Retrospective clinical data from six sites in the USA and Japan were used to externally validate the performance (sensitivity and specificity) of the Us2.ca AI algorithm for detecting cardiac amyloidosis.
Independent pooled dataset of 1,647 patients from six clinical sites across the United States and Japan.
—
Indications for Use
Us2.ca processes acquired transthoracic cardiac ultrasound images to support qualified cardiologists, sonographers, or other licensed professional healthcare practitioners in their diagnosis of cardiac amyloidosis. Us2.ca is intended for use only in adult patients with increased left ventricular wall thickness, defined as an interventricular septal thickness (IVSd) or left ventricular posterior wall thickness (LVPWd) ≥ 12mm. Us2.ca is not intended to provide a diagnosis and does not replace current standards of care. The results from Us2.ca are not intended to exclude the need for further follow-up on cardiac amyloidosis.
Device Story
Us2.ca is a clinical decision support software that analyzes DICOM-format transthoracic echocardiogram images; specifically apical 4-chamber (A4C) clips. It uses a video-based deep-learning algorithm to identify cardiac amyloidosis. The device operates on the Us2.ai platform, hosted on third-party infrastructure. It processes images to generate automated measurements and disease indications. Results meeting confidence thresholds are presented in a report for review by cardiologists, sonographers, or other licensed practitioners. The device is intended to improve clinical workflow and speed of reporting; however, it does not provide a definitive diagnosis, does not replace standard of care, and does not identify specific regions of interest or pathology. Final clinical interpretation and decision-making remain with the clinician. It is validated for use with GE and Philips ultrasound systems.
Clinical Evidence
Retrospective case-control study across six clinical sites (USA/Japan) using 1,647 patients (independent validation cohort). Primary endpoints were sensitivity and specificity for cardiac amyloidosis detection in patients with IVSd/LVPWd ≥ 12mm. In 1,434 patients with confident predictions, sensitivity was 86.9% (95% CI: 84.2%-89.7%) and specificity was 87.4% (95% CI: 85.2%-89.7%). Subgroup analyses performed by age, gender, race, ultrasound vendor (GE/Philips), and clinical site.
Technological Characteristics
SaMD; machine learning-based algorithm; processes DICOM echocardiogram files. Operates on third-party cloud infrastructure. Validated for GE and Philips ultrasound systems. Implements cybersecurity controls per FDA guidance. Software verification conducted at unit, module, and system integration levels.
Indications for Use
Indicated for adult patients with increased left ventricular wall thickness (IVSd or LVPWd ≥ 12mm) undergoing transthoracic echocardiography to support clinicians in the diagnosis of cardiac amyloidosis.
Regulatory Classification
Identification
The adjunctive cardiovascular status indicator is a prescription device based on sensor technology for the measurement of a physical parameter(s). This device is intended for adjunctive use with other physical vital sign parameters and patient information and is not intended to independently direct therapy.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Software description, verification, and validation based on comprehensive hazard analysis must be provided, including:
(i) Full characterization of technical parameters of the software, including any proprietary algorithm(s);
(ii) Description of the expected impact of all applicable sensor acquisition hardware characteristics on performance and any associated hardware specifications;
(iii) Specification of acceptable incoming sensor data quality control measures; and
(iv) Mitigation of impact of user error or failure of any subsystem components (signal detection and analysis, data display, and storage) on accuracy of patient reports.
(2) Scientific justification for the validity of the status indicator algorithm(s) must be provided. Verification of algorithm calculations and validation testing of the algorithm using a data set separate from the training data must demonstrate the validity of modeling.
(3) Usability assessment must be provided to demonstrate that risk of misinterpretation of the status indicator is appropriately mitigated.
(4) Clinical data must be provided in support of the intended use and include the following:
(i) Output measure(s) must be compared to an acceptable reference method to demonstrate that the output measure(s) represent(s) the predictive measure(s) that the device provides in an accurate and reproducible manner;
(ii) The data set must be representative of the intended use population for the device. Any selection criteria or limitations of the samples must be fully described and justified;
(iii) Agreement of the measure(s) with the reference measure(s) must be assessed across the full measurement range; and
(iv) Data must be provided within the clinical validation study or using equivalent datasets to demonstrate the consistency of the output and be representative of the range of data sources and data quality likely to be encountered in the intended use population and relevant use conditions in the intended use environment.
