HeartKey Rhythm is intended for use by qualified healthcare professionals for analysing ECG data from adults. HeartKey Rhythm analyses a single ECG lead recorded from compatible devices used for arrhythmia diagnostics such as Holters, Extended Holters, ECG Patches, Event Recorders, or other similar devices when assessment of the rhythm is necessary. The HeartKey Rhythm application program interface (API) is intended to be integrated into other device software to provide the following capabilities: - ECG signal conditioning, - ECG signal quality assessment, - QRS beat detection, - Ventricular ectopic beat detection, - ECG rhythm analysis HeartKey Rhythm is not intended for use in life supporting, or sustaining systems, or real-time ECG monitors, or cardiac alarm devices, or OTC-use only devices. HeartKey Rhythm analysis results are not intended to be the sole means of diagnosis. It is offered to physicians on an advisory basis only in conjunction with the physician's knowledge of ECG patterns, patient background, clinical history, symptoms, and other diagnostic information.
Device Story
HeartKey Rhythm is a software library integrated via REST API into ECG management or remote patient monitoring platforms. It processes single-lead ambulatory ECG data (Holter, patch, event recorder) to perform signal conditioning, quality assessment, QRS detection, ventricular ectopic beat detection, and rhythm analysis. The device uses a combination of rule-based calculations and AI/ML technology to analyze ECGs. It operates on PCs or servers within home or hospital environments. The output is provided to physicians as advisory information to support clinical decision-making regarding cardiac rhythm, complementing the physician's clinical assessment. It does not provide real-time monitoring or alarms and is not for life-sustaining use.
Clinical Evidence
Bench testing only. Performance evaluated using a proprietary dataset of >1200 ECGs (diverse demographics). Metrics included sensitivity, specificity, and positive predictive values. Verification testing confirmed adherence to design inputs and specifications. No residual anomalies identified. Compliance with ANSI/AAMI EC57:2012, IEC 60601-2-47:2012, and IEC 62304:2006+A1:2015.
Technological Characteristics
Software-only library; REST API interface. Implements rule-based and AI/ML algorithms for ECG signal processing, beat detection, and rhythm classification. Compatible with single-lead ECG inputs (wet electrodes). Deployed on PC/server/cloud platforms. Complies with IEC 62304 (software lifecycle) and IEC 81001-5-1 (cybersecurity).
Indications for Use
Indicated for analysis of single-lead ECG data in adults using compatible ambulatory devices (Holters, patches, event recorders) to assist qualified healthcare professionals in arrhythmia diagnostics.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.
B-Secur Limited Cameron Moore Quality Assurance Manager 11th Floor, City Quays 3 92 Donegall Quay Belfast, BT1 3FE United Kingdom
Re: K233755
Trade/Device Name: HeartKey® Rhythm Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK, DPS Dated: July 19, 2024 Received: July 22, 2024
Dear Cameron Moore:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Kimberly N. Crowley -S
For: Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices
{3}------------------------------------------------
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233755
Device Name HeartKey® Rhythm
#### Indications for Use (Describe)
HeartKey Rhythm is intended for use by qualified healthcare professionals for analysing ECG data from adults. HeartKey Rhythm analyses a single ECG lead recorded from compatible devices used for arrhythmia diagnostics such as Holters, Extended Holters, ECG Patches, Event Recorders, or other similar devices when assessment of the rhythm is necessary.
The HeartKey Rhythm application program interface (API) is intended to be integrated into other device software to provide the following capabilities:
- · ECG signal conditioning,
- · ECG signal quality assessment,
- · QRS beat detection,
- · Ventricular ectopic beat detection,
- ECG rhythm analysis
HeartKey Rhythm is not intended for use in life supporting, or sustaining systems, or real-time ECG monitors, or cardiac alarm devices, or OTC-use only devices. HeartKey Rhythm analysis results are not intended to be the sole means of diagnosis. It is offered to physicians on an advisory basis only in conjunction with the physician's knowledge of ECG patterns, patient background, clinical history, symptoms, and other diagnostic information.
