HeartBeam AIMIGo with 12-L ECG Synthesis Software System
K250258 · Heartbeam, Inc. · DXH · Dec 8, 2025 · Cardiovascular
Device Facts
Record ID
K250258
Device Name
HeartBeam AIMIGo with 12-L ECG Synthesis Software System
Applicant
Heartbeam, Inc.
Product Code
DXH · Cardiovascular
Decision Date
Dec 8, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2920
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The HeartBeam AIMIGo™ System is a portable non-invasive recorder intended to record, store, and transfer a patient's 3-Lead (in three-directions) electrocardiogram (ECG) acquired from 5 electrodes. The device is intended to be used by adult patients in either a clinical setting or at home. The device does not conduct cardiac analysis and can be used with an ECG Viewer software system for manual interpretation of non-life-threatening arrhythmias by a physician or healthcare professional. The HeartBeam 12-L ECG Synthesis Software synthesizes a 12-L ECG from the HeartBeam AIMIGo 3-Leads (in three-directions) recording device, producing a visual 12-L ECG representation that is similar, but not identical, to the same leads of a standard diagnostic 12-L ECG. The synthesized 12-L ECG output is solely intended for manual assessment of normal sinus rhythm and the following non-life-threatening arrhythmias: sinus arrhythmia, sinus tachycardia, sinus bradycardia, atrial premature complexes, atrial fibrillation, and ventricular premature complex. The synthesized 12-L ECG output is not intended for the assessment of any other arrhythmia or conditions (including but not limited to: other atrial arrhythmias, ventricular arrhythmias, hypertrophy, conduction disorders, myocardial infarction or ischemia, pacemaker functions, localization of arrhythmia foci, ECG wave abnormalities, and/or any other disorder). The software does not conduct cardiac analysis and is not intended to replace a standard 12-L ECG. The 12-L ECG Synthesis Software is intended for adult use only.
Device Story
Portable non-invasive recorder captures 3-lead ECG via 5 dry electrodes; transmits signal via Bluetooth to mobile app, then to cloud server. Clinicians access data via portal for manual rhythm interpretation. New 12-L Synthesis Software transforms 3-lead input into synthesized 12-lead representation using personalized transformation matrices derived from patient-specific standard 12-lead ECGs (e.g., EDAN SE-1515 DX-12). Device does not perform automated cardiac analysis; output serves as adjunct information for physician review of non-life-threatening arrhythmias. System enables remote monitoring, reducing need for in-clinic visits for routine arrhythmia assessment.
Clinical Evidence
Two clinical studies conducted. Pivotal study (n=127) validated clinical equivalence between synthesized 12-L ECG and reference standard 12-L ECG for non-life-threatening arrhythmia interpretation. Device Positioning Validation Study (n=45) confirmed output signal stability across varied electrode placements. All endpoints met.
Technological Characteristics
Portable recorder with 5 dry electrodes; rechargeable Li-ion battery; Bluetooth/BLE connectivity to mobile app; cloud-based clinician portal. Synthesis software uses linear transformation matrices. Frequency response 0.5-40 Hz. Software V&V per IEC 62304; usability per IEC 62366-1.
Indications for Use
Indicated for adult patients in clinical or home settings for recording, storing, and transferring 3-lead ECGs. The synthesized 12-L ECG output is indicated for manual assessment of normal sinus rhythm, sinus arrhythmia, sinus tachycardia, sinus bradycardia, atrial premature complexes, atrial fibrillation, and ventricular premature complex. Contraindicated for assessment of life-threatening arrhythmias, myocardial infarction, ischemia, hypertrophy, conduction disorders, or other cardiac conditions.
Regulatory Classification
Identification
A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.
Predicate Devices
HomeTrak Plus EASI Event Recorder System (K982090)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 8, 2025
HeartBeam, Inc.