(5) Labeling must include the following:
(i) The type of sensor data used, including specification of compatible sensors for data acquisition;
(ii) A description of what the device measures and outputs to the user;
(iii) Warnings identifying sensor reading acquisition factors that may impact measurement results;
(iv) Guidance for interpretation of the measurements, including warning(s) specifying adjunctive use of the measurements;
(v) Key assumptions made in the calculation and determination of measurements;
(vi) The measurement performance of the device for all presented parameters, with appropriate confidence intervals, and the supporting evidence for this performance; and
(vii) A detailed description of the patients studied in the clinical validation (
*e.g.,* age, gender, race/ethnicity, clinical stability) as well as procedural details of the clinical study.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
Build Correspondence
Convert to PDF
June 20, 2025
Eko.ai Pte. Ltd. d/b/a Us2.ai
Hui Tay
RAQA Manager
2 College Road, 02-00
Singapore, 169850
Singapore
Re: K250151
Trade/Device Name: Us2.ca
Regulation Number: 21 CFR 870.2200
Regulation Name: Adjunctive Cardiovascular Status Indicator
Regulatory Class: Class II
Product Code: SDJ
Dated: May 21, 2025
Received: May 21, 2025
Dear Hui Tay:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250151 - Hui Tay
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250151 - Hui Tay
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Robert T. Kazmierski -S
for
LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Us2.ca
Page 8 of 54
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K250151 | ? |
| Please provide the device trade name(s). | | ? |
| Us2.ca | | |
| Please provide your Indications for Use below. | | ? |
| Us2.ca processes acquired transthoracic cardiac ultrasound images to support qualified cardiologists, sonographers, or other licensed professional healthcare practitioners in their diagnosis of cardiac amyloidosis. Us2.ca is intended for use only in adult patients with increased left ventricular wall thickness, defined as an interventricular septal thickness (IVSd) or left ventricular posterior wall thickness (LVPWd) ≥ 12mm. Us2.ca is not intended to provide a diagnosis and does not replace current standards of care. The results from Us2.ca are not intended to exclude the need for further follow-up on cardiac amyloidosis. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
Eko.ai Pte. Ltd. d/b/a Us2.ai
510(k)-K250151
Us2.ca
# 510(k) Summary
Table 1. Subject Device Overview.
| Submitter's Name: | Eko.ai Pte. Ltd. (d/b/a Us2.ai) |
| --- | --- |
| Address: | 2 College Road, #02-00, Singapore 169850 |
| Contact Person: | Hui Qun, Tay |
| Title: | Regulatory Affairs and Quality Assurance Manager |
| Telephone Number: | +65 62327857 |
| Fax Number: | +65 62327857 |
| Email: | huiqun@us2.ai |
| Date Summary Prepared: | Jan 17, 2025 |
| Device Proprietary Name: | Us2.ca |
| Model Number: | Us2.ca |
| Common Name: | Us2.ca |
| Regulation Number: | 21 CFR 870.2200 |
| Regulation Name: | Adjunctive cardiovascular statue indicator |
| Product Code: | SDJ |
| Device Class: | Class II |
| Predicate Device | Trade name: EchoGo Cardiac Amyloidosis 1.0
Manufacturer: Ultromics Limited 4630 Kingsgate
Cascade Way,
Oxford Business Park South, Oxford
Oxfordshire, United Kingdom, OX4 2SU
Regulation Number: 21 CFR 870.2200
Regulation Name: Adjunctive Cardiovascular Statue Indiator
Device Class: Class II
Product Code: SDJ
510(k) Number: K240860
510(k) Clearance Date: Nov 15, 2024 |
Page 1 of 10
{5}
Eko.ai Pte. Ltd. d/b/a Us2.ai
510(k)-K250151
Us2.ca
# Device Description
The Us2.ai platform is a clinical decision support tool that analyzes echocardiogram images in order to generate a series of AI-derived measurements. Fully automated, functional reporting with disease indications is also provided, in line with ASE & ESC guidelines.
Echo images are sent to the Us2.ai platform where they are processed, analyzed and measured. Results that meet the confidence threshold for both image quality and measurement accuracy are passed through to a report for review by the clinical users. Report text is also generated and presented with the measurements, providing functional reporting and disease indications. The ultimate clinical decision and interpretation reside solely with the clinician.
