K243705 · DeepHealth, Inc. · QIH · Dec 19, 2024 · Radiology
Device Facts
Record ID
K243705
Device Name
Saige-Density (2.5.0)
Applicant
DeepHealth, Inc.
Product Code
QIH · Radiology
Decision Date
Dec 19, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Breast density category
Deep learning
—
—
Four datasets across various geographic locations within the US
—
Retrospective study as described in K222275
—
Indications for Use
Saige-Density is a software application intended for use with compatible full-field digital mammography (FFDM) and digital breast tomosynthesis (DBT) systems. Saige-Density provides an ACR BI-RADS Atlas 5th Edition breast density category to aid interpreting physicians in the assessment of breast tissue composition. Saige-Density produces adjunctive information. It is not a diagnostic aid.
Device Story
Saige-Density is SaMD using deep learning to process screening/diagnostic mammograms; inputs are 2D DICOM files (FFDM/2D synthetics) from x-ray mammography systems. Software generates study-level breast density category (BI-RADS A-D). Outputs are DICOM Structured Report (SR) and Secondary Capture (SC) summary report viewed on mammography workstations. Used by interpreting radiologists in clinical settings. Provides adjunctive information to assist in breast tissue composition assessment; not a diagnostic aid. Does not replace physician review; clinical decisions must not rely solely on device output. Benefits include standardized, automated breast density categorization to support clinical workflow.
Clinical Evidence
Validation performed via retrospective study (previously submitted in K222275). Data sourced from multiple US clinical sites, distinct from training data, ensuring no overlap. Training data included diverse geographic and racial populations (e.g., NYC, Los Angeles). Verification testing included software unit, integration, and system testing confirming requirements satisfaction.
Technological Characteristics
SaMD; deep learning algorithm. Inputs: DICOM FFDM/2D synthetic images. Outputs: DICOM SR and SC objects. Deployment: On-premise or cloud-enabled environment; single-execution mode. Connectivity: DICOM-compliant. Software version 2.5.0.
Indications for Use
Indicated for female patients aged 35 years or older undergoing mammography to aid interpreting physicians in assessing breast tissue composition.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 19, 2024
DeepHealth, Inc B. Nathan Hunt Head of Quality, Regulatory, and Compliance 212 Elm Street Somerville, Massachusetts 02144
Re: K243705
Trade/Device Name: Saige-Density (2.5.0) Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH Dated: November 29, 2024 Received: November 29, 2024
Dear B. Nathan Hunt:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}------------------------------------------------
the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# YANNA S. KANG -S
Yanna Kang, Ph.D. Assistant Director Mammograph and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K243705
Device Name
Saige-Density (2.5.0)
Indications for Use (Describe)
Saige-Density is a software application intended for use with compatible full-field digital mammography (FFDM) and digital breast tomosynthesis (DBT) systems. Saige-Density provides an ACR BI-RADS Atlas 5th Edition breast density category to aid interpreting physicians in the assessment of breast tissue composition. Saige-Density produces adjunctive information. It is not a diagnostic aid.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "deephealth" in a stylized, lowercase font. The color of the text is a vibrant purple. The letters are connected, giving the word a flowing appearance.
#### 510(k) Summary DeepHealth, Inc Saige-Density
K243705
In accordance with 21 CFR 807.92 the following summary of information is provided, on this date, November 28, 2024:
#### 1. 510(k) SUBMITTER
DeepHealth, Inc 212 Elm St. Somerville, MA 02144 Tel: 443.506.8911
#### Contact Person:
B. Nathan Hunt Head of Quality, Regulatory, and Compliance
DeepHealth, Inc 212 Elm St Somerville, MA 02144 Tel: 443.506.8911
Date Prepared:
November 28, 2024
#### 1. DEVICE
Trade Name of Device: Saige-Density
Device Model: v.2.5.0
Common or Usual Name: Medical Image Software
Regulation Name and Number: Medical Image Management and Processing System (21 CFR 892.2050)
Regulation Class: II
Product Code: QIH
#### 2. PREDICATE DEVICE
Trade Name of Device: Saige-Density
Device Model: v2.0.0
Common Name or Usual Name: Medical Image Software
Regulation Name and Number: Picture Archiving and Communication System (21 CFR 892.2050)
Regulation Class: II
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Deephealth in purple font. Below the logo, the text "Product Code: QIH" is written in bold, black font. The logo is simple and modern, with a clean typeface.
