K251293 · Icardio.Ai · QUO · Nov 21, 2025 · Cardiovascular
Device Facts
Record ID
K251293
Device Name
CardioVision
Applicant
Icardio.Ai
Product Code
QUO · Cardiovascular
Decision Date
Nov 21, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2200
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K251293 · Nov 21, 2025
CardioVision
Icardio.Ai
Retrospective clinical echocardiography studies from 12 independent clinical sites
The device performance was validated using a retrospective cohort of 650 clinical echocardiography studies to assess sensitivity and specificity for detecting moderate or severe aortic stenosis.
Retrospective multi-center performance study; Retrospective, multi-center performance study
Adults over 21 years of age undergoing echocardiographic exams; Sample Size: 650 echocardiography studies from 608 subjects; Number of Sites: 12
Not applicable for this study
Sensitivity, Specificity, and AUROC for detecting moderate or severe aortic stenosis
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Aortic stenosis
Convolutional neural networks
—
Sensitivity 0.896, Specificity 0.872, AUROC 0.945
—
—
Retrospective, multi-center performance study: 650 echocardiography studies from 608 subjects
>1 (Level III echocardiographers)
Indications for Use
The iCardio.ai CardioVision™ AI is an automated machine learning-based decision support system, indicated as a diagnostic aid for patients undergoing an echocardiographic exam consisting of a single PLAX view in an outpatient environment, such as a primary care setting. When utilized by an interpreting clinician, this device provides information that may be useful in detecting moderate or severe aortic stenosis. iCardio.ai CardioVision™ AI is indicated in adult populations over 21 years of age. Patient management decisions should not be made solely on the results of the iCardio.ai CardioVision™ AI analysis. iCardio.ai CardioVision™ AI analyzes a single cine-loop DICOM of the parasternal long axis (PLAX).
Device Story
CardioVision™ is a standalone software for echocardiography review; inputs DICOM-compliant echocardiogram studies (specifically parasternal long-axis (PLAX) cine-loops). Uses convolutional neural networks (CNNs) to estimate likelihood of moderate/severe aortic stenosis; outputs binary classification (none/mild vs. moderate/severe) or indeterminate. Used in outpatient/primary care settings by interpreting clinicians. Integrates with PACS/DICOM viewers. Output serves as adjunctive diagnostic aid; clinicians combine results with patient history/physical exams to inform management. Benefits include automated screening support for aortic stenosis detection.
Clinical Evidence
Retrospective, multi-center study of 650 echocardiography studies (608 subjects) from 12 US sites. Performance compared against board-certified Level III echocardiographer ground truth. Results: AUROC 0.945; Sensitivity 89.6% (95% CI: 84.3-93.2%); Specificity 87.2% (95% CI: 83.8-90.0%). 1.077% rejection rate for poor image quality. Subgroup analyses confirmed consistent performance across age, BMI, and ultrasound manufacturers.
Technological Characteristics
Standalone software; DICOM-based input; CNN-based algorithm; operates on standard PC/server infrastructure; integrates with PACS/DICOM viewers. No hardware components. Fixed (non-adaptive) model architecture.
Indications for Use
Indicated for adult patients >21 years undergoing echocardiographic exams (single PLAX view) in outpatient/primary care settings to aid in detecting moderate or severe aortic stenosis. Not for independent therapy direction.
Regulatory Classification
Identification
The adjunctive cardiovascular status indicator is a prescription device based on sensor technology for the measurement of a physical parameter(s). This device is intended for adjunctive use with other physical vital sign parameters and patient information and is not intended to independently direct therapy.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Software description, verification, and validation based on comprehensive hazard analysis must be provided, including:
(i) Full characterization of technical parameters of the software, including any proprietary algorithm(s);
(ii) Description of the expected impact of all applicable sensor acquisition hardware characteristics on performance and any associated hardware specifications;
(iii) Specification of acceptable incoming sensor data quality control measures; and
(iv) Mitigation of impact of user error or failure of any subsystem components (signal detection and analysis, data display, and storage) on accuracy of patient reports.
