Standalone study: 614 BW images and 684 IOPA images
3 (US licensed dentists)
Carious lesion detection (reader improvement)
Computer vision models with pre-processing, core, and post-processing modules
—
BW AUC 0.848 (aided) vs 0.806 (unaided); IOPA AUC 0.845 (aided) vs 0.807 (unaided)
—
—
MRMC study: 328 digital radiographs (108 BW and 220 IOPA)
29 (US licensed dentists)
Indications for Use
Better Diagnostics Caries Assist is a radiological, automated, concurrent read, CADe software intended to identify and localize carious lesions on bitewings and periapical radiographs acquired from patients aged 18 years or older. Better Diagnostics Caries assist is indicated for use by board licensed dentists. The device is not intended as a replacement for a complete dentist's review or their clinical judgment that takes into account other relevant information from the image, patient history or actual in vivo clinical assessment.
Device Story
BDCA v1.0 is a CADe software for automated detection of carious lesions in bitewing and periapical dental radiographs. Input: digital X-ray images (PNG, BMP, JPG) from dental practice management systems. Processing: cloud-hosted computer vision models perform pre-processing (standardization/type identification), core detection, and post-processing (outlier removal). Output: bounding box annotations on a web-based 'AI Results Screen' or a message stating 'No Suspected: Caries Detected'. Used in dental clinics by dentists; software integrates via API with existing imaging/practice management software. Dentists review AI-annotated images concurrently with original images to assist in diagnosis. Device does not provide treatment recommendations or final diagnosis; serves as an adjunct tool to improve detection sensitivity and specificity. Benefits include enhanced diagnostic accuracy and reduced false positives in caries identification.
Clinical Evidence
Clinical evidence includes a standalone study and a multi-reader, multi-case (MRMC) study. Standalone study (614 BW, 684 IOPA images) showed BW sensitivity 89.2%, specificity 99.5%; IOPA sensitivity 88.2%, specificity 99.1%. MRMC study (328 radiographs, 29 US dentists) compared unaided vs. aided performance. Aided AFROC AUCs (0.848 BW, 0.845 IOPA) significantly outperformed unaided (0.806 BW, 0.807 IOPA, p<0.001). Aided sensitivity and specificity at both image and surface levels showed statistically significant improvements over unaided interpretation.
Technological Characteristics
Cloud-hosted CADe software; processes PNG, BMP, JPG images. Uses computer vision models for binary caries detection and bounding box coordinate generation. Web-based presentation layer (Angular.js/node.js). Complies with IEC 62304 (software lifecycle), IEC 62366-1 (usability), ISO 14971 (risk management), and ISO 15223-1 (symbols). Connectivity via API to dental practice management systems.
Indications for Use
Indicated for board licensed dentists to identify and localize carious lesions on bitewing and periapical radiographs in patients aged 18+ with permanent dentition. Not a replacement for clinical judgment or comprehensive review.
Regulatory Classification
Identification
Medical image analyzers, including computer-assisted/aided detection (CADe) devices for mammography breast cancer, ultrasound breast lesions, radiograph lung nodules, and radiograph dental caries detection, is a prescription device that is intended to identify, mark, highlight, or in any other manner direct the clinicians' attention to portions of a radiology image that may reveal abnormalities during interpretation of patient radiology images by the clinicians. This device incorporates pattern recognition and data analysis capabilities and operates on previously acquired medical images. This device is not intended to replace the review by a qualified radiologist, and is not intended to be used for triage, or to recommend diagnosis.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the image analysis algorithms including a description of the algorithm inputs and outputs, each major component or block, and algorithm limitations.
(ii) A detailed description of pre-specified performance testing methods and dataset(s) used to assess whether the device will improve reader performance as intended and to characterize the standalone device performance. Performance testing includes one or more standalone tests, side-by-side comparisons, or a reader study, as applicable.
(iii) Results from performance testing that demonstrate that the device improves reader performance in the intended use population when used in accordance with the instructions for use. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, predictive value, and diagnostic likelihood ratio). The test dataset must contain a sufficient number of cases from important cohorts (*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, concomitant diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment.(iv) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results; and cybersecurity).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use.