Deborah Castillo
VP of Regulatory Affairs
2118 Walsh Road, Suite 210
Santa Clara, California 95110
Re: K250258
Trade/Device Name: HeartBeam AIMIGo with 12-L ECG Synthesis Software System
Regulation Number: 21 CFR 870.2920
Regulation Name: Telephone Electrocardiograph Transmitter And Receiver
Regulatory Class: Class II
Product Code: DXH, MWJ, DPS
Dated: October 20, 2025
Received: October 20, 2025
Dear Deborah Castillo:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250258 - Deborah Castillo
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}
K250258 - Deborah Castillo
Page 3
See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JENNIFER W. SHIH -S
Jennifer Kozen
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250258 | |
| Device Name | |
| HEARTBEAM AIMIGo™ SYSTEM WITH 12-L ECG SYNTHESIS SOFTWARE | |
| Indications for Use (Describe) HEARTBEAM AIMIGo SYSTEM: The HeartBeam AIMIGo™ System is a portable non-invasive recorder intended to record, store, and transfer a patient's 3-Lead (in three-directions) electrocardiogram (ECG) acquired from 5 electrodes. The device is intended to be used by adult patients in either a clinical setting or at home. The device does not conduct cardiac analysis and can be used with an ECG Viewer software system for manual interpretation of non-life-threatening arrhythmias by a physician or healthcare professional. HEARTBEAM 12-L ECG SYNTHESIS SOFTWARE: The HeartBeam 12-L ECG Synthesis Software synthesizes a 12-L ECG from the HeartBeam AIMIGo 3-Leads (in three-directions) recording device, producing a visual 12-L ECG representation that is similar, but not identical, to the same leads of a standard diagnostic 12-L ECG. The synthesized 12-L ECG output is solely intended for manual assessment of normal sinus rhythm and the following non-life-threatening arrhythmias: sinus arrhythmia, sinus tachycardia, sinus bradycardia, atrial premature complexes, atrial fibrillation, and ventricular premature complex. The synthesized 12-L ECG output is not intended for the assessment of any other arrhythmia or conditions (including but not limited to: other atrial arrhythmias, ventricular arrhythmias, hypertrophy, conduction disorders, myocardial infarction or ischemia, pacemaker functions, localization of arrhythmia foci, ECG wave abnormalities, and/or any other disorder). The software does not conduct cardiac analysis and is not intended to replace a standard 12-L ECG. The 12-L ECG Synthesis Software is intended for adult use only. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) SUMMARY
K250258
# 510(k) Notification K250258
## GENERAL INFORMATION [807.92(a)(1)]
### Applicant:
HeartBeam, Inc.
2118 Walsh Road, Suite 210
Santa Clara, CA, 95050
USA
Phone: 408-899-4443
### Contact Person:
Deborah Castillo
VP of Regulatory Affairs
2118 Walsh Road, Suite 210
Santa Clara, CA, 95050
USA
Phone: 408-899-4443
Date Prepared: November 17, 2025
## DEVICE INFORMATION [807.92(a)(2)]
### Trade Name:
HeartBeam AIMIGo™ System with 12-L Synthesis Software
### Generic/Common Name:
Electrocardiograph
### Classification:
Class II
### Regulation:
21 CFR§870.2920
### Product Codes:
Primary: DXH
Secondary: DPS, MWJ
PAGE 1 OF 11
{5}
510(k) SUMMARY
K250258
# PREDICATE DEVICE(s) [807.92(a)(3)]
Primary Predicate:
- PaceArt Associates, L.P., HomeTrak Plus EASI Event Recorder System (K982090)
Reference Devices:
- HeartBeam AIMIGo™ System (K231424)
# DEVICE DESCRIPTION [807.92(a)(4)]
The HeartBeam AIMIGo™ System captures a 3-lead (in three-directions) electrocardiogram (ECG) recording of patients who have been prescribed a HeartBeam AIMIGo™ System for recording an ECG remotely. The HeartBeam AIMIGo™ System records and transmits the 3-lead ECG signal which can be displayed as rhythm strips on a compatible ECG viewer. The HeartBeam AIMIGo™ System does not provide any analysis of the ECG data nor provide any recommendation toward a clinical diagnosis and is not intended to be used with automated ECG analysis systems. It reports a series of 3-lead ECG rhythm strips for manual interpretation. The hardware platform is designed to be functional and effective with any compatible software designed for the purpose of displaying ECG waveforms for clinical review. The HeartBeam AIMIGo™ System is provided by prescription only.
The system allows the patient (or their caregiver or healthcare provider) the capability to perform the following:
- Record a 3-lead (in three-directions) ECG with the recording device and accompanying mobile Patient Application
- Transmit the recorded 3-lead ECG signal to the cloud server, which can then be accessed by a clinician via the Clinician Portal for review and manual interpretation.
