The Swoop® Portable MR Imaging® System (V2) is a portable, ultra-low field magnetic resonance imaging device for producing images that display the internal structure of the head where full diagnostic examination is not clinically practical. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.
Device Story
Portable, ultra-low field (64.9 mT) MRI system for head imaging at point-of-care (ER, ICU, rehab). Operates via permanent magnet and transmit/receive RF coil. System inputs include MRI k-space data; transforms data using deep learning-based 'Advanced Reconstruction' (denoising, interpolation, geometric correction) to produce high-quality diagnostic images. Operated by clinicians via COTS touch-screen interface for scan setup, execution, and QC. Output viewed on-device or via PACS; assists physicians in diagnosing head pathologies (e.g., stroke, hemorrhage, TBI). Benefits include rapid, accessible imaging in environments where traditional MRI is impractical.
Clinical Evidence
No clinical trials; bench-only performance validation. Advanced Reconstruction validated using 58 subjects (313 images) for robustness/stability and 32 subjects (167 images) for qualitative assessment by 5 ABR-certified radiologists. Metrics included NMSE, SSIM, and CNR. Results showed statistically significant improvement in CNR (p<0.05) and superior qualitative scores (median +2 on 5-point Likert scale) for Advanced vs. Linear reconstruction across all categories (noise, sharpness, contrast, geometric fidelity, artifact).
Technological Characteristics
Ultra-low field (64.9 mT) permanent magnet MRI. Patient bore: 36.0 in. width, 13.4 in. height. Gradient system: X: 33.9 mT/m, Y: 33.2 mT/m, Z: 66.2 mT/m. RF coil: Linear transmit/receive. Connectivity: DICOM output. Software: Deep learning-based image reconstruction (Advanced Reconstruction). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-33, ISO 10993, NEMA MS series.
Indications for Use
Indicated for adult and pediatric patients (≥ 0 years) requiring head imaging where full diagnostic MRI is not clinically practical.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 30, 2025
Hyperfine, Inc.
Christine Kupchick
Sr. Manager, Global Regulatory
351 New Whitfield St
Guilford, Connecticut 06437
Re: K250236
Trade/Device Name: Swoop® Portable MR Imaging® System (V2)
Regulation Number: 21 CFR 892.1000
Regulation Name: Magnetic Resonance Diagnostic Device
Regulatory Class: Class II
Product Code: LNH, MOS
Dated: May 7, 2025
Received: May 7, 2025
Dear Christine Kupchick:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250236 - Christine Kupchick
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250236 - Christine Kupchick
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Daniel M. Krainak, Ph.D.
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250236
Device Name
Swoop® Portable MR Imaging® System (V2)
Indications for Use (Describe)
The Swoop® Portable MR Imaging® System (V2) is a portable, ultra-low field magnetic resonance imaging device for producing images that display the internal structure of the head where full diagnostic examination is not clinically practical. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
HYPERFINE
510(k) Summary
Swoop® Portable MR Imaging® System (V2)
K250236
510(k) SUBMITTER
Company Name: Hyperfine, Inc.
Company Address: 351 New Whitfield St
Guilford, CT 06437
CONTACT
Name: Christine Kupchick
Telephone: (203) 343-3404
Email: ckupchick@hyperfine.io
Date Prepared: May 22, 2025
DEVICE IDENTIFICATION
Trade Name: Swoop® Portable MR Imaging® System (V2)
Common Name: Magnetic Resonance Imaging
Regulation Number: 21 CFR 892.1000
Classification Name: System, Nuclear Magnetic Resonance Imaging Coil, Magnetic Resonance
Product Code: LNH; MOS
Regulatory Class: Class II
PREDICATE DEVICE INFORMATION
The subject Swoop Portable MR Imaging System (V2) is substantially equivalent to the predicate Swoop System (K240944).
{5}
DEVICE DESCRIPTION
The Swoop System (V2) is portable, ultra-low field MRI device that enables visualization of the internal structures of the head using standard magnetic resonance imaging contrasts. The main interface is a commercial off-the-shelf device that is used for operating the system, providing access to scan orders, exam setup, exam execution, viewing MRI image data for quality control purposes, and PACS interactions. The system can generate MRI data sets with a broad range of contrasts. The Swoop system user interface includes touch screen menus, controls, indicators, and navigation icons that allow the operator to control the system and to preview images. The Swoop System (V2) image reconstruction algorithm utilizes deep learning for optimized image quality.
# INDICATIONS FOR USE
The Swoop Portable MR Imaging System (V2) is a portable, ultra-low field magnetic resonance imaging device for producing images that display the internal structure of the head where full diagnostic examination is not clinically practical. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.