(ii) A detailed description of the intended reading protocol.
(iii) A detailed description of the intended user and user training that addresses appropriate reading protocols for the device.
(iv) A detailed description of the device inputs and outputs.
(v) A detailed description of compatible imaging hardware and imaging protocols.
(vi) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality or for certain subpopulations), as applicable.(vii) Device operating instructions.
(viii) A detailed summary of the performance testing, including: test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 11, 2025
Better Diagnostics AI Corp % Rajeev Nohria CEO 29 Alcott Way AVON, CT 06001
Re: K241725
Trade/Device Name: Better Diagnostics Caries Assist (BDCA) Version 1.0 Regulation Number: 21 CFR 892,2070 Regulation Name: Medical Image Analyzer Regulatory Class: Class II Product Code: MYN Dated: February 10, 2025 Received: February 10, 2025
Dear Rajeev Nohria:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K241725
Device Name
Better Diagnostics Caries Assist (BDCA) Version 1.0
#### Indications for Use (Describe)
Better Diagnostics Caries Assist is a radiological, automated, concurrent read, CADe software intended to identify and localize carious lesions on bitewings and periapical radiographs acquired from patients aged 18 years or older. Better Diagnostics Caries assist is indicated for use by board licensed dentists. The device is not intended as a replacement for a complete dentist's review or their clinical judgment that takes into account other relevant information from the image, patient history or actual in vivo clinical assessment.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the logo for BETTERDIAGNOSTICS.AI. The logo consists of a stylized red symbol resembling two intertwined letters, possibly 'b' and 'd'. Below the symbol, the text "BETTERDIAGNOSTICS.AI" is written in a sans-serif font. The text is black and appears to be the name of the company or organization.
Better Diagnostics AI Corp. 29 Alcott Way Avon CT, 06001 USA
# 510(k) Summary
# 1. General Information
| 510(k) Owner | Better Diagnostics AI Corp. |
|-----------------------|-------------------------------------------------------|
| Address | 29 Alcott Way Avon CT, 06001<br>USA |
| Correspondence person | Raj Zaveri<br>Quality and Regulatory Consultant |
| Contact information | Email: rajzaveri10@gmail.com<br>Phone: +91 8369438297 |
| Date prepared | March 23, 2024 |
#### 2. Proposed device
Proprietary Name of Device: Better Diagnostics Caries Assist (BDCA) Version 1.0
Common or Usual Name: Medical image analyzer
Classification Panel: Radiology
Regulation Number: 21 CFR § 892.2070
Regulation Name: Medical image analyzer
Regulation Class: Class II
Product Code: MYN
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for BETTERDIAGNOSTICS.AI. The logo consists of a stylized, interconnected "b" and "d" in red, positioned above the company name. The text "BETTERDIAGNOSTICS.AI" is written in a bold, sans-serif font, with the ".AI" portion slightly smaller than the rest of the text.
#### 3. Predicate Device
Proprietary Name of Device: Overjet Caries Assist Software (Version 1.0)
Premarket Notification: K222746
Common or Usual Name: Medical image analyzer
Classification Panel: Radiology
Regulation Number: 21 CFR § 892.2070
Regulation Name: Medical image analyzer
Regulation Class: Class II
Product Code: MYN
#### 4. Device Description
Better Diagnostics Caries Assist (BDCA) Version 1.0 is a computer-aided detection (CADe) software designed for the automated detection of carious lesions in Bitewings and periapical dental radiographs. This software offers supplementary information to assist clinicians in their diagnosis of potentially carious tooth surfaces. It is important to note that BDCA v1.0 is not meant to replace a comprehensive clinical evaluation by a clinician, which should consider other pertinent information from the image, the patient's medical history and clinical examination. This software is intended for use in identifying carious lesions in permanent teeth of patients who are 18 years or older.
BDCA v1.0 does not make treatment recommendations or provide a diagnosis. Dentists should review images annotated by BDCA v1.0 concurrently with original, unannotated images before making the final diagnosis on a case. BDCA v1.0 is an adjunct tool and does not replace the role of the dentist. The CAD generated output should not be used as the primary interpretation by the dentists. BDCA v1.0 is not designed to detect conditions other than the following: Caries.