Us2.ca is an enhancement to Us2.ai existing Us2.v2 software, adding the capability to detect cardiac amyloidosis. It is an image post-processing analysis software device used for viewing and quantifying cardiovascular ultrasound images in DICOM format. The device is intended to aid diagnostic review and analysis of echocardiographic data, patient record management and reporting.
The primary intended function of Us2.ca is to automatically identify patients who require additional follow-up for cardiac amyloidosis. In doing so, the primary benefit is to improve clinical echocardiographic workflow, enabling clinicians to generate and edit reports faster, with precision and with full control. The final clinical decision of the results still remains with the clinicians.
# Indications for Use
Us2.ca processes acquired transthoracic cardiac ultrasound images to support qualified cardiologists, sonographers, or other licensed professional healthcare practitioners in their diagnosis of cardiac amyloidosis. Us2.ca is intended for use only in adult patients with increased left ventricular wall thickness, defined as an interventricular septal thickness (IVSd) or left ventricular posterior wall thickness (LVPWd) ≥ 12mm. Us2.ca is not intended to provide a diagnosis and does not replace current standards of care. The results from Us2.ca are not intended to exclude the need for further follow-up on cardiac amyloidosis.
# Limitations
Please note the following limitations:
- Our software complements good patient care and does not exempt the user from the responsibility to provide supervision, clinically review the patient, and make appropriate clinical decisions.
- During image acquisition, inappropriate use of the echo machine, use of non-cardiac ultrasound probes, use of suboptimal settings (e.g., gain, contrast, or depth), or lack of electrocardiogram capture may lead to lower accuracy of the software.
- Us2.ca does not identify regions of interest or areas of possible pathology in the images.
- Us2.ca does not differentiate different types of cardiac amyloidosis, e.g., hereditary ATTR, wild-type ATTR, and AL.
Page 2 of 10
{6}
Eko.ai Pte. Ltd. d/b/a Us2.ai
510(k)-K250151
Us2.ca
- Us2.ca was externally validated on cardiac images from GE and Philips ultrasound systems. Use of the device on ultrasound systems other than GE and Philips may lead to lower accuracy of the software.
Page 3 of 10
{7}
Eko.ai Pte. Ltd. d/b/a Us2.ai
510(k)-K250151
Us2.ca
# Summary of Technological Characteristics Comparison
Table 2 shows the similarities and differences between the technological characteristics of the two products. Testing demonstrates that the differences do not raise new questions of safety or effectiveness.
Table 2. Summary of Technological Characteristics Comparison.
| Characteristics | Subject Device(Us2.ca) | Predicate (K240860-EchoGo Cardiac Amyloidosis (1.0)) |
| --- | --- | --- |
| Regulation | 21 CFR 870.2200 | 21 CFR 870.2200 |
| SaMD | Yes | Yes |
| Generic Type Device | Adjunctive cardiovascular status indicator | Adjunctive cardiovascular status indicator |
| Indications for Use | Us2.ca processes acquired transthoracic cardiac ultrasound images to support qualified cardiologists, sonographers, or other licensed professional healthcare practitioners in their diagnosis of cardiac amyloidosis. Us2.ca is intended for use only in adult patients with increased left ventricular wall thickness, defined as an interventricular septal thickness (IVSd) or left ventricular posterior wall thickness (LVPWd) ≥ 12mm. Us2.ca is not intended to provide a diagnosis and does not replace current standards of care. The results from Us2.ca are not intended to exclude the need for further follow-up on cardiac amyloidosis. | EchoGo Amyloidosis 1.0 is an automated machine learning-based decision support system, indicated as a screening tool for adult patients aged 65 years and over with heart failure undergoing cardiovascular assessment using echocardiography. When utilised by an interpreting physician, this device provides information alerting the physician for referral to confirmatory investigations. EchoGo Amyloidosis 1.0 is indicated in adult patients aged 65 years and over with heart failure. Patient management decisions should not be made solely on the results of the EchoGo Amyloidosis 1.0 analysis. |
| Targeting disease | Cardiac Amyloidosis | Cardiac Amyloidosis |
| Intended population | Adults | Adults over the age of 65 |
| Users | Qualified cardiologists, sonographers, or other licensed professional practitioners | Interpreting clinician |
| Nature of device | Clinical decision-making support tool | Diagnostic aid |
| Operating platform | Hosted on Us2.ai's platform on 3rd party infrastructure | Hosted on Ultromic's platform on 3rd party infrastructure |
{8}
