Impacted third molar detection (reader improvement)
Neural network-based computer vision
—
Lesion level wAFROC difference 0.0093
—
—
Fully-crossed MRMC study
>1 (expert readers)
Indications for Use
Second Opinion® Panoramic is a radiological automated image processing software device intended to identify and mark regions, in panoramic radiographs, in relation to suspected dental findings which include: Caries, Periapical radiolucency, and Impacted third molars. It is designed to aid dental health professionals to review panoramic radiographs of permanent teeth in patients 16 years of age or older as both a concurrent and second reader.
Device Story
Second Opinion® Panoramic is a CADe software device for dental panoramic radiographs. It receives digital radiographs via API, processes them using machine learning modules to detect caries, periapical radiolucencies, and impacted third molars, and returns metadata to a client UI. The UI renders these detections as mask overlays on the original radiograph. Used in dental clinics by dental professionals, the device acts as a concurrent and second reader. It does not replace clinical judgment or full patient evaluation. By highlighting potential pathologies, it assists clinicians in identifying findings, potentially improving diagnostic accuracy and patient outcomes.
Clinical Evidence
Clinical evaluation included a standalone study and a fully-crossed MRMC study using 795 panoramic radiographs. Ground truth was established by four board-certified dentists. Primary endpoint was wAFROC figure of merit. Results showed statistically significant improvement (p < 0.0001) in reader performance when aided by the device. wAFROC values were 0.9788 (impacted teeth), 0.8113 (periapical radiolucency), and 0.7211 (caries). Lesion-level sensitivities were 99%, 82%, and 77% respectively. No significant performance degradation was observed across age, gender, or imaging device subgroups.
Technological Characteristics
Software-only device running on Windows OS. Utilizes neural network-based computer vision and machine learning for automated detection. Cloud-based processing environment. Outputs detections as graphical mask overlays on radiographs. No hardware components; non-contact device.
Indications for Use
Indicated for dental health professionals to review panoramic radiographs of permanent teeth in patients 16 years of age or older to identify and mark suspected caries, periapical radiolucency, and impacted third molars.
Regulatory Classification
Identification
Medical image analyzers, including computer-assisted/aided detection (CADe) devices for mammography breast cancer, ultrasound breast lesions, radiograph lung nodules, and radiograph dental caries detection, is a prescription device that is intended to identify, mark, highlight, or in any other manner direct the clinicians' attention to portions of a radiology image that may reveal abnormalities during interpretation of patient radiology images by the clinicians. This device incorporates pattern recognition and data analysis capabilities and operates on previously acquired medical images. This device is not intended to replace the review by a qualified radiologist, and is not intended to be used for triage, or to recommend diagnosis.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the image analysis algorithms including a description of the algorithm inputs and outputs, each major component or block, and algorithm limitations.
(ii) A detailed description of pre-specified performance testing methods and dataset(s) used to assess whether the device will improve reader performance as intended and to characterize the standalone device performance. Performance testing includes one or more standalone tests, side-by-side comparisons, or a reader study, as applicable.
(iii) Results from performance testing that demonstrate that the device improves reader performance in the intended use population when used in accordance with the instructions for use. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, predictive value, and diagnostic likelihood ratio). The test dataset must contain a sufficient number of cases from important cohorts (*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, concomitant diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment.(iv) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results; and cybersecurity).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use.
(ii) A detailed description of the intended reading protocol.
(iii) A detailed description of the intended user and user training that addresses appropriate reading protocols for the device.
(iv) A detailed description of the device inputs and outputs.
(v) A detailed description of compatible imaging hardware and imaging protocols.
(vi) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality or for certain subpopulations), as applicable.(vii) Device operating instructions.
(viii) A detailed summary of the performance testing, including: test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 14, 2025
Pearl Inc.