(2) Scientific justification for the validity of the status indicator algorithm(s) must be provided. Verification of algorithm calculations and validation testing of the algorithm using a data set separate from the training data must demonstrate the validity of modeling.
(3) Usability assessment must be provided to demonstrate that risk of misinterpretation of the status indicator is appropriately mitigated.
(4) Clinical data must be provided in support of the intended use and include the following:
(i) Output measure(s) must be compared to an acceptable reference method to demonstrate that the output measure(s) represent(s) the predictive measure(s) that the device provides in an accurate and reproducible manner;
(ii) The data set must be representative of the intended use population for the device. Any selection criteria or limitations of the samples must be fully described and justified;
(iii) Agreement of the measure(s) with the reference measure(s) must be assessed across the full measurement range; and
(iv) Data must be provided within the clinical validation study or using equivalent datasets to demonstrate the consistency of the output and be representative of the range of data sources and data quality likely to be encountered in the intended use population and relevant use conditions in the intended use environment.
(5) Labeling must include the following:
(i) The type of sensor data used, including specification of compatible sensors for data acquisition;
(ii) A description of what the device measures and outputs to the user;
(iii) Warnings identifying sensor reading acquisition factors that may impact measurement results;
(iv) Guidance for interpretation of the measurements, including warning(s) specifying adjunctive use of the measurements;
(v) Key assumptions made in the calculation and determination of measurements;
(vi) The measurement performance of the device for all presented parameters, with appropriate confidence intervals, and the supporting evidence for this performance; and
(vii) A detailed description of the patients studied in the clinical validation (
*e.g.,* age, gender, race/ethnicity, clinical stability) as well as procedural details of the clinical study.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 21, 2025
iCardio.ai
Joseph Sokol
CEO
8601 Beverly Blvd
Floor 2
West Hollywood, California 90048
Re: K251293
Trade/Device Name: CardioVision™
Regulation Number: 21 CFR 870.2200
Regulation Name: Adjunctive Cardiovascular Status Indicator
Regulatory Class: Class II
Product Code: QUO
Dated: April 25, 2025
Received: April 25, 2025
Dear Joseph Sokol:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251293 - Roman Sandler
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251293 - Roman Sandler
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Stephen C. Browning -S
LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics and Monitoring Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251293 | |
| Device Name CardioVision™ | |
| Indications for Use (Describe) The iCardio.ai CardioVision™ AI is an automated machine learning-based decision support system, indicated as a diagnostic aid for patients undergoing an echocardiographic exam consisting of a single PLAX view in an outpatient environment, such as a primary care setting. When utilized by an interpreting clinician, this device provides information that may be useful in detecting moderate or severe aortic stenosis. iCardio.ai CardioVision™ AI is indicated in adult populations over 21 years of age. Patient management decisions should not be made solely on the results of the iCardio.ai CardioVision™ AI analysis. iCardio.ai CardioVision™ AI analyzes a single cine-loop DICOM of the parasternal long axis (PLAX). | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
{13}/{14}
CARDIO.ai
iCardio.ai CardioVision™
K251293 Traditional 510k Summary
Los Angeles, 2025
1. Summary
| Submitter's Name | iCardio.ai |
| --- | --- |
| Address | 8601 Beverly Blvd,
West Hollywood, CA 90048
Re: Joseph Sokol |
| Contact Person | Joseph Sokol |
| Title | CEO |
| Telephone Number | 310-614-7904 |
| Email | joseph@icardio.ai |
| Date Summary Prepared | November 21, 2025 |
| Device Proprietary Name | CardioVision™ |
| Model Number | V 1.0.0 |
| Common Name | CardioVision |
| Regulation Number | 21 CFR §870.2200 |
| Regulation Name | Adjunctive Cardiovascular Status Indicator |
| Product Code | QUO |
| Device Class | Class II |
| Predicate Device | Trade name: EchoGo™Heart Failure
Manufacturer: Ultromics Limited; 4630 Kingsgate Cascade Way, Oxford Business Park South, Oxford, Oxfordshire, United Kingdom, OX4 2SU
Regulation Number: 21 CFR §870.2200
Regulation Name: Adjunctive Cardiovascular Status Indicator
Device Class: Class II
Product Code: QUO
510(k) Number: K222463 510(k)
Clearance Date: November 23, 2022 |
iCardio.ai 510(K) Summary - Page 1 of 10 - 2025 iCardio.ai Corporation
{5}
CARDIQ.ai
iCardio.ai CardioVision™ K251293 Traditional 510k Summary
Los Angeles, 2025
# 2. Device Description
The iCardio.ai CardioVision™ AI is a standalone image analysis software developed by iCardio.ai Corporation, designed to assist in the review of echocardiography images. It is intended for adjunctive use with other physical vital sign parameters and patient information, but it is not intended to independently direct therapy. The device facilitates determining whether an echocardiographic exam is consistent with aortic stenosis (AS), by providing classification results that support clinical decision-making.