BDCA v1.0 comprises four main components:
- . Presentation Layer: This component includes "AI Results Screen" a web-based interface (user interface) that allows users to view AI marked annotations. This is a custom code provided by Better Diagnostics AI Corp to Dental PMS customers. User Interface uses Angular.js and node.js technology to show images on the "AI Results Screen". System can process PNG, BMP and JPG format images. All images are converted into JPEG format for processing. Computer Vision Models
| | Better Diagnostics AI Corp, 510(k) submission | Page 2 of 15 |
|--|-----------------------------------------------|--------------|
|--|-----------------------------------------------|--------------|
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for BetterDiagnostics.ai. The logo consists of a red symbol that looks like two intertwined lowercase letters 'b' and 'd'. Below the symbol, the text "BETTERDIAGNOSTICS.AI" is written in a sans-serif font. The text is black and appears to be the name of the company or organization.
returns AI annotations and co-ordinates to the business layer. Business layer sends coordinates to the presentation layer and bounding boxes are drawn on the image using custom code written in Angular.js and node.js. Dentists can view, accept or reject the annotations based on his evaluation. Better Diagnostics AI provides UI code to customers e.g. dental practice management software and imaging firms for utilization of BCDA v1.0 software.
- Application Programming Interface (API): APIs are a set of definitions and . protocols for building and integrating application software. It's sometimes referred to as a contract between an information provider and an information user. BDCA v1.0 APIs connect the Dental PMS with the business layer. API receives images input from Dental PMS and passes it to the business layer. It also receives annotations and co-ordinates from the business layer and passes it to the presentation layer hosted by Dental PMS.
- . Business Layer: Receives image from the API Gateway and passes it to computer vision models. It also receives the bounding boxes coordinates from the model and retrieves images from the cloud storage. It sends all information to the "AI Results screen" to display rectangle bounding boxes.
- Computer Vision Models (CV Models): These models are hosted on a cloud . computing platform and are responsible for image processing. They provide a binary indication to determine the presence of carious findings. If carious findings are detected, the software will output the coordinates of the bounding boxes for each finding. If no carious lesions are found, the output will not contain any bounding boxes and will have a message stating "No Suspected: Caries Detected"
AI models have three parts:
- о Pre-Processing Module: Standardization of image to specific height and width to maintain consistency for AI model. Finds out the type of image including IOPA, Bitewings or other types. BDCA v 1.0 can only process Bitewings and IOPA images for patients over age 18. All other types of images will be rejected.
- Core Module: This module provides carious lesion annotations O and co-ordinates to draw bounding boxes.
- Post-Processing Module: includes cleanup process to remove o outliers/incorrect annotations from the images.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Betterdiagnostics.ai. The logo consists of the letters 'b' and 'd' intertwined in red. Below the logo, the text "BETTERDIAGNOSTICS.AI" is written in black, all caps. The logo is simple and modern.
#### 5. Intended Use/Indications for Use
Better Diagnostics Caries Assist is a radiological, automated, concurrent read, CADe software intended to identify and localize carious lesions on bitewings and periapical radiographs acquired from patients aged 18 years or older. Better Diagnostics Caries assist is indicated for use by board licensed dentists. The device is not intended as a replacement for a complete dentist's review or their clinical judgment that takes into account other relevant information from the image, patient history or actual in vivo clinical assessment.
#### 6. Intended Patient Population
The intended patient population of the device is patients that have permanent dentition, and who are at least 18 years of age.