% Ashley Brown
Director of Regulatory Affairs
2515 Benedict Canyon Dr.
BEVERLY HILLS, CA 90210
Re: K250525
Trade/Device Name: Second Opinion® Panoramic
Regulation Number: 21 CFR 892.2070
Regulation Name: Medical Image Analyzer
Regulatory Class: Class II
Product Code: MYN
Dated: February 21, 2025
Received: October 17, 2025
Dear Ashley Brown:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250525 - Ashley Brown
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250525 - Ashley Brown
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiologic Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250525 | |
| Device Name Second Opinion® Panoramic | |
| Indications for Use (Describe) | |
| Second Opinion® Panoramic is a radiological automated image processing software device intended to identify and mark regions, in panoramic radiographs, in relation to suspected dental findings which include: Caries, Periapical radiolucency, and Impacted third molars. It is designed to aid dental health professionals to review panoramic radiographs of permanent teeth in patients 16 years of age or older as both a concurrent and second reader. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K250525
510(K) SUMMARY
1. Submitter's Identification
Pearl Inc.
9200 Sunset Blvd
Suite 430
West Hollywood, CA 90069
USA
(239) 450-8829
Contact Person: Ashley Brown; Director of Regulatory Affairs
Date Summary Prepared: Nov. 14th, 2025
2. Trade Name of the Device
Second Opinion® Panoramic
3. Common or Usual Name
Medical Image Analyzer
4. Classification Name, Regulatory Classification & Product Code
Classification Name: Medical Image Analyzer
Regulatory Classification: 21CFR 892.2070, Class II
Product Code: MYN (Analyzer, Medical Image)
5. Predicate Information
Primary Predicate Device: The primary proposed predicate device for clearance of Second Opinion® PR is Denti.AI Detect by Denti.AI Technology, Inc., cleared in (K230144, October 6, 2023), an automated software device classified as a Class II device pursuant to 21 CFR §892.2070, under product code MYN (Analyzer, Medical Image).
Denti.AI Detect is a Computer-Assisted Detection (CADe) software device intended to be used by dental professionals, including dentists and dental specialists, while reading extraoral and intraoral 2D dental radiographs. The device aims to assist in detecting and highlighting uncategorized regions of interest (ROIs) within the teeth area, which include caries and periapical radiolucency, as a second reader.
The Denti.AI Detect device has the same technology (AI software) and intended use as automated image processing to aid dental professionals in the detection of caries and
{5}
periapical radiolucencies in panoramic radiographs.
Secondary Predicate Device: The proposed secondary predicate is the Second Opinion® device, by Pearl Inc., cleared on March 04, 2022 (K210365) and classified as a Class II Medical Image Analyzer pursuant to 21 CFR §892.2070, under product code MYN (Analyzer, Medical Image).
The cleared device is a computer aided detection (CADe) software device, indicated for use by dental health professionals as an aid in their assessment of bitewing and periapical radiographs of permanent teeth in patients 12 years of age or older, as second reader. The device utilizes computer vision technology, developed using machine learning techniques, to identify and mark anatomy & pathology in bitewing and periapical radiographs.
The subject device is substantially equivalent to the predicate as the overall intended use and nature of the software remains the same. The difference with the subject device is the addition of detections in panoramic radiographs.
## 6. Device Description
Second Opinion® PR is a radiological automated image processing software device intended to identify and mark regions, in panoramic radiographs, in relation to suspected dental findings which include: caries, periapical radiolucency, and impacted third molars. It should not be used in lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis.
It is designed to aid dental health professionals to review panoramic radiographs of permanent teeth in patients 16 years of age or older as a concurrent and second reader.
Second Opinion® PR consists of three parts:
- Application Programming Interface (“API”)
- Machine Learning Modules (“ML Modules”)
- Client User Interface (UI) (“Client”)
The processing sequence for an image is as follows:
1. Images are sent for processing via the API
2. The API routes images to the ML modules
3. The ML modules produce detection output
4. The UI renders the detection output
The API serves as a conduit for passing imagery and metadata between the user interface and the machine learning modules. The API sends imagery to the machine learning modules for processing and subsequently receives metadata generated by the machine
{6}
learning modules which is passed to the interface for rendering.
Second Opinion® PR uses machine learning to detect regions of interest. Images received by the ML modules are processed yielding detections which are represented as metadata. The final output is made accessible to the API for the purpose of sending to the UI for visualization. Detected regions of interest are displayed as mask overlays atop the original radiograph which indicate to the practitioner which regions contain which detected potential conditions that may require clinical review. The clinician can toggle over the image to highlight a potential condition for viewing.
# 7. Indications for Use
Second Opinion® Panoramic is a radiological automated image processing software device intended to identify and mark regions, in panoramic radiographs, in relation to suspected dental findings which include: Caries, Periapical radiolucency, and Impacted third molars.