# SUBSTANTIAL EQUIVALENCE DISCUSSION
The table below compares the subject device to the predicate.
| Specification | Subject Swoop Portable MR Imaging System (V2) | Predicate Swoop Portable MR Imaging System (V1) (K240944) |
| --- | --- | --- |
| INTENDED USE | | |
| Intended Use/ Indications for Use | The Swoop Portable MR Imaging System is a portable, ultra-low field magnetic resonance imaging device for producing images that display the internal structure of the head where full diagnostic examination is not clinically practical. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis. | Same |
| Patient Population | Adult and pediatric patients (≥ 0 years) | Same |
| Anatomical Sites | Head | Same |
| Environment of Use | At the point of care in professional health care facilities such as emergency rooms, intensive/critical care units, hospitals, outpatient, or rehabilitation centers. | Same |
| Energy Used and/or delivered | Magnetic Resonance | Same |
{6}
| MAGNET | | |
| --- | --- | --- |
| Field strength | 64.9 mT (nominal) | 63.3 ± 2.0 mT |
| Type | Permanent magnet | Same |
| Patient accessible bore size | 36.0 in. width, 13.4 in. height | 24.0 in. width, 12.4 in. height |
| Magnet weight | 712 lbs. | 705 lbs. |
| GRADIENT SYSTEM | | |
| Maximum gradient amplitude | X: 33.9 mT/m
Y: 33.2 mT/m
Z: 66.2 mT/m | X: 24.3 mT/m
Y: 22.9 mT/m
Z: 38.5 mT/m |
| Rise time (zero to max) | X: 1.8 ms
Y: 1.8 ms
Z: 5.1 ms | X: 2.1 ms
Y: 2.0 ms
Z: 3.8 ms |
| Slew rate | X: 18.8 T/m/s
Y: 18.4 T/m/s
Z: 13.0 T/m/s | X: 24 T/m/s
Y: 22 T/m/s
Z: 21 T/m/s |
| RF COIL | | |
| Type | Transmit/receive | Same |
| Transmit coil design | Linear | Same |
| OTHER | | |
| Patient weight capacity | 1.6kg-200 kg | Same |
| Operation temperature | 15-30 C | Same |
| Warm up time | <3 minutes | Same |
| Temperature control | No | Same |
| Humidity control | No | Same |
| SEQUENCES & IMAGE PROCESSING | | |
| T1W sequences | • T1 (Standard), T1 (Gray/White)
• Advanced Gridding reconstruction | • Same
• Same |
| T2W sequences | • T2, T2 (Fast)
• Advanced Gridding reconstruction | • Same
• Same |
| FLAIR sequences | • FLAIR, FLAIR (Fast)
• Advanced Gridding reconstruction | • FLAIR only
• Same |
| DWI/ADC sequences | • DWI/ADC, DWI/ADC (Fast)
• Advanced Gridding + FISTA | • DWI/ADC only
• FISTA only |
| Image post-processing (All Sequences) | • Advanced Denoising
• Image orientation transform
• Geometric distortion correction
• Receive coil intensity correction
• Advanced Interpolation
• DICOM output | • Same
• Same
• Same
• Same
• n/a
• Same |
The subject device and the predicate device have the same intended use, operating principles, and similar technological characteristics. The subject device differs from the predicate in hardware and software. These differences do not raise new questions of safety and efficacy as compared to the predicate.
{7}
NON-CLINICAL PERFORMANCE
The subject device passed all the testing in accordance with internal requirements and applicable standards to support substantial equivalence to the predicate.