The iCardio.ai CardioVision™ AI takes as input a DICOM-compliant, partial or full echocardiogram study, which must include at least one parasternal long-axis (PLAX) view of the heart and at least one full cardiac cycle. The device uses a set of convolutional neural networks (CNNs) to analyze the image data and estimate the likelihood of moderate or severe aortic stenosis. The output consists of a binary classification of "none/mild" or "moderate/severe," indicating whether the echocardiogram is consistent with moderate or severe aortic stenosis. In cases where the image quality is insufficient, the device may output an "indeterminate" result.
The CNNs and their thresholds are fixed prior to validation and do not continuously learn during standalone testing. These models are coupled with pre- and post-processing functionalities, allowing the device to integrate seamlessly with pre-existing medical imaging workflows, including PACS, DICOM viewers, and imaging worklists. The iCardio.ai CardioVision™ AI is intended to be used as an aid in diagnosing AS, with the final diagnosis always made by an interpreting clinician, who should consider the patient's presentation, medical history, and additional diagnostic tests.
# 3. Indications for use:
The iCardio.ai CardioVision™ AI is an automated machine learning-based decision support system, indicated as a diagnostic aid for patients undergoing an echocardiographic exam consisting of a single PLAX view in an outpatient environment, such as a primary care setting.
When utilized by an interpreting clinician, this device provides information that may be useful in detecting moderate or severe aortic stenosis. iCardio.ai CardioVision™ AI is indicated in adult populations over 21 years of age. Patient management decisions should not be made solely on the results of the iCardio.ai CardioVision™ AI analysis. iCardio.ai CardioVision™ AI analyzes a single cine-loop DICOM of the parasternal long axis (PLAX).
iCardio.ai 510(K) Summary - Page 2 of 10 - 2025 iCardio.ai Corporation
{6}
CARDIQ.ai
iCardio.ai CardioVision™
K251293 Traditional 510k Summary
Los Angeles, 2025
# 4. Substantial Equivalence
As does the predicate device, EchoGo Heart Failure 1.0, the iCardio.ai CardioVision™ software works with digital imaging and communications in medicine (DICOM) echocardiography images to provide adjunctive information on a patient's cardiovascular condition. Both devices are machine learning-based software products that receive DICOM inputs, perform image processing operations, and return results to assist clinicians in diagnosing cardiovascular conditions. The iCardio.ai CardioVision™ AI specifically aids in the detection of aortic stenosis, while the EchoGo Heart Failure 1.0 aids in detecting heart failure with preserved ejection fraction (HFpEF). Both devices are intended to be used by interpreting clinicians as diagnostic aids, where patient management decisions should not be made solely on the results produced by either system.
