Dice Similarity Coefficient > 0.951; 95% Hausdorff Distance 0.98 mm – 7.31 mm
Dice Similarity Coefficient 0.959 [0.955 – 0.963]; 95% Hausdorff Distance 1.367 mm [1.170 mm – 1.563 mm]
—
—
Performance testing dataset: 72 patient images
>1 (US board-certified radiologists)
Indications for Use
FUJIFILM Synapse PACS Software is intended for use as a web-based application on an off-the shelf PC which meets or exceeds minimum specifications and is networked with a FUJIFILM Synapse PACS server. The FUJIFILM Synapse PACS Software can process medical images from DICOM compliant modalities and non-DICOM sources. FUJIFILM Synapse PACS Software provides toolsets for: Performing measurements on DICOM images Regional segmentation Importing and presenting data from modalities (DICOM and non-DICOM), Solving clinical calculations Creating and distributing structured reports FUJIFILM Synapse PACS Software is intended to serve as the primary user interface for the processing of medical images for presentation on displays appropriate to the medical task being performed. It enables the display, comparison, fusion, and volume rendering of studies to aid in reading, interpreting, reporting, and treatment planning. Typical users are radiologists, cardiologists, technologists, sonographers, technicians, nurses, and clinicians. MIP, MPR Fusion, and volume rendering are not intended for mammography use. FUJIFILM Synapse PACS Software can be used to process FUJIFILM’s DICOM MG “For Processing” images and also for the display, manipulation, and interpretation of lossless compressed or non-compressed mammography images that have been received in the DICOM For Presentation format and displayed on FDA-cleared, DICOM compatible displays for mammography.
Device Story
Synapse PACS (7.5) is an enterprise-wide medical image management software (SaMD) running on off-the-shelf PCs. It ingests DICOM and non-DICOM medical images; provides tools for measurement, regional segmentation, clinical calculations, and structured reporting. Key features include volume rendering, 3D tools, 2-point VOI (semi-automatic lesion segmentation), and AI-based bone removal. Used in clinical settings by radiologists, cardiologists, and technologists to facilitate image interpretation and treatment planning. The device integrates algorithms previously cleared in Synapse 3D (K221677). Output is presented on clinical displays; aids clinicians in diagnostic reading and reporting. Benefits include enhanced visualization of vessels via bone masking and streamlined 3D analysis workflows.
Clinical Evidence
Bench testing only. Performance of the bone extraction algorithm was validated using 72 patient images (35 male, 37 female; ages 22-120) from Siemens, GE, and Philips scanners. Primary endpoint: Dice Similarity Coefficient (DSC); secondary: 95% Hausdorff Distance (HD). Mean DSC was 0.959 [0.955–0.963], exceeding the 0.951 threshold. Mean 95% HD was 1.367 mm [1.170–1.563 mm], within the 0.98–7.31 mm range. Ground truth established via dual-reader consensus by board-certified radiologists.
Technological Characteristics
Web-based SaMD running on off-the-shelf PC hardware. Features include DICOM/non-DICOM image processing, volume rendering, 3D tools, and AI-based bone removal. Connectivity via network to Synapse PACS server. Software-based segmentation and rendering algorithms.
Indications for Use
Indicated for use by radiologists, cardiologists, technologists, sonographers, technicians, nurses, and clinicians for processing, displaying, comparing, and interpreting medical images (DICOM/non-DICOM) to aid in treatment planning and reporting. Not intended for mammography use regarding MIP, MPR Fusion, and volume rendering.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
FUJIFILM Healthcare Americas Corporation
Kulkarni Chaitrali
Sr. Regulatory Affairs Specialist
81 Hartwell Ave Suite 300
LEXINGTON, MASSACHUSETTS 02421
June 30, 2025
Re: K243647
Trade/Device Name: Synapse PACS (7.5)
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH
Dated: June 3, 2025
Received: June 3, 2025
Dear Kulkarni Chaitrali:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243647 - Kulkarni Chaitrali
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K243647 - Kulkarni Chaitrali
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Jessica Lamb, PhD
Assistant Director
Imaging Software Team
DHT8B: Division of Radiologic Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
## Submission Number (if known)
K243647
## Device Name
Synapse PACS (7.5)
## Indications for Use (Describe)
FUJIFILM Synapse PACS Software is intended for use as a web-based application on an off-the shelf PC which meets or exceeds minimum specifications and is networked with a FUJIFILM Synapse PACS server.