Eko.ai Pte. Ltd. d/b/a Us2.ai
510(k)-K250151
Us2.ca
| Machine Learning-Based algorithm | Yes | Yes |
| --- | --- | --- |
| DICOM files | Yes | Yes |
{9}
Eko.ai Pte. Ltd. d/b/a Us2.ai
510(k)-K250151
Us2.ca
| Characteristics | Subject Device(Us2.ca) | Predicate (K240860-EchoGo Cardiac Amyloidosis (1.0)) |
| --- | --- | --- |
| containing echocardiogram as input | | |
| Cybersecurity | Post-market Management of Cybersecurity in Medical Devices. Content of Premarket Submissions for Management of Cybersecurity in Medical Devices. Content of Premarket Submissions for Management of Cybersecurity in Medical Device Containing Off-the-Shelf (OTS) Software: Guidance for Industry. | Post-market Management of Cybersecurity in Medical Devices. Content of Premarket Submissions for Management of Cybersecurity in Medical Devices. Content of Premarket Submissions for Management of Cybersecurity in Medical Devices. Cybersecurity for Networked Medical Devices Containing Off-the-Shelf (OTS) Software: Guidance for Industry. |
| Pre-clinical performance testing | No animal studies were conducted. | No animal studies were conducted. |
| Bench Performance Testing | Verification and Validation testings, Regression testing | Technical validation, numerical stability, and regression testing. |
Both Us2.ca and the predicate device use AI to analyze echocardiograms and generate reports that support clinical decision making in detecting cardiac amyloidosis. Given the similarity in both devices, Us2.ca does not significantly affect the safety or effectiveness of the devices, and they operate in a similar manner. Thus, Us2.ca is substantially equivalent to the predicate device.
# Special Controls
Per the regulation 21 CFR 870.2200, there are special controls which the Us2.ca has to fulfil. All supporting documentation and/or data for the implementation of the special controls are found within the actual submission.
(1) Software description, verification, and validation based on comprehensive hazard analysis must be provided, including:
(i) Full characterization of technical parameters of the software, including any proprietary algorithm(s);
(ii) Description of the expected impact of all applicable sensor acquisition hardware characteristics on performance and any associated hardware specifications;
(iii) Specification of acceptable incoming sensor data quality control measures; and
(iv) Mitigation of impact of user error or failure of any subsystem components (signal detection and analysis, data display, and storage) on accuracy of patient reports.
(2) Scientific justification for the validity of the status indicator algorithm(s) must be provided. Verification of algorithm calculations and validation testing of the algorithm using a data set separate from the training data must demonstrate the validity of modeling.
(3) Usability assessment must be provided to demonstrate that risk of misinterpretation of the status indicator is appropriately mitigated.
(4) Clinical data must be provided in support of the intended use and include the following:
(i) Output measure(s) must be compared to an acceptable reference method to Page 6 of 10
{10}
Eko.ai Pte. Ltd. d/b/a Us2.ai
510(k)-K250151
Us2.ca
demonstrate that the output measure(s) represent(s) the predictive measure(s) that the device provides in an accurate and reproducible manner;
(ii) The data set must be representative of the intended use population for the device. Any selection criteria or limitations of the samples must be fully described and justified;
(iii) Agreement of the measure(s) with the reference measure(s) must be assessed across the full measurement range; and
(iv) Data must be provided within the clinical validation study or using equivalent datasets to demonstrate the consistency of the output and be representative of the range of data sources and data quality likely to be encountered in the intended use population and relevant use conditions in the intended use environment.
(5) Labeling must include the following:
(i) The type of sensor data used, including specification of compatible sensors for data acquisition;
(ii) A description of what the device measures and outputs to the user;
(iii) Warnings identifying sensor reading acquisition factors that may impact measurement results;
(iv) Guidance for interpretation of the measurements, including warning(s) specifying adjunctive use of the measurements;
(v) Key assumptions made in the calculation and determination of measurements;
(vi) The measurement performance of the device for all presented parameters, with appropriate confidence intervals, and the supporting evidence for this performance;
(vii) A detailed description of the patients studied in the clinical validation (e.g., age, gender, race/ethnicity, clinical stability) as well as procedural details of the clinical study.