It is designed to aid dental health professionals to review panoramic radiographs of permanent teeth in patients 16 years of age or older as both a concurrent and second reader.
# 8. Summary of Substantial Equivalence:
The primary and secondary predicate devices and subject device are similar devices in the following ways:
1) Intended use: All three devices are intended to be used to aid dental clinicians in their detection of anatomical and pathological findings in radiographs of permanent teeth.
2) Technology characteristics: All devices employ computer vision and machine learning to output detections, use cloud-based environments to conduct processing, and demarcate detections within a user interface with a graphical overlay over radiographs.
3) Safety: As both the candidate and predicate device are automated image processing software systems, neither pose a direct safety hazard to the patient. The primary hazards for all devices, subject and predicates, are potential false positives and false negatives. In the case of each device, users are not meant to rely solely on detection output for clinical decision-making.
4) Clinical Performance: Both devices have undergone clinical studies which demonstrate statistically significant improvement in aided reader performance.
Table 1: Comparison of Second Opinion® PR with the predicate devices.
{7}
| | Subject Device
Second Opinion®
Panoramic | Primary Predicate
Denti.AI Detect
K230144 | Secondary Predicate
Second Opinion®
K210365 |
| --- | --- | --- | --- |
| Manufacturer | Pearl Inc. | Denti.AI Technology, Inc. | Pearl Inc. |
| Classification | 892.2070 | 892.2070 | 892.2070 |
| Product Code | MYN | MYN | MYN |
| Image Modality | Radiograph | Radiograph | Radiograph |
| Intended Use | Dental CADe to aid with the detection of regions of interest in panoramic radiograph review by HCP | Dental CADe to aid with the detection of regions of interest in dental radiograph review by HCP | Dental CADe to aid in dental radiograph review by HCP |
| Full IFU | Second Opinion®
Panoramic is a radiological automated image processing software device intended to identify and mark regions, in panoramic radiographs, in relation to suspected dental findings which include: Caries, Periapical radiolucency, and Impacted third molars. It is designed to aid dental health professionals to review panoramic radiographs of permanent teeth in patients 16 years of age or older as both a concurrent and second reader. | Denti.AI Detect is a Computer-Assisted Detection (CADe) software device intended to be used by dental professionals, comprising dentists and dental specialists, while reading extraoral and intraoral 2D dental radiographs.
The device aims to assist in detecting and highlighting uncategorized regions of interest (ROIs) within the teeth area, which include caries and periapical radiolucency, as a second reader.
The device is also intended to aid in the measurements of mesial and distal bone levels associated with each tooth.
The device is aimed to be used with images from the patients of 22 years age and older without remaining primary dentition. The device is not intended to replace a complete clinician's review or clinical judgment that considers other relevant information from the image or patient history. | Second Opinion® is a computer aided detection (“CADe”) software to identify and mark regions in relation to suspected dental findings which include Caries, Discrepancy at the margin of an existing restoration, Calculus, Periapical radiolucency, Crown (metal, including zirconia & non-metal), Filling (metal & non-metal), Root canal, Bridge and Implants.
It is designed to aid dental health professionals to review bitewing and periapical radiographs of permanent teeth in patients 12 years of age or older as a second reader. |
| Intended body part | Dental | Dental | Dental |
| Technology | Automated, CADe software that utilizes machine learning | Automated, CADe software that utilizes machine learning | Automated, CADe software that utilizes machine learning |
| Device Description | Detection and display of anatomy and pathologies in extraoral radiographs | Detection and display of anatomy and pathologies in intraoral and extraoral radiographs | Detection and display of anatomy and pathologies in intraoral radiographs |
{8}
9. Technological Comparison to Predicate Devices
The fundamental technological principle for both the candidate and predicate devices is the automatic detection of anatomical and pathological findings using machine learning.
The candidate and predicate devices are technologically equivalent as follows:
- All are software devices designed to run on Windows operating systems.
- All devices are designed to process digital radiographs.
- All devices use neural network-based computer vision algorithms for anatomical and pathological detection.
- All devices demarcate detections within the user interface with a graphical overlay on the radiograph.
- All devices produce near-instantaneous detection results.