The FUJIFILM Synapse PACS Software can process medical images from DICOM compliant modalities and non-DICOM sources.
FUJIFILM Synapse PACS Software provides toolsets for:
- Performing measurements on DICOM images
- Regional segmentation
- Importing and presenting data from modalities (DICOM and non-DICOM),
- Solving clinical calculations
- Creating and distributing structured reports
FUJIFILM Synapse PACS Software is intended to serve as the primary user interface for the processing of medical images for presentation on displays appropriate to the medical task being performed. It enables the display, comparison, fusion, and volume rendering of studies to aid in reading, interpreting, reporting, and treatment planning.
Typical users are radiologists, cardiologists, technologists, sonographers, technicians, nurses, and clinicians.
MIP, MPR Fusion, and volume rendering are not intended for mammography use. FUJIFILM Synapse PACS Software can be used to process FUJIFILM’s DICOM MG “For Processing” images and also for the display, manipulation, and interpretation of lossless compressed or non-compressed mammography images that have been received in the DICOM For Presentation format and displayed on FDA-cleared, DICOM compatible displays for mammography.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}
| 510(k) #: K243647 | 510(k) Summary | Prepared on: 2025-06-23 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | FUJIFILM Healthcare Americas Corporation | |
| Applicant Address | 81 Hartwell Ave Suite 300 Lexington MA 02421 United States | |
| Applicant Contact Telephone | 704-517-4886 | |
| Applicant Contact | Ms. Kulkarni Chaitrali | |
| Applicant Contact Email | hcusregulatoryaffairs@fujifilm.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | Synapse PACS (7.5) | |
| Common Name | Medical image management and processing system | |
| Classification Name | Automated Radiological Image Processing Software | |
| Regulation Number | 892.2050 | |
| Product Code(s) | QIH | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K190232 | Synapse PACS | LLZ |
| K221677 | Synapse 3D Base Tools v6.6 | LLZ |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| The Synapse PACS is an enterprise-wide medical information and image management software that runs on standard “off-the-shelf” PC hardware and Software (OS, browser). Synapse is intended for communication, storage, display, manipulation, measurement, printing, and processing of images and information acquired from various medical imaging and information systems. As a Software as a Medical Device (SaMD), Synapse PACS performs these purposes without being part of a hardware medical device. | | |
| Intended Use/Indications for Use | | 21 CFR 807.92(a)(5) |
| FUJIFILM Synapse PACS Software is intended for use as a web-based application on an off-the shelf PC which meets or exceeds minimum specifications and is networked with a FUJIFILM Synapse PACS server.
The FUJIFILM Synapse PACS Software can process medical images from DICOM compliant modalities and non-DICOM sources.
FUJIFILM Synapse PACS Software provides toolsets for:
• Performing measurements on DICOM images
• Regional segmentation
• Importing and presenting data from modalities (DICOM and non-DICOM),
• Solving clinical calculations
• Creating and distributing structured reports | | |
{5}
FUJIFILM Synapse PACS Software is intended to serve as the primary user interface for the processing of medical images for presentation on displays appropriate to the medical task being performed. It enables the display, comparison, fusion, and volume rendering of studies to aid in reading, interpreting, reporting, and treatment planning.
Typical users are radiologists, cardiologists, technologists, sonographers, technicians, nurses, and clinicians.
MIP, MPR Fusion, and volume rendering are not intended for mammography use. FUJIFILM Synapse PACS Software can be used to process FUJIFILM's DICOM MG "For Processing" images and also for the display, manipulation, and interpretation of lossless compressed or non-compressed mammography images that have been received in the DICOM For Presentation format and displayed on FDA-cleared, DICOM compatible displays for mammography.
## Indications for Use Comparison
21 CFR 807.92(a)(5)
There are no differences in the indication for use between the subject device and the primary predicate device.