#### 7. Substantial Equivalence
| | Better Diagnostics AI Corp, 510(k) submission | Page 4 of 15 |
|--|-----------------------------------------------|--------------|
|--|-----------------------------------------------|--------------|
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for BETTERDIAGNOSTICS.AI. The logo features the letters 'bd' in a stylized, interconnected design, rendered in red. Below the logo, the words 'BETTERDIAGNOSTICS.AI' are written in a simple, sans-serif font.
| | Proposed Device | Primary Predicate<br>Device | Substantially<br>equivalent |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| 510(k) Number | K241725 | K222746 | - |
| Applicant | Better Diagnostics<br>AI Corp. | Overjet, Inc. | - |
| Device Name | Better Diagnostics<br>Caries Assist | Overjet Caries Assist | - |
| Classification<br>Regulation | 21 CFR 892.2070 | 21 CFR 892.2070 | Yes |
| Product Code | MYN | MYN | Yes |
| Indications for use | Better Diagnostics<br>Caries Assist is a<br>radiological,<br>automated,<br>concurrent<br>read,<br>CADe<br>software<br>intended to identify<br>and localize on<br>carious lesions on<br>bitewings and<br>periapical<br>radiographs<br>acquired from<br>patients aged 18<br>years or older.<br>Better Diagnostics | Overjet Caries Assist<br>(OCA) is a radiological,<br>automated, concurrent<br>read, CADe software<br>intended to aid in the<br>detection and<br>segmentation of caries<br>on bitewing and<br>periapical radiographs.<br>The device provides<br>additional information<br>for the dentist to use in<br>their diagnosis of a<br>tooth surface suspected<br>of being carious. The<br>device is not intended | Yes1 |
| | Proposed Device | Primary Predicate<br>Device | Substantially<br>equivalent |
| | Caries assist is<br>indicated for use by<br>board licensed<br>dentists. The<br>device is not<br>intended as a<br>replacement for a<br>complete dentist's<br>review or their<br>clinical judgment<br>that takes into<br>account other<br>relevant<br>information from<br>the image, patient<br>history, or actual in<br>vivo clinical<br>assessment. | as a replacement for a<br>complete<br>dentist's<br>review or their clinical<br>judgment that takes into<br>account other relevant<br>information from the<br>image, patient history,<br>or actual in vivo<br>clinical assessment. | |
| Image Modality | Radiograph | Radiograph | Yes |
| Study Type | Bitewing and<br>periapical Images | Bitewing and periapical<br>Images | Yes |
| Reader workflow | Concurrent | Concurrent | Yes |
| Clinical Output | Message indicating<br>if a carious lesion<br>was detected.<br>A bounding box is<br>created to indicate | Caries detection and<br>segmentation on<br>radiograph resulting in<br>outline or fill of<br>suspected caries. | Yes2 |
| | Proposed Device | Primary Predicate<br>Device | Substantially<br>equivalent |
| | suspected carious<br>lesions. | | |
| Patient Population | 18 years or older<br>with permanent<br>teeth | Patients requiring<br>dental services, all<br>sexes, 12 years of age<br>or older with permanent<br>teeth | Yes3 |
| Platform | Web - Edge,<br>Chrome, Firefox | Web - Edge, Chrome,<br>Firefox | Yes |
| OS | Any | Any | Yes |
| User Interface | Mouse, Keyboard,<br>Trackpad | Mouse, Keyboard,<br>Trackpad | Yes |
| Intended User | Dentists | Dentist | Yes |
| Image Source | Images<br>are<br>imported<br>from<br>dental<br>PMS/Imaging<br>software<br>from<br>multiple<br>sensor<br>manufacturers | Images imported from<br>the radiographic device,<br>or from the<br>practice<br>management<br>system<br>from multiple sensor<br>manufacturers | Yes |
| | Proposed Device | Primary Predicate<br>Device | Substantially<br>equivalent |
| Image Format | Accepts image<br>formats JPG,<br>PNG, and BMP | JPG, PNG, EOP, TIFF, DICOM | Yes4 |
| Processing<br>Architecture | 1, APIs - Dental<br>PMS/Imaging<br>software send x-ray<br>image to BDCA<br>2. Computer vision<br>models process the<br>x-ray image and<br>send results<br>presentation layer<br>3. Presentation<br>layers is plots<br>rectangle shape<br>annotations on the<br>x-ray image and it<br>is displayed to<br>dentist using "AI<br>Results screen" | Three layers:<br>1 - The Network layer<br>works with the practice<br>PACS or EMR to<br>transmit the image and<br>meta-data to Overjet.<br>2 - The decision layer<br>processes the image to<br>ensure it is the correct<br>data type, and then<br>annotates it via the<br>algorithm<br>3 - The presentation<br>layer displays the<br>annotated image in a<br>non-diagnostic viewer.<br>The dentist can filter,<br>display, hide, create and<br>edit the annotations<br>presented. | Yes* |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the logo for BETTERDIAGNOSTICS.AI. The logo consists of a red graphic element resembling two intertwined letters, possibly 'b' and 'd'. Below the graphic, the text "BETTERDIAGNOSTICS.AI" is written in a bold, sans-serif font. The text is black, providing a strong contrast against the white background.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image contains the logo for BETTERDIAGNOSTICS.AI. The logo consists of the text "BETTERDIAGNOSTICS.AI" in a sans-serif font. Above the text is a red graphic that appears to be two intertwined letters, possibly "b" and "d". The logo is simple and modern in design.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image contains the logo for BETTERDIAGNOSTICS.AI. The logo consists of a stylized red graphic above the company name. The graphic appears to be two intertwined letters, possibly 'b' and 'd'. The text is in a bold, sans-serif font.