| Test | Test Description | Applicable Standard(s) |
| --- | --- | --- |
| Verification | Verification testing in accordance with the design requirements. | • IEC 62304:2015
• FDA Guidance, “Content of Premarket Submissions for Device Software Functions”
• NEMA MS 1-2008 (R2020)
• NEMA MS 3-2008 (R2020)
• NEMA MS 9-2008 (R2020)
• NEMA MS 12-2016
• NEMA MS 8-2016
• American College of Radiology (ACR) Phantom Test Guidance for Use of the Large MRI Phantom for the ACR MRI Accreditation Program
• American College of Radiology standards for named sequences
• FDA Guidance, “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions” |
| Validation | Validation to ensure the subject device meets user needs and performs as intended. | • FDA Guidance, “Content of Premarket Submissions for Device Software Functions”
• FDA Guidance, “Applying Human Factors and Usability Engineering to Medical Devices
• IEC 62366-1:2015+AMD1:2020 |
| Biocompatibility | Biocompatibility evaluation of patient-contacting materials. | • ISO 10993-1:2018
• ISO 10993-5:2009
• ISO 10993-10:2021
• ISO 10993-23:2021 |
| Reprocessing | Cleaning and disinfection evaluation of patient-contacting materials. | • FDA Guidance, “Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling”
• ISO 17664:2021
• ASTM F3208-20
• AAMI TIR12:2020
• ANSI/AAMI ST98:2022 |
{8}
| Test | Test Description | Applicable Standard(s) |
| --- | --- | --- |
| Safety | Electrical Safety, EMC, and Essential Performance testing. | • IEC 60601-1:2005 Ed.3+A1; A2
• IEC 60601-1-2:2014 Ed.4+A1
• IEC 60601-1-6:2010 Ed.3+A1; A2
• IEC 60601-2-33:2015 Ed. 3.2 |
## ADVANCED RECONSTRUCTION PERFORMANCE ANALYSIS AND VALIDATION
Performance analysis and validation of the subject device Advanced Reconstruction models was performed using a test dataset entirely independent from the dataset used for model training. The test dataset comprises a total of 58 individual subjects and 313 unique images collected using sequence types available on the subject device. For each subject, a subset of the following sequences, chosen appropriately for the indication for imaging, was scanned: T1 Gray-White, T1 Standard, T2, T2 Fast, FLAIR, FLAIR Fast, DWI, DWI Fast (DWI b=0, b=900, ADC). Axial, Sagittal, and Coronal orientations were included; for DWI sequences, only Axial was available. A description of the acceptance criteria and subset of data used for each test is included in the test summaries below.
In all cases, models are trained and validated with MRI data and images as the only inputs and outputs; there are no confounding factors and clinical subgroups are not defined or considered. While gender and age are available for most subjects, age, gender, ethnic background, and pathology are not expected to influence model architecture.
## Performance Analysis:
### Study Design:
Advanced Reconstruction was assessed for robustness, stability, and generalizability over a variety of subjects, design parameters, artifacts, and scan conditions using reference-based metrics.
A set of images including Swoop data, high field images, and synthetic contrast images, was used as ground truth target images. Test input data (synthetic k-space generated from the target images) was reconstructed using both Advanced and Linear Reconstruction, and the similarity to the original ground truth image was compared between the two reconstruction methods. Reconstruction outputs with motion and zipper artifacts were qualitatively assessed.
### Reference Standard and Metrics:
Normalized mean squared error (NMSE) and structural similarity index (SSIM) were used to compare the ability of Advanced Reconstruction to reproduce the ground truth image compared to Linear Reconstruction.
{9}
Dataset and Sample Size per Model:
| Model/Sequence group: T1, T2, FLAIR | | |
| --- | --- | --- |
| Patients | 40 | |
| Images | 111 | |
| Demographics | | |
| Gender | Female | 38% |
| | Male | 55% |
| | Unknown | 7% |
| Age | 18-35 | 5% |
| | 35-60 | 28% |
| | 60+ | 62% |
| | 2+* | 5% |
| Ethnicity Data | Not Recorded | |
| Number of sites | 4 | |
| Equipment Type | Swoop v2† | |
| Pathology | Atrophy, Cerebellar Infarct, Chronic Infarct, Demyelinating diseases, Embolic Infarct, Hydrocephalus, Infarct, Intracerebral Hemorrhage, Intraparenchymal Hemorrhage, Lesion, M1 Occlusion, Mass Effect, Seizure, Stroke, Subdural Hemorrhage, Traumatic Brain Injury, Tumor, White Matter Disease, White Matter Lesion | |
* Partially anonymized
† Contained Swoop Mk1.9 (<1%)
| Model/Sequence group: DWI | | |
| --- | --- | --- |
| # of patient | 29 | |
| # Images | 94 | |
| Demographics | | |
| Gender | Female | 35% |
| | Male | 55% |
| | Unknown | 10% |
| Age | 18-35 | 7% |
| | 35-60 | 21% |
| | 60+ | 65% |
| | 2+* | 7% |
| Ethnicity Data | Not Recorded | |
| Number of sites | 4 | |
| Equipment Type | Swoop v2 | |
| Pathology | Atrophy, Cerebellar Infarct, Chronic Infarct, Demyelinating Diseases, Embolic Infarct, Infarct, Intracerebral Hemorrhage, Intraparenchymal Hemorrhage, Lesion, M1 Occlusion, Mass Effect, Stroke, Subdural Hemorrhage, Traumatic Brain Injury, Tumor, White Matter Disease | |
* Partially anonymized
# Study Results:
For all models and all test datasets NMSE was reduced and SSIM was improved for Advanced Reconstruction test images compared to Linear Reconstruction test images. Advanced Reconstruction preserved the presentation of motion and zipper artifacts, and no unexpected output was observed.