This submission includes a modification to the HeartBeam System involving the addition of a new software feature, the HeartBeam 12-L Synthesis Software. Like the HeartBeam AIMIGo system, the patient-activated recording of a 3-L ECG is obtained and the HeartBeam 12-L Synthesis Software can synthesize a 12-L ECG signal from the 3-lead ECG signal recorded from the HeartBeam AIMIGo Device of the cleared system. The synthesized 12-L ECG output signal is intended to be an adjunct set of ECG information that complements but does not replace a standard 12-L ECG, to aid in the manual assessment of non-life-threatening arrhythmias.
The synthesized 12-L ECG signal has not been studied or evaluated for safe and effective use for other cardiac conditions outside of manual non-life-threatening arrhythmias and is contraindicated for the assessment of any urgent, serious, or life-threatening conditions or arrhythmias.
The new software works with the HeartBeam system, which retains all functionalities and performance characteristics of the previously cleared system. No changes to the hardware, intended use, or fundamental technological characteristics were made to incorporate the new
PAGE 2 OF 11
{6}
510(K) SUMMARY
K250258
software. Additionally, the synthesis software works with the standard 12-L ECG system, the EDAN SE-1515 DX-12. The EDAN SE-1515 DX-12 systems provide a patient's compatible standard 12-L ECG signal, which is used as one of the inputs for the synthesis software to create the personalized transformation matrix for each patient.
## INDICATIONS FOR USE [807.92(a)(5)]
## HeartBeam AIMIGo System:
The HeartBeam AIMIGo™ System is a portable non-invasive recorder intended to record, store, and transfer a patient's 3-Lead (in three-directions) electrocardiogram (ECG) acquired from 5 electrodes.
The device is intended to be used by adult patients in either a clinical setting or at home.
The device does not conduct cardiac analysis and can be used with an ECG Viewer software system for manual interpretation of non-life-threatening arrhythmias by a physician or healthcare professional.
## HeartBeam 12-L Synthesis Software
The HeartBeam 12-L ECG Synthesis Software synthesizes a 12-L ECG from the HeartBeam AIMIGo 3-Leads (in three-directions) recording device, producing a visual 12-L ECG representation that is similar, but not identical, to the same leads of a standard diagnostic 12-L ECG.
The synthesized 12-L ECG output is solely intended for manual assessment of normal sinus rhythm and the following non-life-threatening arrhythmias: sinus arrhythmia, sinus tachycardia, sinus bradycardia, atrial premature complexes, atrial fibrillation, and ventricular premature complex. The synthesized 12-L ECG output is not intended for the assessment of any other arrhythmia or conditions (including but not limited to: other atrial arrhythmias, ventricular arrhythmias, hypertrophy, conduction disorders, myocardial infarction or ischemia, pacemaker functions, localization of arrhythmia foci, ECG wave abnormalities, and/or any other disorder). The software does not conduct cardiac analysis and is not intended to replace a standard 12-L ECG. The 12-L ECG Synthesis Software is intended for adult use only.
## SUBSTANTIAL EQUIVALENCE
The HeartBeam AIMIGo™ System with 12-L ECG Synthesis Software is substantially equivalent to the Predicate Device, the HomeTrak Plus EASI Event Recorder System (K982090). These devices share the same intended use in the sense that the fundamental parts of the system allow the capability to record, store, and transfer a patient's 3-L ECG signal, which is used to create synthesized 12-L ECG signals. Both the 3-L ECGs and synthesized 12-L ECGs are intended for manual assessment of non-life-threatening arrhythmias. Moreover, the predicate and proposed device systems have comparable technological features. The methods used by the new software to achieve 12-L synthesis are similar to the methods used by the predicate devices (i.e. linear transformation matrices). Performance testing data support the software's capabilities
PAGE 3 OF 11
{7}
510(K) SUMMARY
K250258
to achieve its intended use and demonstrate that the technological differences do not raise new questions about device safety or effectiveness.