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image contains a logo for a company called BETTERDIAGNOSTICS.AI. The logo consists of a stylized, red-colored graphic that resembles two intertwined letters, possibly 'b' and 'd', or an infinity symbol. Below the graphic, the company name "BETTERDIAGNOSTICS.AI" is written in a simple, sans-serif font. The text is black and appears to be the primary identifier for the company.
BDCA v1.0 and Overjet Caries Assist are both class II devices with the same product classification code - MYN. Both devices are radiological, automated, concurrent read, CADe software that are used to detect caries on bitewing and periapical radiographs. BDCA v1.0 has minor differences compared to the predicate device in terms of the following:
- . Yes! - BDCA v1.0 detects caries by drawing coloured bounding boxes on bitewing and periapical radiographs whereas Overjet caries assist uses segmentation to outline the carious regions on bitewing and periapical radiographs. These are two different approaches prevalent for automated caries detection. Choosing either approach has no impact whatsoever on the outcome. This is supported by the results of the Multi-Reader, Multi-Case (MRMC) study, where BDCA v1.0 demonstrated robust performance in detecting caries lesions with sensitivity improvements as discussed in section 8.a.ii given below. Additionally, the standalone study showed that BDCA significantly improved cavity detection capability without compromising specificity, with consistent performance across various clinical settings and variables.
- . Yes2 - BDCA v1.0 localizes caries via bounding boxes on bitewing and periapical radiographs whereas Overjet caries assist uses segmentation to outline or fill the carious regions on bitewing and periapical radiographs. Both lead to the same clinical outcome i.e detection of caries.
- . Yes3 - BDCA v1.0 is intended to mark carious regions on radiographs acquired from patients aged 18 years or above, whereas Overjet carries assist marks carious regions on radiographs acquired from patients aged 12 years or above. This additional functionality on the Overjet's part has no impact whatsoever on the outcome of BDCA v1.0. Further, the software verification and validation performed as per IEC 62304 validates the performance of BDCA v 1.0.
- Yes4 BDCA v1.0 supports BMP, JPG, and PNG and Overjet supports JPG, . PNG, EOP, TIFF, DICOM. The additional formats supported by Overjet has no impact whatsoever on the outcome of BDCA v1.0. Further, the software verification and validation performed as per IEC 62304 validates this claim as well as the performance of BDCA v1.0.
# 8. Performance Testing
Safety and performance of BDCA v1.0 has been evaluated and verified in accordance with software specifications and the following applicable
| | | Better Diagnostics AI Corp, 510(k) submission | Page 9 of 15 |
|--|--|-----------------------------------------------|--------------|
|--|--|-----------------------------------------------|--------------|
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image contains the logo for Betterdiagnostics.ai. The logo consists of a stylized, interconnected "bd" in red, positioned above the text "BETTERDIAGNOSTICS.AI" in black. The logo is simple and modern, with a focus on the company's name and domain.
performance standards through software verification and validation. identification and mitigation of device-related hazards via cybersecurity and risk management, labeling validation, human factors testing, standalone and clinical performance testing :
- . IEC 62304 Edition 1.1 2015-06 Medical device software - Software life cycle processes
- IEC 62366-1:2015: Medical devices-Part 1: Application of usability . engineering to medical devices.