## Technological Comparison
21 CFR 807.92(a)(6)
The differences between the features in the subject device and the primary predicate device are Volume Rendering Techniques, 3D Tools, Bone Removal, and 2-Point VOI. Although the primary predicate device does not have the features listed, the reference device have the same features. Because the predicate device and reference device contain the same features, the addition of the features in version 7.5 do not affect the safety or efficacy of the subject device.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
Non-clinical testing result:
The differences between the features in the subject device and the primary predicate device are Volume Rendering Techniques including 3D Tools, 2-Point Volume of Interest (VOI), and Bone Removal.
The key addition to Synapse PACS 7.5.0 is the ability to perform volume rendering and 3D volume viewing for CT and MR. The algorithm is already cleared and marketed for Synapse 3D (K221677, reference device). The volume rendering algorithms were integrated unchanged into Synapse PACS 7.5.0 (subject device).
The 2-point VOI performs is a semi-automatic segmentation of a lesion in CT and MRI images using the line drawn by the user on the lesion. The algorithm implemented is the same as already cleared and marketed algorithms for Synapse 3D (K221677, reference device). In Synapse3D, it is called Tumor Boundary Segmentation algorithm. The algorithm has been integrated into Synapse PACS unchanged compared to Synapse 3D. Comparison with Synapse 3D (K221677, reference device) provides the same results in Synapse PACS 7.5.0 (subject device).
Bone Removal is a tool that enhances the visibility of vessels in 3D rendered images by masking out bone regions. The tool is based on an AI algorithm cleared and marketed for Synapse 3D (K221677, reference device). It was improved for Synapse PACS 7.5.0. Comparison with Synapse 3D (K221677, reference device) showed that the improved algorithm does not affect the safety or efficacy of Synapse PACS 7.5.0 (subject device). Please refer to the Performance Bench Testing for testing with the improved algorithm.
Performance Testing Summary
A. Summary test statistics and acceptance criteria:
Dice Similarity Coefficient (DSC) and 95% Hausdorff Distance (HD) were used as primary and secondary endpoints to validate the performance of bone extraction algorithm. A pre-defined acceptance threshold of 0.951 was used for DSC. For 95% HD, we pre-define acceptance range 0.98 mm – 7.31 mm. The thresholds are determined after reviewing existing bone removal models in the literature.
The mean observed dice overlap coefficient and 95% HD with 95% confidence intervals were 0.959 [0.955 – 0.963] and 1.367 mm [1.170 mm – 1.563 mm], respectively, exceeding the predefined acceptance thresholds.
Across the subgroups—region, sex, age band, vendor, slice-thickness bin, body region, contrast—Dice values remained tightly clustered between 0.94 and 0.97 and 95% HD never exceeded 2.57 mm. The 5 mm cohort shows a slightly lower Dice (0.942) because thick slices compress bone detail into fewer voxels, so tiny absolute deviations inflate the overlap metric; the boundary error remains modest (HD = 2.6 mm), confirming clinically acceptable performance.
{6}
B. The number of individual patient's images were collected from.
Images from 72 patients were collected, each patient had one image collected.
C. Demographic Distribution
Gender:
35 Males, 37 Females
Ages:
16 patients from ages 22-34
27 patients from ages 35-64
29 patients from ages 65-120
Regions:
20 patients from Midwest
20 patients from Southwest
32 patients from the Southeast
D. Information about image regions and confounders present in the dataset
Head/Neck: 16 images
Chest/Abdomen: 25 images
Abdomen/Pelvic: 19 images
Extremities: 6 images
Vascular/Others: 6 images
E. Information about equipment and protocols used to collect images
Images were collected from the following manufacturers:
28 SIEMENS, 22 GE, and 22 Philips
Slice thickness:
≤1.25mm: 29 images
1.25-2.5mm: 20 images
2.5-5.0mm: 23 images
≥5mm: 6 images
Contrast agent:
Contrast: 50 images
Non-contrast: 22 images
F. Information about how the reference standard was derived from the dataset (i.e., the "truthing" process)
An initial bone mask was created by certified technologist, then subjected to an independent dual-reader consensus review: two U.S. board-certified radiologists independently evaluated the mask, recorded any discrepancies, and iteratively reconciled them until consensus was achieved. The resulting consensus mask serves as the definitive ground truth for performance testing.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.