- All devices are considered to be of basic documentation level.
- All devices passed all verification and validation testing requirements.
The candidate and predicate devices are technologically different as follows:
- Second Opinion localizes detections in bitewing and periapical radiographs only whereas Denti.AI Detect and Second Opinion® PR localize detections in panoramic radiographs as well. In addition, Second Opinion® PR includes the detection of impactions whereas the predicates do not.
10. Assessment of Benefit-Risk, Safety and Effectiveness, and Substantial Equivalence to Predicate Device
Pearl demonstrated the benefits of the device through a standalone clinical study and fully-crossed MRMC study. The results of the study showed the system is both safe and effective for its intended use. When the probable benefits and probable risks of Second Opinion® PR are weighed against one another, the weight of benefits significantly exceeds that of risks. This judgment can be made based on review of the submitted materials showing that Second Opinion® PR meets the design verification and validation requirements. It is thus concluded that Second Opinion® PR can be considered safe and effective such that the device will aid users in the indicated user population in their radiographic review of caries, periapical radiolucencies, and impacted third molars.
11. Cybersecurity
Pearl developed Security controls and processes in accordance with FDA Guidance - Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions dated September 2023. These processes are used in both the development of Second Opinion® PR and in post-market surveillance to ensure the product upholds the highest standards of privacy and security.
{9}
# 12. Discussion of Non-Clinical Tests Performed
The device is a software-only device, so most testable characteristics common to other device types, including Biocompatibility/Materials, Shelf Life/Sterility, Electromagnetic Compatibility and Electrical Safety, Magnetic Resonance (MR) Compatibility, are not applicable to this device.
# 13. Discussion of Clinical Tests Performed
Clinical evaluation of Second Opinion® PR was performed to validate the efficacy of the system in detecting caries, periapical radiolucencies, and impacted third molars in panoramic radiographs. Second Opinion® PR was clinically tested as a standalone device in comparison to the predicate device, Second Opinion, using both a standalone clinical study and a fully-crossed MRMC study. The test dataset includes a diverse range of conditions aligned with the device's intended use, including primary and secondary caries, as well as cysts, bone loss and caries present within or around an impaction.
The Weighted Alternative Free-Response Receiver Operating Characteristic (wAFROC) paradigm was used as the metric of efficacy for the study. The ground truth (GT) was established using the consensus approach based on agreement among at least three out of four expert readers. The enriched regionally balanced image set of 795 images were independently reviewed by four board-certified dentists each with a minimum of five years practice experience as ground truth readers to establish the location of any caries, periapical radiolucencies, and impactions for each radiograph in the set using polygons for PRs and caries, and bounding boxes for impactions. These ground truth readers were not aware of any outputs from the enrichment process. Each GT reader independently marked areas on any radiograph wherein they identified a caries, periapical radiolucency lesion, or impaction. Readers marked detected features using CVAT, an open-source annotation software tool, allowing them to circumscribe each detected caries, periapical radiolucency lesion, or impaction within a polygon or bounding box respectively.
The GT dataset for the clinical evaluation of the Second Opinion® PR system is characterized by a diverse distribution of radiographs across various geographical regions, genders, ages, imaging devices, and image types. Geographically, with respect to the United States, the included dataset of geographic distribution is seen in Table 2 below.
Table 2 Geographic distribution
| Region/Geo | Impacted Teeth N=420 | Periapical Radiolucency N=496 | Caries N=511 |
| --- | --- | --- | --- |
| Northwest | 95 | 102 | 96 |
| Northeast | 99 | 101 | 117 |
{10}
In terms of gender distribution, the table below describes the gender distribution of the ground truth data by dental feature.
Table 3 Gender distribution
| Gender | Impacted Teeth N=420 | Periapical Radiolucency N=496 | Caries N=511 |
| --- | --- | --- | --- |
| Female | 186 | 231 | 234 |
| Male | 191 | 226 | 225 |
| Unknown | 43 | 39 | 52 |
In addition, in terms of age, the table below describes the age distribution of the ground truth data by dental feature.
Table 4 Age distribution
| Age Group | Impacted Teeth N=420 | Periapical Radiolucency N=496 | Caries N=511 |
| --- | --- | --- | --- |
| >16 - 34 years | 178 | 157 | 208 |
| 35 - 64 years | 156 | 208 | 203 |
| >= 65 years | 86 | 131 | 100 |
The table below summarizes the image characteristics of the ground truth datasets by dental feature.