- ISO 14971 Third Edition 2019-12 Medical Devices Application of . risk management to medical devices.
- ISO 15223-1:2021: Medical devices Symbols to be used with information to be supplied by the manufacturer.
Additionally, the software validation activities were performed in accordance with the FDA Guidance documents, "Content of Premarket Submissions for Device Software Functions" and "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions".
Performance data from standalone and clinical performance assessments are summarized below.
#### a. Standalone and clinical performance assessments
Better Diagnostic has conducted performance testing according to FDA's "Guidance for Industry and Food and Drug Administration Staff Computer-Assisted Detection Devices Applied to Radiology Images and Radiology Device Data - Premarket Notification [510(k)] Submissions Document" and the "Clinical Performance Assessment: Considerations for Computer-Assisted Detection Devices Applied to Radiology Images and Radiology Device Data in Premarket Notification [510(k)] Submissions" Guidance, as part of the development process of the caries model. Performance testing included standalone testing and a clinical reader evaluation. All testing demonstrated that BDCA met prespecified requirements.
#### i. Standalone Testing
In the standalone study, BW and IOPA images were analyzed separately. A patient had at most one BW and one IOPA image included in the analysis dataset. For BW images, out of 614 images, 310 indicated the presence of cavities. Among the 15.687 surfaces examined within these images, 585 were positively identified with cavities, while 15,102 were
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image contains the logo for BetterDiagnostics.ai. The logo consists of a red symbol that looks like two interlocking letters, possibly 'b' and 'd'. Below the symbol, the text "BETTERDIAGNOSTICS.AI" is written in a sans-serif font. The text is black and appears to be the name of the company or organization.
negative. Similarly, for IOPA images, of the 684 images analyzed, 367 showed indications of cavities. Out of 9,253 surfaces reviewed, 618 showed positive results for cavities, and 8,635 were negative. The Standalone analysis demonstrated the use of Intersection over Union (IOU) 0.5.
Ground truth was determined through the consensus of two out of three experienced, licensed dentists, each with over 10 years of professional experience. These dentists examined and labeled dental surfaces, agreeing on the final labels for analysis when at least two dentists identified a surface as carious. The comprehensive analyses targeted both primary and secondary endpoints, focusing on the sensitivity and specificity at both the surface and image levels.
The primary analysis in the standalone study was conducted at the surface level, comparing the performance metrics-BW Surface Level Sensitivity, BW Surface Level Specificity, IOPA Surface Level Sensitivity, and IOPA Surface Level Specificity-against predefined performance goals. As a maximum of one BW and one IOPA image was considered for each patient, avoiding inter-image correlation. However, multiple cavities (surfaces) within an image might lead to potential positive correlations, hence unadjusted 95% CIs are reported at the surface level, with adjusted CIs provided using the bootstrapping method. The analysis at the image level was conducted under two definitions, namely Conservative Definition and Optimistic Definition (per the FDA). Under Conservative Definition: A TP is defined as an image being classified as TP if all lesions marked by the GT within the image are also accurately marked by the AI device. Under Optimistic Definition: A TP is defined as an image being classified as TP if at least one lesion within the image is labeled as "present" (indicating a cavity) by both the device prediction and the GT.
The BDCA demonstrated robust performance in detection of caries lesions.
# Primary Analysis:
BW Surface Level: The BDCA achieved a sensitivity of 89.2% with an adjusted 95% CI of [86.15%, 92.13%], and a specificity of 99.5% with a CI of [99.32%, 99.57%]. These figures significantly surpass the predefined performance goals of 0.74 and 0.95 respectively, illustrating the device's high accuracy in identifying caries at the surface level.
· IOPA Surface Level: The device reached a sensitivity of 88.2% with a CI of [85.27%, 90.78%] and a specificity of 99.1%
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image contains the logo for Betterdiagnostics.ai. The logo consists of the letters 'bd' in a stylized, interconnected design, rendered in red. Below the logo, the text 'BETTERDIAGNOSTICS.AI' is written in a simple, sans-serif font.
with a CI of [98.88%, 99.31%], exceeding the goals of 0.76 and 0.95 respectively. This indicates a robust precision in detecting surface caries within IOPA images, supported by statistically significant outcomes.