All special controls listed below have been implemented within Us2.ca.
## Performance Data
Us2.ca was developed and tested in accordance with Us2.ai’s Design Control processes. The device has been subject to extensive safety and performance testing. Non-clinical verification and validation test results established that the device meets its design requirements and intended use. Specifically, software verification was conducted at unit, module, and system integration levels. Risk management analysis generated multiple risk mitigation measures and verification activities. A Cybersecurity Analysis and data security testing were conducted to verify that data and patient protected health information security measures are included in the design of the software. A Human Factors/Usability Engineering study performed according to the principles of AAMI/ANSI HE75 was performed to validate the device’s usability within the intended user population.
Us2.ca uses a video-based deep-learning approach for CA detection. The model analyzes Apical 4-Chamber (A4C) clips and identifies the presence of CA. The primary objective of the Us2.ca performance test protocol is to validate the performance of Us2.ca, developed for the detection of cardiac amyloidosis in apical four-chamber (A4C) echocardiographic images of adult patients with IVSd or LVPWd ≥ 12mm. The testing data involved two cohorts: Cardiac Amyloidosis Group (CA Group) and Control Group.
The primary performance metrics were sensitivity and specificity. The benchmark used in deriving the acceptance criteria of Us2.ca was made with reference to current standards of care and existing relevant publications.
Us2.ca device performance was evaluated in accordance to a retrospective case-control study across various sites in the USA and Japan. Data was obtained using a range of PACS system to
Page 7 of 10
{11}
Eko.ai Pte. Ltd. d/b/a Us2.ai
510(k)-K250151
Us2.ca
represent the types of machines used.
# Results for Us2.ca
Us2.ca was developed using a training dataset comprising 4,371 patients and externally validated on an independent pooled dataset of 1,647 patients. The training and external validation datasets were sourced from entirely separate data providers, ensuring no overlap and maintaining full independence between development and evaluation. The external validation cohort included a demographically and geographically diverse population, collected from six clinical sites across the United States and Japan. All echocardiographic studies were retrospectively obtained from routine clinical evaluations and provided in standard DICOM format. Ultrasound data were obtained from GE, Philips, and Siemens systems to train the machine learning-based software device, but the testing data were only obtained from GE and Philips systems.
Summary Demographic Data:
| Clinical Characteristic | Training Data (N=4371) | | Testing Data (N=1647) | |
| --- | --- | --- | --- | --- |
| | CA Cases (N=2241) | Control Cases (N=2130) | CA Cases (N=664) | Control Cases (N=983) |
| Age, years, n (%) | | | | |
| <=50 | 48 (2.1%) | 487 (22.9%) | 23 (3.5%) | 143 (14.5%) |
| 51-60 | 138 (6.2%) | 384 (18%) | 43 (6.5%) | 173 (17.6%) |
| 61-70 | 454 (20.3%) | 445 (20.9%) | 136 (20.5%) | 225 (22.9%) |
| 71-80 | 989 (44.1%) | 453 (21.3%) | 271 (40.8%) | 263 (26.8%) |
| >80 | 612 (27.3%) | 357 (16.8%) | 186 (28%) | 175 (17.8%) |
| Not reported | 0 (0%) | 4 (0.2%) | 5 (0.8%) | 4 (0.4%) |
| Sex, n (%) | | | | |
| Female | 382 (17%) | 816 (38.3%) | 151 (22.7%) | 350 (35.6%) |
| Male | 1859 (83%) | 1314 (61.7%) | 508 (76.5%) | 631 (64.2%) |
| Not reported | 0 (0%) | 0 (0%) | 5 (0.8%) | 2 (0.2%) |
| Race, n (%) | | | | |
| African American | 260 (11.6%) | 140 (6.6%) | 240 (36.1%) | 315 (32%) |
| White | 1895 (84.6%) | 396 (18.6%) | 217 (32.7%) | 397 (40.4%) |
| Hispanic or Latino | 0 (0%) | 0 (0%) | 5 (0.8%) | 0 (0%) |
| Asian | 86 (3.8%) | 1594 (74.8%) | 182 (27.4%) | 228 (23.2%) |
| Others/ Not reported | 0 (0%) | 0 (0%) | 20 (3%) | 43 (4.4%) |
| Other Clinical Variables | | | | |
| BMI, kg/m2, mean ± SD | 26.3±4.0 | 26.3±5.6 | 25.8±5.3 | 29.1±7.4 |
Subgroup analyses were conducted across key variables including age, sex, race, ultrasound vendor, and clinical site, which met the acceptance criteria. Us2.ca performance analysis was conducted at the patient level; therefore, the number of samples is equal to the number of patients.