The new software works with the Reference Device, the HeartBeam AIMIGo™ System (K231424) as the 12-L Synthesis Software synthesizes a 12-L ECG signal from the 3-lead ECG signal recorded from the HeartBeam AIMIGo Device of the Reference Device system. The new software works with the HeartBeam system, which retains all functionalities and performance characteristics of the previously cleared system. No changes to the hardware, intended use, or fundamental technological characteristics were made to incorporate the new software, and thus both systems share these identical components.
Based on the above, the HeartBeam AIMIGo™ System with 12-L Synthesis Software is substantially equivalent to the Predicate Device.
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES [807.92(a)(6)]
Table 1 below shows the comparison of intended use and technological characteristics with the Predicate and Reference Devices.
PAGE 4 OF 11
{8}
510(K) SUMMARY
K250258
Table 1: Comparison of Technological Characteristics with the Predicate Device
| Device Name | HeartBeam AIMIGo™ System with 12-L Synthesis Software (Subject Device) | HomeTrak Plus EASI Event Recorder System (Primary Predicate Device) | HeartBeam AIMIGo System (Reference Device) | Rationale for Substantial Equivalence/Comments |
| --- | --- | --- | --- | --- |
| 510(k) Number | K250258 | K982090 | K231424 | - |
| Company | HeartBeam, Inc. | PaceArt Associates, L.P. | HeartBeam, Inc. | - |
| Classification | 21 CFR 870.2920 | 21 CFR 870.2920 | 21 CFR 870.2920 | Same. |
| Device Identification | Telephone electrocardiograph transmitter and receiver | Telephone electrocardiograph transmitter and receiver | Telephone electrocardiograph transmitter and receiver | Same. |
| Product Code | DXH, DPS, MWJ | DXH | DXH, DPS, MWJ | Same. |
| Intended Use / Indications for Use | The HeartBeam AIMIGo™ System is a portable non-invasive recorder intended to record, store, and transfer a patient's 3-Lead (in three-directions) electrocardiogram (ECG) acquired from 5 electrodes.The device is intended to be used by adult patients in either a clinical setting or at home.The device does not conduct cardiac analysis and can be used with an ECG Viewer software system for manual interpretation of non-life-threatening arrhythmias by a physician or healthcare professional.The HeartBeam 12-L ECG Synthesis Software synthesizes a 12-L ECG from the HeartBeam AIMIGo 3-Leads (in three-directions) recording device, producing a visual 12-L ECG representation that is similar, but not identical, to the same leads of a standard diagnostic 12-L ECG.The synthesized 12-L ECG output is solely intended for manual assessment of normal sinus rhythm and the following non-life-threatening arrhythmias: sinus arrhythmia, sinus tachycardia, sinus bradycardia, atrial premature complexes, atrial fibrillation, and ventricular premature complex. The | The HomeTrak Plus Cardiac Twelve-Lead Event Recorder is a patient-activated ambulatory cardiac event monitor and associated central-station receiving equipment which are to be used for recording infrequent and elusive heart arrhythmias and transmitting them by telephone at a later time to a central analysis center. It records three electrocardiographic leads (3-Lead ECG) and derives 12 electrocardiographic leads (12-L ECG) from the three transmitted. | The HeartBeam AIMIGo™ System is a portable non-invasive recorder intended to record, store, and transfer a patient's 3-Lead (in three-directions) electrocardiogram (ECG) acquired from 5 electrodes.The device is intended to be used by adult patients in either a clinical setting or at home.The device does not conduct cardiac analysis and can be used with an ECG Viewer software system for manual interpretation of non-life-threatening arrhythmias by a physician or healthcare professional. | Same.The HeartBeam System with 12-L ECG Synthesis Software has the same intended use as the HomeTrak Plus Cardiac 12-Lead Event recorder in that both systems record 3-Lead ECGs and derive a 12-L ECG from these signals. Both systems also transmit the signals (either recorded 3-Lead or derive 12-lead) for display and review by a physician at a later time. Both systems are also intended to be used for arrhythmia assessment only, as detailed in their corresponding Indications for Use statements.The Reference system, K231424, is being updated with the 12-L Synthesis SW feature to allow display of a synthesized 12-L ECG. Both the Reference and Subject device ECG outputs are intended to be used by adult patients in either a clinical setting or at home. Both ECG output signals are intended to be used for manual interpretation of non-life-threatening arrhythmias by a physician or healthcare professional. Neither system conducts cardiac analysis.The minor differences in the indications statement language do not constitute a new intended use, nor raise new |