Table 6 Summary of Image characteristics
| | Impacted Teeth N=420 | Periapical Radiolucency N=496 | Caries N=511 |
| --- | --- | --- | --- |
| Overall status, n (%) | | | |
| Normal | 225 (53.6) | 294 (59.3) | 292 (57.1) |
| Abnormal | 195 (46.4) | 202 (40.7) | 219 (42.9) |
| Number of dental features on abnormal images | | | |
| Total, n | 459 | 309 | 654 |
| Mean (SD) | 2.35 (1.1) | 1.53 (1.1) | 2.99 (3.3) |
| Median (range) | 2 (1, 4) | 1 (1, 11) | 2 (1, 25) |
{11}
| Dental feature size on abnormal images (percent of image pixels) | | | |
| --- | --- | --- | --- |
| Mean (SD) | 0.83 (0.22) | 0.09 (0.1) | 0.06 (0.07) |
| Median (range) | 0.8 (0.23, 1.59) | 0.05 (0.0, 0.83) | 0.03 (0.0, 0.42) |
In addition, SOPR was assessed using a meaningful number of cases for each of the major U.S. panoramic manufacturers, including but not limited to Carestream, Gendex, Ray, Planmeca, Sirona, and Vatech.
# Standalone Testing
The results of this standalone clinical performance study demonstrate that Second Opinion® Panoramic is both safe and effective for its intended use in aiding dental professionals in the detection and localization of impacted third molars, periapical radiolucencies, and caries on panoramic radiographs.
Across all analytic levels—lesion, surface, and image—the device achieved high accuracy, sensitivity, and localization consistency. Both lesion and image level wAFROC metrics show diagnostic performance comparable to or exceeding that of published benchmarks for AI-assisted radiographic analysis. The segmentation and localization metrics (Dice ≥ 0.68, Jaccard ≥ 0.62) confirm that the device accurately delineates lesion boundaries, an important factor for clinical usability and diagnostic confidence.
The standalone evaluation of Second Opinion® Panoramic demonstrated high efficacy in detecting and localizing dental pathologies across all three target lesion types — impacted third molars, periapical radiolucency, and caries — when compared with consensus ground truth established by expert dental radiologists.
The study met the primary efficacy endpoint, with the weighted alternative free-response receiver operating characteristic (wAFROC) figure of merit (FOM) exceeding the prespecified performance thresholds for all evaluated dental features. Specifically, the device achieved wAFROC values of 0.9788, 0.8113, and 0.7211 for impacted third molars, periapical radiolucency, and caries, respectively, surpassing the minimum benchmarks derived from published clinical literature (0.78, 0.71, and 0.70).
These results were statistically significant $(p < 0.0001)$ and indicate that the device consistently discriminates true-positive lesion regions while maintaining low false-positive rates. The high wAFROC values (0.99, 0.91, and 0.90 across features) reinforce the model's diagnostic reliability at the image level.
The secondary efficacy measures, including lesion-level sensitivity, false positives per image (FPPI), and Dice/Jaccard indices, support the robustness of the primary findings. Second Opinion®
{12}
Panoramic achieved lesion-level sensitivities of 99%, 82%, and 77% across the three features, indicating strong detection ability even for small or low-contrast lesions.
Subgroup analyses by age, gender, geography, and imaging device showed consistent trends, confirming that model performance generalizes well across demographics and image acquisition conditions. No significant outliers or bias were observed, suggesting adequate representation and algorithmic robustness across populations.
In total, these findings confirm that Second Opinion® Panoramic provides clinically meaningful detection performance that meets or exceeds the expectations for dental CADe systems in panoramic radiography.
Importantly, subgroup evaluations across age, gender, imaging device type, and confounding factors demonstrated stability in performance, suggesting broad generalizability and low bias potential. The test dataset includes a diverse range of conditions aligned with the device's intended use, including primary and secondary caries, as well as cysts, bone loss and caries present within or around an impaction. No subgroup exhibited statistically significant degradation in efficacy.
