 |
| Classification Name | Radiological Computer-Assisted Detection And Diagnosis Software |
| Regulation | 21 CFR 892.2090 |
| Classification | Class II |
| Product code | QDQ |
{5}
Page 2/8
Lunit
Lunit Inc.
4-8 F, 374, Gangnam-daero, Gangnam-gu,
Seoul, 06241, Republic of Korea
www.lunit.io
# Predicate Device
| Name of Device | Lunit INSIGHT DBT |
| --- | --- |
| Version | v1.1 |
| Legal manufacturer | Lunit Inc. |
| 510(k) number | K242652 |
| Classification Name | Radiological Computer Assisted Detection/Diagnosis Software For Suspicious Lesions For Cancer |
| Regulation | 21 CFR 892.2090 |
| Classification | Class II |
| Product code | QDQ |
# 3. Device Description
Lunit INSIGHT DBT is a computer-assisted detection/diagnosis (CADe/x) Software as a Medical Device that provides information about the presence, location and characteristics of lesions suspicious for breast cancer to assist interpreting physicians in making diagnostic decisions when reading digital breast tomosynthesis (DBT) images. The software automatically analyzes digital breast tomosynthesis slices via artificial intelligence technology that has been trained via deep learning.
For each DBT case, Lunit INSIGHT DBT generates artificial intelligence analysis results that include the lesion type, location, lesion-level/case-level score, and outline of the regions suspected of breast cancer. This peripheral information intends to augment the physician's workflow to better aid in detection and diagnosis of breast cancer.
# 4. Indication for Use
Lunit INSIGHT DBT is a computer-assisted detection and diagnosis (CADe/x) software intended to be used concurrently by interpreting physicians to aid in the detection and characterization of suspected lesions for breast cancer in digital breast tomosynthesis (DBT) exams from compatible DBT systems. Through the analysis, the regions of soft tissue lesions and calcifications are marked with an abnormality score indicating the likelihood of the presence of malignancy for each lesion. Lunit INSIGHT DBT uses screening mammograms of the female population.
Lunit INSIGHT DBT is not intended as a replacement for a complete interpreting physician's review or their clinical judgment that takes into account other relevant information from the image or patient history.
{6}
Page 3/8
Lunit®
Lunit Inc.
4-8 F, 374, Gangnam-daero, Gangnam-gu,
Seoul, 06241, Republic of Korea
www.lunit.io
# 5. Summary of Substantial Equivalence
| Item | Subject Device | Predicate Device |
| --- | --- | --- |
| | Lunit INSIGHT DBT v1.2 | Lunit INSIGHT DBT v1.1 |
| Classification Name | Radiological Computer Assisted Detection/Diagnosis Software For Suspicious Lesions For Cancer | Radiological Computer Assisted Detection/Diagnosis Software For Suspicious Lesions For Cancer |
| Regulation | 21 CFR 892.2090 | 21 CFR 892.2090 |
| Regulatory Class | Class II | Class II |
| Product Code | QDQ | QDQ |
| Indication for Use | Lunit INSIGHT DBT is a computer-assisted detection and diagnosis (CADe/x) software intended to be used concurrently by interpreting physicians to aid in the detection and characterization of suspected lesions for breast cancer in digital breast tomosynthesis (DBT) exams from compatible DBT systems. Through the analysis, the regions of soft tissue lesions and calcifications are marked with an abnormality score indicating the likelihood of the presence of malignancy for each lesion. Lunit INSIGHT DBT uses screening mammograms of the female population.
Lunit INSIGHT DBT is not intended as a replacement for a complete interpreting physician’s review or their clinical judgment that takes into account other relevant information from the image or patient history. | Lunit INSIGHT DBT is a computer-assisted detection and diagnosis (CADe/x) software intended to be used concurrently by interpreting physicians to aid in the detection and characterization of suspected lesions for breast cancer in digital breast tomosynthesis (DBT) exams from compatible DBT systems. Through the analysis, the regions of soft tissue lesions and calcifications are marked with an abnormality score indicating the likelihood of the presence of malignancy for each lesion. Lunit INSIGHT DBT uses screening mammograms of the female population.
Lunit INSIGHT DBT is not intended as a replacement for a complete interpreting physician’s review or their clinical judgment that takes into account other relevant information from the image or patient history. |
| Target patient population | Women undergoing mammography | Women undergoing mammography |
| Intended user | Physicians interpreting screening mammograms | Physicians interpreting screening mammograms |
| Input Image Source | DBT | DBT |
| Fundamental Technological Basis | Lunit INSIGHT DBT is powered by artificial intelligence/machine learning-based software algorithm | Lunit INSIGHT DBT is powered by artificial intelligence/machine learning-based software algorithm |
{7}
Page 4/8
Lunit
Lunit Inc.
4-8 F, 374, Gangnam-daero, Gangnam-gu,
Seoul, 06241, Republic of Korea
www.lunit.io
## 6. Comparison with Predicate Device
The subject device, Lunit INSIGHT DBT v1.2, maintains the same indications for use and core technological characteristics as the predicate device, Lunit INSIGHT DBT v1.1 (K242652). Both devices are radiological computer assisted detection and diagnostic software and use artificial intelligence technologies and deep learning techniques to fulfill its intended purpose to detect and characterize lesions suspected of breast cancer. Both devices analyze DBT scans and outputs of both devices augments the interpreting physicians in the diagnosis of asymptomatic patients.