iCardio.ai 510(K) Summary - Page 3 of 10 - 2025 iCardio.ai Corporation
{7}
#
CARDIO.ai
iCardio.ai CardioVision™
K251293 Traditional 510k Summary
Los Angeles, 2025
| Characteristic | Subject Device: iCardio.ai CardioVision | Predicate Device: EchoGo Heart Failure (K222463) |
| --- | --- | --- |
| Regulation | 21 CFR 870.2200 | 21 CFR 870.2200 |
| Product Code | QUO | QUO |
| Intended Use | Providing adjunctive information on a patient's cardiovascular condition (diagnostic aid for aortic stenosis). | Providing adjunctive information on a patient's cardiovascular condition (diagnostic aid for Heart Failure with Preserved Ejection Fraction (HFpEF)). |
| Indications for use | The iCardio.ai CardioVision™ AI is an automated machine learning-based decision support system, indicated as a diagnostic aid for patients undergoing limited screening echocardiographic exams in an outpatient environment, such as a primary care setting. When utilized by an interpreting clinician, this device provides information that may be useful in detecting moderate or severe aortic stenosis. iCardio.ai CardioVision™ AI is indicated in adult populations over 21 years of age. Patient management decisions should not be made solely on the results of the iCardio.ai CardioVision™ AI analysis. iCardio.ai CardioVision™ AI takes as input a single cine-loop dicom of the parasternal long axis (PLAX) view of the heart, or a full echocardiographic study that includes a PLAX cine loop. | EchoGo Heart Failure 1.0 is an automated machine learning-based decision support system, indicated as a diagnostic aid for patients undergoing routine functional cardiovascular assessment using echocardiography. When utilized by an interpreting clinician, this device provides information that may be useful in detecting heart failure with preserved ejection fraction (HFpEF). EchoGo Heart Failure 1.0 is indicated in adult populations over 25 years of age. Patient management decisions should not be made solely on the results of the EchoGo Heart Failure 1.0 analysis. EchoGo Heart Failure 1.0 takes as input an apical 4-chamber view of the heart that has been captured and assessed to have an ejection fraction ≥50%. |
| Anatomical Site | Identical | Cardiovascular |
| Tissue | Tissue | Tissue |
iCardio.ai 510(K) Summary - Page 4 of 10 - 2025 iCardio.ai Corporation
{8}
#
CARDIO.ai
iCardio.ai CardioVision™
K251293 Traditional 510k Summary
Los Angeles, 2025
| Modality | Identical | Ultrasound |
| --- | --- | --- |
| Intended Users | Identical | The clinician interpreting the report produced by EchoGo Heart Failure 1.0 and making a diagnostic decision. |
| Intended Patient Population | Adults over age 21 | Adults over age 25 |
| Hardware Component? | No | No |
| Machine learning-based algorithm | Yes | Yes |
| Operates on DICOM clips | Yes | Yes |
| Echocardiogram images on device report | Yes | Yes |
| Auto-view classification | Yes | No Information |
| Conditions diagnosed | 1. Aortic stenosis | 1. heart failure with preserved ejection fraction (HFpEF) |
| Operating platform | Hosted on iCardio platform or on third party infrastructure | Hosted on Ultromics' platform or on third party infrastructure |
| Pre-clinical Performance Testing | Identical | No animal studies were conducted |
| Bench Performance Testing | Identical | Technical validation, numerical stability, and regression testing. |
| Clinical Performance Testing | Validated on a US cohort population, comprising 6+ independent clinical sites representative of the intended use population. | Validated on a US cohort population, comprising 8 independent clinical sites representative of the intended use population. |
iCardio.ai 510(K) Summary - Page 5 of 10 - 2025 iCardio.ai Corporation
{9}
CARDIQ.ai
iCardio.ai CardioVision™
K251293 Traditional 510k Summary
Los Angeles, 2025
# 5. Performance Data
## Nonclinical Testing/Software Assessment:
The CardioVision™ software was developed and tested in accordance with iCardio.ai's design control processes and has undergone rigorous safety and performance evaluations. Verification and validation testing were conducted to demonstrate substantial equivalence to the predicate device. Testing was executed under an automated prospective Validation Plan utilizing a traceability matrix to map Design Input Requirements (e.g., RR-series) directly to verification results. This risk-based validation spanned Unit, Intercomponent, and End-to-End testing, subjecting the system to normal use as well as abnormal and edge-case scenarios to demonstrate that the software consistently fulfills its operating parameters.