{10}
# Contrast-to-Noise Ratio Validation
## Study Design:
Regions of interest (ROI) encompassing pathologies were annotated and reviewed by two American Board of Radiology (ABR) certified radiologists. The contrast-to-noise of hyper- and hypo- intense pathologies were measured with respect to healthy white matter tissue from the same image. The inclusion criterion for images used for this study was at least one visible pathology.
## Reference Standard and Metrics:
Linear Reconstruction was used as the reference standard for the comparison. Contrast-to-Noise Ratio (CNR) between pathology and healthy tissues was measured to quantify how accurately pathology features are preserved by Advanced Reconstruction.
The mean CNR of Advanced Reconstruction was required to be greater than the mean CNR of the baseline Linear Reconstruction at statistical significance level of 0.05 for each sequence type.
## Dataset and Sample Size:
A minimum of 16 images per model type (sequence group) were included for lesion annotation. Pathologies spanning multiple slices were used multiple times. All annotated images were then reviewed, and inaccurate ROI annotations were excluded from the analysis. The data meeting inclusion criteria are described below.
| Patients | 15 | |
| --- | --- | --- |
| Images | 46 | |
| Pathologies | 58 | |
| ROIs | 464 | |
| Demographics and other Variability | | |
| Gender | Female | 28% |
| | Male | 70% |
| | Unknown | 2% |
| Age | 35-60 | 28% |
| | 60+ | 70% |
| | 2+* | 2% |
| Ethnicity Data | Not Recorded | |
| Number of sites | 2 | |
| Equipment | Swoop v2 | |
| Pathology | Atrophy, Demyelinating diseases, Embolic Infarct, Infarct, Intracerebral Hemorrhage, Intraparenchymal Hemorrhage, Lesion, Resection, Stroke, Thrombectomy, Tumor, White Matter Disease, White Matter Hyperintensity, White Matter Lesion | |
*Partially anonymized
## Study Results:
In all cases, CNR of Advanced Reconstruction was greater than or equal to Linear Reconstruction for both hyper- and hypo-intense pathologies. The study result demonstrates that Advanced Reconstruction does not unexpectedly modify, remove, or reduce the contrast of pathology features.
{11}
# Advanced Reconstruction Image Validation
# Study Design:
Five external, ABR-certified radiologists representing clinical users were asked to review side-by-side clinical image sets taken with the subject Swoop System, reconstructed with both Advanced and Linear Reconstruction. The reviewers rated the images using a five-point scale for image quality and the consistency of diagnosis using both methods in the categories of noise, sharpness, contrast, geometric fidelity, artifact, and overall image quality.
# Reference Standard and Metrics:
Linear Reconstruction was used as the reference standard for the comparison. Advanced Reconstruction was required to perform at least as well as Linear Reconstruction in all categories (median score $\geq 0$ on Likert scale) and perform better ( $\geq 1$ on Likert scale) in at least one of the quality-based categories.
# Dataset and Sample Size:
A sample size of at least 16 was used per sequence. Within the sample dataset at least four cases for each sequence-available image orientation (axial, sagittal, coronal) were used.
| Patients | 32 | |
| --- | --- | --- |
| Images | 167 | |
| Demographics and other Variability | | |
| Gender | Female | 28% |
| | Male | 70% |
| | Unknown | 2% |
| Age | 18-35 | 1% |
| | 35-60 | 25% |
| | 60+ | 35% |
| | 2+* | 39% |
| Ethnicity Data | Not Recorded | |
| Number of sites | 3 | |
| Equipment | Swoop v2 | |
| Pathology | Included pathology: Atrophy, Chronic Infarct, Demyelinating diseases, Embolic Infarct, Hemorrhage, Hydrocephalus, Infarct, Intracerebral Hemorrhage, Intraparenchymal Hemorrhage, Lesion, M1 Occlusion, Mass Effect, Stroke, Subdural Hemorrhage, Traumatic Brain Injury, Tumor, White Matter Disease | |
*Partially anonymized
# Test Results:
Advanced Reconstruction achieved a median score of 2 (the most positive rating scale value) in all categories. This scoring indicates reviewers found Advanced Reconstruction improved image quality while maintaining diagnostic consistency relative to Linear Reconstruction.
{12}
CONCLUSION
Based on the intended use, technological characteristics, performance results, and comparison to the predicate, the subject Swoop Portable MR Imaging System (V2) has been shown to be substantially equivalent to the predicate and there are no new questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.