In terms of safety, the retrospective, non-patient-contact nature of the study and the absence of device-related adverse events reaffirm the device's favorable risk profile. Any residual risks (false-positive/false-negative) are mitigated through proper user training and clear labeling emphasizing that the software serves as an assistive, not autonomous, diagnostic tool.
In conclusion, the results provide substantial evidence of safety and effectiveness, supporting the use of Second Opinion® Panoramic as an adjunctive diagnostic aid in routine clinical practice. The device demonstrates a strong benefit-to-risk ratio and aligns with regulatory expectations for CADe systems intended for dental panoramic radiography.
## MRMC Testing
The fully crossed MRMC study successfully demonstrated that use of SOPR significantly improves reader performance for detection of dental findings in panoramic radiographs. The primary endpoint was met, with statistically significant increases in wAFROC area under the curve (at both a lesion level and image level) for all three evaluated dental features—impacted third molars, periapical radiolucencies, and caries—when readers were aided by SOPR.
Key findings include:
- Periapical radiolucency detection showed the most pronounced improvement, with a lesion level wAFROC difference of 0.0705 (95% CI: 0.04–0.10; p < 0.00001), image level wAFROC difference of 0.0715 (95% CI: 0.07, 0.07; p < 0.00001) and an increase in lesion-level sensitivity of 0.2045 (95% CI: 0.17–0.24; p < 0.00001).
{13}
- Caries detection improved with a lesion level wAFROC difference of 0.0306 (95% CI: 0.00–0.06; p = 0.0195), image level wAFROC difference of 0.0176 (95% CI: 0.02, 0.02; p < 0.00001) and a lesion-level sensitivity gain of 0.1169 (95% CI: 0.08–0.15).
- Impacted teeth detection demonstrated a lesion level wAFROC difference of 0.0093 (95% CI: 0.00–0.02; p = 0.0326), image level wAFROC difference of 0.0715 (95% CI: 0.07, 0.07; p < 0.00001) and a lesion-level sensitivity gain of 0.0192 (95% CI: 0.01–0.03).
At the lesion level, SOPR improved reader sensitivity, wAFROC, and localization accuracy, particularly for periapical radiolucencies, confirming enhanced lesion detection precision.
At the surface level, high specificity was maintained (≥0.97) across features, demonstrating minimal over-marking on normal regions.
At the image level, improvements in wAFROC metrics, coupled with stable FPPI values, indicate that the device improves reader diagnostic performance without clinically meaningful increases in false positives.
Across all endpoints, the improvements in detection accuracy were achieved without statistically significant increases in false positives per image (FPPI) or reductions in specificity. Localization accuracy, as measured by the Dice coefficient and Jaccard Index, also improved consistently across all features, confirming that SOPR enhances the precision with which readers identify the true location of dental findings.
Exploratory subgroup analyses by age, gender, geography, and imaging device demonstrated no clinically meaningful variability in device performance, suggesting consistent effectiveness across patient populations and imaging platforms.
In conclusion, the MRMC study supports that the Second Opinion Panoramic Radiographs (SOPR) system performs safely and effectively within its intended use. The results demonstrate statistically significant and clinically meaningful improvement in diagnostic performance compared to unaided reading, without increasing user or patient risk.
These findings, together with the device's technological equivalence and established predicate safety profile, provide strong evidence that no new questions of safety or effectiveness are raised by the use of SOPR on panoramic radiographs.
## 14. Comparison to Predicate Clinical Outcomes
Based on the above considerations, Pearl has concluded that Second Opinion® PR is substantially equivalent to the primary predicate device Denti.AI Detect (K230144) and secondary predicate Second Opinion® (K120365), as it has similar software,
{14}
technological, operating characteristics and intended use. Standalone performance testing has demonstrated Second Opinion® PR is as safe and effective as the legally marketed predicate devices and does not raise different questions of safety and effectiveness than the predicate devices.
## 15. Conclusions
Based on the information presented above, Second Opinion® PR and its predicate devices, are deemed to have similar intended uses as devices which aid in the detection of anatomical and pathological findings that can appear in dental radiographic imagery. Second Opinion® PR's clinical trial results demonstrate that the device effectively performs as well as the primary predicate device.
Second Opinion® PR raises no new or different questions of safety or effectiveness, performs in accordance with its specifications, meets user needs, meets the intended use and therefore was found substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.