## Clinical Testing
The primary objective of the standalone study was to evaluate the software's ability to detect aortic stenosis using predefined success criteria for sensitivity, specificity, and AUROC. Performance was assessed by comparing the software's outputs against ground truth annotations established by board-certified echocardiographers.
This retrospective, multi-center performance study evaluated 650 echocardiography studies from 608 subjects sourced from 12 independent clinical sites across the United States. The dataset reflects a broad distribution of patient ages, heights, weights, and BMI values. Subgroup analyses confirmed consistent performance across available demographic and imaging-related subgroups, including age, weight, BMI, clinical site, ultrasound machine manufacturers, and disease severity (moderate/severe). Subgroup analyses were unable to be performed with gender or race demographics due to anonymization of data. No data from these sites were used in the training or tuning of the algorithm. The dataset included scans acquired on ultrasound systems from GE, Teratech, Siemens, and Mindray. Scans were not included from handheld ultrasound systems.
Ground truth annotations were derived from echocardiographic assessments performed by experienced Level III echocardiographers, with a majority vote approach used in cases of
iCardio.ai 510(K) Summary - Page 6 of 10 - 2025 iCardio.ai Corporation
{10}
#
CARDIO.ai
iCardio.ai CardioVision™ K251293 Traditional 510k Summary
Los Angeles, 2025
disagreement. The study evaluated the software's performance in distinguishing between None/Mild and Moderate/Severe aortic stenosis.
A human factors validation study was conducted to evaluate the usability of CardioVision™ AI for detecting moderate or severe aortic stenosis. All participants successfully completed the critical task of results interpretation without errors.
# Study Results
The key performance metrics to evaluate the ability for CardioVision™ to produce an output that is consistent with moderate/severe and normal/mild readings are summarized in Table 1. Seven (7) out of 650 studies (1.077%) were rejected by the device for poor image quality. The full confusion matrix is shown in Figure 1.

Figure 1: CardioVision™ Confusion Matrix
iCardio.ai 510(K) Summary - Page 7 of 10 - 2025 iCardio.ai Corporation
{11}
CARDIQ.ai
iCardio.ai CardioVision™ K251293 Traditional 510k Summary
Los Angeles, 2025
| Metric | Value |
| --- | --- |
| Dataset Size | 643 |
| Rejection Rate (%) | 1.077 |
| AUROC | 0.945 |
| Sensitivity | 0.896 (95% Wilson score CI: [0.8427, 0.9321]) |
| Specificity | 0.872 (95% Wilson score CI: [0.8384, 0.8995]) |
| PPV | 0.734 (95% Wilson score CI: [0.673, 0.787]) |
| NPV | 0.955 (95% Wilson score CI: [0.931, 0.971]) |
Table 1: CardioVision™ Performance Results Summary.
Sensitivity quantifies the proportion of true Moderate/Severe aortic stenosis cases correctly identified by the algorithm, representing the software's ability to minimize false-negative classifications. Specificity, conversely, measures the algorithm's capacity to correctly classify individuals without Moderate/Severe aortic stenosis, thereby limiting false-positive results. Without indeterminate outputs, the sensitivity was estimated to be $89.6\%$ (95% Wilson score CI: 84.3%-93.2%), and the specificity was estimated to be $87.2\%$ (95% Wilson score CI: 83.8%-90.0%). When including indeterminate outputs (i.e. rejections) in the sensitivity and specificity calculations, the sensitivity was estimated to be 0.876 (95% Wilson score CI: [0.8213, 0.9162]) and the specificity was estimated to be 0.866 (95% Wilson score CI: [0.8324, 0.8943]). The combination of high sensitivity and specificity reinforces its substantial equivalence to the predicate device. These predictive values reflect expected performance at this prevalence; real-world performance may vary depending on the underlying disease prevalence in the population.
iCardio.ai 510(K) Summary - Page 8 of 10 - 2025 iCardio.ai Corporation
{12}
#
CARDIO.ai
iCardio.ai CardioVision™ K251293 Traditional 510k Summary
Los Angeles, 2025
These levels of sensitivity and specificity exceed the predefined success criteria and those of the predicate device, supporting the claim of substantial equivalence.