# Results for Us2.ca:
When tested on a total of 1647 patients, Us2.ca was able to generate a confident prediction for 1434 patients (CA group: 574 patients; Controls group: 860 patients), in which Us2.ca
Page 8 of 10
{12}
Eko.ai Pte. Ltd. d/b/a Us2.ai
510(k)-K250151
Us2.ca
performed with a sensitivity of 86.9% (95% CI: 84.2%-89.7%) and specificity of 87.4% (95% CI: 85.2%-89.7%). The overall yield, which refers to the proportion of total tested patients for whom the device was able to generate a confident prediction, was sufficiently high at 87.1%.
The outcomes of the subgroup analysis are provided in the following breakdown,
**Stratified by Age**
| Subgroup | Primary Analyses | |
| --- | --- | --- |
| Age | Sensitivity (%) (95% CI) | Specificity (%) (95% CI) |
| < 65 years | 78.02 (69.51, 86.53) | 88.63 (85.27, 91.99) |
| >= 65 years | 88.7 (85.87, 91.54) | 86.58 (83.63, 89.52) |
**Stratified by Gender**
| Subgroup | Primary Analyses | |
| --- | --- | --- |
| Gender | Sensitivity (%) (95% CI) | Specificity (%) (95% CI) |
| Female | 85.61 (79.62, 91.59) | 87.83 (84.15, 91.5) |
| Male | 87.41 (84.3, 90.52) | 87.21 (84.43, 89.99) |
**Stratified by Race**
| Subgroup | Primary Analyses | |
| --- | --- | --- |
| Race | Sensitivity (%) (95% CI) | Specificity (%) (95% CI) |
| Asian | 92.31 (88.13, 96.49) | 85.29 (80.43, 90.15) |
| African American | 85.38 (80.62, 90.13) | 87.5 (83.57, 91.43) |
| Hispanic or Latino | 100 (100, 100) | - |
| White | 83.7 (78.36, 89.03) | 88.73 (85.4, 92.06) |
| Others | 83.33 (53.51, 113.15) | 90.91 (73.92, 100) |
{13}
Eko.ai Pte. Ltd. d/b/a Us2.ai
510(k)-K250151
Us2.ca
# Stratified by Vendor
Us2.ca is externally validated on data from Philips and GE vendors.
| Subgroup | Primary Analyses | |
| --- | --- | --- |
| Vendor | Sensitivity (%)
(95% CI) | Specificity (%)
(95% CI) |
| Philips | 84.17
(80.58, 87.76) | 88.42
(85.31, 91.54) |
| GE | 93.14
(89.4, 96.89) | 86.41
(83.24, 89.58) |
# Stratified by Data Source
| Subgroup | Primary Analyses | |
| --- | --- | --- |
| Data Source | Sensitivity (%)
(95% CI) | Specificity (%)
(95% CI) |
| Center 1 | 86.55
(82.63, 90.48) | 87.91
(85.25, 90.57) |
| Center 2 | 78.21
(69.04, 87.37) | 92.11
(83.53, 100) |
| Center 3 | 86.27
(76.83, 95.72) | 86.67
(76.73, 96.6) |
| Center 4 | 92.26
(88.05, 96.47) | 85.35
(80.43, 90.28) |
# Substantial Equivalence Conclusion
Us2.ca is an image processing software which has similar intended use and indications for use statement as the predicate device. The two devices have similar technological characteristics: both use machine learning algorithms to automate the measurement of transthoracic cardiac images. As both the subject and predicate devices support the detection of cardiac amyloidosis, this does not in and of itself produce different questions of safety and effectiveness. This 510(k) submission includes information on the Us2.ca technological characteristics, as well as performance data and verification and validation activities. Given the similarity of both devices, the enclosed information demonstrates that Us2.ca is as safe and effective as the predicate, and does not raise different questions of safety and effectiveness.
Page 10 of 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.