The primary modifications in Lunit INSIGHT DBT v1.2 include expanded compatibility with additional imaging modalities, enabling the software to process input images from Siemens and Fujifilm systems in addition to those from Hologic and GE Healthcare. Lunit INSIGHT DBT v1.2 also introduces an ordinal Case Abnormality Level output that presents the abnormality level of a case using pre-defined likelihood categories for suspicious findings. Pre-populated report feature can be enabled to automatically generate a report for Case Abnormality Level based on the minimum likelihood level.
The new version further includes user-selected threshold operating points, enabling clinicians to choose from two additional sensitivity levels to help reduce false positive cases as appropriate. Additional auxiliary functions include a Current-Prior Comparison capability for reviewing interval changes over multiple years of patient imaging, and an optional integration with external software to display volumetric breast density information for each breast. These modifications do not change the intended use or the fundamental scientific technology of the device.
## 7. Performance Data
### 7.1. Non-clinical Testing Summary
Testing was conducted in accordance with Lunit’s design control processes and in compliance with the following FDA-recognized consensus standards:
- IEC 62304: 2006/A1: 2016, Medical device software – software life-cycle processes
- IEC 62366-1:2015+AMD1:2020, Medical devices – Part 1: Application of usability engineering to medical devices.
Based on results of verification, Lunit INSIGHT DBT demonstrated that it fulfilled the software requirements.
{8}
Page 5/8
# 7.2. Performance Testing
Standalone performance tests were conducted to demonstrate substantial equivalence with the predicate device. Total of 3,277 DBT exams of female adults were collected at multiple imaging facilities in the US healthcare institutions to broadly cover the US population and maintain balanced demographic and cancer characteristic distributions. DBT images were obtained from Hologic, GE Healthcare, Siemens, and FujiFilm 3D mammography equipment.
For the reference standard, each ground truther classified each DBT exam into non-cancer group or cancer group, then annotated the malignant lesion location in the 3D planes of cancer cases. The effectiveness of standalone performance in detection and diagnosis of breast cancer in 3D mammography was examined as comparing the results between standalone test and the ground truthing on each DBT exam. The scoring method that calculates the intersection-over-union (IoU) of the reference standards' ROI and heatmap region detected by the device, only when they are aligned along the same z-axis, indicating their presence in the same slice was implemented. In addition, the device performance for the characterization of the lesion type was evaluated by comparing the device analysis results with the ground truther's interpretation for the lesion type (mass, calcification, or mixed).
The primary goal of this standalone performance test was to demonstrate that the lower bound of $95\%$ CI of device's ROC AUC in standalone performance was greater than 0.903 and p-value was less than the significance level of $5\%$ (0.05). ROC AUC in the standalone performance analysis was 0.9388 (95% CI: 0.9304, 0.9472) with statistical significance $(p < 0.05)$ . Thus, the primary endpoint was achieved.
For the secondary endpoints, the result of the JAFROC AUC was 0.9206 (95% CI: 0.9117, 0.9295). Sensitivity at the default operating point (0.1) was $91.11\%$ (95% CI: 89.66, 92.57) and specificity was $77.62\%$ (95% CI: 75.70, 79.54), respectively. Sensitivity at the supplementary '0.3' operating point was $88.38\%$ (95% CI: 86.74, 90.02) and specificity was $83.68\%$ (95% CI: 81.98, 85.38), respectively. Sensitivity at the supplementary '0.6' point was $81.48\%$ (95% CI: 79.49, 83.47) and specificity was $93.44\%$ (95% CI: 92.30, 94.58), respectively. For lesion type agreement analysis, the matching proportion of the agreement between CAD and ground truther in 3-way classification of lesion type agreement was $75.61\%$ (95% CI: 73.40, 77.80).
# 7.2.1 Demographic distribution
To broadly cover the US population, the data has been comprised of various demographic and clinical information. All clinical data including patients' demographic information such as age, ethnicity, race as well as previous breast cancer history was collected from imaging facilities in the United States.
For baseline demographics information, a total of 3277 cases are female with a mean age of $60.94 \pm 13.99$ . Among the cases with available ethnicity information, the majority were categorized as 'Not Hispanic or Latino' (617, $18.83\%$ ). 2462 cases ( $75.13\%$ ) are 'White', and 647 cases ( $19.74\%$ ) are 'Non-White' in race category.
{9}
Page 6/8
# 7.2.2 Clinical subgroups and confounders present in the dataset
- Breast composition categorized according to the density categories defined by American College of Radiology BI-RADS.
- Distribution of BI-RADS assessment categories, with BI-RADS 0 representing the largest proportion of cases, followed (in decreasing frequency) by BI-RADS 1, 4, 5, 2, and 3.
- Lesion type categorized as "mass only," "calcification only," or "both."