| Type of Use (Select one or both, as applicable) |
|------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="padding-right: 5px;"> <b> <span style="font-size: 16px;">☑</span> </b> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span style="padding-right: 5px;"> <b> <span style="font-size: 16px;">☐</span> </b> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) Summary
| Date Prepared: | 19 July 2024 |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Submitter/Applicant: | B-Secur Ltd<br>City Quays 3<br>92 Donegall Quay<br>Belfast<br>BT1 3FE<br>United Kingdom |
| Official Contact: | Cameron Moore, Quality Assurance Manager |
| Primary Contact Person: | Cameron Moore<br>Quality Assurance Manager, B-Secur Limited.<br>Email: cameron.moore@b-secur.com<br>Phone: +44 (0)28 9568 5489 |
| Secondary Contact Person: | Prabhu Raghavan<br>Regulatory Consultant for B-Secur Limited.<br>Email: prabhu@mdqr.solutions<br>Phone: 408-316-5707 |
| Proprietary or Trade Name: | HeartKey® Rhythm |
| Common/Usual Name: | Electrocardiograph |
| Classification Name: | 21CFR 870.1425<br>DQK - Programmable diagnostic computer<br>Class II |
| Predicate Device: | K210822 - Medicalgorithmics S.A. - DeepRhythmAI |
### Device Description:
HeartKey Rhythm is intended to automatically analyse ECG data and provide ECG signal processing and assessment of cardiac arrhythmias. The software processes the following types of ambulatory ECGs:
- Standard Holter: Lead I, II, III electrode configuration using gel electrodes e.g., Ag/AgCl,
- Patch: Modified Lead II, Lead III, sternum using gel electrodes e.g., Ag/AgCl .
HeartKey Rhythm is capable of analysing both continuous ECGs as well as event recorders. The device provides the following analysis:
{6}------------------------------------------------
| Feature | Description |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ECG Signal<br>Processing Engine<br>(SPE) | Signal Conditioning (SC): Reduce the noise artefact present in<br>the input ECG signal for physician review of NSR, AF and VE<br>beats.<br><br>Beat Detector (BD): Detect the QRS complexes of the input<br>ECG signal.<br><br>Signal Quality (SQ): Identify and classify the ECG segments'<br>signal quality from the input ECG signal. |
| ECG Arrhythmia<br>Analysis (AA) | Rhythm Analysis:<br>Normal Sinus Rhythm<br>Atrial Fibrillation<br>Cardiac Pause<br>Tachycardia<br>Bradycardia<br>Inconclusive<br>Unreadable<br>Beat Analysis:<br>Normal (N)<br>Ventricular (V) Ectopic<br>Beat Clustering:<br>Beat Cluster Template<br>Beat Cluster Count<br>QRS Location<br>Beat Classification<br>Heart Rate |
The Software Library may be integrated into ECG management systems, ECG analysis platforms or remote patient monitoring platforms. The filtered and enhanced ECG is not intended to be used by other automated analysis algorithms.
### Principle of Operation:
It is intended that the library is integrated into other software such as ECG management systems, ECG analysis platforms or remote patient monitoring platforms. Such applications could be written to run on a PC or server. All features are accessed via a documented rest application programming interface (API).
HeartKey Rhythm analyses the input ECG using a combination of rules-based calculations and AI / ML technology. The software is compiled and remains static, with no further modification of the code carried out once deployed.
### Indications for Use:
HeartKey Rhythm is intended for use by qualified healthcare professionals for analysing ECG data from adults. HeartKey Rhythm analyses a single ECG lead recorded from compatible
{7}------------------------------------------------
devices used for arrhythmia diagnostics such as Holters, Extended Holters, ECG Patches, Event Recorders, or other similar devices when assessment of the rhythm is necessary.
The HeartKey Rhythm application program interface (API) is intended to be integrated into other device software to provide the following capabilities:
- ECG signal conditioning, ●
- ECG signal quality assessment,
- QRS beat detection, ●
- . Ventricular ectopic beat detection,
- ECG rhythm analysis
HeartKey Rhythm is not intended for use in life supporting, or sustaining systems, or real-time ECG monitors, or cardiac alarm devices, or OTC-use only devices. HeartKey Rhythm analysis results are not intended to be the sole means of diagnosis. It is offered to physicians and clinicians on an advisory basis only in conjunction with the physician's knowledge of ECG patterns, patient background, clinical history, symptoms, and other diagnostic information.
## Patient Population:
Adults.
Environments of use: Home or hospital environment.
{8}------------------------------------------------
Table 3.1. outlines the proposed device vs. the predicate.
As part of the comparison, the following shall be presented & discussed:
- Indications for Use
- Technology and Principle of Operation •
- . Performance and Specifications
Table 3.1 is a comparison - Subject Device vs. the Predicate, K210822 - Medicalgorithmics S.A. - DeepRhythmAl.
Note: The comparison to the predicate is based upon available technical specifications vs. comparative testing.