Subgroup analyses were performed across patient demographics including age, BMI, and weight as well as ultrasound manufacturer, PLAX subview, and clinical site. See sensitivity and specificity point estimates in Figure 3.
| Subgroup | N | Sensitivity | Specificity |
| --- | --- | --- | --- |
| B-Mode View | | | |
| PLAX Standard | 432 | 0.874 | 0.888 |
| PLAX Aortic Cusps | 127 | 0.943 | 0.824 |
| PLAX Paracardial | 84 | 0.885 | 0.845 |
| Age | | | |
| 21-44 | 165 | 1.000 | 0.994 |
| 45-64 | 162 | 0.769 | 0.926 |
| 65-114 | 316 | 0.915 | 0.706 |
| Weight (lb) | | | |
| 70-174 | 237 | 0.875 | 0.860 |
| 175-219 | 239 | 0.962 | 0.888 |
| 220-549 | 166 | 0.860 | 0.862 |
| BMI | | | |
| 16-25 | 186 | 0.875 | 0.877 |
| 26-31 | 251 | 0.928 | 0.857 |
| 32-76 | 203 | 0.877 | 0.891 |
| Manufacturer | | | |
| GE | 311 | 0.883 | 0.845 |
| Siemens | 43 | 1.000 | 0.947 |
| Teratech | 285 | 0.909 | 0.881 |
| Mindray | 4 | - | - |
| Site | | | |
| OMM | 311 | 0.895 | 0.843 |
| PRO | 120 | 0.944 | 0.961 |
| JAC | 16 | - | - |
| VIE | 10 | - | - |
| VIR | 63 | 0.667 | 0.870 |
| SAN | 97 | 0.897 | 0.824 |
| PEN | 2 | - | - |
| IMA | 22 | 1.000 | 0.812 |
| PRI | 1 | - | - |
| HAR | 1 | - | - |
Figure 3: Subgroup Analysis Results
iCardio.ai 510(K) Summary - Page 9 of 10 - 2025 iCardio.ai Corporation
{13}
CARDIQ.ai
iCardio.ai CardioVision™
K251293 Traditional 510k Summary
Los Angeles, 2025
No hazardous situations were observed during the study, and no safety risks to subjects were identified.
**Study Conclusion:** The statistical analysis of the standalone study confirmed that CardioVision™ met the predefined success criteria for sensitivity, specificity. The sensitivity and specificity levels demonstrate that the device is substantially equivalent to the predicate device.
## 6. Conclusion
iCardio.ai CardioVision™ is substantially equivalent to the predicate device, EchoGo Heart Failure 1.0. The subject device has the same intended use, similar technological characteristics, and principles of operation as the predicate device. The primary difference between the subject and predicate device is the pathology being assessed: iCardio.ai CardioVision™ is intended for detecting aortic stenosis, while EchoGo Heart Failure 1.0 is intended for detecting heart failure with preserved ejection fraction (HFpEF). These differences in the assessed conditions do not alter the intended use of the device and do not raise new or different questions regarding its substantial equivalence when used as labeled. The software verification and validation testing, including clinical performance data, support that iCardio.ai CardioVision™ is substantially equivalent to the predicate in the specified use conditions. Therefore, iCardio.ai CardioVision™ is substantially equivalent to the predicate.
iCardio.ai 510(K) Summary - Page 10 of 10 - 2025 iCardio.ai Corporation
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.