- Cancer type categorized as invasive breast cancer (including invasive ductal carcinoma [IDC] and invasive lobular carcinoma [ILC]) and non-invasive breast cancer (including ductal carcinoma in situ [DCIS]).
- Slice thickness distribution including cases with 1 mm slice thickness and cases with slice thickness less than 1 mm.
# 7.2.3 Equipment information
Out of the total 3277 cases, 1617 (49.34%) DBT exams were taken using Hologic equipment, 585 (17.85%) DBT exams were taken using GE Healthcare equipment, 469 (14.31%) DBT exams were taken using Siemens equipment and 606 (18.49%) DBT exams were taken using Fujifilm equipment
# 7.2.4 Truthing process
After completion of the dataset screening, each exam will have its own ground truthing by expert breast imaging radiologists who called as a 'Ground Truther' in the study. The ground truthers define the reference standard for every DBT exam enrolled in the study. Depending on the dataset, ground truthing will be conducted by either two or three qualified breast imaging radiologists following the same methodology as described in the following.
In datasets where three ground truthers are involved, two ground truthers independently perform the initial review, and the final truther, who is the most experienced, determines the final reference standard considering the results of the other two.
In datasets where two ground truthers are involved, the first truther independently completes the review, and the final truther, who is more experienced, makes the final decision considering the results of the other truther.
Each ground truther classified each DBT exam into non-cancer group or cancer group [STEP A] then annotated the malignant lesion location in the 3D planes of cancer cases [STEP B].
To set the reference standard, the ground truther reviewed the collected study exams using relevant clinical supporting data such as radiology reports and pathology reports acquired from the investigational institution and defined the reference standard based on the radiologic and pathologic clinical evidence. Especially for the biopsy-proven cancer exams, the ground truther can refer to the relevant pathology report containing
{10}
Page 7/8
Lunit®
Lunit Inc.
4-8 F, 374, Gangnam-daero, Gangnam-gu,
Seoul, 06241, Republic of Korea
www.lunit.io
the cancer characteristic information (i.e., cancer location, size, shape, presence of calcification, pathologic results, etc.) for the ground truthing.
## 7.2.5 Independence of test data from training data
The test set used for the clinical validation was completely independent from the datasets used for training, tuning, or calibrating the algorithm.
## 7.2.6 AI Standalone Vs. Unaided Reader Performance Comparison
In an additional comparative analysis to evaluate the AI standalone performance against the average unaided reader performance, a testing dataset consisting of 258 cases (128 negative cases, 65 benign cases, and 65 cancer cases) with 4 views was utilized. The reading panel included 15 American Board of Radiology and MQSA-certified radiologists.
The AI standalone performance was compared with the average radiologist using the Obuchowski-Rockette (OR) method for a single-treatment, random-reader random-case (RRRC) CAD-vs-radiologist design. The results demonstrated that the AI standalone AUROC (0.9430) was significantly superior to the average reader AUROC (0.8983), with a difference of 0.0446 (95% CI: [0.0115, 0.0777], p = 0.0083).
Even at the highest threshold (Score 60), the AI standalone sensitivity was 86.2%, which is equivalent to the average reader's sensitivity of 85.4% (p = 0.8912). Furthermore, the AI standalone specificity at this threshold was 95.9%, demonstrating a statistically significant improvement over the average reader's specificity of 77.3% (p < 0.001). These results conclude that the ROC AUC of the device is consistently and significantly superior to the average radiologist across thresholds.
## 7.2.7 Statistical Methods for Confidence Intervals
To ensure accurate assessment of the device's reliability, all confidence intervals reported for performance metrics in this summary were calculated using statistically appropriate methods that do not rely on normal distribution assumptions. Specifically, AUROC confidence intervals were estimated using the stratified bootstrap percentile method to provide robust interval estimation. For binomial proportion metrics, including sensitivity, specificity, and lesion agreement, the Clopper-Pearson exact method was utilized. Additionally, the confidence interval for AFROC AUC was calculated using the Dorfman-Berbaum-Metz (DBM) method for a Fixed-Reader Random-Case (FRRC) variance analysis.
## 8. Assessment of Benefit-Risk, General Safety and Effectiveness
Risk management of the subject device is conducted via hazard analysis which identifies and mitigates existing and potential hazards. Hazards were controlled throughout the software lifecycle with control measures with
{11}
Page 8/8
Lunit®
Lunit Inc.
4-8 F, 374, Gangnam-daero, Gangnam-gu,
Seoul, 06241, Republic of Korea
www.lunit.io
regards to software development, verification, and validation. Furthermore, labeling information consists of instructions for use with necessary cautionary statements for safe and effective use of the software. Lunit finds the use of the software has a positive balance in terms of probable benefits versus foreseeable and identified risks.
# 9. Conclusion
Lunit INSIGHT DBT v1.2 is substantially equivalent to the predicate device because it has the same intended use and shares the same technological and performance characteristics. The newly introduced features do not change the device's intended use and do not raise new questions of safety or effectiveness. In addition, performance testing demonstrates that the Lunit INSIGHT DBT v.1.2 is as safe and effective as the predicate device in detecting suspicious lesions in DBT exams from compatible DBT systems. Therefore, substantial equivalence has been established.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.