{9}------------------------------------------------
| | Predicate | Subject device |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | DeepRhythmAl | HeartKey® Rhythm |
| K# | K210822 | K233755 |
| Product Code | DQK and DPS | DQK and DPS |
| CFR | 870.1425 | 870.1425 |
| Classification | Class II | Class II |
| Regulation Number(s) | 21 CFR §870.1425 | 21 CFR §870.1425 |
| | 21 CFR §870.2340 | 21 CFR §870.2340 |
| Classification name | Programmable<br>Diagnostic Computer,<br>Electrocardiograph | Programmable<br>Diagnostic Computer,<br>Electrocardiograph |
| Intended Use | DeepRhythmAI is a cloud-based<br>software for the assessment of<br>cardiac arrhythmias using two<br>lead ECG data in adult patients.<br>It is intended for use by a<br>healthcare solution integrator to<br>build web, mobile or another<br>types of applications to let<br>qualified healthcare<br>professionals review and<br>confirm the analytic result. | HeartKey Rhythm is intended<br>for use for signal processing and<br>analysis of ambulatory ECGs. It<br>is intended for non-active<br>patient monitoring or for non-<br>urgent clinical decision making. |
| Indications for Use | DeepRhythmAI is a<br>cloud-based software for the<br>assessment of<br>cardiac arrhythmias<br>using two lead ECG data in<br>adult patients.<br>It is intended for use by a<br>healthcare solution integrator to<br>build web, mobile or another<br>types of applications to let<br>qualified healthcare<br>professionals review and<br>confirm the analytic result. The<br>product supports downloading<br>and analyzing data<br>recorded in the<br>compatible formats from<br>dedicated ambulatory ECG<br>devices such as | HeartKey Rhythm is intended<br>for use by qualified healthcare<br>professionals for analysing ECG<br>data from adults. HeartKey<br>Rhythm analyses a single ECG<br>lead recorded from compatible<br>devices used for arrhythmia<br>diagnostics such as Holters,<br>Extended Holters, ECG Patches,<br>Event Recorders, or other<br>similar devices when assessment<br>of the rhythm is necessary.<br>The HeartKey Rhythm<br>application program interface<br>(API) is intended to be<br>integrated into other device<br>software to provide the<br>following capabilities: ECG<br>signal conditioning, ECG signal |
| | similar devices when the<br>assessment of the rhythm is<br>necessary.<br>The product can be<br>electronically interfaced<br>and perform analysis<br>with data transferred<br>from other computer based ECG<br>systems, such as an ECG<br>management system.<br>DeepRhythmAI can be<br>integrated into medical devices.<br>In this case, the medical device<br>manufacturer will identify the<br>indication for use depending on<br>the application of their device.<br>DeepRhythmAI is not for use in<br>life-supporting or sustaining<br>systems or ECG Alarm devices.<br>Interpretation results are not<br>intended to be the sole means of<br>diagnosis. It is offered to<br>physicians and clinicians on an<br>advisory basis only in<br>conjunction with the physician's<br>knowledge of ECG patterns,<br>patient background, clinical<br>history, symptoms and other<br>diagnostic information. | beat detection, ECG rhythm<br>analysis.<br>HeartKey Rhythm is not<br>intended for use in life<br>supporting, or sustaining<br>systems, or real-time ECG<br>monitors, or cardiac alarm<br>devices, or OTC-use only<br>devices. HeartKey Rhythm<br>analysis results are not intended<br>to be the sole means of<br>diagnosis. It is offered to<br>physicians and clinicians on an<br>advisory basis only in<br>conjunction with the physician's<br>knowledge of ECG patterns,<br>patient background, clinical<br>history, symptoms, and other<br>diagnostic information. |
| Level of<br>Concern/Documentation<br>Level | Moderate | Enhanced |
| Components | Software only: 1) A web API 2)<br>An automated proprietary<br>algorithm. | Software only: 1) A web API 2)<br>An automated proprietary<br>algorithm. |
| Interface | Web application<br>programming interface<br>(API) | Web application<br>programming interface<br>(API) |
| Part responsible<br>for ECG signal<br>analysis | The automated<br>proprietary deep-learning<br>algorithm, which<br>measures and analyzes ECG<br>data to provide<br>qualified healthcare<br>professional with<br>supportive information | The automated<br>proprietary machine-learning<br>algorithm, which<br>measures and analyzes ECG<br>data to provide<br>qualified healthcare<br>professional with<br>supportive information |
| | for review. | for review. |
| Number of leads<br>for a received<br>ECG signal | Two lead ECG data | A single lead from one of the<br>following leads:<br>Wet Electrode Lead I, II, and III<br>Wet Electrode Modified Leads II<br>and III<br>Wet Electrode Sternum Lead |
| Display or<br>Graphical User<br>Interface (GUI) | No primary display or<br>GUI | No primary display or<br>GUI |
# Table 3.1 – Comparison – Subject vs. Predicate
{10}------------------------------------------------
{11}------------------------------------------------
### Substantial Equivalence Discussion
HeartKey Rhythm has the same general intended use and similar indications, technological characteristics, and principles of operation as the predicate Medicalgorithmics DeepRhythmAI, K210822.