# Secondary Analysis:
· BW Image Level: Sensitivity under conservative conditions was reported at 81.0%, with a CI of [76.15%, 85.18%], and under optimistic conditions, it improved to 91.9%, with a CI of [88.33%, 94.71%]. Specificity remained consistent at 98.4% across definitions with a CI of [96.20%, 99.44%]. These results validate the BDCA's capacity to maintain high diagnostic accuracy across varied operational settings, emphasizing the reliability of its detection capabilities at the image level.
· IOPA Image Level: Sensitivity was remarkably high at 83.1% with a CI of [78.87%, 86.80%] under conservative conditions, and under optimistic conditions, it improved to 91.8%, with a CI of [88.54%, 94.42%], substantially exceeding the target goal of 0.75. Specificity was also impressive at 98.4% with a CI of [96.20%, 99.44%]. These metrics significantly surpass the performance thresholds, confirming the device's superior performance in accurately delineating caries within IOPA images.
# Subgroup Analyses
Subgroup analyses were also performed for patient demographics, sensor manufacturer,at the both image and surface levels (including cavity type) for both BW and IOPA images. These analyses demonstrated that the BDCA toolmaintains consistent diagnostic performance across various demographic and technical subgroups, respectively.
#### ii. Clinical Evaluation-Reader Improvement
The MRMC study evaluates the BDCA v1.0 Software in a multi-reader, fully crossed format to assess its effectiveness on 328 digital radiographs, including 108 BW and 220 IOPA. 72 out of 108 of BW images and 91 out of 220 of IOPA images demonstrated cavities. At the surface level, cavities were positively identified in 221 out of 2716 surfaces in BW images, and 160 out of 2967 surfaces in IOPA images. Twenty-nine United States (US) licensed dentists interpreted these radiographs twice: once with and once without BDCA Software assistance. The ground truth was established through consensus of two among three licensed US dentists.
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image contains a logo for a company called BETTERDIAGNOSTICS.AI. The logo consists of the company name in black, sans-serif font, with the ".AI" portion slightly smaller. Above the name is a red graphic element that appears to be a stylized representation of the letters "b" and "d" intertwined. The overall design is clean and modern, suggesting a tech-focused or innovative company.
Dentists were randomly assigned to two groups: Group 1 started with BDCA v1.0 assistance, and Group 2 started without. To mitigate recollection bias, a 30-day washout period was included in the study. Post washout period, group 2 assessed radiographs with BDCA's assistance and group 1 without. The study measures diagnostic performance of the model using Alternative Free Response Operating Characteristic (AFROC), sensitivity and specificity at the image and surface levels for comparing the performance between readers aided by BDCA v1.0 and reader unaided by BDCA v1.0.
## Unassisted vs. Assisted AFROC Score
The MRMC AFROC analysis across various subgroups for BW and IOPA images convincingly demonstrates that AI assistance significantly enhances diagnostic accuracy in radiographic interpretations. The results from the aided groups, displaying AUCs of 0.848 for BW and 0.845 for IOPA images, significantly surpass those of the unaided groups, which recorded AUCs of 0.806 and 0.807, respectively. The statistically significant differences in AUCs, marked by p-values less than 0.001, affirm that the use of BDCA v1.0 improves the precision with which readers identify targets in the images.
Subgroup analyses were also performed for dentist experience, patient demographics, sensor, and cavity type. Subgroup analyses of the BDCA v1.0 tool revealed significant improvements in detection accuracy across various demographic and technical variables.