{9}
510(K) SUMMARY
K250258
| Device Name | HeartBeam AIMIGo™ System with 12-L Synthesis Software (Subject Device) | HomeTrak Plus EASI Event Recorder System (Primary Predicate Device) | HeartBeam AIMIGo System (Reference Device) | Rationale for Substantial Equivalence/Comments |
| --- | --- | --- | --- | --- |
| | synthesized 12-L ECG output is not intended for the assessment of any other arrhythmia or conditions (including but not limited to: other atrial arrhythmias, ventricular arrhythmias, hypertrophy, conduction disorders, myocardial infarction or ischemia, pacemaker functions, localization of arrhythmia foci, ECG wave abnormalities, and/or any other disorder). The software does not conduct cardiac analysis and is not intended to replace a standard 12-L ECG. The 12-L ECG Synthesis Software is intended for adult use only. | | | questions about device safety and effectiveness. |
| Patient Population | Adults Only | Adults Only | Adults Only | Same. |
| Use Environment | Clinical or Home Setting | Clinical or Home Setting | Clinical or Home Setting | Same. |
| Number of Electrodes | The HeartBeam AIMIGo™ Device with embedded 5-dry electrodes ("AIMIGo Device") | 5 wet/gel electrodes | The HeartBeam AIMIGo™ Device with embedded 5-dry electrodes ("AIMIGo Device") | Similar.
Both the HeartBeam systems use dry electrodes to acquire and record the 3-L ECGs.
While the Predicate Device, HomeTrak System, uses wet-electrodes to acquire and record the 3-L ECGs, both the Proposed and Predicate systems used clinical and non-clinical testing to validate the output performance to ensure signal quality and noise is accounted for. The testing conducted on the HeartBeam system with 12-L synthesis software demonstrate the accuracy of the synthesized 12-L ECG output signals is acceptable for the same intended use within the labeled specifications. Thus, the differences do not raise different questions of safety or effectiveness. |
| Electrode Placement/Position on Patient Body | 5 electrodes: Two (2) Chest & two (2) Left Hand Fingers to Right Hand Fingers and (1) RLD – right leg drive electrode. | 5 electrodes – Using the EASI configuration:
E (Electrode on Chest): Placed over the lower part of the sternum at the level of the fifth intercostal space. | 5 electrodes: Two (2) Chest & two (2) Left Hand Fingers to Right Hand Fingers and (1) RLD – right leg drive electrode. | Similar.
Both the HeartBeam systems use the identical electrode placement/positioning on the body to acquire and record the 3-L ECGs. |
PAGE 6 OF 11
{10}
510(K) SUMMARY
K250258
| Device Name | HeartBeam AIMIGo™ System with 12-L Synthesis Software (Subject Device) | HomeTrak Plus EASI Event Recorder System (Primary Predicate Device) | HeartBeam AIMIGo System (Reference Device) | Rationale for Substantial Equivalence/Comments |
| --- | --- | --- | --- | --- |
| | | A (Anterior Left Electrode): Positioned on the left mid-axillary line at the same horizontal level as the E electrode (fifth intercostal space).
S (Superior Electrode on Chest): Located at the manubrium of the sternum (near the sternal notch).
I (Inferior Right Electrode): Positioned on the right mid-axillary line at the level of the fifth intercostal space.
Ground Electrode: Typically placed in a neutral location (e.g., right lower Abdomen, or Right torso) | | While the HomeTrak System uses the specific EASI electrode configuration, both systems have upper and lower chest electrodes, left and right limb placement electrodes, and a fifth, neutral electrode. Both systems also used clinical and non-clinical testing to validate the output performance. The testing conducted on the HeartBeam system with 12-L synthesis software demonstrate the accuracy of the synthesized 12-L ECG output signals from the recorded 3-L ECG from this electrode placement. Therefore, the performance is acceptable for the same intended use within the labeled specifications. Thus, the differences do not raise different questions of safety or effectiveness. |
| ECG Viewer Leads Display Capability | 3-Lead ECG or synthesized 12-Lead ECG in Cloud ECG Viewer | 3-L ECG or derived 12-L ECG via Central Analysis Center | 3-Lead ECG only in Cloud ECG Viewer | Similar.