### Intended Use/ Indications for Use
The indications for use for HeartKey Rhythm and the predicate Medicalgorithmics DeepRhythmAI, K210822 both are rest API based software libraries which are intended to be integrated into computer-based ECG systems. Both devices allow for the processing and extraction of beats from an ECG signal, providing an accurate heart rate measurement, in an ambulatory environment.
Both the proposed and the predicate allow for ECG rhythm analysis and beat classification. The indications for use are substantially equivalent.
### Environment of Use and Target Population
HeartKey Rhythm is intended for prescriptive use and for adults over 18. The predicate DeepRhythmAl has the same environment of use and population.
Both HeartKey Rhythm and DeepRhythmAl are intended for use in a home or hospital environment.
### Technological Characteristics and Principles of Operation
HeartKey Rhythm is a library collection of callable functions that have been compiled into native machine code of the computer on which they will execute. A rest application programming interface (API) calls some or all the available functions within the library to securely exchange information. This can be deployed to a variety of cloud platforms or to a virtual or physical machine.
HeartKey Rhythm is intended to be integrated into different ECG platforms such as ECG management systems, ECG analysis platforms or remote patient monitoring platforms. Medicalgorithmics DeepRhythmAl software has the same intended use and users as the predicate.
{12}------------------------------------------------
HeartKey Rhythm provides the applications/platforms it is intended to be deployed in, with ECG signal processing and beat extraction in order to provide a heart rate reading and arrhythmia analysis to the device user. The predicate DeepRhythmAI (K210822) software assesses ambulatory electrocardiogram (ECG) rhythms from adult subjects.
HeartKey Rhythm and DeepRhythmAI are based on the following technological characteristics:
- Process ECG signals ●
- . Detect peaks in ECG signals
- . Output heart rate measurements to the user
- . Rhythm analysis
- . Beat classification
HeartKey Rhythm and the predicate are substantially equivalent in relation to their technological characteristics.
### Non-clinical Testing
Nonclinical performance testing was conducted in compliance with the appropriate recognised consensus standards, to demonstrate that HeartKey Rhythm satisfies its design and software requirements as well as demonstrate that the ECG analysis performance is substantially equivalent to the predicate device (DeepRhythmAI, K210822). Testing described in this 510(k) consisted of verification of all design input requirements and product specifications. B-Secur evaluated HeartKey Rhythm against a large independent proprietary set of ECGs and standard databases, comparing its analysis output against a known reference. The proprietary dataset included more than 1200 ECGs with approximately 25% females, 25% males and 50% gender unknown, approximately 16% whites, 24% non-whites, and 60% race unknown, and approximately 66% of ECGs recorded in the US. Performance testing included the evaluation of metrics such as sensitivity, specificity, and positive values. These outputs were evaluated against a clinically relevant acceptance criteria to demonstrate HeartKey Rhythm's effectiveness. Unit, integration and system level testing conducted identified no residual anomalies during verification software tests. Cybersecurity testing was conducted in which no vulnerabilities were identified and all software requirements were satisfied. Overall, the software verification & validation testing was completed successfully and met all requirements. Testing demonstrated that the subject device performance was deemed to be acceptable.
Verification and validation testing was completed in compliance with the following standards and guidance documents:
- ANSI/AAMI EC57:2012, Testing and Reporting Performance Results of Cardiac Rhythm . And ST-Segment Measurement Algorithms.
- ANSI/AAMI/IEC 60601-2-47:2012. Medical electrical equipment Part 2-47: Particular . requirements for the basic safety and essential performance of ambulatory electrocardiographic systems.
- IEC 62304:2006+A1:2015, Medical device software Software life cycle processes. ●
{13}------------------------------------------------
- . IEC 81001-5-1:2022, Health software and health IT systems safety, effectiveness and security - Part 5-1: Security - Activities in the product life cycle.
- . General Principles of Software Validation; Final Guidance for Industry and FDA Staff (January, 2002)
- Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions; Final Guidance for Industry and FDA Staff (September, 2023)
### Substantial Equivalence Conclusion
In conclusion, HeartKey Rhythm has the same intended use as the predicate device, and any differences in technological characteristics do not raise different questions of safety or effectiveness. Differences between the subject device and the predicate have been tested to ensure that the device meets its intended use. Therefore, HeartKey Rhythm is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.