510(K) No.: K222275
# 3. DEVICE DESCRIPTION
Saige-Density is Software as a Medical Device that processes screening and diagnostic digital mammograms using deep learning techniques and generates outputs that serve as an aid for interpreting radiologists in assessing breast density. The software takes as input a single x-ray mammogram study and processes all acceptable 2D image DICOM files (FFDM and/or 2D synthetics) and generates a single study-level breast density category. Two DICOM files are outputted as a result: 1) a structured report (SR) DICOM object containing the case-level breast density category and 2) a secondary capture (SC) DICOM object containing a summary report with the study-level density category. Both output files contain the same breast density category ranging from "A" through "D" following Breast Imaging Reporting and Data System (BI-RADS) 5th Edition reporting guidelines. The SC report and/or the SR file may be viewed on a mammography viewing workstation.
# 4. INDICATIONS FOR USE
Saige-Density is intended for use with compatible full-field digital mammography (FFDM) and digital breast tomosynthesis (DBT) systems. Saige-Density provides an ACR BI-RADS Atlas 5th Edition breast density category to aid interpreting physicians in the assessment of breast tissue composition. Saige-Density produces adjunctive information. It is not a diagnostic aid.
# Intended User Population
The intended users of Saige-Density are interpreting physicians qualified to read mammography exams.
# Intended Patient Populations
The device is intended to be used on female patients thirty-five (35) years of age or older undergoing mammography.
### Warnings and Precautions
Saige-Density is an adjunct tool and is not intended to replace a physician's own review of a mammogram. Decisions should not be made solely based on analysis by Saige-Density.
# 5. PREDICATE DEVICE COMPARISON
Saige-Density and the predicate device have the same indications for use, patient population, technical characteristics, and principles of operation. The differences between Saige-Density and the predicate device do not alter the suitability of the subject device for its intended use, and do not raise different questions of safety or effectiveness.
The devices are intended to be used by physicians to aid in the assessment of breast density for mammograms. The devices are not intended to be used as a replacement of a physician's own clinical judgment.
The design of Saige-Density is similar to that of the predicate device. Both devices are compatible with FFDM and DBT mammograms, utilize deep learning to produce an output of a
{6}------------------------------------------------
patient-level breast density category, and are not diagnostic aids that are intended to be used by physicians interpreting either screening or diagnostic mammograms. As both devices use the same proprietary algorithm, there are no differences in the algorithmic components, only minor differences in the technological aspects. Compared to the predicate device, the subject device includes a new deployment option (single-execution mode), added consideration of deployment in a cloud-enabled environment, as well as upgrades to improve speed and memory use. Non-clinical and clinical testing has been completed as part of the prior K222275 submission ensuring the safety and effectiveness of the proposed subject device.
# 6. PERFORMANCE DATA
Verification testing included software unit testing, software integration testing, and system testing. Testing confirmed that the software, as designed and implemented, satisfies the software requirements.
Validation of the software was performed using a retrospective study as described in K222275. The data used in the validation testing was obtained from different clinical sites than those used to develop the Saige-Density algorithm. DeepHealth ensured that there was no overlap between the data used to train and test the Saige-Density algorithm. The data used to train the Saige-Density algorithm consisted of four datasets across various geographic locations within the US, including racially diverse regions such as New York City and Los Angeles.
# 7. CONCLUSION
Verification and Validation testing conducted to support this submission confirm that Saige-Density is safe and effective for its intended use. The minor technological differences between subject and predicate versions of Saige-Density do not alter the intended use of the device and do not affect its safety and effectiveness when used as labeled. Therefore, the information presented in this 510(k) submission demonstrates that Saige-Density is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.