In the Predicate, the HomeTrak System, the central analysis center serves as the access point for a physician to view and assess either the 3-L ECG or derived 12-L ECG signals. The differences do not raise different questions of device safety or effectiveness, as the usability is the same (i.e. to view and assess the ECG signals). |
| Handheld HeartBeam AIMIGo Device Charging Method & Power Source | The HeartBeam AIMIGo™ Charging Dock ("Charging Dock") to charge a (1) rechargeable, Lithium-ion Battery | Not specified. | The HeartBeam AIMIGo™ Charging Dock ("Charging Dock") to charge a (1) rechargeable, Lithium-ion Battery | Similar.
Both the HeartBeam systems use the same charging dock and rechargeable battery. While the HomeTrak system does not appear to need or use a battery or charging dock, given the same battery and charging dock have already been cleared and tested under the Reference Device system, the differences in this technological features do not raise new questions of safety and effectiveness. |
PAGE 7 OF 11
{11}
510(K) SUMMARY
K250258
| Device Name | HeartBeam AIMIGo™ System with 12-L Synthesis Software (Subject Device) | HomeTrak Plus EASI Event Recorder System (Primary Predicate Device) | HeartBeam AIMIGo System (Reference Device) | Rationale for Substantial Equivalence/Comments |
| --- | --- | --- | --- | --- |
| Number of Channels/Leads Recorded | 3 channels/Leads | 3 channels/Leads | 3 channels/Leads | Same. |
| Frequency Response | 0.5 Hz to 40 Hz ± 3dB | Not specified. | 0.5 Hz to 40 Hz ± 3dB | Similar.
Although not specified by the Predicate System, this feature is identical between the Subject and Reference Device. Testing of the system for both the Subject and Predicate Device show acceptable performance for their intended use and thus supports substantial equivalence. |
| Recording Duration | 30 seconds | Not specified. | 30 seconds | Similar.
Although not specified by the Predicate System, the duration of recording does not impact device safety or effectiveness. This feature is identical between the Subject and Reference Device. Testing of the system for both the Subject and Predicate Device show acceptable performance for their intended use and thus support substantial equivalence. |
| Synthesized Leads | All 12-Leads | All 12-Leads | Feature Not Available | Same.
Both the Subject and Predicate systems have a 12-L ECG Synthesis SW feature that uses the recorded reduced lead-set 3-L ECG signals to derive all 12-Lead ECG output signals. Both systems derive all 12-leads. Through the acceptable test methods of SW V&V and full clinical validation test results, HeartBeam demonstrates that this new feature does not raise different questions of safety and effectiveness, and the performance of the synthesized 12-L ECG signal of the HeartBeam System is substantially equivalent to the Predicate device for the same intended use and indications. |
| Clinician User Interface | Cloud-Based Clinician Portal | Central Analysis Center | Cloud-Based Clinician Portal | Similar.
The Predicate Device uses the Central Analysis Center for display and review of |
PAGE 8 OF 11
{12}
510(K) SUMMARY
K250258
| Device Name | HeartBeam AIMIGo™ System with 12-L Synthesis Software (Subject Device) | HomeTrak Plus EASI Event Recorder System (Primary Predicate Device) | HeartBeam AIMIGo System (Reference Device) | Rationale for Substantial Equivalence/Comments |
| --- | --- | --- | --- | --- |
| | | | | the transmitted 3-L ECG and derived 12-L ECG for physician review. The HomeTrak System uses the central analysis center to serve as the access point for a physician to view and assess either the 3-L ECG or derived 12-L ECG signals. In a similar fashion, the Subject Device uses the Clinical Portal for this purpose. The differences do not raise different questions of device safety or effectiveness, as the usability is the same (i.e. to view and assess the ECG signals).
The cloud-base Clinical Portal in the Subject device is the same as the Clinician Portal of the Reference Device system, which has been updated to accommodate access to the synthesized 12-L ECG recordings and display for clinical review. |
| Data Transfer Method Between AIMIGo Device and Mobile Application | Wireless (Bluetooth/ BLE) communication between the patient Mobile Application and the AIMIGo device | Cable-based connection. | Wireless (Bluetooth/BLE) communication between the patient Mobile Application and the AIMIGo device. | Similar.