#### Unassisted vs. Assisted Sensitivity at the Image Level
MRMC sensitivity performance analysis at the image level clearly demonstrates the advantages of AI assistance in radiographic |interpretations under both the Optimistic and Conservative definitions. The findings indicate that BDCA v1.0 significantly enhances the ability of readers to detect lesions more effectively when compared to unaided conditions. Under the Optimistic Definition, which considers any lesion detection as successful, aided conditions showed a marked improvement in sensitivity, with the majority of readers experiencing increased detection rates. Specifically, for BW images, the sensitivity improved from 0.857 to 0.890, and for IOPA images from 0.761 to 0.809 under aided conditions. The Conservative Definition establishes that BDCA v1.0 assistance aids in the comprehensive detection and accurate marking of all identified lesions, contributing to an overall increase in diagnostic precision.
{17}------------------------------------------------
Image /page/17/Picture/0 description: The image contains the logo for BetterDiagnostics.ai. The logo consists of a stylized, red "bd" monogram above the company name, which is written in a simple, sans-serif font. The logo is clean and modern, with the red color providing a bold contrast against the white background.
For BW images, the aided sensitivity is quantified at 0.509, an increase from the unaided sensitivity of 0.444. Similarly, for IOPA images, aided sensitivity rises to 0.619 from 0.564 in unaided conditions. The fact that these improvements are consistent across most readers under such rigorous evaluation criteria underscores the effectiveness of BDCA v1.0 assistance in enhancing diagnostic accuracy.
## Unassisted vs. Assisted Specificity at the Image Level
MRMC specificity analysis for both BW and IOPA images at the image level indicates a general trend of improvement in specificity when BDCA v1.0 is employed. For BW images, the overall specificity increased slightly from 0.634 in unaided conditions to 0.682 in aided conditions. This improvement underscores the potential of BDCA v1.0 to enhance the accuracy of negative lesion identifications in radiographic images. On the other hand, the specificity analysis for IOPA images shows a more pronounced improvement under aided conditions, with specificity increasing from 0.813 to 0.844. This difference is statistically significant, suggesting that BDCA v1.0 is more reliably beneficial in enhancing the specificity of IOPA image interpretations.
#### Unassisted vs. Assisted Sensitivity at the Surface Level
MRMC sensitivity performance analysis at the surface level provides clear evidence that AI assistance substantially enhances diagnostic sensitivity. This improvement is consistent across almost all readers and demonstrates a notable enhancement in the ability to detect lesions at the surface level when compared to unaided conditions. For BW Images, the overall sensitivity under aided conditions is significantly higher (0.763) compared to unaided conditions (0.707), with a statistically significant mean increase of 0.056. This improvement indicates a robust benefit provided by BDCA in enhancing lesion detection capabilities. For IOPA Images, aided conditions show a higher overall sensitivity (0.746) than unaided conditions (0.691), with a significant mean increase of 0.055. This indicates that AI assistance consistently enhances the detection of lesions, providing clear benefits in clinical diagnostics.
#### Unassisted vs. Assisted Specificity at the Surface Level
MRMC specificity analysis at the surface level for both BW and IOPA images reveals that AI assistance leads to statistically significant improvements in specificity. For BW Images, the overall specificity under aided conditions shows an increase to 0.980 from 0.974 in unaided
{18}------------------------------------------------
Image /page/18/Picture/0 description: The image contains a logo for BETTERDIAGNOSTICS.AI. The logo consists of a stylized, interconnected red symbol resembling the letters 'b' and 'd' stacked on top of each other. Below the symbol, the text "BETTERDIAGNOSTICS.AI" is written in a sans-serif font, with the ".AI" portion slightly smaller than the rest of the text.
conditions. This slight but significant improvement reflects the capability of AI to enhance the precision of radiographic diagnostics by minimizing false positives. Similarly for IOPA Images, the overall specificity under aided conditions is slightly improved at 0.983 compared to 0.979 in unaided conditions.
#### 9. Conclusion
Based on the information submitted in this premarket notification, and based on the indications for use, technological characteristics and performance testing, the Better Diagnostics Caries Assist (BDCA) Version 1.0 raises no new questions of safety and effectiveness and is substantially equivalent to the predicate device Overjet Caries Assist Software (Version 1.0) (K222746) in terms of safety, effectiveness, and performance.
| | Better Diagnostics AI Corp. 510(k) submission | Page 15 of 15 |
|--|-----------------------------------------------|---------------|
|--|-----------------------------------------------|---------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.