The Predicate HomeTrak device uses electrode cables to acquire signals. The HeartBeam system uses Wi-Fi connection identical to the Reference Device system, and HeartBeam has ensured that proper transmission have been evaluated and tested to meet current standards and regulations. |
| Single Use | No. | No. | No. | Same. |
| Prescription Use/Over the Counter | Prescription Use. | Prescription Use. | Prescription Use. | Same. |
PAGE 9 OF 11
{13}
510(k) SUMMARY
K250258
# PERFORMANCE DATA [807.92(b)]:
All necessary bench testing was conducted on the HeartBeam AIMIGo™ System with 12-L Synthesis Software to support a determination of substantial equivalence to the Predicate Device.
# Nonclinical Testing Summary [807.92(b)(1)]:
The nonclinical bench testing included:
- Software verification and validation per the requirements of EC 62304:2006 (Medical device software – Software life cycle processes) and in accordance with FDA Guidance Document entitled, “General Principles of Software Validation,” issued January 11, 2002
- Human Factors and Usability testing in accordance with IEC 62366-1 (Medical devices – Part 1: Application of usability engineering to medical devices) and FDA’s Human Factors Guidance, Applying Human Factors and Usability Engineering to Medical Devices, issued on February 3, 2016.
- Cybersecurity Testing has been conducted in accordance with section 524B(b)(2) of the Federal Food, Drug, and Cosmetic Act, and FDA guidance “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions,” issued on June 27, 2025.
- General system-level verifications were conducted to demonstrate that the subject device performs in accordance with its design specifications.
The collective results of the nonclinical testing demonstrate that the design of the HeartBeam AIMIGo™ System with 12-L Synthesis Software meet the established specifications necessary for consistent performance during its intended use. In addition, the collective bench testing demonstrates that the HeartBeam AIMIGo™ System with 12-L Synthesis Software does not raise different questions of safety or effectiveness when compared to the Predicate Device.
# Clinical Testing Summary [807.92(b)(2)]:
Clinical testing was conducted to support device performance in this 510(k) submission. The studies below were conducted in a population consistent with the subject device’s indications for use:
- Pivotal Study: The objective of this U.S. study was to validate the AIMIGo 12-L ECG Synthesis Software by demonstrating the clinical equivalence between the synthesized 12-L ECG waveforms generated by the AIMIGo 12-L ECG Synthesis Software (investigational device) and reference standard 12-L ECG for interpretation of non-life-threatening arrhythmia. All study endpoints were met, and the information provided supports that device performance is substantially equivalent to the predicate. A total of 127 patients were included in the study population. This included data from 41.7% Females, 58.3% Males, and approximately 74% White/Caucasian, 3% Hispanic/Latino, 0.8% Asian, 18% Black, and 4% other ethnic group. The age distribution consisted of approximately 29% of patients under the age of 65 years old, 71% of patients that were 65 years or older, and a median age of 69 years old. For BMI measures, approximately 20% of patients had a BMI of less than 25, and 80% of patients had a BMI of 25 or greater.
PAGE 10 OF 11
{14}
510(K) SUMMARY
K250258
- Device Positioning Validation Study: This was a prospective single-arm supplemental clinical trial to validate that within a predetermined range of distance, direction and orientation with respect to the recommended position per labeling, the output signal characteristics of the AIMIGo 12-L ECG Synthesis Software remain unaffected when compared to a standard 12L ECG signal recorded simultaneously, for interpretation of non-life-threatening arrhythmias. All study endpoints were met, and the information provided supports that device performance is substantially equivalent to the predicate. A total of 45 patients were included in the study population. This included data from approximately 36% Females, 64% Males, and approximately 86% White/Caucasian, 7% Asian, and 7% Black. The age distribution consisted of approximately 78% of patients under the age of 65 years old, 22% of patients that were 65 years or older, and a median age of 51 years old. For BMI measures, approximately 47% of patients had a BMI of less than 25, and 53% of patients had a BMI of 25 or greater.
## CONCLUSIONS [807.92(b)(3)]:
Based on the results from the tests performed in support of the HeartBeam AIMIGo™ System with 12-L Synthesis Software, it is concluded that the Subject Device is substantially equivalent to the legally marketed Predicate Device.
PAGE 11 OF 11
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.